K192414 · Medxl, Inc. · NGT · Feb 13, 2020 · General Hospital
Device Facts
Record ID
K192414
Device Name
Praxiject 0.9% NaCl
Applicant
Medxl, Inc.
Product Code
NGT · General Hospital
Decision Date
Feb 13, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The Praxiject™ 0.9% NaCl prefilled syringe with 0.9% Sodium Chloride Injection, USP, is intended only for flushing vascular access devices. May be placed on a sterile field.
Device Story
Praxiject 0.9% NaCl is a single-use, prefilled plastic piston syringe containing 0.9% Sodium Chloride Injection, USP. It features a Luer lock connection and is capped with a non-vented, female Luer lock tip cap. The device is terminally sterilized via gamma irradiation. It is used by healthcare providers in clinical settings to flush vascular access devices. The syringe may be placed on a sterile field. The device provides a standardized, ready-to-use saline volume to maintain catheter patency, reducing the risk of occlusion and potential infection associated with manual preparation.
Clinical Evidence
Bench testing only. Testing included visual inspection of pouch seals (ASTM F1886), bubble emission (ASTM D3078), syringe damage inspection, luer lock leakage testing, label integrity (ASTM F2250), USP monograph formulation testing (assay, pH, identification, particulate matter, bacterial endotoxins, elemental impurities, iron), and distribution cycle validation (ASTM D4169). All tests conformed to acceptance criteria.
Technological Characteristics
Single-use plastic piston syringe; barrel/plunger: polypropylene; piston: bromobutyl rubber; tip cap: ABS. Luer lock fitting. Sterilization: gamma irradiation (10^-6 SAL). Shelf life: 2 years. No software or electronic components.
Indications for Use
Indicated for flushing vascular access devices in patients requiring such maintenance.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 13, 2020
MedXL Inc. Premala Premanathan Regulatory Affairs Associate 285 Av Labrosse Pointe-Claire, H9R 1A3 Ca
Re: K192414
Trade/Device Name: Praxiject 0.9% NaCl Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: NGT Dated: January 16, 2020 Received: January 21, 2020
Dear Premala Premanathan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
K192414
Device Name Praxiject™ 0.9% NaCl
Indications for Use (Describe)
The Praxiject™ 0.9% NaCl prefilled syringe with 0.9% Sodium Chloride Injection, USP, is intended only for flushing vascular access devices. May be placed on a sterile field.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MEDXL. The letters "MED" are in a dark blue color, while the "XL" is in red. A red line extends from the "D" in MED, going through the "XL" and ending in a heartbeat-like pattern.
………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………
# 510(k) Summary K192414
February 3, 2020
| Device Trade Name: | Praxiject™ 0.9% NaCl |
|-------------------------|---------------------------------------------------|
| Common Name: | Saline Flush Syringe |
| Classification Name: | Saline, Vascular Access Flush |
| Product Code: | NGT |
| Regulation: | 21 CFR §880.5200 |
| Regulatory Class: | Class II |
| Classification Panel: | General Hospital |
| Submitter/Manufacturer: | MedXL Inc. |
| | 285 Av. Labrosse |
| | Pointe Claire, Quebec H9R 1A3, Canada |
| | Tel: (+1) 514.695.7474 Fax: (+1) 514.695.1511 |
| Regulatory Contact: | Premala Premanathan, Regulatory Affairs Associate |
Predicate Device
| Device Trade Name: | Praxiject™ 0.9% NaCl |
|--------------------|-----------------------------------------------------------|
| 510(k) Number: | K171109 |
| Classification: | Class II, 21 CFR §880.5200, Saline, Vascular Access Flush |
| Product Code: | NGT |
## Device Description
The Praxiject™ 0.9% NaCl prefilled syringe is a single use plastic piston syringe with a Luer lock connection fitting, prefilled to labeled volume with 0.9% Sodium Chloride Injection, USP, with no preservatives (normal saline), and capped with a plastic tip cap. Each prefilled syringe is individually packaged in a heat-sealed pouch and terminally sterilized by gamma irradiation.
## Device Modification
The reason for this submission is the extension to the Praxiject™ 0.9% NaCl prefilled syringe line to include an additional 10 mL prefilled syringe model individually packaged in an aluminum foil pouch (as compared to the plastic peel pouch used for the predicate device models). The subject device model is distributed in four double bags with 115 syringes per double bags (as compared to six cases with 100 syringes (10 cc) or 120 syringes (5 cc) per case of the predicate device models) within a standard size shipping carton.
## Intended Use / Indications for Use
The Praxiject™ 0.9% NaCl prefilled syringe with 0.9% Sodium Chloride Injection, USP, is intended only for flushing vascular access devices. May be placed on a sterile field.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MEDXL in blue and red. The MED is in blue, and the XL is in red. There is a red line going through the middle of the letters, and the line extends to the right of the XL in a wave pattern.
..............................................................................................................................................................................
# Technological Characteristics Comparison Table
Shown below is a side-by-side comparison of key device characteristics between the subject device and the predicate device.
| Device<br>Characteristic | Predicate Device<br>Praxiject™ 0.9% NaCl (K171109) | Subject Device<br>Praxiject™ 0.9% NaCl (K192414) | Comparison |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for Use | The Praxiject™ 0.9% NaCl prefilled<br>syringe with 0.9% Sodium Chloride<br>Injection, USP, is intended only for<br>flushing vascular access devices.<br>May be placed on a sterile field. | The Praxiject™ 0.9% NaCl prefilled<br>syringe with 0.9% Sodium Chloride<br>Injection, USP, is intended only for<br>flushing vascular access devices.<br>May be placed on a sterile field. | Identical |
| Design | Prefilled plastic piston syringe with<br>Luer lock connection fitting and non-<br>vented, female Luer lock tip cap. | Prefilled plastic piston syringe with<br>Luer lock connection fitting and non-<br>vented, female Luer lock tip cap. | Identical |
| Syringe Size and<br>Fill Volumes | 3 mL in 5 cc syringe<br>5 mL in 5 cc syringe<br>3 mL in 10 cc syringe<br>5 mL in 10 cc syringe<br>10 mL in 10 cc syringe | 3 mL in 5 cc syringe<br>5 mL in 5 cc syringe<br>3 mL in 10 cc syringe<br>5 mL in 10 cc syringe<br>10 mL in 10 cc syringe | Identical |
| Fill Volume<br>Graduations | On syringe label | On syringe label | Identical |
| Syringe Content | 0.9% Sodium Chloride Injection, USP | 0.9% Sodium Chloride Injection, USP | Identical |
| Labeled Non-<br>pyrogenic | Yes | Yes | Identical |
| Single Use Only | Yes | Yes | Identical |
| Sterile | Yes | Yes | Identical |
| Use on Sterile Field | Yes | Yes | Identical |
| Sterilization Method | Terminally sterilized by gamma<br>radiation, 10-6 SAL | Terminally sterilized by gamma<br>radiation, 10-6 SAL | Identical |
| Shelf Life | 2 years | 2 years | Identical |
| Syringe<br>Material | Barrel and Plunger: Polypropylene<br>Piston: Bromobutyl rubber<br>(Not made with natural rubber latex)<br>Tip Cap: ABS with white colorant | Barrel and Plunger: Polypropylene<br>Piston: Bromobutyl rubber<br>(Not made with natural rubber latex)<br>Tip Cap: ABS with white colorant | Identical |
| Device<br>Characteristic | Predicate Device<br>Praxiject™ 0.9% NaCl (K171109) | Subject Device<br>Praxiject™ 0.9% NaCl (K192414) | Comparison |
| Syringe Packaging | Plastic peel pouch (printed on one<br>side, clear on the other) - all sizes and<br>fill volumes | Plastic peel pouch (printed on one<br>side, clear on the other) - all sizes and<br>fill volumes<br>OR<br>Aluminum foil pouch (printed on one<br>side) – 10 mL in 10 cc syringe | Different |
| Content of Syringe<br>Package | One syringe per pouch | One syringe per pouch | Identical |
| Shipping Package<br>Configuration | 100 syringes (10 cc) or 120 syringes<br>(5cc) per case / 6 cases per shipping<br>carton – all sizes and fill volumes | 100 syringes (10 cc) or 120 syringes<br>(5cc) per case / 6 cases per shipping<br>carton – all sizes and fill volumes in<br>plastic peel pouch<br>OR<br>115 syringes per double bag /<br>4 double bags per shipping carton -<br>10 mL in 10 cc syringe in aluminum<br>foil pouch | Different |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for MEDXL. The logo is in blue and red. The letters MED are in blue, and the letters XL are in red. There is a red line that extends from the top of the E to the top of the X, and a red line that extends from the bottom of the X to the right, forming a heart rate line.
The subject device, Praxiject™ 0.9% NaCl prefilled syringe, compares identical to the predicate device in intended use, indications for use, fundamental technology, product design, operating principles, and performance characteristics. The only difference between the subject device and the predicate device is in the primary protective packaging and shipping package configuration of the additional 10 mL prefilled syringe model.
# Summary of Non-Clinical Testing
Shown below is a summary of the non-clinical testing that was performed with the subject device. The testing below was performed to demonstrate that the subject device meets the identified standards for this device type.
| Test Methodology / Standard | Purpose | Acceptance Criteria | Results |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Visual inspection of pouch seals per<br>ASTM F1886 | To verify pouch seal<br>integrity | No visible unsealed areas,<br>channels/pathway across width<br>of seal, tears/holes: Reject on<br>detection* | Conforms |
| | | Undersealed areas, oversealed<br>areas (hard/brittle seal), narrow<br>seals (thinning along length of<br>seal): non defective seal width<br>should be greater than 67% of<br>nominal seal width* | Conforms |
| Bubble emission test of pouch per<br>ASTM D3078 | To verify the integrity of<br>the product packaging<br>(pouch) | No leaks (no bubbles emitted<br>upon immersion, no fluid<br>inside the package) | Conforms |
| Test Methodology / Standard | Purpose | Acceptance Criteria | Results |
| Visual inspection of prefilled syringe<br>for damage | To ensure the prefilled<br>syringe is not damaged in<br>a way that would prevent<br>its use | No critical damage (consistent<br>with instructions for use); no<br>leaks | Conforms |
| Test for liquid leakage and resistance<br>of luer lock fitting | To ensure the prefilled<br>syringe is adequately<br>sealed | No leaks; no cracks | Conforms |
| Test for integrity of printed label per<br>ASTM F2250 | To verify the integrity of<br>the printed information is<br>adequate for the intended<br>use environment | Print must remain defined and<br>legible, color must not lighten,<br>ink must not run<br>(Slight smudging of ink or<br>transfer to swab is allowable) | Conforms |
| Test of formulation per Sodium<br>Chloride Injection USP monograph:<br>- Assay of Sodium Chloride (per<br>USP monograph)<br>- pH per USP<791><br>- Identification of Sodium and<br>Chloride per USP <191><br>- Particulate matter per USP <788><br>- Bacterial Endotoxins per USP <85><br>- Elemental Impurities (Heavy<br>Metals) per USP<232>/<233><br>(Class I elements)<br>- Iron per USP <24> | To verify formulation of<br>the Sodium Chloride<br>solution meets the USP<br>requirements | See below for individual test<br>acceptance criteria:<br>0.855 to 0.945% NaCl<br>4.5 to 7.0<br>Successful identification<br>≥ 10μm: ≤ 6000 part/syringe<br>≥ 25μm: ≤ 600 part/syringe<br>≤ 0.5 EU/mL<br>Arsenic: ≤ 1.5μg/g<br>Cadmium: ≤ 0.2μg/g<br>Mercury: ≤ 0.3μg/g<br>Lead: ≤ 0.5μg/g<br>Iron: ≤ 2ppm | Conforms<br>(See below)<br>Conforms<br>Conforms<br>Conforms<br>Conforms<br>Conforms<br>Conforms |
| Distribution cycle (Transport)<br>Validation per ASTM D4169,<br>Distribution Cycle 13 | To ensure the packaging<br>maintains product<br>integrity under anticipated<br>shipping and handling<br>conditions | Packaging and syringe integrity<br>per ISO 11607-1 | Conforms |
| Sterilization Validation | To establish the minimum<br>irradiation dose required<br>to render the product<br>sterile | 10-6 SAL per ISO 11137-2 | Conforms |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for MEDXL. The logo is in two colors, blue and red. The letters "MED" are in blue, and the letters "XL" are in red. A red line goes through the middle of the letters "EDX", and a red heartbeat line is to the right of the letters "XL".
## MedXL inc. | 285 Labrosse, Pointe-Claire (Qc), Canada, H9R1A3 T : +1 514.695.7474 | F : +1 514.695.1511 | www.medxl.com
# Conclusion
The conclusions drawn from the nonclinical testing for the subject device, Praxiject™ 0.9% NaCl prefilled syringe, demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.