Southern Implants MAX implant is intended for implantation in the maxillary or mandibular molar region where bone exists and the surgeon has determined that the placement of a narrower diameter implant would increase the probability of failure due to poor primary stability, or increased surgical procedures leading to complications. This MAX implant provides support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses or full arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function.
Device Story
Threaded root-form dental implants; external hex, internal tri-lobe, or internal hex interfaces; diameters 6-9mm; lengths 6-11mm. Used by dental surgeons in clinical settings to replace missing teeth; provide support for fixed/removable prostheses; enable immediate loading. Implants made of unalloyed titanium (ASTM F67); grit-blasted surface; some coronal portions unroughened. Includes various abutments (healing, titanium cylinder, cosmetic, gold cylinder, passive, compact conical) and screws. Surgeon selects implant based on bone availability and stability requirements; implants placed in maxillary/mandibular molar regions. Output is a stable foundation for dental restorations; restores chewing function; improves patient quality of life. Non-clinical testing includes engineering, dimensional, and pull-out analysis; biocompatibility established via predicate comparison.
Clinical Evidence
No clinical data included. Substantial equivalence supported by non-clinical bench testing: biocompatibility (referenced), engineering analysis, dimensional analysis, sterilization validation (ISO 11137, ISO 17665), bacterial endotoxin testing (USP 39-NF34), pull-out tests, and surface area analysis.
Technological Characteristics
Threaded root-form dental implants; unalloyed titanium (ASTM F67) or titanium alloy (ASTM F136); grit-blasted surface; external hex, tri-lobe, or internal hex interfaces. Dimensions: 6-9mm diameter, 6-11mm length. Sterilization: sterile (gamma/moist heat) or moist heat for specific abutments. No software or electronic components.
Indications for Use
Indicated for patients requiring dental implants in the maxillary or mandibular molar region where bone is present, specifically when narrower implants risk failure due to poor primary stability or surgical complications. Supports fixed or removable prostheses (single, partial, or full arch) and allows immediate loading on single or splinted restorations when primary stability and occlusal conditions are met.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 9, 2019
Southern Implants (Pty) Ltd Lauranda Breytenbach Head of Regulatory Affairs and Quality 1 Albert Road Irene, Gauteng 0062 SOUTH AFRICA
Re: K191054
Trade/Device Name: Southern Implants MAX Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 6, 2019 Received: September 10, 2019
Dear Lauranda Breytenbach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191054
Device Name Southern Implants MAX Implant System
#### Indications for Use (Describe)
Southern Implants MAX implant is intended for implantation in the maxillary or mandibular molar region where bone exists and the surgeon has determined that the placement of a narrower diameter implant would increase the probability of failure due to poor primary stability, or increased surgical procedures leading to complications. This MAX implant provides support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses or full arch prostheses. It further adds the option for immediate loading on single unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <b> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary Southern Implants MAX Implant System Southern Implants (Pty) Ltd December 9, 2019
## K191054
#### ADMINISTRATIVE INFORMATION Manufacturer Name
Southern Implants (Pty) Ltd 1 Albert Road Irene, Gauteng, 0062 South Africa +27 12 667 1046 Telephone Fax +27 12 667 1029
Official Contact
Lauranda G. Breytenbach Head of Regulatory Affairs and Quality Email: lauranda.b@southernimplants.com
#### DEVICE NAME AND CLASSIFICATION Trade/Proprietary name Common name
Classification name Classification regulation Product Code
Classification Panel Reviewing Branch
Southern Implants MAX Implant System Dental Implant
Endosseous dental implant 21 CFR 872.3640, Class II DZE, NHA
Dental Products Panel Dental Devices Branch
PREDICATE DEVICE INFORMATION The primary predicate device is K071161 The reference devices are K163634, K070905 and K180465
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
Southern Implants MAX implant is intended for implantation in the maxillary or mandibular molar region where bone exists and the surgeon has determined that the placement of a narrower diameter implant would increase the probability of failure due to poor primary stability, or increased surgical procedures leading to complications. This MAX implant provides support for fixed or removable dental prostheses in a single tooth, partially edentulous prostheses or full arch prostheses. It further adds the option for immediate loading on single and splinted multiple unit restorations when good primary stability is achieved and with appropriate occlusal loading, to restore chewing function.
## SUBJECT DEVICE DESCRIPTION
This submission includes threaded root-form dental implants with an external hex and internal tri-lobe interface and mating abutments. The implants are provided in diameters of 6, 7, 8 and 9mm and lengths of 6, 7, 9 and 11mm. Specifically:
- MAX Implants with External Hex Connection .
- o Ø6 mm at lengths of 6, 7, 9, 11 mm
- o Ø7 mm at lengths of 7, 9, 11 mm
- Ø8 mm at lengths of 7, 9, 11 mm O
- Ø9 mm at lengths of 7, 9, 11 mm O
- TRIMAX Implants with Tri-lobe Connection
- Ø7 mm at lengths of 7, 9, 11 mm o
- Ø8 mm at lengths of 7, 9, 11 mm O
- Ø9 mm at lengths of 7, 9, 11 mm O
- PROMAX Implants with Internal Hex Connection ●
- Ø6 mm at lengths of 7, 9, 11 mm O
- Ø7 mm at lengths of 7, 9, 11 mm O
- Ø8 mm at lengths of 7, 9, 11 mm O
- Ø9 mm at lengths of 7, 9, 11 mm O
This submission also includes: Coverscrews in two diameters and implant connection interfaces: Healing Abutments in two implant connection interfaces (tri-lobe and internal hex) with lengths between 3 and 6mm and diameters between 6 and 7.8mm; Titanium Cylinder abutments for temporary restorations in two implant connection interfaces (tri-lobe and internal hex) in engaging and non-engaging configurations; Cosmetic Abutments for permanent restorations in two engaging implant connection interfaces (tri-lobe and internal hex); Gold Cylinder Abutments for permanent restorations in three implant connection interfaces (external hex, tri-lobe and internal hex) in engaging and non-engaging configurations; Passive Abutments for permanent restorations in two implant connection interfaces (trilobe and internal hex) in engaging and non-engaging configurations; Compact Conical abutments in an internal hex connection for multi-unit restorations; and abutment screws. The Gold Cylinder Abutments and Passive Abutments are UCLA castable abutments which interface with a plastic, burn-out sleeve used to fabricate a prosthesis that is bonded directly to the top of the abutment.
All MAX implants are manufactured from unalloyed titanium conforming to ASTM F67. The implant surface is grit blasted with varying lengths of unroughened coronal portions. The Cover Screw, Healing Abutment, Titanium Cylinder Abutment, Cosmetic Abutment and Passive Abutment are manufactured from unalloyed titanium conforming to ASTM F67. The Gold Cylinder Abutment is manufactured from a gold alloy. The Compact Conical Abutments manufactured from titanium alloy conforming to ASTM
{6}------------------------------------------------
F136. The abutment screws are manufactured from titanium alloy conforming to ASTM F136, or goldplatinum alloy. All subject device components are manufactured in the same facilities using the same materials and manufacturing processes as used for the Southern Implants devices previously cleared in K071161, K163634, K070905 and K180465.
### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility (referenced from K071161 and K180465); engineering analysis; dimensional analysis; sterilization validation according to ISO 11137-1, ISO 11137-2, ISO 17665-1, ISO TS 17665-2; bacterial endotoxin according to USP 39-NF34; sterile barrier shelf life (referenced from K071161). Pull out tests and surface area analysis comparing the worst-case subject device to the predicate device were performed. No clinical data were included in this submission.
## EQUIVALENCE TO MARKETED DEVICE
Southern Implants (Pty) Ltd submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed devices: K071161, Endosseous Dental Implant, Southern Implants (Pty) Ltd
K163634, External Hex Implants, Southern Implants (Pty) Ltd
K070905, Endosseous Dental Implant, Southern Implants (Pty) Ltd
K180465, Provata Implant System, Southern Implants (Pty) Ltd
The primary predicate device is K071161.
The reference devices are K163634, K070905 and K180465
A comparison of the technological characteristics of the subject device and the primary predicate device K071161 is provided in the following table.
| Comparison | Subject device | Primary Predicate device |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use statement | Southern Implants MAX implant is<br>intended for implantation in the<br>maxillary or mandibular molar<br>region where bone exists and the<br>surgeon has determined that the<br>placement of a narrower diameter<br>implant would increase the<br>probability of failure due to poor<br>primary stability, or increased<br>surgical procedures leading to<br>complications. This MAX implant<br>provides support for fixed or<br>removable dental prostheses in a<br>single tooth, partially edentulous<br>prostheses or full arch prostheses.<br>It further adds the option for<br>immediate loading on single and<br>splinted multiple unit restorations | K071161, Endosseous Dental<br>Implant, Southern Implants (Pty)<br>Ltd<br>NSI MAX implant is intended for<br>implantation in the maxillary or<br>mandibular molar region where<br>bone exists and the surgeon has<br>determined that the placement of a<br>narrower diameter implant would<br>increase the probability of failure<br>due to poor primary stability, or<br>increased surgical procedures<br>leading to complications. This<br>MAX implant provides support for<br>fixed or removable dental<br>prostheses in a single tooth,<br>partially edentulous prostheses or<br>full arch prostheses. It further adds<br>the option for immediate loading on<br>single and splinted multiple unit<br>restorations when good primary |
| | when good primary stability is<br>achieved and with appropriate<br>occlusal loading, to restore chewing<br>function. | stability is achieved and with<br>appropriate occlusal loading, to<br>restore chewing function. |
| Implant Design | | |
| Implant Diameter | 6, 7, 8, 9mm | 8, 9mm |
| Implant Length | 6, 7, 9, 11mm | 7, 9, 11, 13mm |
| Implant Platform Diameter | Ø4.5, 5.5, 5.7, 6.5, 7.5mm | Ø6.5, 7.5mm |
| Implant Prosthetic Diameter | Ø3.575, 4, 5, 5.575, 6, 7mm | Ø6, 7mm |
| Implant Interface | External hex<br>Tri-lobe<br>Internal Hex | External hex |
| Material | Unalloyed titanium (ASTM F67) | Unalloyed titanium (ASTM F67) |
| Implant endosseous surface | Grit-blasted | Grit-blasted |
| Cover Screw | | |
| Platform diameter | Ø5.575, 6mm | Ø7mm |
| Maximum diameter | Ø5.6, 6mm | Ø7mm |
| Material | Unalloyed titanium (ASTM F67),<br>anodized | Unalloyed titanium (ASTM F67) |
| Healing Abutment | | |
| Collar height | 3, 5, 6mm | 4, 5.5, 7mm |
| Collar diameter | 6, 7, 7.8mm | 8mm |
| Material | Unalloyed titanium (ASTM F67),<br>anodized | Unalloyed titanium (ASTM F67) |
| Titanium Cylinder Abutment | | |
| Connection configurations | Engaging (single-unit) and Non-<br>engaging (Multi-unit) | Engaging (single-unit) and Non-<br>engaging (Multi-unit) |
| Prosthesis Attachment | Cement-retained | Cement-retained |
| Collar height | 1, 3mm | 1mm |
| Collar diameter | 6.33, 6.55mm | 7.33mm |
| Abutment Angle | 0° | 0° |
| Abutment Material | Unalloyed titanium (ASTM F67),<br>anodized | Unalloyed titanium (ASTM F67) |
| Abutment Screw Material | Titanium Alloy (ASTM F136) or<br>Gold Alloy | Titanium Alloy (ASTM F136) or<br>Gold Alloy |
| Cosmetic Abutment | | |
| Connection configurations | Engaging | N/A |
| Prosthesis Attachment | Cement-retained | N/A |
| Collar height | Varying from 1.6mm to 2.9mm | N/A |
| Collar diameter | Varying from 6.86 mm to 7.92mm | N/A |
| Abutment Angle | 0° | N/A |
| Abutment Material | Unalloyed titanium (ASTM F67),<br>anodized | N/A |
| Abutment Screw Material | Titanium Alloy (ASTM F136) or<br>Gold Alloy | N/A |
| Gold Cylinder Abutment | | |
| Connection configurations | Engaging (single-unit) and Non-<br>engaging (Multi-unit) | N/A |
| Prosthesis Attachment | Screw-retained | N/A |
| Collar height | 0.5, 0.56, 0.64mm | N/A |
| Collar diameter | 4.3, 6.15, 6.45, 7.15mm | N/A |
| Abutment Angle | 0° | N/A |
| Abutment Material | Gold Alloy | N/A |
| Abutment Screw Material | | N/A |
| | Titanium Alloy (ASTM F136) or<br>Gold Alloy | |
| Passive Abutment | | |
| Connection configurations | Engaging (single-unit) and Non-<br>engaging (Multi-unit) | Engaging (single-unit) and Non-<br>engaging (Multi-unit) |
| Prosthesis Attachment | Screw-retained | Screw-retained |
| Collar height | 0.3, 0.4mm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.