K191723 · Alphatec Spine, Inc. · PDQ · Oct 18, 2019 · Ear, Nose, Throat
Device Facts
Record ID
K191723
Device Name
ATEC IOM Accessory Instruments
Applicant
Alphatec Spine, Inc.
Product Code
PDQ · Ear, Nose, Throat
Decision Date
Oct 18, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
Device Story
ATEC IOM Accessory Instruments are surgical tools providing electrical stimulation to locate/identify nerves during spine surgery. Instruments include drills, taps, awls, probes, screwdrivers, dilators, guidewires, and needles. Devices connect to common FDA-cleared neuromonitoring consoles via compatible clips or probes. During surgery, the surgeon uses the instruments to deliver electrical stimulation; the neuromonitoring system measures nerve proximity and integrity. This feedback assists the surgeon in targeting, bone preparation, and implant placement (e.g., pedicle screws, fusion cages). Used in open or percutaneous minimally invasive approaches. Benefits include improved nerve identification and reduced risk of nerve injury during spinal instrumentation. Reusable instruments are steam-sterilized; single-use instruments are provided sterile.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing, including electrical safety (IEC 60601-1), reliability testing, and functional performance testing (insulation effectiveness and electrical resistance). Biocompatibility testing was conducted per ISO 10993-1, covering cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, and hemocompatibility.
Technological Characteristics
Instruments include drills, taps, awls, probes, screwdrivers, dilators, guidewires, and needles. Materials are consistent with legally marketed devices. Electrical stimulation principle; compatible with standard neuromonitoring consoles. Compliant with IEC 60601-1 (Clause 56.3(c) protected pin design). Biocompatibility per ISO 10993-1. Sterilization via steam (reusable) or EtO (single-use).
Indications for Use
Indicated for use in spine surgical procedures to assist in nerve location during or after preparation and placement of intervertebral fusion cages and pedicle screw fixation devices in open and percutaneous minimally invasive approaches.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{1}------------------------------------------------
October 18, 2019
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Alphatec Spine, Inc. Ms. Ruby Zheng Regulatory Affairs Specialist 5818 El Camino Real Carlsbad, California 92008
Re: K191723
Trade/Device Name: ATEC IOM Accessory Instruments Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDQ, ETN Dated: September 16, 2019 Received: September 17, 2019
Dear Ms. Ruby Zheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191723
Device Name ATEC IOM Accessory Instruments
### Indications for Use (Describe)
The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> × Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|------------------------------------------------------------------------------------------------------------------|
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Atec. The logo consists of a green square with a white stylized "a" inside of it. To the right of the square is the word "tec" in white font, with a trademark symbol in the upper right corner.
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.
| I. | SUBMITTER: | Alphatec Spine, Inc.<br>5818 El Camino Real<br>Carlsbad, CA 92008<br>Phone: (760) 431-9286<br>Fax: (760) 431-0289 |
|----|------------------------|-------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Ruby Zheng<br>Regulatory Affairs Specialist |
| | Date Summary Prepared: | September 16, 2019 |
#### II. DEVICE
| Trade or Proprietary Name: | ATEC IOM Accessory Instruments |
|----------------------------|------------------------------------|
| Common Name: | Surgical nerve stimulator/locator. |
| Classification Name: | Neurosurgical Nerve Locator |
| Regulation Number: | 21 CFR 874.1820 |
| Classification: | Class II |
| Product Code: | PDQ, ETN |
#### III. LEGALLY MARKETED PREDICATE DEVICES
Predicate Device(s):
| 510(k) | Product Name | Clearance Date |
|---------|----------------------------------------------------|----------------|
| K182617 | ATEC IOM Accessory Instruments | May 9, 2019 |
| K132138 | Rhythmlink Disposable Concentric Stimulating Probe | April 11, 2014 |
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Atec logo. The logo consists of a green square with rounded corners and a white letter "a" inside. To the right of the square is the word "tec" in white font, with a trademark symbol in the upper right corner. The background is a dark blue color.
#### DEVICE DESCRIPTION IV.
The ATEC IOM Accessory Instruments are surgical instruments that provide electrical stimulation to the body to locate and identify nerves in either open, minimally invasive, or percutaneous procedures. These surgical instruments are compatible with common FDA cleared neuromonitoring platforms as they are connected via a compatible clip or probe depending on the system. The neuromonitoring capability provides the surgeon with spinal nerve location, proximity, and integrity information. This information assists the surgeon during targeting, bone preparation, and placement of orthopedic implants such as intervertebral fusion devices (e.g., interbodies) and bone screws (e.g., pedicle screws). All sterile instruments are single use only and all reusable instruments are offered non-sterile to be steam sterilized by the end user.
#### V. INDICATIONS FOR USE
The ATEC IOM Accessory Instruments are utilized in spine surgical procedures to assist in location of the nerves during or after preparation and placement of implants (intervertebral fusion cages and pedicle screw fixation devices) in open and percutaneous minimally invasive approaches.
#### TECHNOLOGICAL COMPARISON TO PREDICATES VI.
The technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. See Table 1 below.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Atec. The logo consists of a green square with a white design inside, followed by the word "atec" in white. The background of the logo is a dark blue color. The logo is simple and modern, and the colors are eye-catching.
| | Primary Predicate<br>K182617 | Additional Predicate<br>K132138 | Subject Device |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | ATEC<br>IOM Instruments | Rhythmlink Disposable<br>Concentric Stimulating Probe | ATEC<br>IOM Instruments |
| Indications for Use | The ATEC IOM Accessory<br>Instruments are utilized in spine<br>surgical procedures to assist in<br>location of the nerves during or<br>after preparation and placement of<br>implants (intervertebral fusion<br>cages and pedicle screw fixation<br>devices) in open and percutaneous<br>minimally invasive approaches. | The Rhythmlink Disposable<br>Concentric Stimulating Probe is<br>used to perform Intended Use(s)<br>localized stimulation of neural<br>tissue and to locate, identify and<br>monitor cranial motor nerves,<br>peripheral nerve and spinal nerve<br>roots during surgery.<br>The Rhythmlink Disposable<br>Concentric Stimulating Probe is a<br>single patient use device | The ATEC IOM Accessory<br>Instruments are utilized in spine<br>surgical procedures to assist in<br>location of the nerves during or<br>after preparation and placement of<br>implants (intervertebral fusion<br>cages and pedicle screw fixation<br>devices) in open and percutaneous<br>minimally invasive approaches. |
| Regulation Number.<br>Product Code, & | 21 CFR 874.1820<br>PDQ, ETN | 21 CFR 874.1820<br>ETN | 21 CFR 874.1820<br>PDQ, ETN |
| Classification | Class II | Class II | Class II |
| IEC 60601 Compliant | Yes<br>Protected Pin Design via<br>compatible clip/probe<br>Clause 56.3(c) | Yes<br>DIN 42 802 touch proof connectors<br>Clause 56.3(c) | Yes<br>Protected Pin Design via<br>compatible clip/probe<br>Clause 56.3(c) |
| Instrument Type<br>(Description) | Drills/Taps, Awls, Probes,<br>Screwdrivers, Dilators (Sleeves),<br>Guidewires, and Needles | Probes | Drills/Taps, Awls, Probes,<br>Screwdrivers, Dilators (Sleeves),<br>Guidewires, and Needles |
| Biocompatibility Patient<br>Contact Duration | Limited patient duration contact<br>(≤ 24 hours) | Limited patient duration contact<br>(≤ 24 hours) | Limited patient duration contact<br>(≤ 24 hours) |
| Biocompatible | Yes | Yes | Yes |
| Surgical approach | Open or Percutaneous/Minimally<br>Invasive | Not specified | Open or Percutaneous/Minimally<br>Invasive |
| | Primary Predicate<br>K182617 | Additional Predicate<br>K132138 | Subject Device |
| Attribute | ATEC<br>IOM Instruments | Rhythmlink Disposable<br>Concentric Stimulating Probe | ATEC<br>IOM Instruments |
| Sterility | Non-sterile<br>Non-sterile devices are provided<br>with validated steam sterilization<br>parameters to assure an SAL of 10-6 | Sterile via EtO | Sterile via EtO and non-sterile<br>Non-sterile devices are provided<br>with validated steam sterilization<br>parameters to assure an SAL of 10-6 |
| Reusable/Single Use | Guidewires, Targeting Needles and<br>Dilators - Single Use<br>Awls, Drills/Taps, Probes, Dilators<br>(Sleeves), and Screwdrivers -<br>Reusable | Single Use | Guidewires, Targeting Needles and<br>Dilators - Single Use<br>SafeOp Ball Tip Probe - Single<br>Use<br>Awls, Drills/Taps, Probes, Dilators<br>(Sleeves), and Screwdrivers -<br>Reusable |
| Compatible with Common<br>Neuromonitoring Consoles<br>& Software | Yes | Yes | Yes |
| Minimum exposed surface<br>area | 8.6 mm2 | 14.6 mm2 | 8.6 mm2 |
## Table 1: Summary of Technological Comparison to Predicates
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Atec. The logo features a green square with a white alpha symbol inside of it. To the right of the square is the word "tec" in white font with a trademark symbol in the upper right corner. The background of the logo is a dark blue color.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Atec. The logo consists of a green square with a white design inside, followed by the word "tec" in white. The background of the logo is dark blue. There is a trademark symbol after the word "tec".
#### VII. PERFORMANCE DATA
Performance testing includes IEC 60601-1 testing, reliability testing, and functional testing on insulation effectiveness and electrical resistance. Table 2 below summarizes the testing which was performed on the Subject Devices to show substantial equivalence to the Testing results demonstrated the subject ATEC IOM Accessory predicate device. Instruments are appropriate for neuromonitoring applications and do not introduce a new worst case compared to other legally marketed devices cleared by FDA.
| Test | Test Method Summary | Results |
|----------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Electrical Safety Testing<br>and/or Evaluation | Evaluation and testing was performed on<br>the subject devices in accordance with IEC<br>60601-1: 2005. | All samples passed<br>acceptance criteria |
| Reliability Testing | Testing was performed to verify that there<br>is no adverse effect on the safety or<br>effectiveness of the subject devices based<br>on the intended environment and storage<br>conditions. | All samples passed<br>acceptance criteria |
| Functional Performance<br>Testing and Verification<br>Analysis | Insulation Effectiveness Electrical Resistance | All samples passed<br>acceptance criteria |
## Table 2: Summary of Performance Testing
## Clinical Information
Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.
#### BIOCOMPATIBILITY DATA VIII.
A risk analysis was performed taking into account nature of body contact and duration to categorization the use of existing data, end-specific testing, and endpoint assessment to cover the identified test methods. Additionally, data was leveraged by other means (e.g., authorized use of Master File, predicate and reference devices, well known and characterized materials) to support the biocompatibility of the subject devices.
Biocompatibility testing conducted per ISO 10993-1.
- Cytotoxicity
- Sensitization
- Irritation/Intracutaneous
- Reactivity ●
- Acute Systemic Toxicity ●
- Material Mediated Pyrogenicity ●
- . Hemocompatibility
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Atec. The logo consists of a green square with a white "a" inside of it, followed by the word "tec" in white. The background of the logo is dark blue.
In conclusion, the ATEC IOM Accessory Instruments are manufactured from the same materials as other legally US-marketed devices.
#### CONCLUSION IX.
Based upon the information provided in this 510(k) submission, it has been determined that the subject devices are substantially equivalent to legally marketed devices in regards to indications for use, intended use, design, technology, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.