K191283 · Suzhou Melodicare Medical Technology Co., Ltd. · FLL · Oct 18, 2019 · General Hospital
Device Facts
Record ID
K191283
Device Name
Infrared Thermometer, Breeze model
Applicant
Suzhou Melodicare Medical Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Oct 18, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Intended Use
The infrared thermometer is intended for the measurement and monitoring of human body temperature by physician or lay person. The device is not intended for infants less than three months.
Device Story
Breeze Infrared Thermometer measures human body temperature via infrared energy radiation from forehead or ear; utilizes Fresnel lens to focus infrared energy onto internal sensors. Device comprises infrared sensors, signal receiving processor, buttons, buzzer, and LCD display. Powered by 2x 1.5V AAA batteries. Used in home settings and medical institutes by physicians or lay persons. User selects forehead or ear mode; ear mode requires manufacturer-provided ear cover. Temperature data displayed on LCD; device stores up to 32 measurements. Provides non-invasive, intermittent temperature monitoring to assist clinical decision-making regarding patient health status.
Clinical Evidence
Clinical study conducted per ASTM E 1965-98 (2009). Randomized, simple blind homologous control study with minimum 144 subjects across age groups defined in ISO 80601-2-56. Non-clinical data includes biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2), and performance effectiveness (ISO 80601-2-56).
Technological Characteristics
Infrared radiation detection; Fresnel lens focusing. Materials: ABS, PE. Dimensions: 161.5mm x 35.8mm x 55.9mm. Power: 2x 1.5V AAA batteries. Measurement range: 34.0°C–42.4°C. Accuracy: ±0.2°C. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ASTM E1965-98, ISO 10993-5, ISO 10993-10, ISO 80601-2-56. Non-sterile.
Indications for Use
Indicated for measurement and monitoring of human body temperature by physicians or lay persons. Contraindicated for infants less than three months old.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
K100953 — MICROLIFE NON-CONTACT INFRARED FOREHEAD THERMOMETER, MODEL FR1DZ1 · Microlife Intellectual Property GmbH · May 17, 2010
K213079 — Ear Thermometer, Model ET001 · Guangzhou Berrcom Medical Device Co., Ltd. · Feb 18, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 18, 2019
Suzhou Melodicare Medical Technology Co.,Ltd. Xiaoqian Zou General Manager Building Room 301, No 32 Hongxi Road, Suzhou High-Tech Zone Suzhou. 215163 China
Re: K191283
Trade/Device Name: Infrared Thermometer, Breeze model Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: September 11, 2019 Received: September 20, 2019
Dear Xiaoqian Zou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191283
Device Name Infrared thermometer, model Breeze
Indications for Use (Describe)
The infrared thermometer is intended for the measurement and monitoring of human body temperature by physician or lay person. The device is not intended for infants less than three months.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Suzhou Melodicare Medical Technology Co., Ltd.<br>Building Room 301, no. 32 Hongxi road, Suzhou high-tech zone, Jiangsu province, China<br>Tel.: +86 -791-85216384 |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Xiaoqian Zou |
| Prepare date: | October 17, 2019 |
| 2. Device name and classification: | Device Name: Infrared Thermometer<br>Models: Breeze<br>Classification Name:<br>21 CFR 880.2910<br>Clinical Electronic Thermometers<br>Product code: FLL<br>Regulatory Class: Class II |
| 3. Reason for Submission | New Application. |
| 4. Predicate Device(s): | Shenzhen Brav Electronic Technologies Co., Ltd., EFT-165 Infrared Thermometer /K171214 |
| 5. Device Description: | Breeze Infrared Thermometer measure the body temperature through receiving infrared energy radiation via the ear or forehead. These thermometers have the capability to measure temperature via forehead temperature mode or ear temperature mode, and the temperature is directly shown on the LCD display.<br>The device is intended to be used in the home setting and medical institutes.The product is mainly composed of infrared temperature sensors, signal receiving processor, buttons, buzzer, LCD display. It is powered by 2 X 1.5 AAA batteries.It focuses the infrared energy received from the human's forehead or ear by using the Fresnel lens of the thermometer.When measuring the ear temperature, the ear cover provided by the manufacturer must be used. |
| 6. Indications for Use: | The infrared thermometer is intended for the measurement and monitoring of human body temperature by physician or lay person. The device is not intended for infants less than three months. |
{4}------------------------------------------------
# 7. Predicate Device Comparison
In comparison to the predicate device, the subject device has the same intended use, similar product design, same performance effectiveness, and performance safety as the predicate device.
Please refer to the following table that notes differences between the subject device and predicate device. All the differences do not affect the basic design principle, usage, effectiveness and safety of the subject device. No new questions are raised regarding effectiveness and safety.
| ITEM | Proposed Device<br>Breeze | Predicate Device<br>EFT-165/ K171214 | Comparison<br>Result |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacture | Suzhou Melodicare Medical<br>Technology Co., Ltd. | Brav Electronic Technologies<br>Co., Ltd. | --- |
| Indications for<br>Use | The infrared thermometer is<br>intended for the measurement<br>and monitoring of human body<br>temperature by physician or<br>lay person. The device is not<br>intended for infants less than<br>three months. | The infrared thermometer is<br>intended for the measurement<br>and monitoring of human body<br>temperature by doctor or<br>customers in the hospital or<br>home. | Different¹ |
| Operational Specifications | | | |
| Operational<br>Principle | Infrared radiation detection | Infrared radiation detection | Same |
| Measuring<br>Mode | Forehead and ear | Forehead and ear | Same |
| Measurement<br>Range | 34.0℃~42.4℃<br>(93.2℉~108.3℉) | 32.0℃~42.9℃<br>(89.6℉~109.2℉) | Different ² |
| Measurement<br>Distance | 0 cm | 0 cm | Same |
| Accuracy | ±0.2℃(0.4℉) | ±0.2℃(0.4℉) | Same |
| Memory Data<br>Limit | memorize 32 measurements<br>automatically | memorize 20 measurements<br>automatically | Same |
| Product<br>configuration | It is mainly composed with<br>infrared sensor, signal<br>receiving processor, buttons,<br>buzzer, LCD display, battery<br>and etc. | It is mainly composed with<br>infrared sensor, signal<br>receiving processor, buttons,<br>buzzer, LCD display, battery<br>and etc. | Same |
| Temperature<br>unit and<br>conversion | Dual temperature units "℃"<br>and "℉" optional, and the two<br>units can convert by the<br>conversion key automatically | Dual temperature units "℃"<br>and "℉" optional, and the two<br>units can convert by the<br>conversion key automatically | Same |
| Bluetooth<br>Compatible | NO | YES | Different 3 |
| Applicable<br>Standards | IEC 60601-1, IEC 60601-1-2,<br>IEC 60601-1-11, and ASTM<br>E1965-98, ISO 10993-5, ISO<br>10993-10, ISO 80601-2-56 | IEC 60601-1, IEC 60601-1-2,<br>IEC 60601-1-11, and ASTM<br>E1965-98, ISO 10993-5, ISO<br>10993-10, ISO 80601-2-56 | Same |
| Display | 0.1℃/℉ LCD | 0.1℃/℉ LCD | Same |
| Operating<br>Environment | 15°C ~ 35°C ( 59°F to 95 °F)<br>≤85% moisture condensation | 15°C ~ 35°C ( 59°F to 95 °F)<br>≤85% moisture condensation | Same |
| Storage<br>Environment | -20°C~55°C (-4°F~131°F)<br>≤90% moisture condensation | -20°C~55°C (-4°F~131°F)<br>≤90% moisture condensation | Same |
| Power supply | 2 X 1.5V AAA | 2 X 1.5V AAA | Same |
| Physical Specifications | | | |
| Weight | 62 g (without battery) | 72 g (without battery) | Different 4 |
| Dimensions<br>(length x width x height) | 161.5mm *35.8mm *55.9mm | 152mm * 44 mm *30mm | |
| Biological Specifications | | | |
| Patient<br>Contacting<br>Materials | ABS, PE | ABS, PE | Same |
| Patient<br>Contacting | Surface-contacting, less than 24 h | Surface-contacting, less than 24 h | Same |
Table 1 Comparison between the main predicate device EFT-165 and the subject device
{5}------------------------------------------------
#### Justification for the differences:
- 1) Different Indications for Use
As indicated in the comparison table, the subject device can be used both in hospital and home, which the predicate can be used in both scenarios. The difference is just the different language, so the different will not introduce any new questions of safety and effectiveness. The age range of the patients for the subject device is based on the conducted clinical testing.
- 2) Different Measurement Range
The measurement range of the subject device is 34.0℃~42.4℃, minor different from that of the predicate device 32.0℃~42.9℃. The difference will not affect the clinical safety and effectiveness since the range can cover the temperature range during intended application scenarios, and the subject device has been verified to meet the accuracy requirements defined in the standards ASTM E 1965-98 and ISO 80601-2-56.
- 3) Different Bluetooth function
The subject device is not equipped with the Bluetooth function while the predicate does, but absence of this function will not raise new safety and effectiveness.
- 4) Different Physical Specifications
The subject and the predicate are of different size but proximity. Moreover, such engineering design has been verified during the international standards, so such minor different ill not raise any safety and effectiveness questions.
As seen in the comparison tables, the subject and predicate devices have almost the same design features and performance specifications. The only differences between the subject and predicate devices are the indications for use (which link specific age groups with specific modes of temperature reading) and the appearance and operating/storage environment, which do not raise different questions of safety or effectiveness. Moreover, as demonstrated in the non-clinical and clinical testing, the different indications and technological characteristics do not affect the safety and effectiveness of the Breeze Infrared thermometer system.
{6}------------------------------------------------
### 8. Performance Testing:
Performance data includes "Non-Clinical Data" and "Clinical Data". Summary of the data provided is below:
#### Non-Clinical Data:
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility testinq
The biocompatibility evaluation for the Breeze Infrared thermometer was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered tissue-contacting for a duration of less than 24 hours. And Cytotoxicity, Skin Sensitization and Skin Irritation tests were conducted.
#### Reprocessing: Cleaning and Disinfection
The sterilization and shelf life information provided are acceptable because the subject devices are non-sterile devices. The validation testing results for cleaning and disinfection of the subject devices are adequate and acceptable.
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Breeze Infrared thermometer device. consisting of all the modules and accessories in the system complies with IEC60601-1: 2012, Medical electrical equipment Part 1: General requirements for basic safety and essential performance for safety, IEC 60601-1-11: 2015 MEDICAL ELECTRICAL EQUIPMENT –Part 1-11: General requirements for basic safety and essential performance -Collateral Standard: Requirements for medical electrical equipment and medical systems used in the home healthcare environment, and the IEC 60601-1-2: 2014 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances – Requirements and tests standard for EMC.
#### Bench Testing
Bench testing was conducted on the Breeze Infrared thermometer device, consisting of all the accessories in the system. The system complies with ISO 80601-2-56: 2009 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement for performance effectiveness.
#### Software Verification and Validation Testing
Software documentation including verification & validation was provided in accordance with FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern.
#### Risk Analysis
A risk analysis was conducted in accordance with ISO 14971: 2007 – Medical devices - Application of risk management to medical devices.
#### Clinical data:
Clinical testing was conducted per ASTM E 1965-98 (2009) Standard Specification for Infrared Thermometers For Intermittent Determination Of Patient Temperature. This clinical study was randomized, with a simple blind homologous control, which consisted of a minimum of 144 subjects with the various age groups as referenced in Section 201.102 of ISO 80601-2-56.
{7}------------------------------------------------
## 9. Substantial Equivalence Conclusion:
Differences between the indications for use and technological characteristics of the subject device compared to the predicate do not raise different questions of safety and effectiveness. The performance of the device is supported by non-clinical testing, and risk management activities. The Suzhou Melodical Technology Co., Ltd. Breeze Infrared Thermometer is Substantially Equivalent (SE) to the Shenzhen Brav Electronic Technologies Co., Ltd., EFT-165 Infrared Thermometer, cleared under K171214.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches