K193432 · 3D Systems · OYK · Nov 9, 2020 · Orthopedic
Device Facts
Record ID
K193432
Device Name
Vantage PSI System
Applicant
3D Systems
Product Code
OYK · Orthopedic
Decision Date
Nov 9, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3110
Device Class
Class 2
Indications for Use
The Vantage PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Vantage PSI System is intended for use with Exactech's Vantage Total Ankle System and its cleared indications for use.
Device Story
System uses patient CT imaging data to create patient-specific surgical guides and anatomical models; guides fit distal tibial and proximal talar anatomy. Physician uses guides intraoperatively to assist in positioning total ankle replacement components and guiding bone cuts. System includes stainless steel fluoroscopic inserts for alignment. Outputs include surgical guides, models, and case reports. Designed to improve accuracy of implant placement compared to standard instrumentation. Used in clinical settings by orthopedic surgeons.
Clinical Evidence
Bench and cadaveric testing performed. Cadaver study with three surgeons compared subject device to standard instrumentation. 2D analysis (fluoroscopic images) measured AP tibia varus/valgus, bearing-to-bearing, and lateral tibia. 3D analysis (post-op CT segmentation/STL alignment) measured translations and rotations across three axes. Results demonstrated performance equivalent or better than standard instrumentation. Additional testing included process qualification, cleaning validation (AAMI TIR 30), steam sterilization (ISO 17665-1), biocompatibility (ISO 10993-1), packaging (ASTM/NMFC), shelf life (ASTM F1980-16), and debris validation.
Technological Characteristics
Patient-specific guides manufactured via 3D printing from biocompatible polyamide. Accessories are stainless steel. System utilizes CT-based image transformation and manipulation software. Sterilization via steam (dynamic-air-removal cycle).
Indications for Use
Indicated for patients undergoing total ankle replacement requiring patient-specific surgical planning and instrumentation to assist in component positioning and bone cutting. Used in conjunction with the Exactech Vantage Total Ankle System.
Regulatory Classification
Identification
An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
PROPHECY® INVISION® Preoperative Navigation Alignment Guides, Wright Medical Technology, Inc. (K170968)
Reference Devices
VSP® Orthopedics System, 3D Systems Inc. (K190044)
{0}------------------------------------------------
November 9, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
3D Systems Kim Torluemke Vice President, Quality & Regulatory, Healthcare 5381 South Alkire Circle Littleton, Colorado 80127
## Re: K193432
Trade/Device Name: Vantage PSI System Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: OYK Dated: November 5, 2020 Received: November 6, 2020
Dear Kim Torluemke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, PhD, RAC Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K193432
Device Name Vantage PSI System
Indications for Use (Describe)
The Vantage PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Vantage PSI System is intended for use with Exactech's Vantage Total Ankle System and its cleared indications for use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span>☑</span><span> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div style="display:flex; align-items:center;"><span>☐</span><span> Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# 1. INTRODUCTION
This document contains the 510(k) summary for the Vantage™ PSI System. The content of this summary is based on the requirements of 21 CFR 807.92.
## 2. SUBMITTER
3.
| Name: | Medical Modeling - A 3D Systems Company |
|----------------------|------------------------------------------------------------------------------------------------------|
| Address: | 5381 South Alkire Circle<br>Littleton, CO 80127, USA<br>Phone: (720) 643-1001<br>Fax: (720) 643-1009 |
| Official Contact: | Kim Torluemke<br>Vice President, Quality and Regulatory, Healthcare |
| Date Prepared: | November 5, 2020 |
| DEVICE | |
| Trade Name: | Vantage PSI System |
| Common Name: | Ankle surgical guide system |
| Classification Name: | Ankle Arthroplasty Implantation System |
| Classification: | Class II, 21 CFR 888.3110 |
Product Code: ОУК
#### 4. PREDICATE DEVICES
Predicate device:
- · PROPHECY® INVISION® Preoperative Navigation Alignment Guides, Wright Medical Technology, Inc.(K170968)
Reference devices:
- VSP® Orthopedics System, 3D Systems Inc. (K190044)
### 5. DESCRIPTION OF THE DEVICE
3D System's Vantage PSI System is patient-specific guides created to fit the contours of the patient's distal tibial and proximal talar anatomy. The guides and models are designed and manufactured from patient imaging data (CT) and are made from biocompatible nylon. The surgical guides in combination with Exactech Vantage Total Ankle reusable instruments, facilitate the positioning of
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the 3D Systems logo. The logo consists of a stylized cube with the number "3D" on its faces, followed by the text "3D SYSTEMS" in a simple, sans-serif font. The cube is rendered in a metallic silver color, giving it a modern and technological appearance. The text is in a similar color, creating a cohesive and professional brand identity.
Vantage Total Ankle Implants. 3D System's Vantage PSI System produces a variety of patient specific outputs including surgical guides, anatomic models, and case reports for use with Exactech's Vantage Total Ankle System (K152217 and K183343).
#### 6. INTENDED USE
The Vantage PSI System is intended to be used as patient specific surgical planning and instrumentation to assist in the positioning of total ankle replacement components intraoperatively, and in guiding bone cutting. The Vantage PSI System is intended for use with Exactech's Vantage Total Ankle System and its cleared indications for use.
#### 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The intended use and technological characteristics of the subject device (Vantage PSI System) are either identical or substantially equivalent to the predicate device (PROPHECY® Preoperative Navigation Alignment System):
The principles of operation and technological characteristics are all substantially equivalent between the Vantage PSI System and the identified predicate device. Specifically, the predicate device includes 3D printing of patient specific anatomical models and surgical guides with polyamide materials. Both have stainless steel accessories produced from a common subtractive manufacturing process.
The Vantage PSI System uses the same fundamental technologies as its predicate device. These include image transformation and manipulation via software that is subsequently used for 3D printing of anatomical models and guides and for surgical planning, only potentially differing in the software applications.
The outputs of the Vantage PSI System include materials commonly used in surgical instruments (stainless steel) and for surgical guides and anatomical models (biocompatible polyamide). These materials are substantially equivalent to those employed by the predicate, which also uses a polyamide material and stainless steel surgical instruments.
The Vantage PSI System Accessories are standardized stainless steel fluoroscopic inserts that are inserted into patient-specific guides to assist in alignment operations. The accessories are similar to those included with the predicate device, but differ in shape. The predicate device utilizes standard surgical pins. The subject device includes a specific single use tool to be used exclusively with the system. These accessories are sterilized by the healthcare facility with the same cycle as the other system outputs. The minor differences in accessory geometry does not change the intended use of the system, and does not raise new questions of safety and effectiveness.
{5}------------------------------------------------
#### 8. SUMMARY OF PERFORMANCE TESTING
The testing outlined below was intended to show that the output of the design and development process demonstrated compliance with the device specifications. Bench and cadaveric testing was conducted to prove the subject device performs in accordance with its intended use and is substantially equivalent to the listed predicate device.
The following testing was conducted in accordance with the listed consensus standards for the Vantage PSI System:
### · Process Qualification
Process qualification was conducted to assess the manufacturing process as well as operator repeatability within the digital workflow. Cases utilized for testing were representative of ankle reconstruction procedures within the subject device's intended use. Both digital and physical outputs from all manufacturing processes were verified against design specifications. All test method acceptance criteria were met.
#### · Cleaning Validation
Cleaning validations were performed in accordance with AAMI TIR 30. Following soiling and cleaning of the subject devices, bioburden, protein levels, and hemoglobin levels were examined. All test method acceptance criteria were met.
#### • Sterilization Validation
Steam sterilization validations were performed for the dynamic-air-removal cycle in accordance with ISO 17665-1:2006 to a sterility assurance level (SAL) of 10* using the biological indicator (BI) overkill method. All test method acceptance criteria were met.
### • Biocompatibility Validation
Biocompatibility endpoints were evaluated in accordance with ISO 10993-1. The battery of cytotoxicity, sensitization, irritation, and acute systemic toxicity testing conducted on the subject devices manufactured from polyamide and titanium alloy are within the pre-defined acceptance criteria. The results of the testing adequately address biocompatibility for the output devices and their intended use.
### • Packaging Validation
Packaging validation was performed on all packaging configurations of the system. Testing was performed in accordance with the 3D Systems Transportation Test Standard (based on ASTM and National Motor Freight Classification standards). All test method acceptance criteria were met.
### • Shelf Life Validation
Shelf life validation was conducted in accordance with ASTM F1980-16 to verify the stated shelf life duration. All test method acceptance criteria were met.
### • Debris Validation
Debris generation validation was performed utilizing standard orthopedic instrumentation. Debris generation was evaluated and compared to published data. The results of the testing perform equivalent or better than the values found in literature.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the 3D Systems logo. The logo consists of a stylized 3D cube on the left and the text "3D SYSTEMS" on the right. The text is in a sans-serif font and is gray in color. The logo is clean and modern in appearance.
#### • Cadaver Study / Design Validation
Cadaveric and design validation studies were performed to compare the subject device to Vantage Total Ankle standard instrumentation. The cadaver and design validation studies utilized three surgeon users. 2D analysis was conducted by measuring AP tibia varus/valgus, bearing-to-bearing, and lateral tibia from intra-operative fluoroscopic images, and compared to the pre-operative plan. 3D analysis was conducted by comparing final implant placement to pre-operative implant placement. The 3D analysis consisted of post-operative CT image segmentation, STL alignments, and measurements. Translations were analyzed across three axis (medial/lateral, anterior/posterior, and distal/proximal) and rotations were analyzed across three axis (varus/valgus, plantarflexion/dorsiflexion, and internal/external). The results of both 2D and 3D analysis demonstrate the subject device performing equivalent or better than the Vantage Total Ankle standard instrumentation.
#### Summary
All design, process, and other verification and validation testing, which were conducted as a result of risk analyses and design impact assessments, showed conformity to pre-established specifications and acceptance criteria. The acceptance criteria was established in support of device performance, and testing demonstrated substantial equivalence of the system to the predicate device.
#### 9. CONCLUSION
Based on a comparison of the intended use and technological characteristics, the Vantage PSI System is substantially equivalent to the identified predicate device. Minor differences in the indications for use, technological and performance characteristics do not raise new or different questions of safety and effectiveness. Additionally, the non-clinical testing supports that the system performs in accordance with its intended use and is as safe, as effective, and performs as well as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.