DenMat Orthodontic Aligners

K192470 · Denmat Holdings, LLC · NXC · Dec 20, 2019 · Dental

Device Facts

Record IDK192470
Device NameDenMat Orthodontic Aligners
ApplicantDenmat Holdings, LLC
Product CodeNXC · Dental
Decision DateDec 20, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The DenMat Orthodontic Aligner is intended to be used under the supervision of a dentist/orthodontist after a comprehensive dental exam has been completed and the patient has been cleared for orthodontic treatment. The dentist/orthodontist takes a scan/impressions that are then sent to the lab where aligners are fabricated and sent back to the dentist to confirm fit and function and then used by a patient in the home/work/leisure environment. The aligner is to be used 20-22 hours per day for 2 weeks, then be replaced by the next aligner in the series, until prescribed tooth movement is achieved. The patient is instructed in the care of the aligner to maintain clarity of plastic and continued force on teeth and instructed to return to their dentist/orthodontist for any issues with aligner fit or pain.

Device Story

DenMat Orthodontic Aligners are a series of clear plastic trays used for orthodontic tooth movement. A dentist/orthodontist performs a comprehensive exam and takes scans/impressions. These are processed using FDA-cleared software (3Shape Ortho System) to design a treatment series. The software outputs files to a 3D printer to create resin models. These models are used to thermoform Essix Ace plastic aligners. The patient wears each aligner for 2 weeks (20-22 hours/day) in home/work/leisure environments. The aligners apply continuous gentle force to move teeth sequentially. The dentist monitors progress and fit. The device benefits patients by correcting misalignment and malocclusion.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

System consists of thermoformed Essix Ace plastic trays (K062828). Manufacturing utilizes 3D-printed resin models (tested for cytotoxicity) and vacuum forming. Software-driven design via 3Shape Ortho System (K171634/K152086). Non-sterile, prescription-only device.

Indications for Use

Indicated for the alignment of permanent teeth through orthodontic treatment of misalignment and malocclusion.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 20, 2019 DenMat Holdings, LLC % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114 Re: K192470 Trade/Device Name: DenMat Orthodontic Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: December 3, 2019 Received: December 4, 2019 Dear Prithul Bom: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K192470 Device Name DenMat Orthodontic Aligner Indications for Use (Describe) The DenMat Orthodontic Aligners are in the alignment of permanent teeth through orthodontic treatment of misalignment and malocclusion. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ #### 510(k) Summary 5 In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, DenMat Holdings LLC is hereby submitting this 510(k) summary. # Submitter [510(k) owner] DenMat Holding, LLC 1017 W. Central Avenue Lompoc. CA 93436 # Company Contact Tom Juarez Director RA/QA DenMat P: (805) 346-3780 C: (805) 345-0632 tjuarez@denmat.com # Submitted Device Information Trade Name: DenMat Orthodontic Aligner Common Name: Aligner, sequential Classification Name: Orthodontic Plastic Bracket, Dental Devices # Classification Information Classification: Class II Classification Regulation: 21 CFR 872.5470 Classification Product Code: NXC # Legally Marketed Predicate Devices The DenMat Orthodontic Aligner manufactured by DenMat Holdings, LC (DenMat) is substantially equivalent to the following device currently in commercial use: - Device Trade Name: Clear Image Aligners Specialty Appliance Works, Inc. - Address: 4905 Hammond Industrial Drive, Cumming, GA 30041 - 510(k) number: Manufacturer: ● - K183643 NXC - . Product Code: # Submitted Device Description The DenMat Orthodontic Aligner consists of a series of clear plastic trays, which fit over a patient's teeth, and move the teeth by external force a small amount daily. This material is the only patient contacting material, there are no lubricants, {4}------------------------------------------------ coatings, or colorants. The dental professional determines the ultimate movement needed, and an FDA-cleared software design tool is used to produce a treatment series of aligners to achieve this goal. The trays are produced when the software design tool outputs a file to a 3-D printer, which produces a resin model. The resin used to manufacture the model has been tested for cytotoxicity and has been shown to be non-cytotoxic. The positive mold is cleaned with isopropyl alcohol prior to vacuum forming the aligner. The resin model is then used with the FDA-cleared plastic material to thermoform the plastic aligner. The aligners, once formed, are rinsed with distilled water only, and air dried. Depending on the prescription, 10 to 15 aligners are produced for a single patient set. Each aligner is worn for 2 weeks, and the dental professional checks progress during treatment. # Intended Use The DenMat Orthodontic Aligner is intended to be used under the supervision of a dentist/orthodontist after a comprehensive dental exam has been completed and the patient has been cleared for orthodontic treatment. The dentist/orthodontist takes a scan/impressions that are then sent to the lab where aligners are fabricated and sent back to the dentist to confirm fit and function and then used by a patient in the home/work/leisure environment. The aligner is to be used 20-22 hours per day for 2 weeks, then be replaced by the next aligner in the series, until prescribed tooth movement is achieved. The patient is instructed in the care of the aligner to maintain clarity of plastic and continued force on teeth and instructed to return to their dentist/orthodontist for any issues with aligner fit or pain. ## Indications for Use: The DenMat Orthodontic Aligners are indicated for use in the alignment of permanent teeth through orthodontic treatment of misalignment and malocclusion. ## Substantial Equivalence The DenMat Orthodontic Aligner is substantially equivalent to the predicate device, in which the basic features and intended uses are essentially the same. The identical 3-D design software and plastic thermoform material are used by both manufacturers. The software and material have both been cleared by the FDA under their own 510(k)s. The DenMat Orthodontic Aligner system is substantially equivalent in design, manufacturing materials, intended use, principles of operation, and technical characteristics to the Clear Image Aligners, and raises no new issues of safety or {5}------------------------------------------------ ### effectiveness. # COMPARISION OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The DenMat Orthodontic Aligner system is substantially equivalent in design, manufacturing materials, intended use, principles of operation, and technical characteristics to the Clear Image Aligners, and raises no new issues of safety or effectiveness. The similarities and differences between the predicate and proposed aligners are: | Property or<br>Characteristic | Proposed Device<br>DenMat Holdings, LLC<br>DenMat Orthodontic<br>Aligner | Predicate Device<br>Specialty Appliance Works,<br>Inc.<br>Clear Image Aligners<br>K183643 | Comments | |---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------| | Device Classification<br>Name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket | Same | | Product Code | NXC | NXC | Same | | Classification | II | II | Same | | Indications for Use | The DenMat Orthodontic<br>Aligners are indicated for use<br>in the alignment of permanent<br>teeth through orthodontic<br>treatment of misalignment<br>and malocclusion. | The Clear Image Aligners<br>are indicated for use in the<br>alignment of permanent<br>teeth through orthodontic<br>treatment of misalignment<br>and malocclusion. | Same | | Mode of Action | Alignment of teeth by<br>application of continuous<br>gentle force, by sequential<br>use of preformed plastic<br>trays. | Alignment of teeth by<br>application of continuous<br>gentle force, by sequential<br>use of preformed plastic<br>trays. | Same | | Method of Use | Each preformed plastic tray<br>is worn by the patient as<br>prescribed by the<br>orthodontist/dentist prior to<br>using the next sequential<br>aligner tray. | Each preformed plastic tray<br>is worn by the patient as<br>prescribed by the<br>orthodontist/dentist prior to<br>using the next sequential<br>aligner tray. | Same | | Material | Essix Ace plastic<br>(K062828) | Essix Ace plastic<br>(K062828) | Same | | Prescription or OTC | Prescription | Prescription | Same | | Software Used during<br>manufacturing | Use of 3Shape Ortho<br>System (K171634/K152086) | Use of 3Shape Ortho<br>System<br>(K171634/K152086) | Same | | Provided Non-Sterile | Yes | Yes | Same | {6}------------------------------------------------ ### Summary of Substantial Equivalence: Based on the information presented in this submission, DenMat Holdings, LLC concludes that the DenMat Orthodontic Aligners are substantially equivalent to the predicate device in regard to indications for use, design and technology.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...