E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS

K200011 · Microport Orthopedics, Inc. · LPH · Oct 1, 2021 · Orthopedic

Device Facts

Record IDK200011
Device NameE-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS
ApplicantMicroport Orthopedics, Inc.
Product CodeLPH · Orthopedic
Decision DateOct 1, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DYNASTY® Dual Mobility Inserts and Liners, when used with compatible acetabular shells and femoral heads, are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients. In revision arthroplasties, all devices associated with the wear couple must be removed and replaced.

Device Story

Dual mobility hip system consisting of Vitamin E cross-linked polyethylene (UHMWPE) inserts and cobalt chrome (CoCr) liners; designed for use with compatible acetabular shells and femoral heads in total hip arthroplasty. System creates two articulating interfaces in the acetabular joint space to improve hip function and reduce pain. Inserts provide primary articulation for 28mm non-skirted femoral heads; CoCr liners provide secondary articulation against the insert via a taper locking mechanism. Used in clinical settings by orthopedic surgeons for primary or revision arthroplasty. Implants are single-use and intended for uncemented fixation. System benefits patients by addressing dislocation risks and restoring joint function in degenerative or traumatic hip conditions.

Clinical Evidence

No clinical data. Substantial equivalence is supported by extensive non-clinical bench testing, including range of motion analysis (EN ISO 21535), impingement testing (ASTM F2582-14, F1820-13), static disassociation testing (ASTM F1820-13), high-angle edge loading/fatigue (ASTM F2003-02, ISO 7206-6), fretting and corrosion assessment, wear testing (ISO 14242 series), and push-out/lever-out/torque-out testing (ASTM F1820-13). MRI safety was evaluated per ASTM F2052, F2119, and F2182. UHMWPE material characterization was performed per various ASTM and ISO standards.

Technological Characteristics

Materials: Vitamin E cross-linked polyethylene (ASTM F2695) and cobalt chrome alloy (ASTM F1537 Type 1). Design: Dual-articulation hip system with taper locking mechanism. Dimensions: Inserts 28mm ID, 38-56mm OD; CoCr liners 38mm+ ID. Connectivity: None (mechanical implant). Sterilization: Not specified. Software: None.

Indications for Use

Indicated for skeletally mature patients requiring total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), correction of functional deformity, revision procedures, dislocation risks, and treatment of non-union, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement unmanageable by other techniques.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The text reads "U.S. FOOD & DRUG ADMINISTRATION". October 1, 2021 Microport Orthopedics, Inc. Matthew Paul Sr. Regulatory Affairs Project Manager 5677 Airline Road Arlington, Tennessee 38002 Re: K200011 Trade/Device Name: E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH. KWY. LZO. OOG Dated: September 2, 2021 Received: September 3, 2021 Dear Matthew Paul: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Limin Sun, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K200011 Device Name E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS Indications for Use (Describe) The DYNASTY® Dual Mobility Inserts and Liners, when used with compatible acetabular shells and femoral heads, are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients. In revision arthroplasties, all devices associated with the wear couple must be removed and replaced. Indications for Use: 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2) inflammatory degenerative joint disease such as rheumatoid arthritis; 3) correction of functional deformity; and; 4) revision procedures where other treatments or devices have failed; 5) dislocation risks; 6) treatment of non-union, femoral neck fractures of the proximal femur with head involvement which are unmanageable by other techniques Dual Mobility Inserts and Liners are single use implants intended for uncemented arthroplasty. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------| | <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K200011 Page 1 of 3 | | 510(k) Summary | |----------------------------|---------------------------------------------------------------------------------------------------------| | Submitted By: | MicroPort Orthopedics Inc.<br>5677 Airline Road<br>Arlington, TN 38002 | | | Telephone Number: (901) 290-5175 | | Date Prepared: | October 1, 2021 | | Contact Person: | Matthew Paul | | | Sr. Regulatory Affairs Project Manager | | | Telephone number: (901) 867-4350 | | | Email: matt.paul@ortho.microport.com | | Name of Device: | E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL<br>MOBILITY LINERS | | Common Name: | Dual Mobility Inserts and Dual Mobility Liners | | Device Classification Name | 21 CFR 888.3358 - Hip joint metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis | | And Reference: | Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented | | Device Class: | Class II | | Panel Code: | Orthopedics/87 | | Product Code: | LPH, KWY, LZO, OQG | | Predicate Device: | ZIMMER BIOMET G7® DUAL MOBILITY SYSTEM (K150522) | | Additional Predicate: | ZIMMER BIOMET G7® DUAL MOBILITY SYSTEM (K161190) | | Reference Devices: | MEDACTA VERSAFITCUP® (K083116) | | | ORTHOMET BIPOLAR HIP SYSTEM (K892398) | | | PROCOTYL® PRIME® E-CLASS® XLPE LINERS (K171181) | | | DYNASTY® COCR ACETABULAR LINERS (K061844) | #### Device Description: The Dual Mobility (DM) Inserts and Liners are designed for use with compatible DYNASTY® acetabular shells, MicroPort Orthopedics non-skirted femoral heads and PROFEMUR® femoral stems, to create a system with two articulating interfaces in the acetabular joint space of the hip. The E-CLASS® DUAL MOBILITY INSERTS are manufactured from Vitamin E cross linked polyethylene conforming to ASTM F2695 and are available with an inner diameter of 28 mm and outer diameters from 38 mm to 56 mm in 2 mm increments. The E-Class® DM Inserts are compatible with and provide a primary articulating surface for all 28 mm, non-skirted femoral heads. DM CoCr Liners are manufactured from cobalt chrome (CoCr) alloy conforming to ASTM F1537 (Type 1) and are available with inner diameters from 38 mm in 2 mm increments. The DM CoCr Liners are designed to mate with all DYNASTY® acetabular shells via a taper locking mechanism and provide the secondary articulating surface for the E-Class® DM Inserts. {4}------------------------------------------------ #### Intended Use Statement: The DYNASTY® Dual Mobility Inserts and Liners, when used with compatible acetabular shells and femoral heads, are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in sketally mature patients. In revision arthroplasties, all devices associated with the wear couple must be removed and replaced. #### Indications for Use - non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, 1) protrusio acetabuli, and painful hip dysplasia: - 2) inflammatory degenerative joint disease such as rheumatoid arthritis; - 3) correction of functional deformity; and, - 4) revision procedures where other treatments or devices have failed; - 5) dislocation risks; - treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur 6) with head involvement which are unmanageable by other techniques. Dual Mobility Inserts and Liners are single use implants intended for uncemented arthroplasty. #### Comparison of Technological Characteristics with the Predicate device: Device comparison described in this premarket notification demonstrates that the subject E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS are substantially equivalent to the identified predicate ZIMMER BIOMET G7® DUAL MOBILITY SYSTEM (K150522) cleared by the FDA for commercial distribution in the United States. The subject devices were shown to be substantially equivalent and have similar technological characteristics to their predicate and reference devices in areas including design, intended use, material composition, operational principles and function. ## Discussion of the Non-clinical Testing/Performance Data: ## Mechanical Testing: MicroPort has evaluated the subject E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS and data demonstrates substantial equivalence to the identified predicate ZIMMER BIOMET G7® DUAL MOBILITY SYSTEM (K150522). The testing and engineering analyses performed for the subject E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS included: - 1. Range of Motion (ROM) Analysis utilizing EN ISO 21535 (2009) - 2. Impingement Testing utilizing ASTM F2582-14, ASTM F1820-13, ISO 14242-1 (2014), ISO 14242-2 (2016), - 3. Static Disassociation Testing of a femoral head from a dual mobility insert, utilizing ASTM F1820-13 - 4. High Angle Edge Loading (Fatigue Testing) utilizing ASTM F2003-02(2015), ISO 7206-6:2013 - 5. Fretting and Corrosion Observational Assessment utilizing ASTM F2003-02(2015), ISO 7206-6:2013 - 6. Wear Testing and Engineering Analysis utilizing ISO 14242-1 (2014), ISO 14242-2 (2016), ISO 14242-3 (2009)/Amd 1 (2019) - 7. Push-out, Lever-out, Torque-out (PO/LO/TO) Testing of a dual mobility liner from a shell, utilizing ASTM F1820-13 - 8. Magnetic Resonance Imaging (MRI) Safety Evaluation utilizing ASTM F2052-6. ASTM F2119-7. ASTM F2182-11a - 9. Characterization of Ultrahigh Molecular Weight Polyethylene (UHMWPE) Rationale utilizing ASTM F2003-02(2015), ASTM F2102-13, ASTM F2183-02, ASTM F2214-02, ASTM F2565-13, {5}------------------------------------------------ ASTM F2625-10, ASTM F2695-12, ASTM F2381-10, ISO 5834-2 (2011), ISO 5834-3 (2005), ISO 5834-4 (2005), ISO 5834-5 (2005), ISO 11542-2 (1998), ISO 527-3 (1995) The data demonstrate that the subject E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS are substantially equivalent to the legally marketed predicate, ZIMMER BIOMET G7® DUAL MOBILITY SYSTEM (K150522). Having met all acceptance criteria for mechanical testing performed on worst case constructs, the subject E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS do not introduce new or modified risks for safety and effectiveness compared to the predicate or other reference MicroPort Hip Systems and acetabular components which utilize comparable test methods and acceptance criteria. #### Conclusion: Based on the design features, the use of established well-known materials, feature comparisons, indications for use, principle of operations and results of the mechanical testing, the subject E-CLASS® DUAL MOBILITY INSERTS and DYNASTY® DUAL MOBILITY LINERS have shown to be substantially equivalent to the legally marketed predicate device cited in this premarket notification.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%