The ES2 Neuromonitoring instruments (Awls, Taps, Screwdriver and LITe Y-NEEDLE 200, 300 and 400) can be used by the surgeon to assist in location of the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws in open and percutaneous minimally invasive posterior surgical approaches of the non-cervical spine.
Device Story
ES2 Neuromonitoring Accessory Instruments are surgical tools (awls, taps, screwdrivers, and LITe Y-Needles) used in non-cervical spine surgery. The devices connect to external, FDA-cleared neuromonitoring consoles via clips or probes to deliver electrical stimulation to surrounding tissue. By measuring the electrical response, the surgeon receives proximity information regarding the location of spinal nerves. This helps guide pedicle targeting, guidewire placement, bone preparation, and bone screw insertion. The instruments are used in OR settings by surgeons. The LITe Y-Needles are single-use and sterile-packed, while other instruments are reusable. The device provides real-time feedback to the surgeon, allowing for safer navigation near neural structures during spinal procedures.
Clinical Evidence
Bench testing only. Performance data included biocompatibility (ISO 10993-1), sterility/packaging validation (Gamma radiation), and electrical safety (IEC 60601-1). Functional testing verified stationary potential, electrical resistance, current variance, and signal conduction (amplitude change) against predefined acceptance criteria. No clinical data was required.
Technological Characteristics
Instruments manufactured from surgical grade stainless steel, ABS, and Radel. Sensing/actuation principle: electrical stimulation via conductive electrodes. Dimensions/form factor: manual surgical instruments (awls, taps, screwdrivers) and needles (200, 300, 400). Connectivity: wired connection to external neuromonitoring consoles. Sterilization: Gamma radiation for single-use needles; reusable instruments provided non-sterile with validated parameters. Compliant with IEC 60601-1.
Indications for Use
Indicated for use by surgeons to assist in locating spinal nerves by providing proximity information during bone preparation and bone screw placement in open and percutaneous minimally invasive posterior surgical approaches of the non-cervical spine.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
Predicate Devices
ES2 Spinal System Neuromonitoring Accessory Instruments (K140400)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three human profiles facing right. The profiles are stacked on top of each other, creating a sense of depth and unity. Encircling the caduceus are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 18, 2017
Stryker Corporation Mr. Nikin Desai Regulatory Affairs Specialist 2 Pearl Court Allendale, NJ 07401
Re: K171807
Trade/Device Name: ES2 Neuromonitoring Accessory Instruments Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: PDQ Dated: June 14, 2017 Received: June 19, 2017
Dear Mr. Desai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
## Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171807
Device Name
ES2 Neuromonitoring Accessory Instruments
Indications for Use (Describe)
The ES2 Neuromonitoring instruments (Awls, Taps, Screwdriver and LITe Y-NEEDLE 200, 300 and 400) can be used by the surgeon to assist in location of the spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws in open and percutaneous minimally invasive posterior surgical approaches of the non-cervical spine.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CER 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black, and the background is white. There is a horizontal line below the word.
| 510(k) Summary: ES2® Neuromonitoring Accessory Instruments | | |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Stryker Spine | |
| Manufacturer/Submitter: | 2 Pearl Ct. | |
| | Allendale, NJ 07401 | |
| | Name: Nikin Desai | |
| Contact Person : | Phone: (201) 749-8176 | |
| | Fax: (201) 831-3000 | |
| | Email:Nikin.Desai@stryker.com | |
| Date Prepared: | 07/18/2017 | |
| Trade Name: | ES2 Neuromonitoring Accessory Instruments | |
| Common Name: | Surgical Nerve Stimulator/Locator | |
| Proposed Class: | Class II | |
| Classification Name: | Surgical Nerve Stimulator/Locator (21 CFR §874.1820) | |
| Product Code: | PDQ | |
| Predicate Devices: | Primary Predicate: | |
| | ES2 Spinal System Neuromonitoring Accessory Instruments | |
| | (K140400) | |
| | | |
| Device Description: | The purpose of this submission is to introduce a line extension to the ES2 | |
| | Neuromonitoring system to include the LITe Y-Needles. | |
| | | |
| | The ES2 Neuromonitoring instruments (Awl, Taps, Screwdriver and LITe® Y- | |
| | NEEDLE 200, 300 and 400) are accessory instruments to be used with FDA | |
| | cleared neuromonitoring systems to deliver electrical stimulation to assist in | |
| | location of the spinal nerves during intraoperative neurological monitoring of the | |
| | non-cervical spine in open and percutaneous minimally invasive posterior surgical | |
| | approaches. The instruments are manufactured from surgical grade stainless steel | |
| | and Acrylonitrile Butadiene Styrene (ABS). The Awls, taps and screwdriver are | |
| | provided non-sterile. The LITe Y-Needles are sterile packed and single use devices | |
| | that must be discarded after use. | |
| | | |
| | The neuromonitoring application is a surgical option that allows the surgeon to | |
| | locate the spinal nerves by providing proximity information during targeting, bone | |
| | preparation, and placement/insertion of bone screws. In addition to the | |
| | neuromonitoring function: The LITe Y-Needles assist in targeting the pedicle and<br>placing guidewires, the awl and taps facilitate bone preparation, and the | |
| | screwdriver facilitates bone screw placement/insertion. The surgical accessories | |
| | are compatible with commercially available FDA cleared neuromonitoring | |
| | consoles/systems and associated electrodes. The nerves are stimulated using | |
| | electrodes attached to the subject accessory devices. The screwdriver can be used | |
| | with or without a powered option for bone screw placement. | |
| | | |
| 510(k) Summary: ES2® Neuromonitoring Accessory Instruments | | |
| Indications for Use: | The ES2 Neuromonitoring instruments (Awls, Taps, Screwdriver and LITE Y-<br>NEEDLE 200, 300 and 400) can be used by the surgeon to assist in location of the<br>spinal nerves by providing proximity information before, during or after bone<br>preparation and placement of bone screws in open and percutaneous minimally<br> | |
| Summary of the<br>Technological<br>Characteristics | The Stryker Spine ES2 Neuromonitoring Accessory Instruments are substantially<br>equivalent to the predicate devices in terms of design, function, principals of<br>operation, technological characteristics, and indications and intended uses. See<br>table below for comparisons to predicate ES2 Neuromonitoring Accessory<br>Instruments. | |
| Summary of the<br>Performance Data | See table below. | |
| Conclusion | The risk analysis performed on the proposed addition of 200, 300 and 400 LITE Y-<br>Needles to ES2 Neuromonitoring and subsequent verification and validation<br>activities demonstrate the line extension does not impact the ES2<br>Neuromonitoring Instrument's safety and effectiveness. Furthermore, the design<br>verifications and validations performed as a result of the risk analysis and<br>presented herein demonstrate the proposed device does not raise new questions<br>of safety or effectiveness. Thus, the proposed modification and the predicate<br>device are considered substantially equivalent. | |
| Summary of the Performance Data | | |
| Test | Test Method Summary | Results |
| Biocompatibility | A material biocompatibility evaluation was<br>performed on the subject device in line with FDA<br>guidance dated June 16, 2016 "Use of<br>International Standard ISO 10993-1" and in<br>accordance with ISO 10993-1:2009 and, Biological<br>evaluation of medical devices – Part 1: Evaluation<br>and Testing. The LITe Y-Needle is categorized as:<br>Externally communicating patient<br>contacting with tissue/bone/dentin<br>contact;limited exposure ≤24hrs | All samples passed acceptance criteria<br>testing on subject device. The device<br>demonstrated it had an acceptable<br>biocompatibility profile commensurate<br>with anticipated exposure. |
| Sterility and<br>Packaging | Testing appropriate for the sterile packaged LITE<br>Y-Needles device was performed as per the 2016<br>FDA Guidance "Submission and Review of Sterility<br>Information in Premarket Notification (510(k))<br>Submissions for Devices Labeled as Sterile", are<br>sterilized using an "Established Category A"<br>sterilization method, Gamma Radiation. This<br>testing included but not limited to the following:<br>Physical stressing | All samples passed acceptance criteria<br>testing on subject device and<br>demonstrated that packaging and<br>sterilization was appropriate for single<br>use, sterile packaged devices. |
| Summary of the Performance Data | | |
| Test | Test Method Summary | Results |
| | Visual inspection Seal strength Bubble emission Packaging aging | |
| EMC and Safety | Stryker Spine conducted Safety and EMC testing<br>to evaluate the electrical performance and safety<br>of the ES2 accessory instruments used for<br>neuromonitoring applications. Safety tests were<br>conducted and/or evaluated per IEC 60601-1. | EMC risk evaluation per IEC 60601-1<br>demonstrated that the subject device<br>did not present a new worst case and<br>met all applicable UL Safety and EMC<br>tests. |
| Stationary<br>Potential<br>Generation Test | The purpose of this test is to verify that a potential<br>stationary potential(s) produced when applying<br>electrical current to the LITe Y-300 and Y-400<br>needles are not sufficiently large so false positive is<br>not produced when used for neuromonitoring.<br><br>The LITe® Y-Needle was placed in a biological<br>simulant and connected to the stimulation probe.<br>Recording probes and thermocouples were placed<br>in the simulant.<br><br>The LITE® Y-Needle was stimulated at clinical<br>current levels for 60 seconds. Using a fixed sweep<br>trace capture, the average stationary potential was<br>obtained for the full duration. | The passing results of the stationary<br>potential testing demonstrated that the<br>LITe® Y-Needles generated a stationary<br>potential less than or equal to the<br>predefined acceptance criteria. |
| Electrical<br>Resistance | The purpose of this test is to ensure the resistive<br>and conduction sections of the needles are<br>appropriate to allow for current conduction and<br>current insulation during neuromonitoring.<br><br>Probes were connected to the conductive and<br>insulating portions of the device. The resistance of<br>the conductive portion was measured. The<br>resistance of the insulating portion was measured. | The passing results of the resistance test<br>demonstrated that the LITe® Y-Needles<br>experienced a resistance measurement<br>less than or equal to the predefined<br>acceptance criteria for conductivity and<br>greater than or equal to the predefined<br>acceptance criteria for the resistivity. |
| Current Variance | The purpose of this test is to verify the degree of<br>current deviation along the axial length of the<br>LITe® Y-Needle. This testing ensures the signal is<br>conducted in a consistent manner as delivered by<br>the neuromonitoring device.<br><br>The LITe® Y-Needles were connected to<br>stimulation probes at the conductive locations of<br>the device. Clinical current levels were applied to<br>the device for 10seconds, 100seconds, and 500<br>seconds. The transmitted current was recorded<br>and compared to the applied current. | The passing results of the current<br>deviation test demonstrated that the<br>LITe® Y-Needles experienced a current<br>deviation less than or equal to the<br>predefined acceptance criteria. |
| Summary of the Performance Data | | |
| Test | Test Method Summary | Results |
| Signal Conduction<br>(amplitude<br>change) | The purpose of this test is to verify the degree of<br>signal loss the axial length of the LITe® Y-Needle.<br>This testing ensures the signal is conducted in a<br>consistent manner as delivered by the<br>neuromonitoring device.<br><br>The LITe® Y-Needles were connected to a<br>stimulation probe at the conductive locations of<br>the device. The device was stimulated using clinical<br>current levels and the percentage difference of the<br>power input was compared to the transmitted<br>signal wave. | The passing results of the signal loss test<br>demonstrated that the LITe® Y-Needles<br>experienced a signal loss less than or<br>equal to the predefined acceptance<br>criteria. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and connected, giving the word a sense of unity. The color of the text is black, and the background is white, creating a high contrast that makes the word stand out.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. The word is written in lowercase letters.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and black. The word is centered and takes up most of the frame.
Г
| Summary of the Technological Characteristics | | | |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|--|
| | | Primary Predicate | |
| Characteristics | Subject device | ES2 Spinal System<br>Neuromonitoring Accessory<br>Instruments | |
| Manufacturer/OEM | Stryker Spine | Stryker Spine | |
| 510(k) Number | K171807 | K140400 | |
| Product Code<br>(Regulation number) | PDQ (21 CFR 874.1820) | ETN (21 CFR 874.1820) | |
| Neuromonitoring<br>Accessories<br>Instruments | Awl, Taps, Screwdriver and 200,300 and 400<br>LITe Y-Needles | Awl, Taps, Screwdriver | |
| Materials | Stainless steel, ABS and Radel | Awl, Taps, and Screwdriver:<br>Surgical Grade Stainless Steel<br>Dilators and Tap Sleeve- RADEL | |
| Indications for use | The ES2 Neuromonitoring instruments (Awls,<br>Taps, Screwdriver and LITe Y-NEEDLE 200, 300<br>and 400) can be used by the surgeon to assist in<br>location of the spinal nerves by providing<br>proximity information before, during or after<br>bone preparation and placement of bone screws<br>in open and percutaneous minimally invasive<br>posterior surgical approaches of the non-<br>cervical spine. | Identical | |
| Summary of the Technological Characteristics | | | |
| | | Primary Predicate | |
| Characteristics | Subject device | ES2 Spinal System<br>Neuromonitoring Accessory<br>Instruments | |
| Surgical Approach | Open or Percutaneous/Minimally Invasive | Identical | |
| How supplied | 200,300 and 400 LITe Y-Needles- Sterile | Awls, Taps, and Screwdriver- Non- | |
| (Sterile/Non-sterile) | Awls, Taps, and Screwdriver- Non-Sterile | Sterile | |
| Sterilization | The 200,300 and 400 LITe Y-Needles are<br>provided as single-use sterile packaged devices.<br>Non-sterile devices are provided with validated<br>sterilization parameters to assure an SAL of 10-<br>6. | Non-sterile devices are provided with<br>validated sterilization parameters to<br>assure an SAL of 10-6. | |
| Reusable or Single<br>use? | 200,300,and 400 LITe Y-Needles-Single use<br>Awls, Taps, and Screwdriver- Reusable | Awls, Taps, and Screwdriver-<br>Reusable | |
| Use of Dilators | Dilators or Needle depth stop | Identical | |
| Compatible with<br>Common<br>Neuromonitoring<br>Consoles & Software | Compatible with FDA cleared neuromonitoring<br>systems. | Identical | |
| Connection to<br>Neuromonitoring unit | Clip or Probe (based on Neuromonitoring<br>System used)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.