The QuickVue Influenza A+B Test allows for the rapid, qualitative detection of influenza type A and type B antigens directly in nasal swab and nasopharyngeal swab specimens from symptomatic patients. The test is intended for use as an aid in the rapid differential diagnosis of acute influenza type A and type B viral infections. The test is not intended to detect influenza C antigens. A negative test is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not be used as the sole basis for treatment or other patient management decisions. The test is intended for professional and laboratory use. Performance characteristics for influenza A were established during the 2017/2018 influenza seasons when influenza A/H3N2 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Device Story
QuickVue Influenza A+B Test is a lateral flow immunoassay for rapid, qualitative detection of influenza A and B viral nucleoproteins. Clinical specimens (nasal/nasopharyngeal swabs) are placed in an extraction reagent tube to disrupt viral particles and expose internal nucleoproteins. The test strip is then inserted into the tube; extracted antigens react with reagents on the strip. Positive results appear as a pink-to-red test line at specific locations for influenza A or B, alongside a blue procedural control line. Negative results show only the blue control line. Used in professional and laboratory settings by healthcare personnel. Results are interpreted visually within 10 minutes. The test aids in rapid differential diagnosis, allowing clinicians to make timely patient management decisions, though negative results require confirmation by culture or molecular assay due to the presumptive nature of the test.
Clinical Evidence
Prospective multi-center study (n=1178 evaluable specimens) across 5 US clinical sites. Compared modified QuickVue Influenza A+B Test to FDA-cleared molecular assay. Influenza A: PPA 81.5% (95% CI: 76.1%-86.0%), NPA 97.8% (95% CI: 96.6%-98.5%). Influenza B: PPA 80.9% (95% CI: 72.6%-87.2%), NPA 99.1% (95% CI: 98.3%-99.5%). Performance stratified by age groups (<5, 5-<18, >=18 years) showed consistent agreement. No invalid results reported for the subject device.
Technological Characteristics
Lateral flow immunoassay; qualitative visual read; utilizes extraction reagent to disrupt viral particles and expose nucleoproteins; test strip with specific locations for influenza A and B detection; blue procedural control line; 10-minute read time.
Indications for Use
Indicated for symptomatic patients requiring rapid, qualitative detection of influenza A and B viral antigens in nasal or nasopharyngeal swab specimens. Aids in differential diagnosis of acute influenza A and B infections. Not for influenza C detection. Negative results are presumptive and require confirmation by culture or molecular assay.
Regulatory Classification
Identification
An influenza virus antigen detection test system is a device intended for the qualitative detection of influenza viral antigens directly from clinical specimens in patients with signs and symptoms of respiratory infection. The test aids in the diagnosis of influenza infection and provides epidemiological information on influenza. Due to the propensity of the virus to mutate, new strains emerge over time which may potentially affect the performance of these devices. Because influenza is highly contagious and may lead to an acute respiratory tract infection causing severe illness and even death, the accuracy of these devices has serious public health implications.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) The device's sensitivity and specificity performance characteristics or positive percent agreement and negative percent agreement, for each specimen type claimed in the intended use of the device, must meet one of the following two minimum clinical performance criteria:
(i) For devices evaluated as compared to an FDA-cleared nucleic acid based-test or other currently appropriate and FDA accepted comparator method other than correctly performed viral culture method:
(A) The positive percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The negative percent agreement estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(ii) For devices evaluated as compared to correctly performed viral culture method as the comparator method:
(A) The sensitivity estimate for the device when testing for influenza A must be at the point estimate of at least 90 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 80 percent. The sensitivity estimate for the device when testing for influenza B must be at the point estimate of at least 80 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 70 percent.
(B) The specificity estimate for the device when testing for influenza A and influenza B must be at the point estimate of at least 95 percent with a lower bound of the 95 percent confidence interval that is greater than or equal to 90 percent.
(2) When performing testing to demonstrate the device meets the requirements in paragraph (b)(1) of this section, a currently appropriate and FDA accepted comparator method must be used to establish assay performance in clinical studies.
(3) Annual analytical reactivity testing of the device must be performed with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria:
(i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA-designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains.
(ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate.
(iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or
(B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
(4) If one of the actions listed at section 564(b)(1)(A)-(D) of the Federal Food, Drug, and Cosmetic Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain:
(i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus.
(ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by:
(A) Placing the results directly in the device's § 809.10(b) of this chapter compliant labeling that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or
(B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public Web site where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's Web site that discusses the device, must provide a prominently placed hyperlink to the Web page containing this information and must allow unrestricted viewing access.
{1}
1
# SPECIAL 510(K): DEVICE MODIFICATION OIR DECISION SUMMARY
510(k) Number: K180288
This 510(k) submission contains information/data on modifications made to the applicant’s own class II device requiring 510(k). The following items are present and acceptable:
1. The name and 510(k) number of the applicant’s previously cleared device.
| 510(k) Number | Device Name | Clearance Date | Primary Reason for 510(k) Submission |
| --- | --- | --- | --- |
| K031899 | QuickVue Influenza A+B Test | 9/11/2003 | Initial 510(k) – the data used for this submission was primarily from the clinical and analytical QuickVue Influenza (A/B) test studies (reference K991633). The primary difference between the two assays is that QuickVue Influenza A+B Test provides separate results for influenza A and influenza B. Whereas, QuickVue Influenza (A/B) test only provides one positive result which could be influenza A or B. Specimen types: nasal swab, nasal wash, nasal aspirate. |
| K053146 | QuickVue Influenza A+B Test | 12/14/2005 | This 510(k) included a clinical study performed with QuickVue Influenza A+B Test. Clinical data for nasal swab, nasopharyngeal swab, and frozen nasal wash specimens was submitted. |
| K092698 | QuickVue Influenza A+B Test | 9/15/2009 | Provided analytical reactivity data for A/California/04/2009 (A 2009 H1N1 Virus) |
| K131619 | QuickVue Influenza A+B Test | 6/28/2013 | Provided analytical reactivity data for the A/Anhui/1/2013 (H7N9 Virus) |
2. Applicant’s statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed instructions for use.
3. A description of the device MODIFICATION(S) in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
1) QuickVue Influenza A+B Test was originally developed and manufactured using standard manual lateral flow techniques. This manual manufacturing process can result in quality lateral flow immunoassays, but can yield product with a high degree of variability due to the process parameters not being under tight control.
Quidel embarked on improving product consistency and performance by focusing on improving the way components were processed through automation, using a proprietary process called Web manufacturing. Web manufacturing is a continuous process that holds critical processing steps in tight control. The Web system ensures that every foot of product on the reel is exposed to conditions within specification limits. Any product that fails to meet product specifications are marked and removed from the lot prior to further processing. The better control of the manufacturing process parameters via the
{2}
implementation of the Web system led to significant improvements in product performance.
2) The product instructions for use (package insert) was updated to add the new clinical performance data generated using the modified QuickVue Influenza A+B Test (see Section 6 below) and to remove the old clinical performance data generated using the QuickVue Influenza A+B Test as previously FDA-cleared in K031899 and K053146.
3) The product instructions for use (package insert) was updated to remove nasal aspirate and nasal wash specimens from the list of claimed upper respiratory specimen types, and to add the relevant up to date FDA recommended warning and limitation statements
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device.
| Item | Previously Cleared Device | Modified Device | Predicate Device |
| --- | --- | --- | --- |
| Features | QuickVue Influenza A+B Test (K031899, K053146) | QuickVue Influenza A+B Test (K180288) | Sofia Influenza A+B FIA (K162438) |
{3}
| Item | Previously Cleared Device | Modified Device | Predicate Device |
| --- | --- | --- | --- |
| Features | QuickVue Influenza A+B Test (K031899, K053146) | QuickVue Influenza A+B Test (K180288) | Sofia Influenza A+B FIA (K162438) |
| Intended Use | The QuickVue Influenza A+B Test allows for the rapid, qualitative detection of influenza type A and type B antigens directly from nasal swab, nasopharyngeal swab, nasal aspirate, and nasal wash specimens. The test is intended for use as an aid in the rapid differential diagnosis of acute influenza type A and type B viral infections. The test is not intended to detect influenza C antigens. Negative results should be confirmed by cell culture; they do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions. The test is intended for professional and laboratory use. | The QuickVue Influenza A+B Test allows for the rapid, qualitative detection of influenza type A and type B antigens directly in nasal swab and nasopharyngeal swab specimens from symptomatic patients. The test is intended for use as an aid in the rapid differential diagnosis of acute influenza type A and type B viral infections. The test is not intended to detect influenza C antigens. A negative test is presumptive, and it is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not be used as the sole basis for treatment or other patient management decisions. The test is intended for professional and laboratory use.
Performance characteristics for influenza A were established during the 2017/2018 influenza seasons when influenza A/H3N2 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. | The Sofia Influenza A+B FIA employs immunofluorescence to detect influenza A and influenza B viral nucleoprotein antigens in direct nasal swab, nasopharyngeal swab, and nasopharyngeal aspirate/wash specimens and nasopharyngeal swab and nasopharyngeal aspirate/wash specimens in transport media from symptomatic patients. This qualitative test is intended for use as an aid in the rapid differential diagnosis of acute influenza A and influenza B viral infections. The test is not intended to detect influenza C antigens. A negative test is presumptive and it is recommended these results be confirmed by viral culture or an FDA-cleared influenza A and B molecular assay. Negative results do not preclude influenza virus infections and should not be used as the sole basis for treatment or other patient management decisions. This test is intended for professional and laboratory use. The Sofia Influenza A+B FIA may be used with Sofia or Sofia 2.
Performance characteristics for influenza A and B were established during February through March 2011 when influenza viruses A/California/7/2009 (2009 H1N1), A/Perth/16/2009 (H3N2), and B/Brisbane/60/2008 (Victoria-Like) were the predominant influenza viruses in circulation according to the Morbidity and Mortality Weekly Report from the CDC entitled “Update: Influenza Activity--United States, 2010-2011 Season, and Composition of the 2011-2012 Influenza Vaccine.” Performance characteristics may vary against other emerging influenza viruses.
If infection with a novel influenza virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, samples should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture samples. |
{4}
| Item | Previously Cleared Device | Modified Device | Predicate Device |
| --- | --- | --- | --- |
| Features | QuickVue Influenza A+B Test (K031899, K053146) | QuickVue Influenza A+B Test (K180288) | Sofia Influenza A+B FIA (K162438) |
| Read Results | Visual | Visual | Reader |
| Specimen Types | Nasal swab, nasopharyngeal swab, nasal aspirate, and nasal wash | Nasal swab, nasopharyngeal swab | Direct nasal swab, nasopharyngeal swab, and nasopharyngeal aspirate/wash specimens and nasopharyngeal swab and nasopharyngeal aspirate/wash specimens in transport media |
| Read Result Time | 10 minutes | 10 minutes | 15 minutes |
| External Controls | Test kit contains Positive and Negative Control swabs | Test kit contains Positive and Negative Control swabs | Test kit contains Positive and Negative Control swabs |
| Manufacturing Method | Manual Process | Automated Web Process | Automated Web Process |
# 5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis.
The Risk Assessment process used was based on an internal Quidel Risk Management Procedure (QIN002), which follows the requirements in ISO 14971.
Using this procedure, the primary failures/risks associated with transferring QuickVue Influenza $\mathrm{A + B}$ to the Web manufacturing process, as well as the process controls were analyzed.
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied.
Based on the Risk Analysis, Quidel followed its established internal procedures when developing, validating, and transferring QuickVue Influenza A+B to the Web manufacturing process. Several component level validations were completed. The primary validation encompassed testing multiple lots of components to demonstrate that the process controls were effective and that the entire product when manufactured on Web equipment met the manufacturing specifications.
In addition to this validation, a Stability Study was conducted that gathered both real-time and accelerated data.
Furthermore, Quidel conducted a new prospective clinical study using the modified devices to demonstrate that the performance of the modified QuickVue Influenza A+B Test meets the Class II performance special control specified in 21 CFR 866.3328. (See Section 6 below).
{5}
# 6. Clinical Performance
The clinical performance of the modified device was evaluated in a prospective multi-center field clinical study during two distinct influenza seasons in the United States, February through May 2017 and October 2017 through January 2018. Performance for influenza A was established when influenza A/H3 and A/H1pdm09 were the predominant influenza A viruses in circulation in the United States.
In this prospective clinical study, the performance of the modified QuickVue Influenza A+B Test was compared to an FDA-cleared influenza A and B molecular assay. This study was conducted at five (5) clinical sites in the USA which represent CLIA waived testing environments comprised of Urgent Care, Pediatric, and General Practice offices. A total of forty-eight (48) operators from the five intended user sites participated in the study.
Two (2) nasal or two (2) nasopharyngeal swab specimens were collected from each of the 1183 patients enrolled in the study. All clinical samples were collected from symptomatic patients meeting the inclusion and exclusion criteria. Thirty-nine percent $(39\%)$ of the patients tested were $< 5$ years of age, forty-one percent $(41\%)$ $5 - < 18$ years of age, and twenty percent $(20\%) \geq 18$ years of age. Forty-eight percent $(48\%)$ were male and fifty-two percent $(52\%)$ were female.
On-site testing of one nasal swab or nasopharyngeal swab specimen in the QuickVue Influenza A+B Test was performed by CLIA waived test operators within one (1) hour of specimen collection. This swab specimen was incubated for one (1) minute with the Extraction Reagent Solution before addition of the dipstick. The other swab specimen was placed in viral transport media and stored at $2^{\circ}\mathrm{C}$ to $8^{\circ}\mathrm{C}$ prior to testing with the FDA-cleared influenza molecular assay.
Out of the 1183 nasal or nasopharyngeal swab specimens tested, there were no invalid QuickVue Influenza A+B Test results, but there were five (5) invalid comparator test results; therefore, 1178 evaluable specimens, including 924 nasopharyngeal swab specimen and 254 nasal swab specimens, were included in the performance analysis below.
Table 1: QuickVue Influenza A+B Test Nasal Swab and Nasopharyngeal Swab Positive Persent Agreement (PPA) and Negative Percent Agreement (NPA) versus the FDA-cleared Influenza A and B Molecular Assay (Comparator) (All Age Groups)
| | Comparator Results | | | |
| --- | --- | --- | --- | --- |
| QuickVue Influenza A+B Test | Flu A Positive | Flu A Negative | Total | Performance |
| Flu A Positive | 190 | 21 | 211 | PPA: 81.5%95% CI: 76.1%-86.0% |
| Flu A Negative | 43 | 924 | 967 | NPA: 97.8%95% CI: 96.6%-98.5% |
| Total | 233 | 945 | 1178 | |
| | Comparator Results | | | |
| --- | --- | --- | --- | --- |
| QuickVue Influenza A+B Test | Flu B Positive | Flu B Negative | Total | Performance |
| Flu B Positive | 89 | 10 | 99 | PPA: 80.9%95% CI: 72.6%-87.2% |
| Flu B Negative | 21 | 1058 | 1079 | NPA: 99.1%95% CI: 98.3%-99.5% |
| Total | 110 | 1068 | 1178 | |
{6}
Table 2: QuickVue Influenza A+B Nasal Sawb and Nasopharyngeal Swab Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) versus the FDA-cleared Influenza A and B Molecular Assay (Comparator) (by Age Group)
| | <5 years of age N=458 | | 5-<18 years of age N=480 | | ≥18 years of age N=240 | |
| --- | --- | --- | --- | --- | --- | --- |
| | PPA and 95% CI | NPA and 95% CI | PPA and 95% CI | NPA and 95% CI | PPA and 95% CI | NPA and 95% CI |
| Flu A | 90.0% (63/70) | 98.2% (381/388) | 80.4% (90/112) | 97.8% (360/368) | 72.5% (37/51) | 96.8% (183/189) |
| | 80.8%-95.1% | 96.3%-99.1% | 72.0%-86.7% | 95.8%-98.9% | 59.1%-82.9% | 93.2%-98.5% |
| Flu B | 84.2% (16/19) | 99.3% (436/439) | 80.3% (61/76) | 98.8% (399/404) | 80.0% (12/15) | 99.1% (223/225) |
| | 62.4%-94.5% | 98.0%-99.8% | 70.0%-87.7% | 97.1%-99.5% | 54.8%-93.0% | 96.8%-99.8% |
# 7. Conclusion
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the applicant's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The applicant has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.