The Plexus SCD 110 DVT Sleeve is a prescription device intended to be used preventatively to increase venous blood flow in patients at risk of deep vein thrombosis due to the associated risk factors for thrombus formation during: trauma, critical care, general medicine, general surgery, as well as neurological, orthopedic, obstetric conditions and treatments.
Device Story
Plexus SCD110 is a non-sterile, single-chamber sequential compression sleeve for lower leg DVT prevention. Device consists of soft, non-woven foam-felt garment with internal PVC air bladder. Used with commercially available, single-chamber DVT compression pumps; connector design prevents attachment to multi-chamber systems. Operation: sleeve wraps around patient calf; pump inflates bladder to 40 mmHg; relief valve triggers deflation; cycle repeats (~50-second cycle) to stimulate venous blood flow. Used in clinical settings (trauma, surgery, critical care) under physician prescription. Output is pneumatic pressure on calf muscle/veins. Benefits patient by increasing venous flow velocity, reducing thrombus formation risk. Healthcare providers use device to manage DVT risk in immobile or high-risk surgical/medical patients.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included ISO 10993-5:2009 (cytotoxicity), ISO 10993-10:2010 (intracutaneous reactivity and skin sensitization), sleeve burst test, pressure accuracy test, and pull force test. All tests passed.
Technological Characteristics
Single-chamber pneumatic compression sleeve. Materials: non-woven foam-felt exterior, PVC internal bladder. Dimensions: calf-wrap form factor. Energy source: external pneumatic pump (not included). Connectivity: mechanical connector specific to single-chamber systems. Biocompatibility: ISO 10993-5, ISO 10993-10. Non-sterile.
Indications for Use
Indicated for patients at risk of deep vein thrombosis (DVT) during trauma, critical care, general medicine, general surgery, neurological, orthopedic, obstetric, and urologic conditions/treatments. Contraindicated for severe atherosclerosis, ischemic vascular disease, suspected/known acute DVT, severe congestive cardiac failure, pulmonary edema, pulmonary embolisms, extreme limb deformity, local skin/tissue conditions (gangrene, infected wounds, recent skin grafts, dermatitis), malignancy in legs, untreated limb infections (cellulitis), lymphangiosarcoma, or where increased blood/lymph flow is undesirable.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 22, 2019
Midway Medical, LLC. Jim Tobin Vice President of Business Development 3004 Gill Street Bloomington, Illinois 61704
Re: K190975
Trade/Device Name: Plexus SCD110 Sequential Compression Device Sleeve Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: April 3, 2019 Received: April 15, 2019
Dear Jim Tobin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190975
Device Name
Plexus SCD110 Sequential Compression Device Sleeve
Indications for Use (Describe)
The Plexus SCD 110 DVT Sleeve is a prescription device intended to be used preventatively to increase venous blood flow in patients at risk of deep vein thrombosis due to the associated risk factors for thrombus formation during: trauma, critical care, general medicine, general surgery, as well as neurological, orthopedic, obstetric conditions and treatments.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Summary Preparation Date:
April 3, 2019
# Revision:
July 15, 2019
# Submitter / 510(k) Sponsor
Midway Medical, LLC. 3004 Gill Street Bloomington, IL 61704
## Contact Person
Jim Tobin Vice President of Business Development Phone: 484-695-6209 Email: jim@midway-medical.com
## Device Name
Device Name: Disposable Sequential Compression Device (SCD) Sleeve Proprietary Trade Name: Plexus SCD110 Sequential Compression Device Sleeve Common Name: Compressible Limb Sleeve
## Product Classification
Product Classification: Compressible Limb Sleeve, 21 CFR 870.5800 Product Code: JOW Device Class: II
## Primary Predicate Device:
| Product | Manufacturer | 510(k) Number |
|-------------------------|------------------------|---------------|
| Cirona 6200 DVT Sleeves | Devon Medical Products | K141578 |
## Reference Devices:
| Product | Manufacturer | 510(k) Number |
|--------------------------------|---------------------------------|---------------|
| Excel Flowtron DVT Garments | Arjo Huntleigh | K133119 |
| Aircast VenoFlow Elite Sleeves | DJO, LLC | K122499 |
| Venodyne Advantage Sleeves | Ecolab, Inc. (Microtek Medical) | K110358 |
| VasoPress Compression Sleeve | Compression Therapy Concepts | K101915 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the words "MIDWAY MEDICAL" stacked on top of each other. The words are in a bold, sans-serif font and are colored in blue. To the left of the words are two vertical gray bars.
510(k) Summary
#### Device Description
The Plexus SCD110 Sequential Compression Device is a non-sterile sequential compression sleeve intended to be used with commercially-available compression pumps*. The device is used to prevent the onset of deep vein thrombosis (DVT) by stimulating blood flow and increasing venous flow velocity. The device consists of a pair of soft pressure garment sleeves with an internal air bladder. Each package of Plexus SCD110 has two compression sleeves – one for the right calf and one for the left calf.
The Plexus SCD110 is a single-bladder sleeve that provides a preset pattern of intermittent pressure to the deep vein in the lower leg. To use the device, the garment sleeve is wrapped around a patient's calf. The compression sleeve is connected to a commercially available DVT compression pump intended to be used with single-chamber air bladder sleeves. The connector of the sleeve is designed such that it will not connect to a pump intended for multi-chamber sleeves. When the compression pump is turned on, air flows from the air pump into the air bladder inside the compression sleeve, providing pneumatic pressure on the calf muscle and veins underneath.
Once the compression pump reaches an air pressure of 40 mmHg), a relief valve is triggered within the DVT pump to let air escape from the sleeve and reduces the pressure within the system. With the relief valve opened, the compression sleeve completely deflates and allows blood to re-enter the deep veins of the patient during this cycle. After a pre-set deflation period (~50-second cycle), the relief valve will close and the air pump will start pumping again to 40 mmHg. This cycle of inflation of the air bladder will continue until the compression pump device is turned off.
The Plexus SCD110 sleeve is made of a soft, lightweight, non-woven foam-felt material with a singlechamber bladder made of polyvinyl chloride (PVC). The Plexus SCD sleeve is intended to be compatible with single-chamber DVT pump systems only and the connectors will not allow the sleeve to be attached to any other compression systems (i.e. multi-chamber DVT systems, lymphedema compression systems)
* All testing was completed using the following DVT compressions pumps: Cirona™ 6200 (Devon Medical), Flowtron® Excel (ArjoHuntleigh), VenaFlow® Elite System (DJO Global/Aircast), VasoPress® Compression Therapy Pumps (Zimmer Biomet / Compression Therapy Concepts), and Venodyne Advantage (EcoLab Inc. / Microtek Medical, Inc.)
#### Device Indications for Use:
The Plexus DVT sleeve is a prescription device intended to be used preventatively to increase venous blood flow in patients at risk deep vein thrombosis due to the associated risk factors for thrombus formation during: trauma, critical care, general medicine, general surgery, as well as neurological, orthopedic, urologic, obstetric conditions and treatments
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Midway Medical. The logo is composed of two lines of text. The first line says "MIDWAY" and the second line says "MEDICAL". The text is in a sans-serif font and is blue. To the left of the text is a gray vertical bar.
# 510(k) Summary
#### Contraindications:
The Plexus DVT sleeves are contraindicated for use by patients with the following conditions:
- Severe atherosclerosis or other ischemic vascular diseases
- Suspected or known acute deep vein thrombosis
- Severe congestive cardiac failure
- Existing pulmonary edema
- Existing pulmonary embolisms
- Extreme deformity of the limbs
- Any local skin or tissue condition in which the garments would interfere including gangrene, untreated or infected wounds, recent skin graft, and/or dermatitis
- Known presence of malignancy in the legs
- Limb infections, including cellulitis, that have not received antibiotic coverage
- Presence of lymphangiosarcoma
- In situations where increased lymph and blood flow is undesirable
#### Technical Characteristics
The technological characteristics of the Plexus are substantially equivalent to the sleeves of the predicate and other single-chamber reference devices. The overall design, materials, mode of operation, and performance characteristics are substantially equivalent to the predicate and reference devices.
#### Perfomance Data
To validate that the Plexus DVT sleeve meets the intended specifications, a number of non-clinical tests were conducted by a third-party test facility.
#### Biocompatibility Tests
| Standards and Regulations Applied | Results | Equivalence to Predicate Devices* |
|--------------------------------------------|---------|-----------------------------------|
| ISO10993-5:2009 In-Vitro Cytotoxicity | Passed | Equivalent |
| ISO10993-10:2010 Intracutaneous Reactivity | Passed | Equivalent |
| ISO10993-10: 2010 Skin Sensitization | Passed | Equivalent |
{6}------------------------------------------------
## Comparative and Compatibility Performance Tests
| Test | Results | Equivalence to Predicate Devices* |
|------------------------|---------|-----------------------------------|
| Sleeve Burst Test | Passed | Equivalent |
| Pressure Accuracy Test | Passed | Equivalent |
| Pull Force Test | Passed | Equivalent |
The biocompatibility testing and functional testing confirm that the Plexus DVT sleeves perform as intended and are substantially equivalent to the proposed predicate device. No new or additional risks are presented with the Plexus DVT sleeves.
* All testing was completed using the following DVT compressions pumps: Cirona™ 6200 (Devon Medical), Flowtron® Excel (ArjoHuntleigh), VenaFlow® Elite System (DJO Global/Aircast), VasoPress® Compression Therapy Pumps (Zimmer Biomet / Compression Therapy Concepts), and Venodyne Advantage (EcoLab Inc. / Microtek Medical, Inc.)
## Conclusion
Based on the assessment of the functional testing and biocompatibility testing performed, the Plexus DVT sleeve is substantially equivalent to the predicate device in material content, function and indications for use. As such, these sleeves can be concluded to be as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.