VV330 Continuous ECG Platform is a wireless recording system intended for use by healthcare professionals for record and display of physiological data within healthcare settings or at home. This includes electrocardiogram (ECG), accelerometer data, R-R Interval and heart rate. Data is transmitted wirelessly to a separate location (such as a mobile phone) for storage and display. The VV330 Continuous ECG Platform can be configured by Authorized Persons to modify or merge or ignore any of the operational alerts, but not to set new alerts related to physiological data. The device is not intended to be used on critical care patients and is intended to supplement vital signs recording for later viewing by healthcare professionals, not to replace current standards of care. The device is an ambulatory, continuous recording system intended for use on general care patients and on patients who are 18 years of age or older.
Device Story
Wireless, single-channel, rechargeable, ambulatory ECG recorder; uses disposable adhesive patch with hydrogel for skin contact. Collects ECG, accelerometer, R-R interval, and heart rate data. Transmits data via Bluetooth Low Energy (BLE) to mobile devices for storage and display. Used in healthcare settings or home by healthcare professionals; data viewed by professionals on mobile devices. Provides operational alerts (lead on/off, battery, system status); does not provide physiological alarms. Supplements vital signs recording; does not replace standard of care. Benefits patients by enabling continuous, ambulatory monitoring outside of critical care environments.
Clinical Evidence
Bench testing only. No clinical data presented. Verification included software, hardware, mechanical, biocompatibility, packaging, and electromagnetic compatibility testing per FDA-recognized standards (ISO 10993, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-47).
Technological Characteristics
Single-channel ECG recorder; rechargeable battery; tri-axial accelerometer; BLE connectivity. Dimensions: 90mm x 28mm x 7.9mm; weight 7.5g. Biocompatible materials per ISO 10993-1, -5, -10. Conforms to IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), and IEC 60601-2-47 (ambulatory ECG).
Indications for Use
Indicated for general care patients 18 years of age or older for ambulatory, continuous recording of ECG, accelerometer data, R-R interval, and heart rate. Not for use on critical care patients. Intended to supplement vital signs recording for later viewing by healthcare professionals.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 7, 2020
VivaLNK, Inc. Quoi Huynh Vice President of Operations 51 E. Campbell Ave. Suite 160 Campbell, California 95008
Re: K191870
Trade/Device Name: VV330 Continuous ECG Platform. VivaLNK Adhesive Patch Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II Product Code: DRG Dated: December 5, 2019 Received: December 9, 2019
Dear Quoi Huynh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Jessica Paulsen Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K191870
Device Name VV330 Continuous ECG Platform
#### Indications for Use (Describe)
VV330 Continuous ECG Platform is a wireless recording system intended for use by healtheare professionals for record and display of physiological data within healthcare settings or at home. This includes electrocardiogram (ECG), accelerometer data, R-R Interval and heart rate. Data is transmitted wirelessly to a separate location (such as a mobile phone) for storage and display.
The VV330 Continuous ECG Platform can be configured by Authorized Persons to modify or merge or ignore any of the operational alerts, but not to set new alerts related to physiological data.
The device is not intended to be used on critical care patients and is intended to supplement vital signs recording for later viewing by healthcare professionals, not to replace current standards of care. The device is an ambulatory, continuous recording system intended for use on general care patients who are 18 years of age or older.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Section 5. 510(k) Summary
# 1. General Information
| 510(k) Sponsor | VivaLNK, Inc. |
|-----------------------|-------------------------------------------------------|
| Address | 51 East Campbell Ave, Suite#160<br>Campbell, CA 95008 |
| Correspondence Person | Quoi Huynh |
| Contact Information | Qh@vivaLNK.com<br>408 205-4202 |
| Date Prepared | July 10th, 2019 |
# 2. Subject device:
| Proprietary Name | VV330 Continuous ECG Platform |
|---------------------|---------------------------------------------------------------------|
| Classification Name | Transmitters and Receivers, Physiological Signal,<br>Radiofrequency |
| Regulation Number | 870.2910 |
| Product Code | DRG |
| Regulatory Class | II |
#### 3. Primary Predicate Device:
| Proprietary Name | VitalConnect Platform |
|------------------------|-------------------------------------------------------------------------------------------------------|
| Premarket Notification | K132447 |
| Classification Name | Transmitters and Receivers, Physiological Signal,<br>Radiofrequency<br>Detector and alarm, arrhythmia |
| Regulation Number | 870.2910<br>870.1025 |
| Product Code | DRG, DSI |
| Regulatory Class | II |
# 4. Secondary Predicate Device:
| Proprietary Name | VitalConnect Platform |
|------------------------|---------------------------------------------------------------------|
| Premarket Notification | K141167 |
| Classification Name | Transmitters and receivers, physiological signal,<br>radiofrequency |
{5}------------------------------------------------
| | Detector and alarm, arrhythmia |
|-------------------|---------------------------------------------------------------|
| | Monitor, physiological, patient (with arrhythmia detection or |
| | alarm) |
| Regulation Number | 21 C.F.R. 870.2910 |
| | 21 C.F.R. 870.1025 |
| | 21 C.F.R. 870.1025 |
| Product Code | DRG, DSI, MHX |
| Regulatory Class | II |
#### Device Description 5.
The VV330 Continuous ECG Platform is a single-channel, rechargeable, re-usable, ambulatory medical-grade continuous ECG recorder, intended for use by healthcare professionals for record and display of physiological data within healthcare settings or at home.
The device is not intended to be used on critical care patients and is intended to supplement vital signs recording for later viewing by healthcare professionals, not to replace current standards of care. The device is an ambulatory, continuous recording systems intended for use on general care patients and on patients who are 18 years of age or older.
The VV330 Continuous ECG Platform is an ECG acquisition, storage, and transmission devices that utilizes disposable adhesive to maintain contact with the patient's skin and a hydrogel to promote electrical connectivity. The Platform has a re-usable sensor/recorder module allowing collection, storage and transfer of physiological data such as electrocardiogram (ECG), Accelerometer data, R-R Interval (RRI) and Heart Rate (HR). The Platform has a mobile data-display application built from a proprietary Software Library. And it also allows Authorized Persons to build their own applications using the same Software Library.
User's physiological are not transferred to healthcare professionals directly in any way; healthcare professionals can only view the data on the intended device (such as a mobile phone). User may view the physiological data in real time from the smart phone (or similar devices) while it records the data.
The Subject device provides operational alarms such as lead on/off status detection, battery monitoring, Over the Air (OTA) updates, real time clock, and power management. The operational alarms are intended to notify users of any interruption in data and of the overall operation status of the Recorder. The Subject device, being just a Transmitters and Receivers of Physiological Signal, does not provide any alarms based on physiological data setting.
{6}------------------------------------------------
#### 6. Indications for Use
VV330 Continuous ECG Platform is a wireless recording system intended for use by healthcare professionals for record and display of physiological data within healthcare settings or at home. This includes electrocardiogram (ECG), accelerometer data, R-R Interval and heart rate. Data is transmitted wirelessly to a separate location (such as a mobile phone) for storage and display. The VV330 Continuous ECG Platform can be configured by Authorized Persons to modify or merge or ignore any of the operational alerts, but not to set new alerts related to physiological data.
The device is not intended to be used on critical care patients and is intended to supplement vital signs recording for later viewing by healthcare professionals, not to replace current standards of care. The device is an ambulatory, continuous recording system intended for use on general care patients and on patients who are 18 years of age or older.
# 7. Comparison of Indications for Use and Technological Characteristics with the Primary and Secondary Predicates
| Feature/<br>Function | Subject device: | Primary Predicate Device: | Secondary Predicate Device: |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | VV330 Continuous ECG<br>Platform | VitalConnect Platform<br>(K132447) | VitalConnect Platform<br>(K141167) |
| Indications for Use | | | |
| Indications for Use | VV330 Continuous ECG<br>Platform is a wireless recording<br>system intended for use by<br>healthcare professionals for<br>record and display of<br>physiological data within<br>healthcare settings or at home.<br>This includes electrocardiogram<br>(ECG), accelerometer data, R-R<br>interval and heart rate. Data is<br>transmitted wirelessly to a<br>separate location (such as a<br>mobile phone) for storage and<br>display. The VV330 Continuous<br>ECG Platform can be configured<br>by Authorized Persons to modify<br>or merge or ignore any of the<br>operational alerts, but not to set<br>new alerts related to<br>physiological data. | The VitalConnect Platform is a<br>wireless monitoring system<br>intended for use by healthcare<br>professionals for unattended<br>surveillance of physiological data<br>within healthcare settings. This<br>includes heart rate,<br>electrocardiography (ECG), heart<br>rate variability, respiratory rate,<br>skin temperature, activity<br>(including step count), and<br>posture (body position relative to<br>gravity including fall). Data is<br>transmitted wirelessly to a central<br>location where it is stored for<br>analysis.<br><br>The VitalConnect Platform can be<br>configured by Authorized Persons<br>to notify healthcare professionals | The VitalConnect Platform is a<br>wireless remote monitoring system<br>intended for use by healthcare<br>professionals for continuous<br>collection of physiological data in<br>home and healthcare settings. This<br>includes heart rate,<br>electrocardiography (ECG), heart<br>rate variability (R-R interval),<br>respiratory rate, skin temperature,<br>activity (including step count), and<br>posture (body position relative to<br>gravity including fall). Data is<br>transmitted wirelessly to a central<br>location where it is stored for<br>analysis.<br><br>The VitalConnect Platform can be<br>configured by Authorized Persons<br>to notify healthcare professionals<br>when physiological data falls<br>outside selected parameters. |
| Feature/<br>Function | Subject device:<br>VV330 Continuous ECG<br>Platform | Primary Predicate Device:<br>VitalConnect Platform<br>(K132447) | Secondary Predicate Device:<br>VitalConnect Platform<br>(K141167) |
| The device is not intended to be<br>used on critical care patients and<br>is intended to supplement vital<br>signs recording for later viewing<br>by healthcare professionals, not<br>to replace current standards of<br>care. The device is an<br>ambulatory, continuous<br>recording system intended for<br>use on general care patients and<br>on patients who are 18 years of<br>age or older. | when physiological data falls<br>outside selected parameters.<br>The device is not intended to be<br>used on critical care patients and<br>is intended to supplement vital<br>signs monitoring by healthcare<br>professionals, not to replace<br>current standards of care. The<br>device is intended for use on<br>general care patients and on<br>patients who are 18 years of age<br>or older. It is not intended for<br>home use. | The device is intended for use on<br>general care patients who are 18<br>years of age or older as a general<br>patient monitor, to provide<br>physiological information. The data<br>from the VitalConnect Platform is<br>intended for use by healthcare<br>professionals as an aid to diagnosis<br>and treatment. It is not intended for<br>use on critical care patients. | |
| Regulations and<br>Product Code(s) | 21 CFR 870.2910, DRG-<br>Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency | 21 CFR 870.2910, DRG-<br>Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency<br>21 CFR 870.1025, DSI- Detector<br>and alarm, arrhythmia | 21 CFR 870.2910, DRG-<br>Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency<br>21 CFR 870.1025, DSI- Detector<br>and alarm, arrhythmia<br>21 CFR 870.1025, MHX- Monitor,<br>physiological, patient (with<br>arrhythmia detection or alarm) |
| Intended Use | Wireless recording and display<br>of physiological data | Wireless monitoring, recording,<br>and analysis of physiological data | Wireless monitoring, recording, and<br>analysis of physiological data |
| Intended Users | Healthcare Professionals | Healthcare Professionals | Healthcare Professionals |
| Intended Population | General care patients 18 years of<br>age or older | General care patients 18 years of<br>age or older | General care patients 18 years of<br>age or older |
| Intended Use<br>Environment | For home use and healthcare<br>setting | For healthcare setting | For home use and healthcare setting |
| | Hardware | | |
| Device Placement on<br>Human Body | Left upper chest area | Left upper chest area | Left upper chest area |
| Reuse | ECG Recorder: Re-usable<br>Adhesive Patch: Single use | Sensor Module: Single use<br>Adhesive Patch: Single use | Sensor Module: Single use<br>Adhesive Patch: Single use |
| Feature/<br>Function | Subject device:<br>VV330 Continuous ECG<br>Platform | Primary Predicate Device:<br>VitalConnect Platform<br>(K132447) | Secondary Predicate Device:<br>VitalConnect Platform<br>(K141167) |
| Duration of<br>Continuous Use | 3 days of continuous<br>recording once fully charged | 4 days of continuous<br>monitoring. Dispose after<br>single usage | 4 days of continuous monitoring.<br>Dispose after single usage |
| Battery | Rechargeable battery | Single use, non-rechargeable<br>battery | Single use, non-rechargeable<br>battery |
| Accelerometer | Tri-axial accelerometer | Tri-axial accelerometer | Tri-axial accelerometer |
| Size and Weight | Device Size:<br>90mm x 28mm x7.9mm<br>Adhesive Patch Size:<br>110mm x 48mm x 0.9mm<br>(without the release liner)<br>Device Weight: 7.5g | Device Size:<br>~21mm x 12mm x 4mm<br>Adhesive Patch Size:<br>~111mm x 47mm x 6mm<br>(without the release liner)<br>Device Weight: 11g | Device Size:<br>~21mm x 12mm x 4mm<br>Adhesive Patch Size:<br>~111mm x 47mm x 6mm<br>(without the release liner)<br>Device Weight: 11g |
| Relay Server | No | Yes | Yes |
| Signal Transmission | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) |
| Operating<br>Temperature | 10-45 degree C | 10-43 degree C | 10-43 degree C |
| Electrical Safety | Conformed to IEC 60601-1 | Conformed to IEC 60601-1 | Conformed to IEC 60601-1 |
| Electromagnetic<br>Compatibility | Conformed to IEC 60601-1-2 | Conformed to IEC 60601-1-2 | Conformed to IEC 60601-1-2 |
| | | Software | |
| Heart Rate Detection<br>Range | 40-300 Bpm (beats per<br>minute) | 30-200 Bpm (beats per<br>minute) | 30-200 Bpm (beats per minute) |
| Heart Rate Accuracy | 95% accuracy against true<br>heart rate value under<br>ambulatory conditions | < +/- 5 or 10% Beats per<br>minute, whichever is greater…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.