K183643 · Specialty Appliance Works, Inc. · NXC · Jan 3, 2019 · Dental
Device Facts
Record ID
K183643
Device Name
Clear Image Aligners
Applicant
Specialty Appliance Works, Inc.
Product Code
NXC · Dental
Decision Date
Jan 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Clear Image Aligners are indicated for use in the alignment of permanent teeth through orthodontic treatment of misalignment and malocclusion.
Device Story
Clear Image Aligners are patient-specific, removable, sequential plastic trays prescribed by orthodontists or dentists to treat tooth misalignment and malocclusion. A trained technician uses 3Shape orthodontic software to design the treatment plan based on clinician recommendations. The device applies gentle, continuous force to teeth through minor, sequential changes in tray geometry. Aligners are provided non-sterile; patients wear them as directed to gradually reposition teeth. The system relies on established orthodontic principles for sequential tooth movement.
Clinical Evidence
No clinical data. Substantial equivalence is supported by established performance of sequential aligners (product code NXC) since 1998 and biocompatibility testing per ISO 10993-1.
Technological Characteristics
Sequential thermoformed polyurethane aligners (Essix Ace plastic, K062828). Design utilizes 3Shape orthodontic software (K171634/K152086). Non-sterile, patient-specific, removable form factor. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for alignment of permanent teeth in patients requiring orthodontic treatment for misalignment and malocclusion.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
Predicate Devices
Derby Dental Laboratory Custom Clear Aligner System (K173785)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Specialty Appliance Works, Inc. % Mark Job Official Correspondent Regulatory Technology Services, LLC 1394 25th Street, Nw Buffalo, Minnesota 55313
Re: K183643
Trade/Device Name: Clear Image Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic plastic bracket Regulatory Class: Class II Product Code: NXC Dated: December 21, 2018 Received: December 26, 2018
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
January 3, 2019
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Mary Mary S. = S. Runner -S3 19.01.03 - 5.3 - Date: 2019.01.03
> Runner - 5.3 - Date: 2019.01.03
> Runner - 5.3 - 09:06:24 -05'00' For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for Specialty Appliances. The logo features a circular design composed of several smaller circles of varying sizes and shades of gray. To the right of the circular design, the word "Specialty" is written in a bold, sans-serif font. Below "Specialty", the word "Appliances" is written in a smaller, lighter font.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Form Approved: OMB No. 0910-0120 Food and Drug Administration Expiration Date: 06/30/2020 Indications for Use See PRA Statement below. 510(k) Number (if known)
Device Name Clear Image Aligners
Indications for Use (Describe)
The Clear Image Aligners are indicated for use in the alignment teeth through orthodontic treatment of misalignment and malocclusion.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information, Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
FSC Publiching Seessour (306) 443-678)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Specialty Appliances. The logo consists of a cluster of gray circles on the left, arranged in a circular pattern. To the right of the circles, the word "Specialty" is written in a bold, sans-serif font. Below "Specialty", the word "Appliances" is written in a smaller font size.
# 510(k) Summary
| Date Prepared: | November 16, 2018 |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Specialty Appliance Works, Inc.<br>4905 Hammond Industrial Drive<br>Cumming, GA 30041 |
| Contact: | Carolyn Thomas<br>Quality Manager<br>Specialty Appliance Works, Inc.<br>(678) 513-4408<br>Carolyn.Thomas@specialtyappliances.com |
| Proprietary Name: | Clear Image Aligners |
| Common Name: | Sequential Aligners |
| Classification: | 21 CFR 872.5470: Orthodontic plastic bracket; Class II |
| Product Code: | NXC |
| Predicate Device: | K173785, Derby Dental Laboratory Custom Clear Aligner<br>System |
#### Intended Use / Indications:
The Clear Image Aligners are indicated for use in the alignment of permanent teeth through orthodontic treatment of misalignment and malocclusion.
### Device Description:
The Clear Image Aligners are a sequential set of removable aligners prescribed by an orthodontist or dentist, used to gradually move a patient's teeth in order to treat misalignment and malocclusion. Each aligner in the sequential set is created by a trained technician utilizing 3Shape orthodontic software (K171634/K152086) and tooth movement recommendations from the prescribing orthodontist/dentist. Aligners are then fabricated from plastic. Corrective force to progressively reposition the teeth is delivered via minor changes from one aligner to the next within the series.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for "Specialty Appliances". The logo features a circular design on the left, composed of multiple smaller circles arranged in a pattern. To the right of the circular design, the word "Specialty" is written in a bold, sans-serif font. Below "Specialty", the word "Appliances" is written in a smaller font size.
Each aligner set is patient specific and can only be used for an individual patient for whom it is prescribed. Clear Image Aligners are provided non-sterile and are completely removable by the patient and treatment/use may be discontinued at any time.
#### Non-Clinical Testing:
Specialty Appliance Works, Inc. has completed biocompatibility testing per ISO 10993-1 and its applicable parts, as appropriate for the aligner contact and duration. The mechanical properties of the Essix Ace plastic (K062828) has been previously demonstrated by the manufacturer as appropriate for use with aligners, therefore no additional testing was required.
#### Clinical Testing:
The performance of sequential aligners in the clinical environment has been well established since the first aligners were cleared by the FDA under product code NXC in 1998. Therefore, clinical testing was not necessary to demonstrate substantial equivalence of the Clear Image Aligners to the predicate device.
| Property or<br>Characteristic | Proposed Device<br>Specialty Appliance Works,<br>Inc.<br>Clear Image Aligners | Predicate Device<br>Derby Dental Laboratory<br>Custom Clear Aligner<br>System (K173785) | Comments |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Device Classification<br>Name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket | Same as K173785 |
| Product Code | NXC | NXC | Same as K173785 |
| Classification | II | II | Same as K173785 |
| Indications for Use | The Clear Image Aligners<br>are indicated for use in the<br>alignment of permanent<br>teeth through orthodontic<br>treatment of misalignment<br>and malocclusion. | The Custom Clear Aligner<br>System is indicated for use<br>in the alignment of<br>permanent teeth through<br>orthodontic treatment of<br>misalignment and<br>malocclusion. | Same as K173785 |
| Mode of Action | Alignment of teeth by<br>application of continuous | Alignment of teeth by<br>application of continuous | Same as K173785 |
## Predicate Device Comparison:
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Specialty Appliances. The logo features a circular design composed of several smaller circles on the left side. To the right of the circular design, the word "Specialty" is written in a bold, sans-serif font. Below "Specialty", the word "Appliances" is written in a smaller, sans-serif font.
| Property or<br>Characteristic | Proposed Device<br>Specialty Appliance Works, Inc.<br>Clear Image Aligners | Predicate Device<br>Derby Dental Laboratory<br>Custom Clear Aligner System (K173785) | Comments |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Method of Use | gentle force, by sequential<br>use of preformed plastic<br>trays.<br>Each preformed plastic tray<br>is worn by the patient as<br>prescribed by the<br>orthodontist/dentist prior to<br>using the next sequential<br>aligner tray. | gentle force, by sequential<br>use of preformed plastic<br>trays.<br>Each preformed plastic tray<br>is worn by the patient as<br>prescribed by the dental<br>practitioner, usually a few<br>weeks prior to using the<br>next sequential aligner tray. | Same as K173785 |
| | Material | Essix Ace plastic<br>(K062828) | Thin thermoformed<br>polyurethane |
| Prescription or OTC | Prescription | Prescription | Same as K173785 |
| Software Used during<br>manufacturing | Use of 3Shape Ortho<br>System<br>(K171634/K152086) | Use of 3Shape Ortho<br>System (K152086) | Same as K173785 |
| Provided Non-Sterile | Yes | Yes | Same as K173785 |
### Summary of Substantial Equivalence:
Based on the information presented in this submission, Specialty Appliance Works, Inc. concludes that the Clear Image Aligners are substantially equivalent to the predicate device in regard to indications for use, design and technology.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.