K193381 · Lifeline Software, Inc. · MUJ · Dec 31, 2019 · Radiology
Device Facts
Record ID
K193381
Device Name
RadCalc Software
Applicant
Lifeline Software, Inc.
Product Code
MUJ · Radiology
Decision Date
Dec 31, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
RadCalc has the following intended uses: 1. RadCalc performs a secondary dose calculation verification on the treatment plan done by the treatment planning software. This is RadCalo's primary function. Radiation therapy systems typically calculate the monitor units needed to deliver the desired amount of radiation to a point of reference within the patient. In this situation, RadCale will serve to validate those monitor units computed by the primary radiation therapy planning system. Additional verification activities revolve around point dose comparisons, 3D dose evaluation via Gamma analysis, and DVH comparisons. It is not the intention of RadCalc to replace the calculation performed by the primary radiation therapy planning computer but to validate its calculation as a means of quality assurance. The practice of performing a secondary check is recommended by the American Association of Physicis in Medicine (AAPM) Task Group 40 as part of a good quality assurance program. This practice is an important aspect in providing quality patient care. 2. Import data from the treatment planning software and export the treatment planning system to the verify and record system, which is the device that actually controls the radiation beam. This will reduce the number of errors that occur as a result of manually inputting this data. 3. In addition to performing the secondary dose verification calculation, RadCalc can also be used as the primary means of calculating monitor units in situations where the physician does not order the use of a radiation therapy treatment plan. RadCalc can independently calculate the amount of radiation the beam should produce (called the MU or monitor unit) to deliver to the patient the radiation dose the doctor recommends. This function is usually only used in urgent, emergency situations. 4. In addition, RadCalc performs brachytherapy-type calculations. For brachytherapy calculations, High Dose Rate (HDR), Low Dose Rate (LDR), and Permanents can be verified. Verification activities revolve around point dose comparisons, 3D dose evaluation via Gamma analysis, and DVH comparisons. RadCalc is not used as a primary means of calculating patient dose for brachytherapy treatments. 5. Analysis of fluence and dose maps can be percentage difference, distance to agreement, or gamma analysis methodologies. 6. Interoperability with external dose calculation engines (EDCE) by sending them treatment plans and associated information in their necessary format so that the EDCE can perform a 3D dose calculation using its dose calculation algorithm. The computed dose volume is received back and the 3D analysis tools described above are used to compare against the treatment planning system.
Device Story
RadCalc Software (v7.1) is a secondary dose verification tool for radiation therapy. It imports treatment plans from primary planning systems to validate monitor units (MU) and dose distributions; performs point dose comparisons, 3D dose evaluation via Gamma analysis, and DVH comparisons. In urgent cases, it serves as a primary MU calculator. It also facilitates data transfer between planning and record-and-verify systems to minimize manual entry errors. The software supports interoperability with external dose calculation engines (EDCE) for 3D dose verification. Operated by medical physicists or clinicians in clinical settings, the output provides quality assurance metrics to confirm the accuracy of the primary treatment plan before radiation delivery, thereby reducing the risk of treatment errors and improving patient safety.
Clinical Evidence
No clinical data provided; device relies on bench testing and verification of dose calculation algorithms.
Technological Characteristics
Software-based dose verification system; performs point dose, 3D Gamma analysis, and DVH comparisons. Supports interoperability with external dose calculation engines (EDCE). Functions as a secondary check for radiation therapy planning systems.
Indications for Use
Indicated for radiation therapy patients requiring secondary dose verification of treatment plans, independent monitor unit (MU) calculation for emergency/urgent cases, and verification of brachytherapy (HDR, LDR, Permanents) dose distributions. Used for quality assurance in clinical radiation oncology settings.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Related Devices
K090531 — RADCALC SOFTWARE, VERSION 6.0 · Lifeline Software, Inc. · May 27, 2009
K252863 — ClearCalc Model RADCA V2.6 · Radformation, Inc. · Dec 19, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
December 31, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
LifeLine Software, Inc. % Mr. Craig A. Laughton Chief Executive Officer 3304 S Broadway Avenue, Suite 200 TYLER TX 75701
Re: K193381
Trade/Device Name: RadCalc Software, Version 7.1 Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: MUJ, IYE Dated: December 2, 2019 Received: December 5, 2019
Dear Mr. Laughton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193381
Device Name RadCalc Software, Version 7.1
## Indications for Use (Describe) RadCalc has the following intended uses:
1. RadCalc performs a secondary dose calculation verification on the treatment plan done by the treatment planning software. This is RadCalo's primary function. Radiation therapy systems typically calculate the monitor units needed to deliver the desired amount of radiation to a point of reference within the patient. In this situation, RadCale will serve to validate those monitor units computed by the primary radiation therapy planning system. Additional verification activities revolve around point dose comparisons, 3D dose evaluation via Gamma analysis, and DVH comparisons. It is not the intention of RadCalc to replace the calculation performed by the primary radiation therapy planning computer but to validate its calculation as a means of quality assurance. The practice of performing a secondary check is recommended by the American Association of Physicis in Medicine (AAPM) Task Group 40 as part of a good quality assurance program. This practice is an important aspect in providing quality patient care.
2. Import data from the treatment planning software and export the treatment planning system to the verify and record system, which is the device that actually controls the radiation beam. This will reduce the number of errors that occur as a result of manually inputting this data.
3. In addition to performing the secondary dose verification calculation, RadCalc can also be used as the primary means of calculating monitor units in situations where the physician does not order the use of a radiation therapy treatment plan. RadCalc can independently calculate the amount of radiation the beam should produce (called the MU or monitor unit) to deliver to the patient the radiation dose the doctor recommends. This function is usually only used in urgent, emergency situations.
4. In addition, RadCalc performs brachytherapy-type calculations. For brachytherapy calculations, High Dose Rate (HDR), Low Dose Rate (LDR), and Permanents can be verified. Verification activities revolve around point dose comparisons, 3D dose evaluation via Gamma analysis, and DVH comparisons. RadCalc is not used as a primary means of calculating patient dose for brachytherapy treatments.
5. Analysis of fluence and dose maps can be percentage difference, distance to agreement, or gamma analysis methodologies.
6. Interoperability with external dose calculation engines (EDCE) by sending them treatment plans and associated information in their necessary format so that the EDCE can perform a 3D dose calculation using its dose calculation algorithm. The computed dose volume is received back and the 3D analysis tools described above are used to compare against the treatment planning system.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.