Dental Handpiece with models of SG200, SG200L, SA100L, and CA100L, and CA100R are intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, root canal preparations, removal of fillings, processing and finishing tooth preparations for polishing teeth. In addition, Dental Handpiece with models of SG200, SG200L are also intended for implant surgery such as perforating the bone, tapping and threading procedures. Dental Handpiece is designed for use by a trained professional in the field of general dentistry.
Device Story
Dental handpiece (models SG200, SG200L, SA100L, SA100R, CA100L, CA100R) transmits rotational force from an electric micro-motor to a dental bur for general dental procedures and implant surgery. Device connects to dental units via ISO 3964 coupling; receives energy, cooling water, and air. Rotational force is decelerated via internal gear speed reducers. Features include push-button or latch-type chucks for bur attachment. Used by trained dental professionals in clinical settings. Sterilizable via steam autoclave. Provides mechanical cutting/drilling capability; aids in restorative and surgical procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Verification included rotational speed, noise, water supply, light, operation, and appearance testing. Biocompatibility testing included MEM elution, sensitization, intracutaneous irritation, pyrogenicity, and acute systemic injection tests.
Technological Characteristics
Reusable, electrically-powered dental handpiece. Materials biocompatible per ISO 10993-1. Complies with ISO 14457 (handpieces/motors), ISO 3964 (coupling), and ISO 1797 (shanks). Features internal/external spray, gear ratios 20:1/1:1, and optical fiber for illumination (select models). Sterilization: steam autoclave per ISO 17665-1.
Indications for Use
Indicated for trained dental professionals to perform removal of carious material, tooth structure reduction, cavity/crown/root canal preparations, filling removal, and finishing/polishing. Models SG200 and SG200L are additionally indicated for implant surgery (bone perforation, tapping, threading).
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 25, 2020
Micro-NX Co., Ltd. % Kyungyoon Kang CEO K-Bio Solutions 589 Oakwood Drive Santa Clara, California 95054
Re: K192809
Trade/Device Name: Dental Handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I Product Code: EFA Dated: May 20, 2020 Received: May 27, 2020
Dear Kyungyoon Kang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Srinivas "Nandu" Nandkumar, Ph. D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192809
Device Name Dental Handpiece
Indications for Use (Describe)
Dental Handpiece with models of SG200, SG200L, SA100L, and CA100L, and CA100R are intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, root canal preparations, removal of fillings, processing and finishing tooth preparations for polishing teeth.
In addition, Dental Handpiece with models of SG200, SG200L are also intended for implant surgery such as perforating the bone, tapping and threading procedures.
Dental Handpiece is designed for use by a trained professional in the field of general dentistry.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K192809
### 510(k) SUMMARY
Pursuant to Section 510(k) of Chapter V of the Federal Food, Drug, and Cosmetic Act and in accordance with subpart E of Part 807, Title 21 of the Code of Federal Regulations, MICRO-NX Co., Ltd. submits the following information as premarket notification for the proposed device, Dental Handpiece.
### I . SUBMITTER
510(k) Correspondent: Kyungyoon Kang CEO, K-Bio Solutions (Kyungyoon.kang@kbiotechsolutions.com) Tel: 82-2-597-2700 (US: 812-345-7485)
Sponsor: MICRO-NX Co., Ltd. 22 Maeyeo-ro 1 gil, Dong-gu, Daegu Postal code: 41059, Republic of Korea Tel: 82-053-650-1000 Fax: 82-053-650-1005
510(k) Number: K192809 Manufacturer Contact: Sojeong Park (sojeong@micronx.co.kr) Date Prepared: August 25th, 2020
### II . SUBMISSION DEVICE
- Trade Name: Dental Handpiece ●
- Common Name: Handpiece, dental
- Classification Name: Dental handpiece and accessories ●
- Regulation Number: 21 CFR 872.4200 ●
- Regulation Name: Dental handpiece and accessories
- Product Code: EFA (handpiece, belt and/or gear driven, dental) ●
- Review Panel: Dental ●
- . Regulatory Class: Class I
Traditional 510(k) Registration (There were no prior submissions for the proposed, Micro-NX Dental Handpiece, which is labeled for 5 year shelf life)
### II. PREDICATE DEVICES
- SMARTmatic (K163239), Manufacturer: Kaltenbach & Voigt GmbH ●
- Anthogyr Contra Angles and Handpieces (K093894). Manufacturer: Anthogyr ●
These predicate devices have not been subject to a design-related recall.
{5}------------------------------------------------
## IV. DEVICE DESCRIPTION
Dental Handpiece with models of SG200, SG200L, SA100L, SA100R, CA100L, and CA100R are intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, root canal preparations, removal of fillings, processing and finishing tooth preparations, and restorations for polishing teeth.
In addition, Dental Handpiece with models of SG200, SG200L are also intended for implant surgery such as perforating the bone, tapping and threading procedures.
Dental Handpiece is designed for use by a trained professional in the field of general dentistry.
The proposed dental handpiece consists of the main body of a handpiece, which is used for transmitting a rotational force for the general dental treatments as well as accessory components. which are used for channeling water and air for further applications to the dental treatments. The torque transmitted from an electric micro-motor gets further transmitted to the joint part of Dental Handpiece. The rotational force is then decelerated according to the speed reducer of the gear to be transmitted to the head of the dental handpiece. The rotational force is transmitted under the permitted rotation mechanism.
The product is offered as an electrical-powered Dental Handpiece that is reusable and ergonomically shaped. The dental handpiece can be sterilized by the steam autoclave method. Through the tube and the electrical motor connected to a dental unit, the Dental Handpiece equipped with a handpiece connection according to ISO 3964 (Dentistry - Coupling dimensions for handpiece connectors - ISO 3964:1982) receive the energy for the gear, the cooling water and air for cutting treatment. Dental burs and other attachments will be used with the proposed Dental Handpiece per ISO 1797:2017 (Dentistry - Shanks for rotary and oscillating instruments).
Dental Handpiece has been developed with a full range of surgical contra angle intended to be used in implantology. The contra angle-design, size and performance of the proposed Dental Handpiece conform to ISO 14457 (2017) "Dentistry -Handpieces and motors"
The six (6) product codes and models of Dental Handpiece are SG200L, SG200, CA100L, CA100R. SA100L. SA100R.The component names and functional descriptions for critical components of Dental Handpiece are provided in the table below.
| Table IV-1. Component Names and Functional Descriptions for Critical Components of | |
|---------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dental Handpiece: SG200L, SG200, CA100L, CA100R | |
| Component Name | Function |
| Components which are commonly included in the models of SG200L, SG200, CA100L, CA100R | |
| Spindle Assembly | Receives torque from Middle Gear Assembly and transmits torque to final<br>dental bur |
| Middle Gear<br>Assembly | Transmitting the rotational force from being Housing Assembly also<br>delivering the rotational force to the Spindle Assembly |
| Button Assembly | Button for removal and attachment of the hand piece head |
| Head Cap | Case that wraps the Middle Gear Assembly and Housing<br>Assembly, and the Button Assembly is assembled |
| Head<br>Assembly | The component that is assembled Spindle Assembly, Middle Gear Assembly,<br>Button Assembly, Head Assembly |
| Housing<br>Assembly | Transmits the torque received from Joint Assembly to Middle Gear Assembly |
| Joint<br>Assembly | Transmits the torque received from the motor to the Housing Assembly |
| Inner Handle | Case to fix Joint Assembly |
| Planetary Gear<br>Assembly | The component that is assembled Housing Assembly, Joint Assembly, Inner<br>Handle |
| Spray Adapter | Adapter to help oil fill the handpiece |
| Components only available for SG200L | |
| Optic | Optical fiber that allows light to pass from the motor's LED light source to the<br>handpiece head |
| Angle Handle | The case of Planetary Gear Assembly, Optic, Head<br>Assembly |
| Angle Body<br>Assembly | The case is installed in Handle Assembly and has two parts where the light<br>passing through the optical fiber comes out and the part where the Pipe Clip<br>can be fixed |
| Handle Assembly | The component that is assembled Optic, Angle Handle |
| Pipe Clip | Clip to fix Outside Pipe mounted on specific groove of Angle Body Assembly |
| Y-coupling | Supply water from the external water supply line to the spray nozzle and<br>outside pipe |
| Tube Clip | Fix the external water line to the handpiece |
| Inside Pipe | Pipes attached to the handpiece head to supply water |
| Outside Pipe | Pipes fixed to Pipe Clip pipe to supply water |
| Components only available for SG200 | |
| Angle Handle | The case of Planetary Gear Assembly, Optic, Head Assembly |
| Angle Body Assembly | The case is installed in Handle Assembly and has two parts where the light passing through the optical fiber comes out and the part where the pipe clip can be fixed |
| Spray Nozzle | Pipe for water supply mounted on the handpiece head |
| Pipe Clip | Clip to fix Outside Pipe mounted on specific groove of Angle Body Assembly |
| Y-coupling | Supply water from the external water supply line to the spray nozzle and outside pipe |
| Tube Clip | Fix the external water line to the handpiece |
| Outside Pipe | Pipes fixed to Pipe Clip pipe to supply water |
| Components only available for CA100L | |
| Handle Assembly | The component that is assembled Optic, Angle Handle |
| Angle Body Assembly | The case is installed in Handle Assembly and has two parts where the light passing through the optical fiber comes out and the part where the pipe clip can be fixed |
| Angle Handle | The case of Planetary Gear Assembly, Optic, Head Assembly |
| Optic | Optical fiber that allows light to pass from the motor's LED light source to the handpiece head |
| Irrigation Pipe Assembly | Pipes for water supply |
| Components only available for CA100R | |
| Handle Assembly | The component that is assembled Optic, Angle Handle |
| Angle Body<br>Assembly | The case is installed in Handle Assembly and has two parts where the light<br>passing through the optical fiber comes out and the part where the pipe clip can<br>be fixed |
| Angle Handle | The case of Planetary Gear Assembly, Optic, Head<br>Assembly |
| Irrigation Pipe<br>Assembly | Pipes for water supply |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Table IV-2. Component Names and Functional Descriptions for Critical Components of Dental Handpiece: SA100L, SA100R
| Components which are commonly included in the models of SA100L, SA100R | |
|------------------------------------------------------------------------|---------------------------------------------------------------------|
| Spindle Assembly | Receives torque from motor and transmits torque to final dental bur |
| Cam | A Component to fix the bottom of Spindle Assembly |
| Inner Handle | A component that is assembled Spindle Assembly |
| Nose Cap | An outer case assembled the front part of Cam Handle Assembly |
| Cam Handle<br>Assembly | An outer case that is assembled Nose Cap, Angle Handle. |
| Angle Handle | An outer case assembled the back part of Cam Handle Assembly |
| Irrigation Pipe<br>Assembly | Pipes for water supply |
| Handle Fixing Bolt | Assembled and fastened at the bottom of the handpiece |
| Spray Adapter | Adapter to help oil fill the handpiece |
| Components only available for SA100L | |
{9}------------------------------------------------
| Optic | Optical fiber that allows light to pass from the motor's LED light source to the<br>handpiece head |
|-------|----------------------------------------------------------------------------------------------------|
|-------|----------------------------------------------------------------------------------------------------|
### V. INDICATIONS FOR USE/INTENDED USE
Dental Handpiece with models of SG200, SG200L, SA100L, SA100R, CA100L, and CA100R are intended for the removal of carious material, reducing of hard tooth structure, cavity and crown preparations, root canal preparations, removal of fillings, processing and finishing tooth preparations, and restorations for polishing teeth.
In addition, Dental Handpiece with models of SG200, SG200L are also intended for implant surgery such as perforating the bone, tapping and threading procedures.
Dental Handpiece is designed for use by a trained professional in the field of general dentistry.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Fundamental technological characteristics of MICRO-NX dental handpiece are substantially equivalent to the predicate devices, SMARTmatic and ANTHOGYR Contra angles and Handpieces as demonstrated in Table VI-1 below.
| Comparison<br>Category | Proposed Device:<br>Dental Handpiece | Predicate<br>Device:<br>SMARTmatic<br>(K163239) | Predicate<br>Device:<br>ANTHOGYR<br>Contra angles and<br>Handpieces<br>(K093894) | Substantial<br>Equivalence<br>Assessments |
|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>/Indications<br>for Use | Dental Handpiece<br>with models of<br>SG200, SG200L,<br>SA100L, SA100R,<br>CA100L, and<br>CA100R are intended<br>for the removal of<br>carious material,<br>reducing of hard tooth<br>structure, cavity and<br>crown preparations,<br>root canal<br>preparations, removal<br>of fillings, processing<br>and finishing tooth<br>preparations and | The SMARTmatic<br>handpieces are<br>intended for the<br>removal of carious<br>material, reducing<br>of hard tooth<br>structure, cavity<br>and crown<br>preparations, root<br>canal preparations,<br>removal of fillings,<br>processing and<br>finishing tooth<br>preparations,<br>restorations, and<br>for polishing teeth | ANTH-OGYR'S<br>fully autoclavable<br>contra-angles<br>Implantology<br>Impulsion are<br>devices intended<br>for a wide<br>range of dental<br>procedures<br>including:<br>Implant surgery<br>such as perforating<br>the bone, tapping<br>and<br>threading<br>procedures | Intended use and<br>indications for use<br>are the same<br>between Dental<br>Handpiece and the<br>predicate device<br>of SMARTmatic<br>handpieces, for the<br>removal of carious<br>material, reducing<br>of hard tooth<br>structure, cavity<br>and crown<br>preparations, root<br>canal preparations,<br>removal of |
| | restorations for<br>polishing teeth.<br><br>In addition, Dental<br>Handpiece with<br>models of SG200,<br>SG200L are also<br>intended for implant<br>surgery such as<br>perforating the bone,<br>tapping and<br>threading procedures.<br><br>Dental Handpiece is<br>designed for use by a<br>trained professional<br>in the field of general<br>dentistry. | They are designed<br>for use by a trained<br>professional in the<br>field of general<br>dentistry. | fillings,<br>processing and<br>finishing tooth<br>preparations, and<br>restorations for<br>polishing teeth.<br>The intended use<br>of the predicate,<br>ANTHOGYR for<br>implant surgery<br>such as<br>perforating the<br>bone, tapping and<br>threading<br>procedures is the<br>same as the<br>intended use of<br>Micro-NX dental<br>handpiece for<br>models of SG200,<br>SG200L. | |
| Functional<br>Principle | Through the micro<br>motor connected to<br>the dental treatment<br>unit, the straight and<br>contra-angle<br>handpieces equipped<br>with a handpiece<br>connection according<br>to ISO 3964<br>(Dentistry - Coupling<br>dimensions for<br>handpiece connectors<br>- ISO 3964:1982)<br>receive the energy,<br>the cooling water and<br>air for treatment and<br>the light for<br>illumination the<br>operating area. | Through the micro<br>motor connected to<br>the dental<br>treatment unit, the<br>straight and<br>contra-angle<br>handpieces<br>equipped with a<br>handpiece<br>connection<br>according to ISO<br>3964 (Dentistry -<br>Coupling<br>dimensions for<br>handpiece<br>connectors - ISO<br>3964:1982) receive<br>the energy, the<br>cooling water and<br>air for treatment<br>and the light for<br>illumination the<br>operating area. | Through the micro<br>motor connected to<br>the dental<br>treatment unit, the<br>straight and<br>contra-angle<br>handpieces<br>equipped with a<br>handpiece<br>connection<br>according to ISO<br>3964 (Dentistry -<br>Coupling<br>dimensions for<br>handpiece<br>connectors - ISO<br>3964:1982) receive<br>the energy, the<br>cooling water and<br>air for treatment<br>and the light for<br>illumination the<br>operating area. | Same Functional<br>Principle |
| Dimensions | Head size-<br>Height: Up to 14mm<br>Head size-Diameter:<br>Up to 9mm<br>Length: Up to<br>95.80mm | Head size-Height:<br>Up to 13.6 mm<br>Head size-<br>Diameter: Up to<br>9.8 mm<br>Length: Up to 93.4<br>mm | Head size-Height:<br>Up to 13.02mm<br>Head size-<br>Diameter: Up to<br>10.07 mm<br>Length: Up to<br>104.84mm | The head size,<br>height, and<br>diameter of<br>MICRO-NX<br>Dental Handpiece<br>are broadly in<br>alignment with the |
| | | | | dimensions of the |
| | | | | predicate, |
| | | | | SMARTmatic |
| | | | | handpiece. The |
| | | | | minute differences |
| | | | | in the identified |
| | | | | dimensions do not |
| | | | | raise different |
| | | | | questions of safety |
| | | | | and effectiveness. |
| Direct<br>patient-<br>contacting<br>portions of<br>the device | All materials for<br>MICRO-NX Dental<br>Handpiece models are<br>listed in the raw<br>material table below<br>including chemical<br>composition of the<br>waterlines and the<br>patient-contacting<br>portions of the<br>device.<br>Biocompatible<br>according to ISO<br>10993-1 (Biological<br>evaluation of medical<br>devices - Part 1:<br>Evaluation and<br>testing within a risk<br>management system -<br>ISO 10993-1:2009) | All materials for<br>the SMARTmatic<br>models are listed<br>in the tables below<br>including chemical<br>composition of the<br>waterlines and the<br>patient-contacting<br>portions of the<br>device.<br>Biocompatible<br>according to ISO<br>10993-1<br>(Biological<br>evaluation of<br>medical devices -<br>Part 1: Evaluation<br>and testing within<br>a risk management<br>system - ISO<br>10993-1:2009) | Biocompatible<br>according to ISO<br>10993-1<br>(Biological<br>evaluation of<br>medical devices -<br>Part 1: Evaluation<br>and testing within<br>a risk management<br>system - ISO<br>10993-1:2009) | All patient<br>contacting<br>materials of<br>MICRO-NX<br>Handpiece have<br>been tested per the<br>same standards of<br>ISO 10993-1.<br>Given the<br>favorable<br>biocompatibility<br>test results, the<br>differences in<br>patient contacting<br>materials do not<br>raise different<br>questions in terms<br>of safety and<br>effectiveness. |
| Indirect<br>patient-<br>contacting<br>portions of<br>the device<br>(water / air<br>lines) | All materials for<br>MICRO-NX Dental<br>Handpiece models are<br>listed in the tables<br>below including<br>chemical composition<br>of the waterlines and<br>the patient-contacting<br>portions of the<br>device.<br>Biocompatible<br>according to ISO<br>10993-1 (Biological<br>evaluation of medical<br>devices - Part 1:<br>Evaluation and<br>testing within a risk<br>management system -<br>ISO 10993-1:2009) | All materials for<br>the SMARTmatic<br>models are listed<br>in the tables below<br>including chemical<br>composition of the<br>waterlines and the<br>patient-contacting<br>portions of the<br>device.<br>Biocompatible<br>according to ISO<br>10993-1<br>(Biological<br>evaluation of<br>medical devices -<br>Part 1: Evaluation<br>and testing within<br>a risk management<br>system - ISO<br>10993-1:2009) | Biocompatible<br>according to ISO<br>10993-1<br>(Biological<br>evaluation of<br>medical devices -<br>Part 1: Evaluation<br>and testing within<br>a risk management<br>system - ISO<br>10993-1:2009) | All in-direct<br>patient contacting<br>materials of<br>MICRO-NX<br>Handpiece have<br>been tested per the<br>same standards of<br>ISO 10993-1.<br>Given the<br>favorable<br>biocompatibility<br>test results, the<br>differences in<br>patient contacting<br>materials do not<br>raise different<br>questions of safety<br>and effectiveness. |
| | | system - ISO<br>10993-1:2009) | | |
| Chuck<br>Design | Push Button, Latch<br>Type | Push Button,<br>Twist-tension<br>Chuck, Snap-on &<br>Screw-in | Push Button, Latch<br>Type | The chuck design<br>of MICRO-NX<br>Dental Handpiece<br>is the same as the<br>predicate,<br>ANTHOGYR<br>(K093894). |
| Speed Range<br>(RPM's) | Up to 40,000 rpm | Up to 40,000 rpm | Up to 40,000 rpm | Same Speed<br>Range |
| Conformance<br>Standards<br>(Handpieces<br>and Motors) | ISO 14457 (Dentistry<br>- Handpieces and<br>motors - ISO<br>14457:2012) | ISO 14457<br>(Dentistry -<br>Handpieces and<br>motors - ISO<br>14457:2012) | ISO 14457<br>(Dentistry -<br>Handpieces and<br>motors - ISO<br>14457:2012) | Compliant with<br>the same standards<br>for dentistry<br>handpiece and<br>motors. |
| Conformance<br>Standards<br>(Shanks) | ISO 1797 (2017)<br>"Dentistry - Shanks<br>for rotary and<br>oscillating<br>instruments" | ISO 1797-1 (2011)<br>"Dentistry -<br>Shanks for rotary<br>instruments - Part<br>1: Shanks made of<br>metals" | ISO1797-1(1995)<br>"Dental rotatory<br>instruments -<br>Shanks - Par 1:<br>Shanks<br>made of metal" | Compliant with<br>the same standards<br>of dentistry<br>shanks. |
| Conformance<br>Standards<br>(Coupling<br>Dimensions) | ISO 3964 (Dentistry -<br>Coupling dimensions<br>for handpiece<br>connectors - ISO<br>3964:1982) | ISO 3964<br>(Dentistry -<br>Coupling<br>dimensions for<br>handpiece<br>connectors - ISO<br>3964:1982) | ISO 3964<br>(Dentistry -<br>Coupling<br>dimensions for<br>handpiece<br>connectors - ISO<br>3964:1982) | Compliant with<br>the same standards<br>of dentistry<br>coupling<br>dimensions. |
| Sterilization | Sterilisable according<br>to ISO 17665-1<br>(Sterilization of<br>health care products -<br>Moist heat - Part 1:<br>Requirements for the<br>development,<br>validation and routine<br>control of a<br>sterilization process<br>for medical devices | Sterilisable<br>according to ISO<br>17665-1<br>(Sterilization of<br>health care<br>products - Moist<br>heat - Part 1:<br>Requirements for<br>the development,<br>validation and<br>routine control of a | Sterilisable<br>according to ISO<br>17665-1<br>(Sterilization of<br>health care<br>products - Moist<br>heat - Part 1:<br>Requirements for<br>the development,<br>validation and<br>routine control of a | Conform to the<br>same standards for<br>user sterilization. |
| | on the final, finished device - ISO 176651:2006) | sterilization process for medical devices on the final, finished device - ISO 176651:2006) | sterilization process for medical devices on the final, finished device - ISO 176651:2006) | |
| Air / water ports | Internal/ External Spray | No Spray / External Spray | External Spray | Internal spray function added to provide an additional option to the users. Given the favorable test results of the internal spray components, the addition does not raise different questions of safety and effectiveness. |
| Gear Ratio | 20:1/ 1:1 | 8:1/32:1/1:1 | 20:1 | Compared to the predicate device of SMARTmatic device, the addition of a gear ratio of 20:1 does not raise different questions of safety and effectiveness given the favorable test results of performance testing of MICRO-NX Dental Handpiece. |
Table VI-1. Substantial Equivalence Assessment between MICRO-NX Dental Handpiece and Predicate Devices
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
Overall, design verification testing was performed to assess the following critical elements of medical devices used in dentistry applications. The favorable test results of the proposed device which confirmed to meet the FDA recognized standards as well as ISO standard demonstrate Dental Handpiece retains the substantially equivalent profile as the predicate devices of SMARTmatic (K163239) and ANTHOGYR Contra angles and Handpieces (K093894).
- ISO 3964 (2016) "Dental Handpieces – Coupling dimensions" (Recognition List Number: 003 Effective Date: 05/03/1999)
- ISO 14457 (2017)"Dentistry Handpieces and motors" (Recognition List Number: 031 . Effective Date: 09/15/2012)
- ISO 1797 (2017) "Dentistry Shanks for rotary and oscillating instruments" ●
{14}------------------------------------------------
## VII. DESIGN VERIFICATION & PERFORMANCE DATA
The following design verification and performance data are provided in support of the conclusive determination that the proposed dental handpiece is substantially equivalent to the predicate devices of SMARTmatic (K163239), ANTHOGYR Contra angles and Handpieces (K093894).
#### Biocompatibility Testing
In accordance with ISO 10993-1: 2009, Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process, the proposed Dental Handpiece is categorized as a device appropriate for biocompatibility testing required for devices with Tissue/bone/dentin contact with limited <24 hour contact duration. This is the same testing classification of the biocompatibility evaluation as the predicates SMARTmatic(K163239) and ANTHOGYR Contra angles and Handpieces (K093894)
The biocompatibility testing of the Dental Handpiece listed below has been conducted in 2019 in order to ensure FDA's latest consensus standards with respect to biocompatibility evaluations are met for the proposed device. The favorable biocompatibility test results drawn in 2019 testing confirms the biocompatibility profile of the Dental Handpiece.
- ISO MEM Elution Using L-929 Mouse Fibroblast Cells (GLP) ●
- ISO Guinea Pig Maximization Sensitization Test (GLP - 2 Extracts)
- ISO Intracutaneous Irritation Test (GLP 2 Extracts) ●
- ISO Materials Mediated Rabbit Pyrogen (GLP)
- ISO Acute Systemic Injection Test (GLP - 2 Extracts)
### Design Verification and Validation Testing
Design Verification and Validation testing were performed to verify that the proposed Dental Handpiece meets the pre-determined design requirements and demonstrated design input matched with design outputs. Risk management assessment was conducted and appropriate risk mitigation measures were implemented including verifications of the effectiveness of the implemented risk control measures to mitigate the risks identified within the risk management process (per ISO 14971:2012 Medical Devices - Application of Risk Management to Medical Devices).
- Rotational Speed Testing ●
- Noise Testing
- Water Supplying Testing
- Light Testing
- Operation Testing
- . Appearance Testing
The favorable results of the biocompatibility, design verification performance, and validation testing demonstrate conformance of the proposed Dental Handpiece to the applicable, recognized standards of FDA. The testing results further demonstrate the substantially equivalent profile as
{15}------------------------------------------------
the predicate device, the proposed device are subject to the same applicable test standards under FDA's recognized standards including ISO 10993-1 (2009).
### VII. CONCLUSIONS
Overall, the proposed device is assessed substantially equivalent to the predicate devices given the fact that its indications for use are the same, and fundamental technological characteristics are equivalent to those of the predicates of SMARTmatic (K163239) and ANTHOGYR Contra angles and Handpieces (K093894). The favorable results of the aforementioned design verification and performance testing demonstrate conformance to the appropriate standards pertinent to dental handpieces, and demonstrate substantial equivalence to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.