K171109 · Medxl, Inc. · NGT · Dec 21, 2017 · General Hospital
Device Facts
Record ID
K171109
Device Name
Praxiject 0.9% NaCl
Applicant
Medxl, Inc.
Product Code
NGT · General Hospital
Decision Date
Dec 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The Praxiject™ 0.9% NaCl prefilled syringe with 0.9% Sodium Chloride Injection, USP, is intended only for flushing vascular access devices. May be placed on a sterile field.
Device Story
Praxiject™ 0.9% NaCl is a single-use, prefilled plastic piston syringe containing 0.9% Sodium Chloride Injection, USP. Device features Luer lock connection and non-vented tip cap. Available in various fill volumes (3 mL, 5 mL, 10 mL) within 5 cc or 10 cc syringe barrels. Terminally sterilized via gamma irradiation. Used in clinical settings by healthcare professionals to flush vascular access devices; maintains patency of IV catheters and tubing. Syringe barrel/plunger constructed of polypropylene; plunger stopper of bromobutyl rubber. Device is packaged in a plastic peel pouch. Clinical benefit includes standardized, sterile delivery of saline for catheter maintenance, reducing risk of contamination during flushing procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including USP 39-NF34 solution compliance, ISO 7886-1 syringe performance, and comprehensive biocompatibility testing (ISO 10993-1, -4, -5, -10, -11, -18). Testing confirmed non-hemolytic, non-cytotoxic, non-sensitizing, and non-irritant properties. Particulate matter and bacterial endotoxins met USP <788>, <85>, and <161> requirements. Shelf life validated per ISO 11607-1.
Technological Characteristics
Single-use plastic piston syringe; barrel/plunger: polypropylene; plunger: bromobutyl rubber; tip cap: ABS. Luer lock connection. Sterilization: gamma radiation (10-6 SAL). Standards: ISO 7886-1 (syringe performance), ISO 10993 series (biocompatibility), ISO 11607-1 (packaging), USP <788> (particulates), USP <85>/<161> (endotoxins).
Indications for Use
Indicated for flushing vascular access devices in patients requiring intravenous catheter maintenance. Suitable for use on a sterile field.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Excelsior Saline Pre-filled Syringe(s) in Sterile Field Packaging (K082837)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" above the word "ADMINISTRATION".
December 21, 2017
MedXL Inc. Premala Premanathan Regulatory Affairs Associate Contact Address
Re: K171109
Trade/Device Name: Praxiject™ 0.9% NaCl Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: NGT Dated: November 28, 2017 Received: November 30, 2017
Dear Premala Premanathan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
Praxiject™ 0.9% NaCl
Indications for Use (Describe)
The Praxiject™ 0.9% NaCl prefilled syringe with 0.9% Sodium Chloride Injection, USP, is intended only for flushing vascular access devices. May be placed on a sterile field.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-family: Arial, sans-serif;"> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the MEDXL logo in blue and red, followed by the text "510(k) Summary" in black. The MEDXL logo is a stylized version of the company name, with the "XL" portion in red and a red line extending from the "L" to the right. The text "510(k) Summary" is in a simple, sans-serif font and is positioned to the right of the logo.
December 20, 2017
| Trade Name: | Praxiject™ 0.9% NaCl |
|-------------------------|-----------------------------------------------------------------------------------------------------------|
| Common Name: | Saline Flush Syringe |
| Classification Name: | Saline, Vascular Access Flush |
| Product Code: | NGT |
| Regulation: | 21 CFR §880.5200 |
| Regulatory Class: | Class II |
| Submitter/Manufacturer: | MedXL Inc. |
| Address: | 285 Av Labrosse<br>Pointe Claire, Quebec H9R 1A3, Canada<br>Tel: (+1) 514.695.7474 Fax: (+1) 514.695.1511 |
| Contact: | Premala Premanathan, Regulatory Affairs Associate |
This 510(k) Summary has been prepared in accordance with 21 CFR §807.92.
## Predicate Device
K082837, Excelsior Saline Pre-filled Syringe(s) in Sterile Field Packaging - Excelsior Medical Corp.
## Device Description
The Praxiject™ 0.9% NaCl prefilled syringe is a single use plastic piston syringe with a Luer lock connection fitting, prefilled to labeled fill volume with 0.9% Sodium Chloride Injection, USP, with no preservatives (normal saline), and capped with a plastic tip cap. Each prefilled syringe is individually packaged in a plastic peel pouch and terminally sterilized by gamma irradiation.
### Intended Use / Indications for Use
The Praxiject™ 0.9% NaCl prefilled syringe with 0.9% Sodium Chloride Injection, USP, is intended only for flushing vascular access devices. May be placed on a sterile field.
### Comparison of Technological Characteristics
The Praxiject™ 0.9% NaCl prefilled syringe is substantially equivalent to the predicate device in fundamental technology, product design and materials, operating principles, and performance characteristics for the intended use. The differences between the two devices have been addressed by material and device component qualification, process and packaging validation, and finished product release testing in conformance with recognized consensus standards.
{4}------------------------------------------------
| Device<br>Characteristic | Predicate Device (K082837) | Praxiject™ 0.9% NaCl |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | Excelsior Sterile Field Saline Flush<br>Syringe(s) are intended for use in<br>flushing IV catheters and IV tubing. | The Praxiject™ 0.9% NaCl prefilled<br>syringe with 0.9% Sodium Chloride<br>Injection, USP, is intended only for<br>flushing vascular access devices.<br>May be placed on a sterile field. |
| Design | Prefilled plastic piston syringe with<br>Luer lock connection fitting and<br>non-vented, female Luer lock tip cap | Prefilled plastic piston syringe with<br>Luer lock connection fitting and<br>non-vented, female Luer lock tip cap. |
| Syringe Size and<br>Fill Volumes | 10 mL in 10 cc syringe | 3 mL in 5 cc syringe<br>5 mL in 5 cc syringe<br>3 mL in 10 cc syringe<br>5 mL in 10 cc syringe<br>10 mL in 10 cc syringe |
| Fill Volume<br>Graduations | On syringe label | On syringe label |
| Syringe Content | 0.9% Sodium Chloride Injection, USP | 0.9% Sodium Chloride Injection, USP |
| Labeled<br>Non-pyrogenic | Yes | Yes |
| Single Use Only | Yes | Yes |
| Sterile | Yes | Yes |
| Use on Sterile Field | Yes | Yes |
| Sterilization<br>Method | Terminally sterilized by<br>steam, 10-6 SAL | Terminally sterilized by<br>gamma radiation, 10-6 SAL |
| Shelf Life | 2 years | 2 years |
| Syringe Material | - Barrel and Plunger: Polypropylene<br>- Plunger: Isoprene rubber (Not made<br>with natural rubber latex)<br>- Tip Cap: Polypropylene with white<br>colorant | - Barrel and Plunger: Polypropylene<br>- Plunger: Bromobutyl rubber (Not<br>made with natural rubber latex)<br>- Tip Cap: ABS with white colorant |
| Syringe Packaging | Plastic peel pouch (printed on one<br>side, clear on the other) | Plastic peel pouch (printed on one<br>side, clear on the other) |
| Content of Syringe<br>Package | One or two syringes per pouch | One syringe per pouch |
A side-by-side comparison of key device characteristics is presented in the following table.
{5}------------------------------------------------
## Summary of Non-Clinical Testing
The design and manufacturing of Praxiject™ 0.9% NaCl prefilled syringes are subject to verification and validation testing in conformance with regulatory guidance and recognized consensus standards.
| Device / Performance<br>Characteristic | Performance Standard | Results |
|-------------------------------------------|----------------------------------------------------|-----------------------------------------------|
| Solution | Sodium Chloride Injection,<br>USP 39-NF34 | Conforms |
| Plastic syringe | ISO 7886-1, 0 mL average syringe<br>induced reflux | Conforms |
| Biocompatibility | ISO 10993-1 | N/A |
| - Hemolysis | ISO 10993-4 (ASTM F756) | Non-hemolytic |
| - Cytotoxicity | ISO 10993-5 | Non-cytotoxic |
| - Sensitization | ISO 10993-10 | Non-sensitizer |
| - Irritation/Intracutaneous<br>reactivity | ISO 10993-10 | Non-irritant |
| - Acute system toxicity | ISO 10993-11 | No systemic toxicity |
| - Material—mediated<br>pyrogenity | ISO 10993-11 (USP <151>) | No material mediated<br>response observed |
| - Bacterial Endotoxins | USP <85> and USP <161><br>(≤ 0.5 EU/mL) | Conforms |
| - Chemical<br>characterization | ISO 10993-18 | Acceptable<br>extractables/leachables profile |
| - Particulate matter | USP <788> | Conforms |
The shelf life of the final finished sterilized device was evaluated using the recognized consensus standard on the requirements for materials, sterile barrier systems and packaging systems for terminally sterilized medical devices (ISO 11607-1).
## Substantial Equivalence Conclusion
The results of the above described non-clinical testing demonstrate that the Praxiject™ 0.9% NaCl prefilled syringe fulfills the established performance criteria and it is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.