The General Cutting Contra Handpiece is powered by either an air-motor or electronic micromotor for use in general dentistry. The device is intended for cutting and grinding teeth, cavity preparations, tooth and crown preparations, finishing and trimming teeth and filling materials and removal of crowns and filling materials.
Device Story
General Cutting Contra Handpiece is a dental handpiece designed for cutting, grinding, and finishing teeth and restorative materials; powered by air-motor or electronic micromotor; connects via ISO 3964 coupling; accepts friction grip or contra-angle burs (ISO 1797-1). Available in optic and non-optic models with integrated water spray. Used in dental clinics by dentists/dental professionals. Device operates by transmitting motor power to a rotating bur for mechanical tooth/material removal. Benefits include precise tooth structure reduction and restorative material removal. Provided non-sterile; requires steam sterilization (moist heat) before first use and between patients. Healthcare providers use the device to perform restorative procedures; output is the mechanical action of the bur on the tooth surface.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through functional testing against internal specifications and international standards, including ISO 14457 for electrical safety, ISO 17665-1 for steam sterilization, and ISO 10993-1 for biocompatibility.
Technological Characteristics
Dental handpiece with contra-angle design; ISO 3964 coupling; ISO 1797-1 bur compatibility; max motor speed 40,000 min-1; gear ratios ranging from 16:1 reduction to 1:5 increasing; optic/non-optic options; water spray feature; steam sterilization (moist heat).
Indications for Use
Indicated for use in general dentistry for patients requiring tooth/crown preparation, cavity preparation, finishing/trimming of teeth and filling materials, and removal of crowns and filling materials.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
Predicate Devices
E-Type Speed Increaser Contra Angle, ES-95 (K972569)
{1}------------------------------------------------
| Submitter | NAKANISHI INC.<br>700 Shimohinata<br>Kanuma-Shi, Tochigi-Ken Japan 322-8666 | | |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-----------------------------------------------|
| Contact<br>Person | Mr. Kimihiko Satoh<br>General Manager, Regulatory Affairs, R&D Division<br>TEL: +81-289-64-3380<br>FAX: +81-289-62-6665<br>k-satoh@nsk-nakanishi.co.jp | | |
| Date Prepared | April 15, 2019 | | |
| Trade Name | General Cutting Contra Handpiece | | |
| Common<br>Name | Handpiece, Contra-And Right-Angle Attachment, Dental | | |
| Classification<br>Name | Dental Handpiece and Accessories | | |
| Product Code | EGS | Class I | 872.4200 |
| Predicate<br>Device | K972569 | NAKANISHI INC. | E-Type Speed Increaser<br>Contra Angle, ES-95 |
| Reference<br>Devices | K113655 | NAKANISHI, INC. | Ti-Max X Turbine |
| | K073652 | NAKAMURA | ND LOWSPEED<br>AIRMOTOR |
| Device<br>Description | The General Cutting Contra Handpiece features an allowable maximum motor<br>speed at $40,000 min^{-1}$ . The handpieces are provided optic and non-optic and are<br>designed with a contra-angle appearance. Models are available to be connected<br>directly onto motors compliant to ISO 3964. Available handpieces include the<br>S-Max M Contra Series, Ti-Max X Contra Series, Ti-Max Z Contra Series and<br>the Ti-Max nano Series. | | |
| Indications for<br>Use | The General Cutting Contra Handpiece is powered by either an air-motor or<br>electronic micromotor for use in general dentistry. The device is intended for<br>cutting and grinding teeth, cavity preparations, tooth and crown preparations,<br>finishing and trimming teeth and filling materials and removal of crowns and<br>filling materials. | | |
| Summary of<br>Technological<br>Characteristics | As with the predicate devices referenced above, the General Cutting Contra<br>Handpiece is intended for dental treatment. The handpieces are powered by<br>either an air-motor or electronic micro motors compliant to ISO 14457 with<br>couplings compliant to [ISO 3964 (EN ISO 3964)]. Friction Grip Burs and<br>Contra Angle Burs that conform to ISO 1797-1 (EN ISO 1797) are to be used<br>with the General Cutting Contra Handpiece. The allowable maximum motor<br>speed rotation for the handpieces is at $40,000 min^{-1}$ . The handpieces are available | | |
Image /page/0/Picture/2 description: In the image, a hand wearing a white glove is holding a blue plastic tool. The tool has a unique design with sharp edges and points. The words "Powerful Partners" are written at the bottom of the image. The hand is holding the tool in a way that suggests it is being used for a precise task.
{2}------------------------------------------------
with fiber optic and non-optic. All models include a water spray feature. The General Cutting Contra Handpiece is provided non-sterile, so it must be sterilized prior to first use and sterilized after each subsequent use using steam sterilization (moist heat sterilization).
The proposed device shares technological characteristics with the predicate device. The proposed device also has some differences in technological characteristics from those of the predicate device. Any differences in the technological characteristics are minor and reflect market strategy and/or perceived user preferences.
| | Proposed Device:<br>NAKANISHI INC.<br>[K182999] | Predicate Device:<br>NAKANISHI INC.<br>[K972569] | Reference Device:<br>NAKANISHI INC.<br>[K113655] | Reference Device:<br>NAKAMURA<br>DENTAL<br>[K073652] | Coupling<br>Type | ISO 3964 (EN ISO<br>3964) Standard<br>Coupling | ISO 3964 Standard<br>Coupling | Nakanishi (NSK),<br>KaVo, W&H, Bien<br>Air, Sirona | ISO 3964 Standard<br>Coupling |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Trade<br>Name | General Cutting<br>Contra Handpiece | E-Type Speed<br>Increaser Contra<br>Angle, ES-95 | Ti-Max X Turbine<br>Series | ND E-TYPE<br>CONTRAANGLE | Burs | ISO 1797-1 (Type1)<br>ISO 1797-1 (Type3) | ISO-1797-1 (Type<br>3)<br>(1.6mm<br>or .0630in) | ISO 1797-1 Type 3 | EC-20L : ISO 1797-1<br>(Type1)<br>EC-30BL: ISO 1797-1<br>(Type1)<br>EG-20L : ISO 1797-1<br>(Type1)<br>EGG-20L:Unknown |
| Model<br>Numbers | S-Max M:<br>M15L, M15, M25L,<br>M25, M95L, M95<br>Ti-Max X:<br>X10L, X10, X12L,<br>X12, X15L, X15,<br>X20L, X20, X25L,<br>X25, X85, X85L, X95,<br>X95L<br>Ti-Max Z:<br>Z10L, Z15L, Z25L,<br>Z85L, Z95L<br>Ti-Max nano:<br>nano15LS, nano25LS,<br>nano95LS | ES-95 | Miniature Head:<br>X500; X500L;<br>X500K; X500KL;<br>X500SL; X500WL;<br>X500BL<br>Standard Head:<br>X600; X600L;<br>X600K; X600KL;<br>X650L; X650KL;<br>X600SL; X600WL;<br>X600BL;<br>X600WLED;<br>X600BLED<br>Torque Head:<br>X700; X700L;<br>X700K : X700KL:<br>X700SL; X700WL;<br>X700BL; | EC-20L<br>EC-30BL<br>EG-20L<br>EGG-20L | Max<br>Rotation<br>Speed<br>(motor) | $40,000min^{-1}$ max | $40,000min^{-1}$ max | Miniature Head:<br>(500): 390,000 –<br>450,000<br>Standard Head:<br>(600): 380,000 –<br>440,000<br>Torque Head:<br>(700): 300,000 –<br>360,000 | EC-20L : $20,000min^{-1}$<br>EC-<br>30BL : $30,000min^{-1}$<br>EG-20L : $20,000min^{-1}$<br>EGG-20L:Unknown |
| Indication<br>for Use | The General Cutting<br>Contra Handpiece is<br>powered by either an<br>air-motor or electronic<br>micromotor for use in<br>general dentistry. The<br>device is intended for<br>cutting and grinding<br>teeth,<br>cavity<br>preparations, tooth and<br>crown preparations,<br>finishing and trimming<br>teeth and filling<br>materials and removal<br>of crowns and filling<br>materials. | The device is<br>intended for use<br>where high speed<br>such as 200,000 rpm<br>is required in general<br>dentistry with or<br>without use of<br>coolant; such as<br>cutting a tooth for<br>cavity and/or crown<br>preparation and<br>finishing, cutting<br>and/or finishing of<br>dentures, denture<br>bases, crowns, inlays<br>and metal plates.<br>The device is driven<br>by an electrical<br>micrometer or an air<br>motor handpiece that<br>has the ISO-3964<br>coupling. | The Ti-Max X Series<br>is an air-powered<br>dental handpiece<br>intended for removing<br>carious material,<br>reducing hard tooth<br>structure, cavity<br>preparation, finishing<br>tooth preparations and<br>restorations and<br>polishing teeth. | ND e-type contra p is<br>powered by either low<br>speed air-motor or<br>electric micromotor<br>for removing carious<br>material and excess<br>filling material, cavity<br>and crown<br>preparation, root | Gear Ratio<br>Max<br>rotation<br>speed<br>(handpiece) | 16:1 Reduction-<br>$2,500min^{-1}$<br>10:1 Reduction-<br>$4,000min^{-1}$<br>4:1 Reduction-<br>$10,000min^{-1}$<br>1:1 Direct Drive-<br>$40,000min^{-1}$<br>1:5 Increasing-<br>$200,000min^{-1}$ | 1:5 Increasing-<br>$200,000min^{-1}$ | N/A | 16:1 Reduction:<br>Unknown<br>4:1 Reduction:<br>$5,000min^{-1}$<br>1:1 Direct Drive:<br>$30,000min^{-1}$ |
| Weight | 53.5g - 78.5g | 62.5 g | 36.7 – 52.5g | Unknown | Cleaners/<br>Lubricants | NSK PANA SPRAY<br>Plus (K163483) | Mineral Oil | Care3 Plus automatic<br>handpiece<br>cleaning/lubrication<br>tabletop unit<br>[K081811] | YOSHIDA<br>Lubrication Oil Spray |
| Sterilization<br>Method | Steam Autoclave<br>(Moist Heat) | Steam Autoclave<br>(Moist Heat) | Steam Autoclave<br>(Moist Heat) | Steam Autoclave<br>(Moist Heat) | | | | | |
| Sterilization<br>Standard | ANSI/AAMI/ISO<br>17665-1:2006 | ISO 17665-1 | EN ISO 17665-<br>1:2006 | Unknown | | | | | |
Image /page/1/Picture/4 description: The image shows a hand wearing a white glove holding a blue object with the text "Powerful Partners" below. The blue object appears to be a stylized logo or design element. The hand is positioned in a way that suggests it is presenting or showcasing the blue object. The background is plain, drawing attention to the hand and the blue object.
NAKANISHI INC. Headquarters : 700 Shimohihata Kanuma-shi Tochigi 322-8666, Japan www.nsk-inc.com Tokyo Office : MY Building 1-13-3 Ueno Taito-ku Tokyo 110-0005, Japan
{3}------------------------------------------------
NSK
Summary of Performance Testing
The General Cutting Contra Handpiece was developed and is produced under considerations of all applicable technical standards, internal specifications, and FDA guidance documents. The product's conformance with applicable international and internal standards was verified in the course of bench testing. Tests were performed on the handpiece including verification/validation testing to internal functional specifications which demonstrated that the device is substantially equivalent and complies with FDA requirements stated in Guidance Document "Dental Handpieces - Premarket Notification [510(k)] Submissions - Guidance for Industry and FDA Staff'. Sterilization has been validated in conformance to the FDA recognized consensus standard AAMI/ANSI/ISO 17665-1:2006; "Sterilization of Health Care Products – Moist
Image /page/2/Picture/4 description: In the image, a hand wearing a white glove is holding a blue surgical instrument. The instrument has a unique design with sharp edges and angles. The words "Powerful Partners" are visible at the bottom of the image.
Section 5: Page 3 of 4
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for NSK, a Japanese manufacturer of bearings. The logo is white and is set against a blue background. The letters "NSK" are in a bold, sans-serif font, with the top of the "N" extending over the "S" and "K".
Heat - Part 1: Requirements for the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices." Documentation was provided demonstrating that the General Cutting Contra Handpiece complies with the FDA requirements stated in Guidance for the "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling - Guidance for Industry and Food and Drug Administration Staff". Biocompatibility evaluations were selected in accordance with AAMI/ANSI/ISO 10993-1: "Biological evaluation of medical devices - Part 1: Evaluation and testing" and FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process- Guidance for Industry and Food and Drug Administration Staff". In addition, testing for electrical safety was performed in accordance with ISO 14457.
- NAKANISHI, INC. considers the General Cutting Contra Handpiece to be Conclusion substantially equivalent to the predicate devices listed above. This conclusion is based on the similarities in primary intended use, principles of operation, functional design, and established medical use.
Image /page/3/Picture/3 description: The image shows a hand wearing a white glove holding a blue surgical instrument. The instrument has a sleek, modern design with sharp edges and a pointed tip. The words "Powerful Partners" are written in a small font at the bottom of the image.
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182999
Device Name General Cutting Contra Handpiece
# Indications for Use (Describe)
The General Cutting Contra Handpiece is powered by either an air-motor or electronic micromotor for use in general dentistry. The device is intended for cutting and grinding teeth, cavity preparations, tooth and crown preparations, finishing and trimming teeth and filling materials and removal of crowns and filling materials.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
## FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 16, 2019
Nakanishi Inc. % Katie Reneson Regulatory Specialist Ken Block Consulting 800 East Campbell Road, Suite 202 Richardson, Texas 75081
Re: K182999
Trade/Device Name: General Cutting Contra Handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I Product Code: EGS Dated: March 15, 2019 Received: March 18, 2019
Dear Katie Reneson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{7}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.