K191403 · Nonin Medical, Inc. · DQA · Dec 20, 2019 · Cardiovascular
Device Facts
Record ID
K191403
Device Name
Pulse Oximeter
Applicant
Nonin Medical, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Dec 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nonin® Model 9591Onyx® 3 Finger Pulse Oximeter is a small, lightweight, portable and reusable spot-check device indicated for use in measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly perfused. The Respiration Rate parameter provides a non-invasive measurement of respiration rate, in breaths per minute. For %SpO2, and pulse rate, the 9591 is intended for use in hospitals, clinics, long-term care facilities, skilled nursing facilities, and home healthcare services. It is intended for adult and pediatric patients who are well or poorly perfused, with digits that are between 0.3 - 1.0 inch (0.8 - 2.5 cm) thick, under non-motion conditions. For Respiration rate, the 9591 is intended for use in hospitals, clinics, long-term care facilities, and home healthcare services. It is intended for adult who are well perfused, with digits that are between 0.3 - 1.0 inch (0.8 - 2.5 cm) thick, under non-motions. It is not intended for use in high-acuity environments, such as ICU or operating rooms where continuous monitoring is expected.
Device Story
Small, portable, reusable finger pulse oximeter; measures arterial hemoglobin saturation (%SpO2), pulse rate, and respiration rate. Inputs: red and infrared light absorption through perfused tissue. Operation: light passes through digit; device calculates parameters; displays values on integrated color LCD. Used in hospitals, clinics, long-term care, and home settings; operated by clinicians or patients for spot-checks. Provides visual pulse signal and poor pulse quality indicators. Output assists healthcare providers in assessing patient oxygenation and respiratory status; facilitates non-invasive monitoring without high-acuity requirements.
Clinical Evidence
Clinical testing included a respiratory rate accuracy study (n=30) comparing device output to capnography, and SpO2 accuracy testing (70-100% SaO2 range) on healthy subjects compared to co-oximetry. SpO2 accuracy measured via root-mean-squared (ARMS) values. Results demonstrated performance compliance with ISO 80601-2-61 and equivalency to predicate devices.
Technological Characteristics
Materials: Polycarbonate/Polyester enclosure, TPE finger pads. Sensing: Red (660nm) and IR (910nm) light absorption. Power: 3V DC (2x AAA batteries). Connectivity: Bluetooth 4.0. Form factor: Portable finger-clip. Software: Embedded firmware. Safety: Type BF, IP32 ingress protection. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-61, ISO 10993-1.
Indications for Use
Indicated for adult and pediatric patients (well or poorly perfused) for spot-check measurement of %SpO2 and pulse rate. Indicated for adult patients (well perfused) for spot-check measurement of respiration rate. Requires digit thickness 0.3–1.0 inch. Use restricted to non-motion conditions; not for high-acuity environments (ICU/OR).
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Nonin Model 3230 Bluetooth Smart Pulse oximeter (K131021)
Reference Devices
Nellcor Bedside Respiratory Patient Monitoring System (PM1000N) (K141518)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nonin Medical, Inc. Walter Holbein Manager, RA/CA 13700 1st Avenue North Plymouth, Minnesota 55441
Re: K191403
Trade/Device Name: Nonin Onyx 3, Model 9591 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: November 20, 2019 Received: November 25, 2019
## Dear Walter Holbein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191403
Device Name Nonin Onyx 3, Model 9591
#### Indications for Use (Describe)
The Nonin® Model 9591Onyx® 3 Finger Pulse Oximeter is a small, lightweight, portable and reusable spot-check device indicated for use in measuring and displaying function of arterial hemoglobin (%SpO2) and pulse rate of patients who are well or poorly perfused. The Respiration Rate parameter provides a non-invasive measurement of respiration rate, in breaths per minute.
For %SpO2, and pulse rate, the 9591 is intended for use in hospitals, clinics, long-term care facilities, skilled nursing facilities, and home healthcare services. It is intended for adult and pediatric patients who are well or poorly perfused, with digits that are between 0.3 – 1.0 inch (0.8 –2.5 cm) thick, under non-motion conditions.
For Respiration rate, the 9591 is intended for use in hospitals, clinics, long-term care facilities, and home healthcare services. It is intended for adult who are well perfused, with digits that are between 0.3 – 1.0 inch (0.8 - 2.5 cm) thick, under non-motions. It is not intended for use in high-acuity environments, such as ICU or operating rooms where continuous monitoring is expected.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1. 510(k) Summary: K191403
| Submitter: | Nonin Medical, Inc.<br>13700 1st Ave. North<br>Plymouth, MN 55441-5443 |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Walter Holbein<br>RA/CA Manager |
| Phone: | 763-577-5504 |
| Fax: | 763-553-7807 |
| Date Prepared: | 20 May, 2019 |
| Trade Name: | Onyx® 3 Model 9591 Finger Pulse Oximeter |
| Common Name: | Finger Pulse Oximeter |
| Classification Name: | Oximeter |
| Regulation Number: | Class II, 21 CFR 870.2700 |
| Product Code, Panel: | DQA, Anesthesiology |
| Predicate Device(s): | Nonin Model 3230 Bluetooth® Smart Pulse oximeter cleared by the<br>FDA under K131021 on 9/11/2013, and |
| Reference Device: | Nellcor™ Bedside Respiratory Patient Monitoring System<br>(PM1000N) cleared by the FDA under K141518 on 3/5/2015. |
## Device Description
The Model 9591 Finger Pulse Oximeter is a small, lightweight, portable, reusable, digit pulse oximeter that displays numerical values for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate by measuring the absorption of red and infrared (IR) light passing through perfused tissue. The SpO2, pulse rate, and respiration rate are displayed on the LCD display contained within the device. A color LCD provides a visual indication of the pulse signal, while blinking at the corresponding pulse rate. The display will indicate of poor pulse quality that may affect the readings. The Respiration Rate parameter provides a non-invasive measurement of respiration rate, in breaths per minute. It is intended for spot-checking of adult and pediatric patients who are well or poorly perfused with digits that are between 0.3 - 1.0 inch (0.8 - 2.5 cm) thick.
{4}------------------------------------------------
| Intended Use: | Model 9591 |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Nonin® Model 9591Onyx® 3 Finger Pulse Oximeter is a small,<br>lightweight, portable and reusable spot-check device indicated for<br>use in measuring and displaying functional oxygen saturation of |
| | arterial hemoglobin (%SpO2) and pulse rate of patients who are well<br>or poorly perfused. The Respiration Rate parameter provides a non-<br>invasive measurement of respiration rate, in breaths per minute. |
| | For %SpO2, and pulse rate, the 9591 is intended for use in hospitals,<br>clinics, long-term care facilities, skilled nursing facilities, and home<br>healthcare services. It is intended for adult and pediatric patients |
| | who are well or poorly perfused, with digits that are between 0.3 -<br>1.0 inch (0.8 - 2.5 cm) thick, under non-motion conditions. |
| | For Respiration rate, the 9591 is intended for use in hospitals, |
| | clinics, long-term care facilities, skilled nursing facilities, and home |
| | healthcare services. It is intended for adult who are well perfused, |
| | with digits that are between 0.3 - 1.0 inch (0.8 - 2.5 cm) thick, |
| | under non-motion conditions. It is not intended for use in high-<br>acuity environments, such as ICU or operating rooms where<br>continuous monitoring is expected. |
| Testing: | Nonin's Model 9591 Finger Pulse Oximeter has successfully<br>undergone both laboratory and clinical hypoxia accuracy testing in<br>order to ensure that it has appropriate performance, functional<br>features to fully comply with ISO 80601-2-61:2011 and is<br>substantially equivalent to the predicate devices. |
| Functional and<br>Safety Testing: | The results of the testing demonstrate equivalency with the predicate<br>devices and compliance to recognized standards. Table 1<br>summarizes test results for the proposed devices, which met the<br>relevant requirements of the applicable recognized standards. |
{5}------------------------------------------------
| Test | Reference | Result |
|-------------------------------------------|--------------------------------------------------------------------------------------------|--------|
| Electrical Safety | IEC 60601-1 | Pass |
| Temperature and Humidity | IEC 60601-1 | Pass |
| Atmospheric Pressure (Altitude) | IEC 60601-1 | Pass |
| Electromagnetic Immunity and<br>Emissions | IEC 60601-1-2 | Pass |
| Performance | ISO 80601-2-61<br>IEC 60601-1<br>IEC 60601-1-6<br>IEC 60601-1-11<br>IEC 62304<br>ISO 14155 | Pass |
| Ingress Protection | ISO 80601-2-61 | Pass |
| Diaphoretic related ingress | Internal performance<br>characterization | Pass |
| Mechanical Durability | IEC 60601-1<br>ISO 80601-2-61 | Pass |
| Biocompatibility | ISO 10993-1 | Pass |
#### Table 1
#### Clinical Testing: Respiratory Rate Accuracy Verification Clinical (QATR10912)
Respiratory rate accuracy testing was a comparative single-center, randomized study. The respiratory rate output of the Onyx 3 fingertip oximeter was compared to Capnography based respiratory rate. This was a minimal risk study using a non-significant risk device. A total of 30 subjects were enrolled in the study. Up to 3 subjects were enrolled per day. Subject participation lasted up to 2 hours. Demographic and anthropometric data (date of birth, gender, ethnicity, race, height, weight, skin tone and finger dimensions) were collected for all subjects. Accuracy data was calculated using both mean error and root mean square error (RMSE).
### SpO2 Accuracy Testing
SpO2 accuracy testing was conducted at an independent research laboratory on healthy, male and female, non-smoking, light to darkskinned subjects that were 18 years of age and older. The measured oxygen saturation value (SpO2) of the device was compared to simultaneous arterial blood samples as assessed by co-oximetry. The accuracy of the device in comparison to the co-oximeter samples was measured over the SaO2 range of 70-100% in motion and non-motion conditions. Accuracy data was calculated using the root-mean-squared (ARMS value) for all subjects.
{6}------------------------------------------------
# Summary of Substantial Equivalence:
The Model 9591 Onyx® 3 Finger Pulse Oximeter has the following similarities to its respective predicate Nonin device:
- Similar SpO2 and Pulse Rate ranges ●
- . Identical SpO2 and Low Perfusion SpO2 accuracy specifications
- Identical critical optics technology
- Perform equivalently to the similar specifications
The following lists the differences and the rationale for those differences between the proposed device and the predicate devices:
- Indications for Use: the proposed and predicate Nonin devices have identical Indications for Use and identical patient population, with the exception of the respiratory rate feature in the proposed device. The Nellcor device is a tabletop device with the additions of an alarm feature and expanded patient population, which Nonin is not seeking claims for.
- Respiratory Rate: the Nonin predicate device does not have the respiratory rate feature. The Nonin proposed device and the Nellcor predicate have similar respiratory rate ranges.
- SpO2 accuracy with Motion: Nonin is not seeking motion claims similar to the predicate Nellcor device.
- Pulse accuracy with Motion: Nonin is not seeking motion claims similar to the predicate Nellcor device.
- Enclosure Ingress Protection: the proposed and predicate ● Nonin devices are identical and offer more ingress protection than the Nellcor device.
- Battery Input: the proposed and predicate Nonin devices are identical. The Nellcor device is a tabletop-type device with a rechargeable battery (Li-Ion battery). The Nonin devices do not use rechargeable batteries.
{7}------------------------------------------------
| Category | Identical/Similar/<br>Different<br>(rationale) | Predicate: Model 3230<br>(K131021, cleared 9/11/15) | Predicate: Nellcor™ Bedside<br>Respiratory Patient Monitoring<br>System<br>(K141518, cleared 3/5/15) | Proposed: Model 9591 | | | | | |
|----------------------------------------------------------------|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|---------------|--|-----------|--|
| | | Indications for Use | | | | | | | |
| Indications for use and<br>Intended Use from<br>Labeling | Different | The Model 3230 Finger Pulse<br>oximeter is a small, lightweight<br>portable device indicated for use<br>in measuring and displaying<br>functional oxygen saturation of<br>arterial hemoglobin (%SpO2) and<br>pulse rate of patients who are well<br>or poorly perfused. It is intended<br>for spot checking of adult and<br>pediatric patients on digits<br>between 0.3-1.0 inch (0.8-2.5<br>cm) thick. | The Nellcor™ Bedside<br>Respiratory Patient Monitoring<br>System is a portable pulse<br>oximeter intended for prescription<br>use only as a continuous non-<br>invasive monitor of arterial<br>oxygen saturation (SpO₂) and<br>pulse rate of adult, pediatric, and<br>neonatal patients during both no<br>motion and motion conditions,<br>and for patients who are well or<br>poorly perfused. The monitoring<br>system is intended for use in<br>hospitals, hospital-type facilities,<br>and during intra-hospital<br>transport. The OxiMax SPD™<br>Alert (SPD) feature is intended<br>only for facility-use care of<br>adults to detect patterns of<br>desaturation indicative of<br>repetitive reductions in<br>airflow through the upper airway<br>and into the lungs. | The Nonin® Model 9591 Finger<br>Pulse Oximeter is a small,<br>lightweight, portable spot-check<br>device indicated for use in<br>measuring and displaying<br>functional oxygen saturation of<br>arterial hemoglobin (%SpO2) and<br>pulse rate of patients who are well<br>or poorly perfused. The<br>Respiration Rate parameter<br>provides a non-invasive<br>measurement of respiration rate,<br>in breaths per minute. For %<br>SpO2, pulse rate, and respiration<br>rate the 9591 is intended for use<br>in professional healthcare and<br>home healthcare settings in adult<br>and pediatric patients who are<br>well or poorly perfused, with<br>digits that are between 0.3-1.0<br>inch (0.8-2.5 cm) thick. | | | | | |
| SpO₂ Range | Similar | 0% to 100% SpO₂ | 1% to 100% SpO₂ | 0% to 100% SpO₂ | | | | | |
| Pulse Rate Range | Similar | 18-321 BPM | 20 to 250 BPM | 18-321 BPM | | | | | |
| Respiration Rate | Different | NA | 4 to 40 breaths/minute | 3 to 44 breaths/minute | | | | | |
| | | SpO₂ Accuracy | | | | | | | |
| Adult /Pediatric SpO₂ | Identical | $\pm$ 2 digits ( $\pm$ 1 Arms) | 70 to 100% $\pm$ 2 digits | $\pm$ 2 digits ( $\pm$ 1 Arms) | | | | | |
| Low Perfusion SpO₂ | Identical | $\pm$ 2 digits ( $\pm$ 1 Arms) | 70 to 100% $\pm$ 2 digits | $\pm$ 2 digits ( $\pm$ 1 Arms) | | | | | |
| Adult /Pediatric SpO₂<br>with Motion | Different | NA | 70 to 100% $\pm$ 3 digits | NA | | | | | |
| | | Pulse Rate Accuracy | | | | | | | |
| Adult and Pediatric<br>Pulse Rate | Identical | 20 to 250 BPM $\pm$ 3 digits | 20 to 250 BPM $\pm$ 3 digits | 20 to 250 BPM $\pm$ 3 digits | | | | | |
| Low Perfusion Pulse<br>Rate | Similar | 40 to 240 BPM $\pm$ 3 digits | 20 to 250 BPM $\pm$ 3 digits | 40 to 240 BPM $\pm$ 3 digits | | | | | |
| Adult/Pediatric Pulse<br>Rate with Motion | Different | NA | 48 to 127 BPM $\pm$ 5 digits | NA | | | | | |
| | | Device Specifications | | | | | | | |
| Measurement<br>Wavelength | Identical | 660 and 910 nanometers | 660 and 900 nanometers | 660 and 910 nanometers | | | | | |
| Operating Altitude | Similar | Up to 13,123 feet | -1,000 ft. to 15,000 ft. | Up to 13,123 feet | | | | | |
| Operating Temperature<br>Storage/Transportation<br>Temperature | Similar | -5° to +40°C | 5° to +40°C | -5° to +40°C | | | | | |
| Operating Humidity | Similar | -40° to +70°C | -20° to +70°C | -40° to +70°C | | | | | |
| Storage/Transportation<br>Humidity | Similar | 10 - 95% non-condensing | 15 - 95% non-condensing | 10 - 95% non-condensing | | | | | |
| Power Requirements | Similar | 3 volts DC | 7.2 volts DC | 3 volts DC | | | | | |
| Battery Life<br>(Operating) | Similar | 2200 Spot Checks | 6 hours | 2000 Spot checks 25 hours<br>continuous | | | | | |
| Battery Life (Storage) | Different | 1 month, with batteries installed | 4 months | 1 month, with batteries installed | | | | | |
| Electrical - Type and<br>Degree of Protection | Identical | Type BF Internally powered | Type BF Internally powered | Type BF Internally powered | | | | | |
| Enclosure - Degree of<br>Ingress Protection | Different | IP32 | IPX1 | IP32 | | | | | |
| Modes of Operation | Similar | Spot-check | Continuous | Spot-check | | | | | |
| Battery Input | Different | Two 1.5 volt AAA-size batteries | 30 VAC maximum | Two 1.5 volt AAA-size batteries | | | | | |
| 7-Segment 3-Digit<br>Displays | Identical | Multi-pixel 3-Digit Displays<br>(color LCD) | Not specified | Multi-pixel 3-Digit Displays<br>(color LCD) | | | | | |
| Pulse Strength Indicator | | Identical | | Color LCD | | Not specified | | Color LCD | |
| Sensor Fault Indicator | Identical | Poor signal symbol followed by dashes | Not specified…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.