Osom Ultra Plus Flu A&B Test Kit

K192719 · SEKISUI Diagnostics, LLC · PSZ · Apr 3, 2020 · Microbiology

Device Facts

Record IDK192719
Device NameOsom Ultra Plus Flu A&B Test Kit
ApplicantSEKISUI Diagnostics, LLC
Product CodePSZ · Microbiology
Decision DateApr 3, 2020
DecisionSESE
Submission TypeDual Track
Regulation21 CFR 866.3328
Device ClassClass 2

Indications for Use

The OSOM® ULTRA PLUS FLU A&B Test is an in vitro rapid diagnostic immunochromatographic assay intended for the qualitative detection of influenza type A and type B nucleoprotein antigens directly from nasal and nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. This test is not intended for the detection of influenza C viruses. A negative test result is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Performance characteristics for influenza A were established during the US 2018-2019 influenza season when A/H1N1pdm09 and influenza A/H3N2 were the predominant influenza A viruses in circulation, and the influenza B Yamagata and Victoria lineages were in co-circulation. When other influenza A or B viruses are emerging, performance characteristics may vary. If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Story

Rapid immunochromatographic assay for qualitative detection of influenza A and B nucleoprotein antigens. Input: nasal or nasopharyngeal swab specimens. Process: swab solubilized in extraction buffer; sample migrates on nitrocellulose membrane; viral antigens bind to mouse monoclonal IgG antibodies conjugated to colloidal gold; complex captured by rat anti-influenza A or mouse anti-influenza B antibodies on membrane. Output: visual pink/purple test lines (A, B, or both) and control line. Used in clinical laboratories and point-of-care (POC) settings; operated by healthcare professionals or untrained users. Results aid in rapid differential diagnosis of influenza A and B. Negative results are presumptive and require confirmatory testing. Benefits include rapid clinical decision-making for respiratory infections.

Clinical Evidence

Prospective clinical study (n=1210) plus 316 banked samples (total n=1526) conducted at 21 POC sites by untrained users. Compared to FDA-cleared molecular comparator. Influenza A sensitivity: 90.3% (95% CI: 87.0%-92.8%), specificity: 96.7% (95% CI: 95.5%-97.6%). Influenza B sensitivity: 88.0% (95% CI: 81.8%-92.3%), specificity: 99.2% (95% CI: 98.6%-99.6%). Reproducibility study confirmed performance across multiple sites and operators.

Technological Characteristics

Immunochromatographic test stick; nitrocellulose membrane; mouse monoclonal IgG antibodies conjugated to colloidal gold; rat anti-influenza A and mouse anti-influenza B capture antibodies. Extraction buffer contains phosphate buffered salt solution and 0.09% sodium azide. Visual readout. Standalone device; no software or electronic components.

Indications for Use

Indicated for qualitative detection of influenza A and B nucleoprotein antigens in nasal and nasopharyngeal swab specimens from symptomatic patients. Not for influenza C detection. Negative results are presumptive and require confirmation by culture or molecular assay.

Regulatory Classification

Identification

An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria: (i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method: (A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent. (B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent. (ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method: (A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent. (B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent. (2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies. (3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria: (i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains. (ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate. (iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by: (A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or (B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access. (4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain: (i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus. (ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by: (A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or (B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 3, 2020 Sekisui Diagnostics, LLC Nisha Li Principal Regulatory Affairs Specialist 6659 Top Gun Street San Diego, California 92121 Re: K192719 Trade/Device Name: OSOM ULTRA PLUS FLU A&B Test Regulation Number: 21 CFR 866.3328 Regulation Name: Influenza virus antigen detection test system Regulatory Class: Class II Product Code: PSZ Dated: September 24, 2019 Received: September 26, 2019 Dear Nisha Li: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR {2}------------------------------------------------ 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Steven Gitterman, M.D., Ph.D. Deputy Director Division of Microbiology Devices OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K192719 Device Name The OSOM® ULTRA PLUS FLU A&B Test ### Indications for Use (Describe) The OSOM® ULTRA PLUS FLU A&B Test is an in vitro rapid diagnostic immunochromatographic assay intended for the qualitative detection of influenza type B nucleoprotein antigens directly from nasal and nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. This test is not intended for the detection of influenza C viruses. A negative test result is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Performance characteristics for influenza A were established during the US 2018-2019 influenza season when A/ H1N1pdm09 and influenza A/H3N2 were the predominant influenza A viruses in circulation, and the influenza B Yamagata and Victoria lineages were in co-circulation. When other influenza A or B viruses are emerging, performance characteristics may vary. If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size: 14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |---------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 14px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary This 510(k) summary is being submitted in accordance with the requirements 21 CFR 807.92. The assigned 510(k) number is: K192719 ### 1. Sponsor/Applicant Name and Address | Company Name: | Sekisui Diagnostics, LLC | |-----------------|---------------------------------------------------| | Address: | 6659 Top Gun Street<br>San Diego, CA 92121 | | Telephone: | 858-777-2668 | | Contact Person: | Nisha Li, Principal Regulatory Affairs Specialist | Date Summary Prepared: 09/25/2019 ### 2. Device Name and Classification | Trade Name: | OSOM® ULTRA PLUS FLU A&B Test | |---------------------------------------|-----------------------------------------------------------------| | Regulation: | 21 CFR 866.3328 – Influenza virus antigen detection test system | | Classification of Device: | Class II | | Product Code: | PSZ | | Panel: | Microbiology (83) | | Special Conditions for Use Statement: | For prescription use | #### 3. Predicate Device Quidel Corp QuickVue® Influenza A+B Test (K180288) ### Device Description 4. ## Operating Principle The OSOM® ULTRA PLUS FLU A&B Test consists of a test stick that separately detects influenza A and B. The test procedure requires the solubilization of the nucleoproteins from a swab by mixing the swab in Extraction Buffer. The test stick is then placed in the sample mixture, which then migrates along the membrane surface. If influenza A and/or B viral antigens are present in the sample, it will form a complex with mouse monoclonal IgG antibodies to influenza A and/or B nucleoproteins conjugated to colloidal gold. The complex will then be bound by another a rat anti-influenza A and/or mouse anti-influenza B antibody coated on the nitrocellulose membrane. A pink to purple control line must appear in the control region of the stick for results to be valid. The appearance of a second and possibly a third light pink to purple {5}------------------------------------------------ line in the test line region indicates an A, B or A and B positive result. A visible control line with no test line is a negative result. # OSOM® ULTRA PLUS FLU A&B Test Kit Contents OSOM ULTRA PLUS FLU A&B Test Kit contains the following components: - 25 Test Sticks - 25 Sterile Nasal Swabs - 25 Extraction Buffer vials each containing: 0.25 mL phosphate buffered salt solution (with 0.09% sodium azide as a preservative) - 1 Influenza A Positive Control Swab (packaged with a desiccant tablet): coated with non-infectious recombinant influenza A containing 0.05% sodium azide - 1 Influenza B Positive Control Swab (packaged with a desiccant tablet): coated with non-infectious recombinant influenza B containing 0.05% sodium azide - 1 Instructions for Use (IFU) - 1 Quick Reference Guide (QRG) - 1 Workstation Note: (2 extra Test Sticks and Extraction Buffer vials have been included in the kit for External Quality Control Testing) # 5. Indications for Use The OSOM® ULTRA PLUS FLU A&B Test is an in vitro rapid diagnostic immunochromatographic assay intended for the qualitative detection of influenza type A and type B nucleoprotein antigens directly from nasal and nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection. It is intended to aid in the rapid differential diagnosis of influenza A and B viral infections. This test is not intended for the detection of influenza C viruses. A negative test result is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative test results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other patient management decisions. Performance characteristics for influenza A were established during the US 2018-2019 influenza season when A/H1N1pdm09 and influenza A/H3N2 were the predominant influenza A viruses in circulation, and the influenza B Yamagata and Victoria lineages were in co-circulation. When other influenza A or B viruses are emerging, performance characteristics may vary. If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, {6}------------------------------------------------ specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. #### Comparison to Predicate Device 6. The following table provides a comparison of the characteristics of the OSOM® ULTRA PLUS FLU A&B Test to the predicate device, the Quidel® QuickVue Influenza A+B (K180288). | Similarities | | | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Item | Predicate Device:<br>QuickVue® INFLUENZA A+B<br>Test K180288 | Proposed 510(k) Device:<br>OSOM® ULTRA PLUS FLU<br>A&B Test | | Indications for<br>Use | The QuickVue® Influenza A+B<br>Test allows for the rapid,<br>qualitative detection of influenza<br>type A and type B antigens<br>directly in nasal swab and<br>nasopharyngeal swab specimens<br>from symptomatic patients. The<br>test is intended for use as an aid<br>in the rapid differential diagnosis<br>of acute influenza type A and<br>type B viral infections. The test<br>is not intended to detect<br>influenza C antigens. A negative<br>test is presumptive, and it is<br>recommended these results be<br>confirmed by viral culture or an<br>FDA-cleared influenza A and B<br>molecular assay. Negative results<br>do not preclude influenza virus<br>infections and should not be used<br>as the sole basis for treatment or<br>other patient management<br>decisions. The test is intended<br>for professional and laboratory<br>use.<br><br>Performance characteristics for<br>influenza A were established<br>during the 2017/2018 influenza<br>seasons when influenza A/H3N2<br>and A/H1N1 pandemic were the<br>predominant influenza A viruses<br>in circulation. When other | The OSOM® ULTRA<br>PLUS FLU A&B Test is<br>an <i>in vitro</i> rapid<br>diagnostic<br>immunochromatographic<br>assay intended for the<br>qualitative detection of<br>influenza type A and type<br>B nucleoprotein antigens<br>directly from nasal and<br>nasopharyngeal swab<br>specimens from patients<br>with signs and symptoms<br>of respiratory infection.<br><br>It is intended to aid in the<br>rapid differential<br>diagnosis of influenza A<br>and B viral infections.<br>This test is not intended<br>for the detection of<br>influenza C viruses.<br><br>A negative test result is<br>presumptive, and it is<br>recommended these<br>results be confirmed by<br>viral culture or an FDA-<br>cleared influenza A and B<br>molecular assay. Negative<br>test results do not preclude<br>influenza virus infection<br>and should not be used as<br>the sole basis for | | Similarities | | | | Item | Predicate Device:<br>QuickVue® INFLUENZA A+B<br>Test K180288 | Proposed 510(k) Device:<br>OSOM® ULTRA PLUS FLU<br>A&B Test | | | influenza A viruses are<br>emerging, performance<br>characteristics may vary. | treatment or other patient<br>management decisions. | | | If infection with a novel<br>influenza A virus is suspected<br>based on current clinical and<br>epidemiological screening<br>criteria recommended by public<br>health authorities, specimens<br>should be collected with<br>appropriate infection control<br>precautions for novel virulent<br>Influenza viruses and sent to<br>state or local health department<br>for testing. Viral culture should<br>not be attempted in these cases<br>unless a BSL 3+ facility is<br>available to receive and culture<br>specimens. | Performance<br>characteristics for<br>influenza A were<br>established during the US<br>2018-2019 influenza<br>season when<br>A/H1N1pdm09 and<br>influenza A/H3N2 were<br>the predominant influenza<br>A viruses in circulation,<br>and the influenza B<br>Yamagata and Victoria<br>lineages were in co-<br>circulation. When other<br>influenza A or B viruses<br>are emerging,<br>performance<br>characteristics may vary.<br><br>If infection with a novel<br>influenza virus is<br>suspected based on<br>current clinical and<br>epidemiological screening<br>criteria recommended by<br>public health authorities,<br>specimens should be<br>collected with appropriate<br>infection control<br>precautions for novel<br>virulent influenza viruses<br>and sent to state or local<br>health department for<br>testing. Viral culture<br>should not be attempted in<br>these cases unless a BSL<br>3+ facility is available to<br>receive and culture<br>specimens. | | Assay Results | Qualitative | Same | | Similarities | | | | Item | Predicate Device:<br>QuickVue® INFLUENZA A+B<br>Test K180288 | Proposed 510(k) Device:<br>OSOM® ULTRA PLUS FLU<br>A&B Test | | Assay Targets | Influenza A and B antigens | Same | | Test Principle | Immunochromatography | Same | | Sample Types | Nasal swab; nasopharyngeal swab | Same | | Assay Antibodies | Monoclonal antibodies to influenza A and B<br>nucleoproteins | Same | | Read Results | Visual | Same | | Time to Result | 10 minutes | Same | | Intended Users and Use Locations | Clinical laboratory and point of care | Same | | Storage Temperature | Room Temperature | Same | | Assay Controls | Internal procedural control | Same | | Differences | | | | Item | Predicate Device:<br>QuickVue® INFLUENZA A+B<br>Test K180288 | Proposed 510(k) Device:<br>OSOM® ULTRA PLUS FLU<br>A&B Test | | External Controls | Test kit contains separate<br>Positive and Negative Control Swabs | Test kit contains Positive<br>Control Swabs where Influenza<br>A+ Control Swab acts as a<br>negative for influenza B antigen<br>and conversely, Influenza B+<br>Control Swab serves as negative<br>control for influenza A antigen. | | Extraction Reagents | Reagent solution (salt solution)<br>added to lyophilized buffer<br>(detergents and reducing agents) | Extraction buffer containing salt<br>solution. | {7}------------------------------------------------ {8}------------------------------------------------ ### Performance Summary 7. # Expected Values The prevalence of influenza varies year to year typically peaking in the winter months. The rate of positivity in influenza testing is impacted by many factors including specimen collection and handling, test method used, patient age, time of year, geographic location, and local disease prevalence. {9}------------------------------------------------ The overall positivity rate as determined by the OSOM ULTRA PLUS FLU A&B Test during the 2018-2019 clinical study was 33.0% for influenza A and 1.7% for influenza B. The observed results by age are presented in the tables below. | Age Group | Number of<br>Specimens | Number of Influenza A<br>Positives | Influenza A Positivity<br>Rate | |----------------------|------------------------|------------------------------------|--------------------------------| | ≤ 5 years of age | 362 | 127 | 35.1% | | 6 to 21 years of age | 479 | 211 | 44.1% | | ≥ 22 years of age | 369 | 61 | 16.5% | | Total | 1210 | 399 | 33.0% | Influenza A Positives by the OSOM ULTRA PLUS FLU A&B Test per Age Group | Influenza B Positives by the OSOM ULTRA PLUS FLU A&B Test per Age Group | | | | | | | | | | |-------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--| | | | | | | | | | | | | Age Group | Number of<br>Specimens | Number of Influenza B<br>Positives | Influenza B Positivity Rate | |----------------------|------------------------|------------------------------------|-----------------------------| | ≤ 5 years of age | 362 | 5 | 1.4% | | 6 to 21 years of age | 479 | 9 | 1.9% | | ≥ 22 years of age | 369 | 6 | 1.6% | | Total | 1210 | 20 | 1.7% | # Clinical Performance A prospective clinical study to establish the performance characteristics of the OSOM ULTRA PLUS FLU A&B Test in detecting influenza A and B antigens in nasal and nasopharyngeal swab specimens was conducted with specimens collected from January 2019 to May 2019 at 21 pointof-care (POC) sites across the United States. Testing was performed at POC sites representative of CLIA waived settings by untrained operators with no laboratory training or experience. Samples were collected from individuals with influenza-like symptoms who provided informed consent. Two (2) nasal swabs or two (2) nasopharyngeal swabs were collected from the same nostril according to standard collection methods from each subject. One (1) nasal or nasopharyngeal swab was used for immediate testing with the OSOM ULTRA PLUS FLU A&B Test per the test procedure. The other nasal or nasopharyngeal swab of the pair was eluted in 3.0 mL of viral transport media (VTM). The sample eluted in VTM was stored at 2-8°C until transport was made on ice packs to a central reference laboratory. The samples collected in VTM were tested by the reference method, an FDA-cleared molecular test and another FDA-cleared molecular test for discrepant analysis, within the allowable time frames of specimen collection per the product instructions. Nasal or nasopharyngeal swab specimens were collected from 1228 subjects enrolled in the prospective clinical study. Of those, 18 swab samples were unevaluable due to eligibility criteria. sample handling issues, or inconclusive testing results, leaving a total of 1210 prospective evaluable samples. The subject age and gender distribution for the 1210 prospective evaluable samples is presented in the table below. {10}------------------------------------------------ | Age Group | Female | Male | Total | |----------------|--------|------|-------| | ≤ 5 years | 175 | 187 | 362 | | 6 to 21 years | 261 | 218 | 479 | | 22 to 59 years | 107 | 206 | 313 | | ≥ 60 years | 19 | 37 | 56 | | Total | 562 | 648 | 1210 | Age and Gender Distribution Due to the atypically low prevalence of influenza B virus in the U.S. during the 2018-2019 influenza season, 1210 prospective samples (20 influenza B positive samples and 1190 influenza B negative samples) were supplemented with 317 banked samples collected from previous influenza seasons, for a total of 1527 samples tested by untrained users at POC sites. Of those, one (1) banked sample was unevaluable due to sample handling issues, leaving a total of 316 evaluable banked samples. The banked samples were masked as subject samples, randomized, and incorporated into the daily workflow at three (3) CLIA waived sites that participated in the prospective clinical study. A total of 1526 samples (1210 prospective samples and 316 banked samples) were included in the evaluation of the assay performance. For a total of 1526 evaluable tests performed, one (1) was invalid (1/1526) for a 0.07% invalid rate (95%CI: 0.01%-0.37%). The performance of the OSOM ULTRA PLUS FLU A&B Test compared to an FDA-cleared molecular comparator method with prospective samples and banked samples is presented in the tables below. | OSOM ULTRA PLUS FLU<br>A&B Test - Influenza A | | Comparator Method | | |-----------------------------------------------|-----------------------------|-------------------|-------| | | Positive | Negative | Total | | Positive | 362 | 37 a | 399 | | Negative | 39 b | 1088 c | 1127 | | Total | 401 | 1125 | 1526 | | Sensitivity | 90.3% (95% CI: 87.0%-92.8%) | | | | Specificity | 96.7% (95% CI: 95.5%-97.6%) | | | Influenza A Performance - Nasal and Nasopharyngeal Swab Samples ª Flu A was detected in 23/37 false positive specimens using a second FDA-cleared molecular test b Flu A was not detected in 7/39 false negative specimens using a second FDA-cleared molecular test C All banked samples were negative for influenza A [Two (2) samples did not yield valid results on the second FDA-cleared molecular test] | OSOM ULTRA PLUS FLU<br>A&B Test - Influenza B | Comparator Method | | | |-----------------------------------------------|-------------------|----------|-------| | | Positive | Negative | Total | | Positive | 132 | 11 a | 143 | Influenza B Performance - Nasal and Nasopharyngeal Swab Samples {11}------------------------------------------------ | OSOM ULTRA PLUS FLU<br>A&B Test - Influenza B | Comparator Method | | | |-----------------------------------------------|-----------------------------|----------|-------| | | Positive | Negative | Total | | Negative | 18b | 1365 | 1383 | | Total | 150 | 1376 | 1526 | | Sensitivity | 88.0% (95% CI: 81.8%-92.3%) | | | | Specificity | 99.2% (95% CI: 98.6%-99.6%) | | | a Nine (9) of the prospective samples and two (2) of the banked samples were false positive with the OSOM ULTRA PLUS FLU A&B Test. Flu B was detected in 3/11 false positive specimens using a second FDA-cleared molecular test. b Four (4) of the prospective samples and 14 of the banked samples were negative by the OSOM ULTRA PLUS FLU A&B Test. Flu B was not detected in 2/18 false negative specimens using a second FDA-cleared molecular test. # Reproducibility Studies Reproducibility of the OSOM ULTRA PLUS FLU A&B Test, when in the hands of untrained users, was evaluated in a multicenter study. Testing was performed at three (3) of the CLIA waived sites that participated in the prospective clinical study included samples with analyte levels at and below the limit of detection (LoD) for influenza A and influenza B. A panel of swabs including true negative (no virus), high negative (just below the LoD), low positive (at or near the LoD) and moderate positive (at or near 2x the LoD) for influenza A and B were coded, randomized, and masked to the operators. Samples were masked as subject samples and were presented to the intended use operators for testing throughout the course of a normal testing day. The study was conducted with two operators per site over five non-consecutive days. The OSOM ULTRA PLUS FLU A&B Test produces reproducible results when tested by multiple untrained intended users, at multiple sites, over multiple days. The study demonstrated that untrained intended users were able to accurately perform and interpret the OSOM ULTRA PLUS FLU A&B Test at and below the level of the LoD for both influenza A and influenza B. The results are presented in the table below. | Sample Category | Site 1 | Site 2 | Site 3 | Overall | |-------------------------------|---------------|--------------|--------------|---------------| | Influenza A High Negative1 | 100% (30/30) | 100% (30/30) | 100% (30/30) | 100% (90/90) | | Influenza A Low Positive | 96.7% (29/30) | 100% (30/30) | 100% (30/30) | 98.9% (89/90) | | Influenza A Moderate Positive | 100% (30/30) | 100% (30/30) | 100% (30/30) | 100% (90/90) | | Influenza B High Negative1 | 100% (30/30) | 100% (30/30) | 100% (30/30) | 100% (90/90) | | Influenza B Low Positive | 100% (30/30) | 100% (30/30) | 100% (30/30) | 100% (90/90) | | Influenza B Moderate Positive | 100% (30/30) | 100% (30/30) | 100% (30/30) | 100% (90/90) | | True Negative | 100% (30/30) | 100% (30/30) | 100% (30/30) | 100% (90/90) | ### Reproducibility Study Results - Percent Agreement with Expected Results 1 The "Expected Result" for High Negative samples is "not detected". {12}------------------------------------------------ ## Analytical Sensitivity: Limit of Detection The limit of detection (LoD) for the OSOM ULTRA PLUS FLU A&B Test was established in dilution studies performed with 2 influenza A strains and 2 influenza B strains on two lots of the OSOM ULTRA PLUS FLU A&B Test. The LoD represents the concentration of influenza virus that produces consistently positive results > 95% of the time. The approximate LoD concentrations identified for each strain tested are listed in the table below. | Influenza<br>Type | Viral Strain Tested | LoD | |-------------------|--------------------------------------------------|-------------------| | A | Influenza A/Michigan/45/15 (H1N1) | 7.1x101 TCID50/mL | | A | Influenza A/Singapore/INFIMH-16-0019/2016 (H3N2) | 2.2x105 CEID50/mL | | B | Influenza B/Colorado/6/2017 (Victoria) | 3.5x103 TCID50/mL | | B | Influenza B/Phuket/3073/13 (Yamagata) | 1.6x102 TCID50/mL | ## Analytical Reactivity A total of 28 influenza A, B, and C strains were tested with the OSOM ULTRA PLUS FLU A&B TEST A&B Test, at levels at or near the assay limit of detection (LoD). All influenza A isolates gave the expected influenza A positive and influenza B negative results, and all influenza B isolates gave the expected influenza A negative and influenza B positive results. The influenza strain isolates in the table below are listed at the lowest testing concentrations that gave the expected results. *NOTE: The influenza C strain listed below produced the expected influenza A negative and influenza B negative results and is listed at the highest concentration tested. | Influenza Strain | Concentration | Type | Sub Type | Test Result | |-------------------------|--------------------|------|----------|-------------| | A/NY/02/09 | 1.23x10¹ TCID50/mL | A | H1N1pdm | Detected | | A/Mexico/4108/09 | 7.24x10¹ TCID50/mL | A | H1N1pdm | Detected | | A/Singapore/63/04 | 1.57x10³ TCID50/mL | A | H1N1 | Detected | | A/Taiwan/42/06 | 1.15x10³ TCID50/mL | A | H1N1 | Detected | | A/NY/01/09 | 5.24x10¹ TCID50/mL | A | H1N1pdm | Detected | | A/Canada/6294/09 | 2.08x10³ TCID50/mL | A | H1N1pdm | Detected | | A/New Cal/20/99 | 1.77x10² TCID50/mL | A | H1N1 | Detected | | A/Solomon Islands/03/06 | 2.45x10¹ TCID50/mL | A | H1N1 | Detected | | A/NY/03/09 | 7.06 TCID50/mL | A | H1N1pdm | Detected | | A/Brisbane/10/07 | 7.06 TCID50/mL | A | H3N2 | Detected | | A/Victoria/361/11 | 2.94x10¹ TCID50/mL | A | H3N2 | Detected | | A/Perth/16/09 | 1.77x10¹ TCID50/mL | A | H3N2 | Detected | | A/Wisconsin/67/05 | 7.06x10¹ TCID50/mL | A | H3N2 | Detected | | A/Florida/2/2006 | 8.25x10⁴ CEID50/mL | A | H3N2 | Detected | | A/Texas/71/2007 | 3.25x10³ TCID50/mL | A | H3N2 | Detected | | A/Texas/50/2012 | 1.41x10¹ TCID50/mL | A | H3N2 | Detected | | B/Malaysia/2506/04 | 3.53x10¹ TCID50/mL | B | Victoria | Detected | | B/Florida/02/06 | 6.29x10¹ TCID50/mL | B | Victoria | Detected | | B/Massachusetts/2/12 | 3.53x10² TCID50/mL | B | Yamagata | Detected | {13}------------------------------------------------ | Influenza Strain | Concentration | Type | Sub Type | Test Result | |--------------------|-----------------------|------|-----------|---------------| | B/Wisconsin/1/10 | $1.70x10^1$ TCID50/mL | B | Yamagata | Detected | | B/Texas/6/11 | $1.81x10^2$ TCID50/mL | B | Yamagata | Detected | | B/Florida/04/06 | $1.05x10^2$ TCID50/mL | B | Yamagata | Detected | | B/Florida/07/04 | $6.14x10^1$ TCID50/mL | B | Yamagata | Detected | | B/Lee/40 | $1.77x10^1$ TCID50/mL | B | Victoria | Detected | | B/Brisbane/60/2008 | $1.41x10^1$ TCID50/mL | B | Victoria | Detected | | B/Colorado/06/2017 | $2.51x10^6$ EID50/mL | B | Victoria | Detected | | A/Anhui/1/2013 | $1.99x10^7$ EID50/mL | A | A (Avian) | Detected | | C/Taylor/1233/1947 | $2.10x10^5$ CEID50/mL | C | C | Not Detected* | ### Analytical Specificity: Cross-Reactivity and Microbial Interference The OSOM ULTRA PLUS FLU A&B Test was evaluated with 41 organisms (bacterial, viral, fungal) and human DNA, listed below. Bacterial isolates were tested at concentrations of approximately 10° colony forming units per mL (CFU/mL). Chlamydia pneumoniae was tested at a concentration at least 2.0 x 102 CFU/mL. Corynebacterium ulcerans and Streptococcus pyogenes were tested at a concentration of at least 1.0 x 10 CFU/mL. Viral isolates were tested at approximately 105 copy number per mL(CP/mL) or 10 - 108 tissue culture infectious dose 50% per mL (TCID50/mL). Human genomic DNA was diluted to a level greater than the minimum recommended concentration of 104 copies/mL in viral transport media (VTM). No cross-reactivity was observed at the concentrations tested, as all of the microorganisms and human genomic DNA produced negative results. ### Bacterial / Fungal Panel | Bordetella pertussis | Legionella pneumophila | Staphylococcus aureus MRSA | |--------------------------------|----------------------------------------|---------------------------------| | Candida albicans | Moraxella catarrhalis | Staphylococcus aureus MSSA | | Chlamydia pneumoniae | avirulent Mycobacterium tuberculosis | Staphylococcus epidermidis MRSE | | Corynebacterium ulcerans | Mycoplasma hominis | Streptococcus pneumoniae | | Escherichia coli | Mycoplasma pneumoniae | Streptococcus pyogenes | | Haemophilus influenzae | Neisseria meningitides | Streptococcus salivarius | | Klebsiella pneumoniae | Neisseria gonorrhoeae | | | Lactobacillus acidophilus Z048 | Pseudomonas aeruginosa | | | Viral Panel / DNA | | | | Adenovirus type 1 | Human herpes virus 7 (HHV7), SB strain | Metapneumovirus 9 type A1 | | Adenovirus type 7A | Parainfluenza virus 1 | Rhinovirus type 1A | | Coronavirus NL63 | Parainfluenza virus 2 | Enterovirus 68…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...