K161946 · Sun Nuclear Corporation · IYE · Sep 16, 2016 · Radiology

Device Facts

Record IDK161946
Device NameDoseCHECK
ApplicantSun Nuclear Corporation
Product CodeIYE · Radiology
Decision DateSep 16, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Sun Nuclear's DoseCHECK is a software product intended to independently calculate radiotherapy dose to provide a quality check of the planned dose.

Device Story

DoseCHECK is a secondary check QA software for radiotherapy. It ingests DICOM RT Plan, RT Structure Set, and CT Image Set data from a treatment planning system (TPS), alongside linear accelerator characterization and CT-to-ED curves. The device performs an independent 3D volumetric dose calculation using an algorithm distinct from the primary TPS. The resulting dose distribution is compared against the TPS-generated dose distribution. Used in clinical settings by medical physicists or radiation oncologists, the output allows clinicians to detect potential errors in TPS beam modeling, calculation algorithms, or inhomogeneity handling. This independent verification enhances patient safety by ensuring the accuracy of the planned radiation dose before treatment delivery.

Clinical Evidence

Bench testing only. Testing demonstrated that the device performs within design specifications and equivalently to the predicate device.

Technological Characteristics

Software-based secondary check QA system. Inputs: DICOM-RT (Plan, Structure Set, CT Image Set). Outputs: DICOM-RT dose volume. Computes 3D volumetric dose distributions using independent algorithms. Operates via comparison of independent calculation to TPS dose distribution.

Indications for Use

Indicated for use by clinicians to perform independent radiotherapy dose calculations for quality assurance of treatment plans.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 16, 2016 Sun Nuclear Corporation % Ms. Rene' Hardee Regulatory Affairs Specialist III 3275 Suntree Blvd. MELBOURNE FL 32940 Re: K161946 Trade/Device Name: Model 1217028 DoseCHECK Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: July 8, 2016 Received: July 15, 2016 Dear Ms. Hardee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K161946 Device Name Model 1217028 DoseCHECK Indications for Use (Describe) Sun Nuclear's DoseCHECK is a software product intended to independently calculate radiotherapy dose to provide a quality check of the planned dose. | Type of Use ( <i>Select one or both, as applicable</i> ) | |----------------------------------------------------------| |----------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |----------------------------------------------------------------------------------| | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is a logo for Sun Nuclear Corporation. The logo features a blue square with a yellow sun-like design inside on the left. To the right of the square, the words "SUN NUCLEAR" are written in a bold, blue font, with the word "corporation" written in a smaller, blue font underneath. # Section 5 - 510(k) Summary Provided in accordance with 21 CFR 807.92 (c) #### 1 General Provisions Date Prepared: June 10, 2016 Submitted by: Sun Nuclear Corporation 3275 Suntree Blvd. Melbourne, FL 32940 Ph: 321-259-6862 Fax: 321-259-7979 Web: www.sunnuclear.com Contact Person: Rene' Hardee ReneHardee@sunnuclear.com Ph: 321-259-6862 extension 2304 Classification Name: Accelerator, Linear, Medical Common Name: Secondary check QA software Proprietary Names: Model 1217028 DoseCHECK Establishment Registration Number: 1038814 Classification: Regulation Number: 21 CFR 892.5050 Name: Accelerator, Linear, Medical Product code: IYE Class II {4}------------------------------------------------ Predicate Device(s): | Model Name: | Mobius3D | |---------------|-----------------------------| | Common Name: | Secondary check QA software | | 510(k) # | K140660 | | Manufacturer: | Mobius Medical Systems, LP | | Submitted: | May 05, 2014 | To our knowledge, this predicate has not been subject to a designrelated recall. #### 2 Description and Use: DoseCHECK provides the clinician with the opportunity to compute dose using a different algorithm from that used by their treatment planning system (TPS). DoseCHECK utilizes the DICOM RT Plan, RT Structure Set, and CT Image Set from the TPS as inputs for the calculation, along with a linear accelerator characterization and CT-to-ED curve. The resulting dose distribution can be compared to the TPS dose distribution as a means of independent check. This comparison allows for detection of errors or inaccuracies that may occur within the TPS such as with beam modeling, calculation algorithm, and inhomogeneities. #### 3 Intended Use Statement: Sun Nuclear's DoseCHECK is a software product intended to independently calculate radiotherapy dose to provide a quality check of the planned dose. #### ব Technological Characteristics The primary technological characteristics of the Model 1217028 DoseCHECK are the usage of three dimensional volumetric imaging information and beam intensity values in DICOM-RT format to compute a dose volume (also in DICOM-RT format) and a comparison of this independent calculation to the TPS dose distribution. The technological characteristics are believed to be substantially equivalent to the predicate device. #### 5 Performance Data and Comparison with Predicate Model 1217028 DoseCHECK has been tested using appropriate bench testing methods. Test results of the modified device have demonstrated that the device performs within its design specifications and equivalently to the predicate device. #### 6 Summarv Model 1217028 DoseCHECK is believed to be substantially equivalent to the predicate Mobius3D device due to the similarities in function, technology, and performance. The intended use, performance testing, safety and effectiveness reviews demonstrate that Model 1217028 DoseCHECK is as safe, as effective, and performs as well as the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...