TempSure

K190678 · Cynosure, Inc. · GEI · Jul 24, 2019 · General, Plastic Surgery

Device Facts

Record IDK190678
Device NameTempSure
ApplicantCynosure, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateJul 24, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids. The 18mm, 25mm, 30mm, and 60mm handpieces provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The massage device is intended to provide a temporary reduction in the appearance of cellulite The following surgical modes are applicable to the generator: Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed Cutting: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty. Blended Cutting and Coagulation: snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and development of skin flaps. Fulguration: basal cell carcinoma. papilloma. cyst destruction, tumors. verrucae, hemostasis. Bipolar: pinpoint precise coagulation. pinpoint hemostasis in any, field (wet or dry). snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment and turbinate shrinkage

Device Story

Radiofrequency (RF) generator system; delivers RF energy to patient tissue via temperature-sensing handpieces (10-60mm) or surgical accessories; controlled via GUI, footswitch, or fingerswitch. Used in clinical settings by healthcare providers. System transforms electrical energy into thermal energy for tissue heating, coagulation, or cutting. Output affects clinical decision-making by enabling aesthetic skin treatments, pain relief, and surgical hemostasis. Benefits include non-invasive wrinkle reduction, temporary pain/spasm relief, and precise surgical tissue management. New 60mm handpiece and large neutral pad added for expanded body treatment coverage.

Clinical Evidence

Bench testing only. No clinical data submitted. Bench testing confirmed the 60mm handpiece and large neutral pad maintain target tissue temperature for 10 minutes under simulated clinical conditions.

Technological Characteristics

RF generator; 4.0 MHz Sin-wave CW, fully/partially rectified, 1.7 MHz (Bipolar). Materials: Gold-plated aluminum, PVDF, polyetherimide, Loctite, Delrin, polycarbonate, hydrogel. Dimensions: 22.5" x 18" x 12"; Weight: 30 lbs. Connectivity: GUI-controlled. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2. Temperature sensing: <1s response time.

Indications for Use

Indicated for patients requiring non-ablative treatment of mild to moderate facial wrinkles/rhytids, temporary relief of pain, muscle spasms, increased local circulation, temporary reduction in cellulite appearance, and general surgical electrosurgical procedures (coagulation, cutting, hemostasis).

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. July 24, 2019 Cynosure, Inc Michael King Regulatory Affairs Specialist II 5 Carlisle Road Westford, Massachusetts 01886 Re: K190678 Trade/Device Name: TempSure System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, PBX Dated: March 15, 2019 Received: March 18, 2019 Dear Michael King: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190678 Device Name TempSure System #### Indications for Use (Describe) The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids. The 18mm. 25mm. 30mm. and 60mm handpieces provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The massage device is intended to provide a temporary reduction in the appearance of cellulite The following surgical modes are applicable to the generator: Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed Cutting: submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty. Blended Cutting and Coagulation: submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and development of skin flaps. Fulguration: basal cell carcinoma. papilloma. cyst destruction, tumors. verrucae, hemostasis. Bipolar: pinpoint precise coagulation. pinpoint hemostasis in any, field (wet or dry). snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment and turbinate shrinkage Type of Use (*Select one or both, as applicable*) | <div> <object data="checkbox_checked.svg" height="12" type="image/svg+xml" width="12"></object> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <object data="checkbox.svg" height="12" type="image/svg+xml" width="12"></object> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------| |------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------| |X | Prescription Use (Part 21 CFR 801 Subpart D) |__| Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary for Cynosure TempSure A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92. | 807.92(a)(1) Submitter Information | | |--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | Applicant | Cynosure, LLC | | Address | 5 Carlisle Road, Westford MA, 01886 | | Phone Number | 781-993-2454 | | Fax Number | 978-256-6556 | | Establishment Registration Number | 1222993 | | Contact Person | Michael King | | Preparation Date | March 15, 2019 | | 807.92(a)(2) Name of Device | | | Trade or Proprietary Name | TempSure System | | Common or Usual Name | Surgical RF Generator | | Classification Name | Electrosurgical, Cutting & Coagulation & Accessories;<br>Massager, Vacuum, Radio Frequency Induced Heat | | Classification Panel | General & Plastic Surgery | | Regulation | 21 CFR 878.4440 | | Regulatory Class | II | | Product Code(s) | GEI, PBX | | 807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed | | | Predicate Devices | Cynosure TempSure (K182365)<br>Cynosure TempSure (K171262) | {5}------------------------------------------------ | K190678 | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 807.92(a)(4) Device Description | | The Cynosure TempSure™ System is a radiofrequency generator<br>with a variety of applications both aesthetic and surgical procedures.<br>The intended action is achieved through application of radiofrequency<br>energy to the patient which results in minimization of heat dissipation<br>and cellular alteration. Output of energy is controlled via the guided<br>user interface (GUI) and the foot and/or hand-switch.<br><br>An additional 60mm Smart Handpiece is now included for the<br>purpose of tissue heating for selected medical conditions. The 60mm<br>handpiece can also be used with a new massage head for the<br>temporary reduction in the appearance of cellulite. The handpiece has<br>the same maximum output power of 300W in the Smart Handpiece<br>Mode. In addition, a new larger disposable neutral pad is provided for<br>these body treatments. There have been no additional changes to the<br>previous existing handpieces (10, 15, 18, 20, 25, or 30mm) or<br>Surgical modes (COAG, Cut, Blend, Bipolar, Fulgurate) of the<br>system. The TempSure system is still used with the same existing<br>electrosurgical accessories.<br><br>The TempSure™ System includes:<br>TempSure™ Generator Temperature Sensing Handpieces (10, 15, 18, 20, 25, 30, and 60mm) Massage Heads (25, 30, and 60mm) IEC Power Cord Footswitch Disposable/Reusable Neutral Pads Surgical Fingerswitch/Foot Controlled Handpieces Monopolar Cables Disposable/Reusable Electrodes, Forceps New Additions to the TempSure System include:<br>60mm Smart Handpiece (ASY-12745) 60mm Massage Head (ASY-12746) Large Neutral Pad (MEL-01197) | {6}------------------------------------------------ | 807.92(a)(5) Intended Use of the Device | | |-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids. | | | The 18mm, 25mm, 30mm, and 60mm handpieces provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. | | | The massage device is intended to provide a temporary reduction in the appearance of cellulite | | | The following surgical modes are applicable to the generator: | | | <b>Coagulation/Hemostasis:</b> General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed | | | <b>Cutting:</b> snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty. | | | <b>Blended Cutting and Coagulation:</b> snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and development of skin flaps. | | | <b>Fulguration:</b> basal cell carcinoma. papilloma. cyst destruction, tumors. verrucae, hemostasis. | | | <b>Bipolar:</b> pinpoint precise coagulation. pinpoint hemostasis in any, field (wet or dry). snoring. submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment and turbinate shrinkage | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ {7}------------------------------------------------ # 807.92(b)(1) Non-clinical tests submitted The following non-clinical tests have been included in this 510k submission in determination of substantial equivalence between the test device and the referenced predicates. ### Tissue Heating with 60mm Handpiece Bench testing was conducted to show that the new 60mm handpiece and large neutral pad, when used in a simulated clinical condition, was able to heat and maintain temperature of the treatment area for 10 minutes. Refer to Section 18 - Performance Testing, Bench for additional information. ### Electromagnetic Compatibility and Electrical Safety Electrical safety testing for the Cynosure TempSure was also completed to prove the safe use of the device. These test reports are provided in accordance with FDA Guidance "Premarket Notification 510k Submissions for Electrosurgical Devices for General Surgery" - Section XII - Electrical Safety and Electromagnetic Compatibility". The following test reports are available in Section 17 -Electromagnetic Compatibility and Electrical Safety. - . IEC 60601-1, Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performances - IEC 60601-1-2, Medical Electrical Equipment Part 1 -2: General requirements for basic ● safety and essential performance - Collateral standard: Electromagnetic disturbances requirements and tests - . IEC 60601-2-2, Medical electrical equipment - Part 2 -2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories ### 807.92(b)(2) Clinical tests submitted - N/A - No clinical tests submitted ### 807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted The non-clinical tests demonstrate that the TempSure generator with the 60mm handpiece is safe and effective and performs as well or better than the legally marketed predicate device. The Tissue Heating bench test demonstrated that the device is able to maintain temperature for at least the 10minute treatment time when used as intended under clinical conditions at various treatment areas and settings. In addition, the Electrosurgical Compatibility and Electrical Safety testing shows that the device is safe to use and meets required standards. These non-clinical tests show that the TempSure generator using the 60mm handpiece meets design specifications as well as performance requirements. {8}------------------------------------------------ | Characteristic | Cynosure TempSure System<br>(K190678) | Cynosure TempSure System<br>(K182365) | Cynosure TempSure (K171262) | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | The 10mm, 15mm, and 20mm<br>handpieces are indicated for non-<br>ablative treatment of mild to moderate<br>facial wrinkles and rhytids. | The 10mm, 15mm, and 20mm<br>handpieces are indicated for non-ablative<br>treatment of mild to moderate facial<br>wrinkles and rhytids | The 10mm, 15mm, and 20mm handpieces<br>are indicated for non-ablative treatment of<br>mild to moderate facial wrinkles and<br>rhytids | | | The 18mm, 25mm, 30mm, and 60mm<br>handpieces provide heating for the<br>purpose of elevating tissue<br>temperature for selected medical<br>conditions such as temporary relief of<br>pain, muscle spasms, and increase in<br>local circulation. | The 25mm, and 30mm handpieces<br>provide heating for the purpose of<br>elevating tissue temperature for selected<br>medical conditions such as temporary<br>relief of pain, muscle spasms, and<br>increase in local circulation. | The 18mm, 25mm, and 30mm handpieces<br>provide heating for the purpose of<br>elevating tissue temperature for selected<br>medical conditions such as temporary<br>relief of pain, muscle spasms, and<br>increase in local circulation. | | | The massage device is intended to<br>provide a temporary reduction in the<br>appearance of cellulite | The massage device is intended to<br>provide a temporary reduction in the<br>appearance of cellulite | The massage device is intended to provide<br>a temporary reduction in the appearance<br>of cellulite | | | The following surgical modes are<br>applicable to the generator: | The following surgical modes are<br>applicable to the generator: | Coagulation/Hemostasis: Using the<br>surgical handpieces and accessories,<br>general surgical procedures including | | | Coagulation/Hemostasis: General<br>surgical procedures including urologic,<br>thoracic, plastic, reconstructive, and<br>gynecological procedures where | Coagulation/Hemostasis: General<br>surgical procedures including urologic,<br>thoracic, plastic, reconstructive, and<br>gynecological procedures where<br>electrosurgical coagulation of tissue is<br>performed | urologic, thoracic, plastic, reconstructive,<br>and gynecological procedures where | | | electrosurgical coagulation of tissue is<br>performed | Cutting: snoring. submucosal palatal<br>shrinkage, traditional uvulopalatoplasty<br>(RAUP), myringotomy with effective | electrosurgical coagulation of tissue is<br>performed | | | Cutting: snoring. submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty (RAUP), | hemorrhage control. epistaxis treatment,<br>turbinate shrinkage. skin incisions,<br>biopsy, cysts, abscesses. tumors. | | | | | | | | | | | | {9}------------------------------------------------ {10}------------------------------------------------ | Characteristic | Cynosure TempSure System<br>(K190678) | Cynosure TempSure System<br>(K182365) | Cynosure TempSure (K171262) | |-------------------------------------|--------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------| | Rx/OTC | Prescription | Prescription | Prescription | | Energy Type | Radiofrequency | Radiofrequency | Radiofrequency | | Modality | Monopolar, Bipolar | Monopolar, Bipolar | Monopolar | | Temperature<br>Sensing | Temperature-Sensitive Handpiece | Temperature-Sensitive Handpiece | Temperature-Sensitive Handpiece | | Temperature<br>Response Time | <1 second | <1 second | <1 second | | Handpiece Size | 10mm, 15mm, 18mm, 20mm, 25 mm,<br>30 mm, 60mm | 10mm, 15mm, 20mm, 25 mm, 30 mm | 10mm, 15mm, 18mm, 20mm, 25 mm, 30<br>mm | | Treatment<br>Activation | Fingerswitch, Footswitch | Fingerswitch, Footswitch | Fingerswitch, Footswitch | | Aesthetic<br>Optimal<br>Temperature | 39-45°C | 39-45°C | 39-45°C | | Patient<br>Contacting<br>Material | Gold-Plated Aluminum, PVDF,<br>Polyetherimide, Loctite, Delrin,<br>Polycarbonate sabic lexan, Hydrogel | Gold-Plated Brass, Polyetherimide,<br>Loctite, Delrin | Gold-Plated Brass, Polyetherimide,<br>Loctite, Delrin | | Massage Head | Yes | Yes | Yes | | Output<br>Waveform | 4.0 MHz Sin-wave CW, Fully<br>Rectified, Partially Rectified, and 1.7<br>MHz for Bipolar | 4.0 MHz Sin-wave CW, Fully Rectified,<br>Partially Rectified, and 1.7 MHz for<br>Bipolar | 4.0 MHz Sin-wave CW | | Modes | Surgical (Coag, Cut, Blend, Fulgurate,<br>Bipolar)<br>Smart Handpiece Mode | Surgical (Coag, Cut, Blend, Fulgurate,<br>Bipolar)<br>Smart Handpiece Mode | Surgical (Coag only)<br>Portrait Mode (Smart Handpiece) | | Max Power<br>Output | 300W (Surgical)<br>120W (Wrinkles)<br>300W (Tissue Heating) | 300W (Surgical)<br>120W (Wrinkles)<br>300W (Tissue Heating) | 50W (Surgical - Coag only)<br>120W (Wrinkles)<br>300W (Tissue Heating) | | Dimensions | 22.5" x 18" x 12" | 22.5" x 18" x 12" | 18" x 18" x 12" | | Weight | 30 lbs | 30 lbs | 30 lbs |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...