The AGILON® XO Shoulder Replacement System is indicated for use for cementless inverse (reverse) total or hemi shoulder replacement in cases of: - Non- inflammatory degenerative joint disease including osteoarthritis and avascular necrosis, . - . Post-traumatic osteoarthritis. - Fractures, - . Rheumatoid arthritis. The main indications for the implantation of an AGILON® hemi shoulder prosthesis are: - Multifragmental comminuted fractures of the humeral head, - 3- and 4-Fragment-fractures of the proximal humerus, - . Head-splitting fractures, - . Dislocated head-splitting fractures, - . Humeral head depression with more than 40% of joint surface depressed, - . Interlocking chronic dislocation with deep HILL-SACHS lesion, - . Fracture instability following internal fixation attempt in 3-fragment fractures (secondary dislocation, material loosening), - Posttraumatic humeral head necrosis, ● - . Omarthrosis. AGLON® CTA heads are destined for treatment of stable types of rotator cuff tear arthropathy. In order to achieve satisfactory results with the CTA heads the fornix humeri and the subscapularis tendon must be intact. A CTA cap is intended for the use as a hemi-arthroplasty, to treat a patient after an inverse shoulder has failed. It is not combined with a glenoid implant. It can be used in primary and revision cases. The main indications for the implantation of an AGILON® inverse (reverse) shoulder prosthesis are: - . Rotator cuff tear arthropathy, - . Chronic trauma shoulder, - . Decentering of the humeral head after implantation of a humeral head prosthesis. Please note, that the patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary. In case of revision surgery the available bone stock has to be evaluated to allow for implantation of well-fixed stems. Conversion of the system can be performed in revision cases as follows: - From Hemi Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty ● - . From Hemi Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder Arthroplasty - From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi Shoulder Arthroplasty as salvage procedure ● - . From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty as salvage procedure The surgeon decides which version of prosthesis for the individual patient is used. This decision depends on several factors, such as the age and the patient's weight, bone quality, shape of the bone and deformation of the joint. The provision of prostheses is generally indicated only in patients whose skeleton is fully grown.
Device Story
Modular shoulder replacement system; provides hemi and reverse shoulder configurations. Components include humeral heads (caps/CTA caps), humeral stems, glenoid baseplates, and glenospheres. Used in orthopedic surgery for joint replacement; surgeon selects components based on patient anatomy, bone quality, and joint deformation. System replaces damaged joint surfaces; restores function in patients with degenerative disease, fractures, or rotator cuff arthropathy. Enables primary and revision procedures; allows conversion between hemi and reverse configurations. Benefits include pain relief and restoration of shoulder joint function.
Clinical Evidence
No clinical data. Performance characterized through bench testing: metallic bone screw testing (ASTM F543), glenoid loosening/disassociation (ASTM 2028), disassembly testing (ASTM F1820), axial disassembly of taper connections (ASTM F2009), fatigue, fretting, corrosion, metal ion analysis, microstructural analysis, range of motion evaluation, and biocompatibility testing.
Technological Characteristics
Modular shoulder prosthesis; materials include metal and polymer. Testing performed per ASTM F543 (bone screws), ASTM 2028 (glenoid loosening), ASTM F1820 (disassembly), and ASTM F2009 (taper connections). Cementless fixation. System includes humeral stems, heads, glenoid baseplates, and glenospheres.
Indications for Use
Indicated for patients with fully grown skeletons requiring cementless inverse (reverse) total or hemi shoulder replacement. Conditions include non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), post-traumatic osteoarthritis, fractures (multifragmental, 3/4-fragment, head-splitting), rheumatoid arthritis, rotator cuff tear arthropathy, chronic trauma shoulder, and humeral head decentering. Requires intact deltoid muscle and anatomically suitable joint. CTA heads require intact fornix humeri and subscapularis tendon.
Regulatory Classification
Identification
A shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis is a device made of alloys, such as cobalt-chromium-molybdenum. It has an intramedullary stem and is intended to be implanted to replace the articular surface of the proximal end of the humerus and to be fixed without bone cement (§ 888.3027). This device is not intended for biological fixation.
Delta CTA Reverse Shoulder System Humeral Heads (K062116)
Submission Summary (Full Text)
{0}------------------------------------------------
November 13, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Implantcast GmbH Dave McGurl Director, Regulatory Affairs Mrca, LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001 USA
Re: Re: K191433
Trade/Device Name: AGILON® XO Shoulder Replacement System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: HSD, PHX Dated: November 9, 2020 Received: November 9, 2020
Dear Dave McGurl:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 12, 2020. Specifically, FDA is updating the SE Letter as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Michael Owens, OHT6: Office of Orthopedic Devices, 301-796-5650, Michaelc.Owens@fda.hhs.gov.
Sincerely,
Digitally signed by Farzana Farzana Sharmin -S Sharmin -S Date: 2020.11.13 13:19:49 -05'00
For Michael Owens Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
November 12, 2020
Image /page/1/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Implantcast GmbH Dave McGurl Director, Regulatory Affairs MRCA, LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001 USA
### Re: K191433
Trade/Device Name: AGILON® XO Shoulder Replacement System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: KWS, HSD, PHX, MBF Dated: November 9, 2020 Received: November 9, 2020
Dear Dave McGurl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies product submissions. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical
{2}------------------------------------------------
device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Figure/6 description: The image shows a digital signature. The signature is for "Farzana Sharmin -S". The date of the signature is November 12, 2020. The time of the signature is 17:59:26 -05'00'.
For Michael Owens Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K191433
#### Device Name
AGILON® XO Shoulder Replacement System
#### Indications for Use (Describe)
The AGILON® XO Shoulder Replacement System is indicated for use for cementless inverse (reverse) total or hemi shoulder replacement in cases of:
- Non- inflammatory degenerative joint disease including osteoarthritis and avascular necrosis, .
- . Post-traumatic osteoarthritis.
- Fractures,
- . Rheumatoid arthritis.
The main indications for the implantation of an AGILON® hemi shoulder prosthesis are:
- Multifragmental comminuted fractures of the humeral head,
- 3- and 4-Fragment-fractures of the proximal humerus,
- . Head-splitting fractures,
- . Dislocated head-splitting fractures,
- . Humeral head depression with more than 40% of joint surface depressed,
- . Interlocking chronic dislocation with deep HILL-SACHS lesion,
- . Fracture instability following internal fixation attempt in 3-fragment fractures (secondary dislocation, material loosening),
- Posttraumatic humeral head necrosis, ●
- . Omarthrosis.
AGLON® CTA heads are destined for treatment of stable types of rotator cuff tear arthropathy. In order to achieve satisfactory results with the CTA heads the fornix humeri and the subscapularis tendon must be intact. A CTA cap is intended for the use as a hemi-arthroplasty, to treat a patient after an inverse shoulder has failed. It is not combined with a glenoid implant. It can be used in primary and revision cases.
The main indications for the implantation of an AGILON® inverse (reverse) shoulder prosthesis are:
- . Rotator cuff tear arthropathy,
- . Chronic trauma shoulder,
- . Decentering of the humeral head after implantation of a humeral head prosthesis.
Please note, that the patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary.
In case of revision surgery the available bone stock has to be evaluated to allow for implantation of well-fixed stems. Conversion of the system can be performed in revision cases as follows:
- From Hemi Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty ●
- . From Hemi Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder Arthroplasty
- From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi Shoulder Arthroplasty as salvage procedure ●
- . From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty as salvage procedure
{4}------------------------------------------------
The surgeon decides which version of prosthesis for the individual patient is used. This decision depends on several factors, such as the age and the patient's weight, bone quality, shape of the bone and deformation of the joint.
The provision of prostheses is generally indicated only in patients whose skeleton is fully grown.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
. .
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(K) SUMMARY
| Device Trade Name: | AGILON® XO Shoulder Replacement System |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | implantcast, GmbH<br>Lüneburger Schanze 26<br>21614 Buxtehude, Germany |
| Contact: | Ms. Juliane Höppner<br>Head of Regulatory Affairs<br>Phone: +49 4161 744 135<br>Email: j.hoeppner@implantcast.de |
| Prepared by: | Mr. Dave McGurl<br>Director, Regulatory Affairs<br>MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20005<br>Phone: 202.552.5800<br>dmcgurl@mcra.com |
| Date Prepared: | November 12, 2020 |
| Classification: | 21 CFR 888.3660, Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis<br>21 CFR 888.3690, Shoulder joint humeral (hemi-shoulder) metallic<br>uncemented prosthesis |
| Class: | II |
| Product Code: | HSD, PHX |
| Primary Predicate Device: | DePuy Global UNITE Shoulder System (K170748) |
| Additional Predicates: | Stryker ReUnion Systems (K183039)<br>Medacta Shoulder System (K170910, K170452)<br>Delta CTA Reverse Shoulder System Humeral Heads (K062116) |
### Indications for Use:
The AGILON® XO Shoulder Replacement System is indicated for use for cementless inverse (reverse) total or hemi shoulder replacement in cases of:
- Non- inflammatory degenerative joint disease including osteoarthritis and avascular necrosis, .
{6}------------------------------------------------
- Post-traumatic osteoarthritis. ●
- Fractures, ●
- Rheumatoid arthritis. ●
The main indications for the implantation of an AGILON® hemi shoulder prosthesis are:
- Multifragmental comminuted fractures of the humeral head, ●
- 3- and 4-Fragment-fractures of the proximal humerus, ●
- Head-splitting fractures, ●
- Dislocated head-splitting fractures. ●
- Humeral head depression with more than 40% of joint surface depressed, ●
- Interlocking chronic dislocation with deep HILL-SACHS lesion, ●
- Fracture instability following internal fixation attempt in 3-fragment and 4-fragment fractures (secondary dislocation, material loosening),
- Posttraumatic humeral head necrosis, o
- Omarthrosis. ●
AGILON® CTA heads are destined for treatment of stable types of rotator cuff tear arthropathy. In order to achieve satisfactory results with the CTA heads the fornix humeri and the subscapularis tendon must be intact. A CTA cap is intended for the use as a hemi-arthroplasty, to treat a patient after an inverse shoulder has failed. It is not combined with a glenoid implant. It can be used in primary and revision cases.
The main indications for the implantation of an AGILON® inverse (reverse) shoulder prosthesis are:
- Rotator cuff tear arthropathy,
- Chronic trauma shoulder, ●
- Decentering of the humeral head after implantation of a humeral head prosthesis. ●
Please note, that the patient's joint must be anatomically and structurally suited to receive the selected implant(s), and a functional deltoid muscle is necessary.
In case of revision surgery the available bone stock has to be evaluated to allow for implantation of well-fixed stems. Conversion of the system can be performed in revision cases as follows:
- From Hemi Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty ●
- From Hemi Shoulder Arthroplasty to Inverse (Reverse) Total Shoulder Arthroplasty ●
- From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi Shoulder Arthroplasty as ● salvage procedure
- From Inverse (Reverse) Total Shoulder Arthroplasty to Hemi CTA Shoulder Arthroplasty o as salvage procedure
The surgeon decides which version of prosthesis for the individual patient is used. This decision depends on several factors, such as the age and the patient's weight, bone quality, shape of the bone and deformation of the joint.
The provision of prostheses is generally indicated only in patients whose skeleton is fully grown.
{7}------------------------------------------------
#### Device Description:
The AGILON® XO Shoulder Replacement System is a modular shoulder replacement system offering various components that can be combined to replace the shoulder joint with various options depending upon the size and of each patient. The components can be combined into an hemi or reverse shoulder:
### Hemi Shoulder:
- Humeral Head Components (AGILON® XO Cap and AGILON® XO CTA Cap) ●
- o Humeral Stems (Cementless)
#### Reverse Shoulder:
- Humeral Head Components (AGILON® Cap Inverse and AGILON® PE-Inlay) ●
- o Glenoid and Glenosphere Components (AGLION® XO Glenoid Baseplate and AGILON® Glenosphere)
- o Humeral Stems (Cementless)
Metaphyseal components, stem extension pieces, and component connection fixation screws are intended to be used for hemi and reverse shoulder configurations.
#### Substantial Equivalence:
The AGILON® XO Shoulder Replacement System is substantially equivalent to the predicate devices cited on the previous page with respect to intended use, design, and materials.
#### Performance Testing:
All necessary testing has been performed for the worst-case configuration of the AGILON® XO Shoulder Replacement System to assure substantial equivalence to its predicates and to demonstrate the subject devices perform as intended. All testing was performed on test units representative of finished devices. The performance of the AGILON® XO Shoulder Replacement System was characterized through the following tests:
- Testing of Metallic Medical Bone Screws (ASTM F543)
- Torsional characterization, driving torque, and pull-out strength testing o
- Glenoid Loosening or Disassociation (ASTM 2028)
- o Subluxation and cyclic testing
- Disassembly Testing (ASTM F1820) ●
- o Axial disassembly, offset pull-out, and torque out disassembly testing
- Axial Disassembly of Taper Connections (ASTM F2009) ●
- 0 Metaphyseal loosening
- Fatigue, fretting, corrosion, and metal ion analysis testing
- Microstructural Analysis ●
- Range of motion evaluation ●
- Biocompatibility Testing and Risk Assessment ●
{8}------------------------------------------------
### Conclusion:
The AGILON® XO Shoulder Replacement System possesses the same intended use and technological characteristics as the predicate devices. Therefore, the AGILON® XO Shoulder Replacement System is substantially equivalent for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.