The SLP OxSAT 100 Patient Oximeter Module is indicated for use in measuring and transmitting functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and plethysmographic data to a compatible PSG and/or HST device. It is not intended for use with low perfused patients.
Device Story
OxSAT 100 is a pulse oximeter module; interfaces with external sensors to measure blood oxygenation via transmissive photoelectric sensing. Inputs: red (660nm) and infrared (910nm) light transmitted through tissue; sensed by photoelectric cell. Outputs: SpO2, pulse rate, and plethysmographic data transmitted via cable to host PSG or HST device. Used in sleep study settings or home environments; operated by clinicians or patients. Powered by host monitor; no internal display or battery. Benefits: enables continuous monitoring of oxygen saturation and pulse rate during sleep studies. Output used by healthcare providers to assess patient respiratory status.
Clinical Evidence
Bench testing only. Compliance with ISO 80601-2-61 for clinical accuracy, AAMI ANSI ES60601-1 for electrical safety, and IEC 60601-1-2 for electromagnetic compatibility. No human clinical trials were conducted; performance verified through bench testing and comparison to predicate.
Technological Characteristics
Transmissive pulse oximeter module. Wavelengths: 660nm (red), 910nm (infrared); max optical output <1 mW. Powered by host device via cable. Connectivity: wired transmission to host PSG/HST. Standards: AAMI ANSI ES60601-1, IEC 60601-1-2, ISO 80601-2-61. Module is not patient-contacting; uses previously cleared sensors.
Indications for Use
Indicated for pediatric (3+ years) and adult patients requiring measurement and transmission of SpO2, pulse rate, and plethysmographic data to compatible PSG/HST devices. Not for use with low perfused patients.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
S.L.P. Ltd. % Paul Dryden Consultant S.L.P. Ltd. c/o ProMedic, LLC 131 Bay Point Dr. NE. St Petersburg, Florida 67060
Re: K191574
Trade/Device Name: OxSAT 100 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: December 6, 2019 Received: December 9, 2019
# Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K191574
Device Name
OxSAT 100
Indications for Use (Describe)
The SLP OxSAT 100 Patient Oximeter Module is indicated for use in measuring and transmitting functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and plethysmographic data to a compatible PSG and/or HST device. It is not intended for use with low perfused patients.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size: 16px;"> <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 16px;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "
FORM FDA 3881 (7/17)
Page 1 of 1
SC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
| Date Prepared: | 1-Apr-20 |
|----------------------------|--------------------------------------------------------------------------------------------------------|
| Sponsor: | Doron D. Patt<br>S.L.P. Ltd<br>62 Anilewicz Street<br>Tel-Aviv, Israel 67060<br>Tel: +972 (9) 834 6731 |
| Proprietary or Trade Name: | OxSAT 100 |
| Common/Usual Name: | Oximeter |
| Classification Name: | Oximeter<br>DQA, Class II, CFR 870.2700 |
| Predicate Device: | Nonin Model 4100, 510(k) K043359 |
# Device Description:
The SLP OxSAT 100 is a pulse oximeter module that photoelectrically determines the oxygenation of blood in a part of the body, based upon the sensor placement.
The OxSAT 100 contains the electronics to interface to an attached sensor. The red and infrared light are transmitted through oxyhemoglobin and are sensed in the photoelectric cell. The red and infrared light is absorbed in different amounts depending on the oxygenation of the blood.
The OxSAT 100 has the ability to determine both the percent of saturated hemoglobin and the pulse rate. It performs these functions on adult and pediatric patient populations. It is designed for measuring and transmitting functional oxygen saturation of arterial hemoglobin (SpO2), Pulse rate and plethysmographic data to compatible polysomnogram ("PSG") and/or home sleep testing ("HST") devices.
Pulse amplitude is not displayed. The OxSAT 100 is powered by the host monitor. The wavelength of red LED is 660nm and Infrared LED is 910 nm with maximum optical output power of less than 1 mW.
TThe OxSAT 100 Module is not intended for use with skin contact and, therefore, has not been tested for biocompatibility. The OxSAT 100 is either attached to the outside of a belt on the patient or is set off to the side of the bed. An extension cable might be used with the sensors to avoid skin contact.
### Indications for Use:
The SLP OxSAT 100 Patient Oximeter Module is indicated for use in measuring and transmitting functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and plethysmographic data to a compatible PSG and/or HST device. It is not intended for use with low perfused patients.
# Contraindications:
None
### Substantial Equivalence Discussion
Table 1 compares the SLP OxSAT 100 to the predicate device for equivalence of:
{5}------------------------------------------------
## Indications -
The SLP OxSAT 100 Patient Oximeter Module is indicated for use in measuring and transmitting functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, and plethysmographic data to a compatible PSG and/or HST device. It is not intended for low perfused patients.
Discussion - The indications are similar to the predicate Nonin Model 4100 Patient Oximeter Module. The host PSG and/or HST device will dictate the monitoring use.
## Patient Population -
Both are intended for pediatrics (3 years and older) and adult populations.
# Environment of Use -
The environment of use is the same.
# Prescriptive -
Both are prescriptive.
# Design and Technology -
The OxSAT 100 contains the electronics to interface to an attached sensor. The red and infrared light are transmitted through oxyhemoglobin and are sensed in the photoelectric cell. The red and infrared light is absorbed in different amounts depending on the oxygenation of the blood.
The OxSAT 100 has the ability to determine both the percent of saturated hemoglobin and the pulse rate.
Pulse amplitude is not displayed. The OxSAT 100 is powered by a host monitor. The wavelength of red LED is 660nm and Infrared LED is 910 nm with maximum optical output power of less than 1 mW. Discussion -
The proposed device is powered by connection to a compatible device whereas the predicate device is selfpowered and uses Bluetooth to transmit data to a compatible device. In either case no data is displayed by the proposed or predicate device itself.
# Performance Specifications -
Both have equivalent specifications.
### Compliance with standards -
The OxSAT 100 complies with AAMI ANSI ES60601-1, IEC 60601-1-2, and ISO-80601-2-61. The predicate complies with the same standards.
### Performance Testing
### Non-clinical -
We have performed bench tests and found that the OxSAT 100 met all requirements specifications and standards requirements and was found to be equivalent in comparison to the predicate. Testing includes the following:
- Verification Testing
- Testing for compliance to AAMI ES 60601-1
- Testing for compliance to IEC 60601-1-2,
- Testing for compliance to IEC 80601-2-61 ●
{6}------------------------------------------------
The results demonstrate that the device performs as intended and is substantially equivalent to the performance of the predicate and in accordance with applicable standards.
#### Biocompatibility and Materials -
The only patient contacting materials of the OxSAT 100 are the sensors which have been previously cleared. The OxSAT 100 module is not intended for use with skin contact and, therefore, has not been tested for biocompatibility. The OxSAT 100 is either attached to the outside of a belt on the patient or is set off to the side of the bed. An extension cable might be used with the OxSAT 100 to avoid skin contact.
#### Host System Compatibility -
Testing demonstrated that the following host PSG and HST systems are compatible with the OxSAT 100. All host systems are considered as capable of continuous measurements.
| Host System | K# | Environment |
|----------------|---------|-------------------------------|
| Embletta PDS | K041904 | Hospital, Sleep Center |
| Alice 6 | K040595 | Hospital, Sleep Center, Other |
| Alice NightOne | K083874 | Hospital, Sleep Center, Home |
| Embla S4500 | K024322 | Hospital, Sleep Center, Other |
| Embla N7000 | K024322 | Hospital, Sleep Center, Other |
| ApneaLink Air | K143272 | Hospital, Sleep Center, Home |
| Compumedics | K000068 | Hospital |
| Embletta MPR | K122516 | Hospital, Sleep Center, Home |
| Embleeta Gold | K073682 | Hospital, Sleep Center, Home |
### Clinical Testing -
Testing to ensure clinical accuracy of the device in accordance with ISO 80601-2-61. This testing and results showed compliance to the standard.
### Differences –
The identified difference:
- . Data transmission via cable vs. Bluetooth
The OxSAT 100 has been evaluated and tested and confirmed that the difference does not raise different questions of safety or effectiveness when compared to the predicate device for the proposed indications for use.
#### Substantial Equivalence Conclusion
The OxSAT 100 is substantially equivalent to the predicate in: indications for use, patient population, environment of use, technology characteristics, materials, specifications / performance and compliance with international standards.
{7}------------------------------------------------
| CHARACTERISTICS | OxSAT 100 | Nonin Model 4100 Patient Oximeter<br>Module (K043359) | Comments |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Indications for use | The SLP OxSAT 100 Patient Oximeter<br>Module is indicated for use in<br>measuring and transmitting functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2), pulse rate, and<br>plethysmographic data to a compatible<br>PSG and/or HST device. It is not<br>intended for low perfusion patients. | The Nonin® Bluetooth® -enabled Model<br>4100 Patient Oximeter Module is indicated<br>for use in measuring and transmitting<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2), pulse rate, and<br>plethysmographic data to a compatible<br>Bluetooth-enabled device. | Similar indications<br>Both require connection to a<br>compatible device to display<br>collected data |
| Type of use | Continuous based on the host device | Spot checking<br>Continuous | Similar |
| Motion | Non-motion | Non-motion | Similar |
| Patient Population | Pediatrics and Adults | Pediatrics and Adults | Similar |
| Perfusion | Well | Well<br>Poorly | Similar |
| Environment of Use | Sleep Study setting and home use | Sleep study settings and home use | Similar |
| Technology | Transmissive | Transmissive | Similar |
| System components / Configurations | | | |
| Batteries | Powered by compatible device | 2 x 1.5 AAA batteries | Similar |
| SpO2 Display Range | 0% to 100% SpO2 | 0% to 100% SpO2 | Similar |
| Pulse rate declared<br>accuracy range | 20-300 BPM | 18-300 BPM | Similar |
| Sensors | APK - K082846<br>Solaris - K100077 | Nonin sensors | Similar |
# Table 1 Device Comnarison
{8}------------------------------------------------
# 510(k) Summary Page 5 of 5
| CHARACTERISTICS | OxSAT 100 | Nonin Model 4100 Patient Oximeter<br>Module (K043359) | Comments |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|----------|
| Accuracy | | | |
| SpO2 | 70-100% ± 2 digits | 70-100% ± 2 digits | |
| Pulse rate | ± 3 digits | ± 3 digits | Similar |
| Display | On the compatible device | On the compatible device | |
| Data transmission | Direct connection to compatible device | Bluetooth to compatible device | |
| Application site | Digits | Digits | Similar |
| Operation mode | Depends upon the host device | Spot checking or continuous monitoring | Similar |
| LED wavelengths<br>(multiple) | 660 and 910 nm | 660 and 910 nm | Similar |
| Functional and safety<br>testing | ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-61 | ES 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 80601-2-61 | Similar |
| Biocompatibility | The OxSAT 100 Module has not been<br>tested for biocompatibility.<br>Accessory sensors have been tested and<br>are considered<br>Surface Contact Skin<br>Limited duration (<24 hours) | Surface contact<br>Skin<br>Limited duration (<24 hours) | Similar |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.