Penumbra Reperfusion Catheters and Separators As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra 3D Revascularization Device As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment. Penumbra Aspiration Tubing As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump. Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
Device Story
Penumbra System 3D Revascularization Device assists reperfusion catheters in removing thrombus during acute ischemic stroke treatment. Device consists of delivery wire with nitinol 3D lattice; used in conjunction with reperfusion catheter, aspiration tubing, and vacuum pump. Operated by trained interventional neuro-endovascular physicians in clinical settings. Mechanism involves aspiration to draw clot into catheter; 3D device aids in clot engagement/removal. Output is physical removal of intracranial occlusion. Benefits include restored blood flow in stroke patients. Device provided sterile, single-use.
Clinical Evidence
No clinical study conducted. Substantial equivalence supported by bench-top design verification (simulated use, radial pressure, torsion, kink resistance, tensile strength) and literature review of clinical outcomes for technologically equivalent devices.
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, M1/M2 middle cerebral segments) within 8 hours of symptom onset. Candidates include patients ineligible for or failing IV t-PA therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue font.
April 30, 2019
Penumbra, Inc. Michaela Mahl Senior Manager of Regulatory Affairs One Penumbra Place Alameda, California 94502
Re: K190464
Trade/Device Name: Penumbra System (3D Revascularization Device) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: February 25, 2019 Received: February 26, 2019
Dear Michaela Mahl:
This letter corrects our substantially equivalent letter of April 26, 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Xiaolin Zheng -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190464
Device Name
Penumbra System (3D Revascularization Device)
### Indications for Use (Describe)
Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
Penumbra 3D Revascularization Device
As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
### Penumbra Aspiration Tubing
As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
Penumbra Aspiration Pump The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
| Type of Use (Select one or both, as applicable) | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <table style="border:none"><tr><td><span style="font-size:120%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </td></tr><tr><td><span style="font-size:120%">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </td></tr></table> | <span style="font-size:120%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:120%">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <span style="font-size:120%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | | |
| <span style="font-size:120%">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
#### 1 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Penumbra System® 3D Revascularization Device.
#### 1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
#### 1.2 Sponsor Contact Information
Michaela Mahl Senior Manager, Regulatory Affairs Phone: (510) 748-3288 FAX: (510) 217-6414 Email: mmahl@penumbrainc.com
#### 1.3 Date of Preparation of 510(k) Summary
February 22, 2019
#### 1.4 Device Trade or Proprietary Name
Penumbra System® (3D Revascularization Device)
#### 1.5 Device Classification
Regulatory Class: II Classification Panel: Neurology Classification Name: Percutaneous Catheter Regulation Number: 21 CFR §870.1250 Product Code: NRY (Catheter, Thrombus Retriever)
#### 1.6 Predicate and Reference Devices
| 510(k) Number/Clearance Date | Name of Device | Name of Manufacturer |
|-----------------------------------|-----------------------------------------|---------------------------------------------------------------|
| Predicate Device | | |
| K162901 cleared on April 20, 2017 | Penumbra 3D<br>Revascularization Device | Penumbra, Inc.<br>One Penumbra Place<br>Alameda, CA 94502 USA |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the left side of the "P" to the edge of the circle.
#### Predicate Comparison 1.7
| System Name | Penumbra System® | |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| | Predicate Device | Subject Device |
| Device Name | Penumbra 3D Revascularization Device | Penumbra 3D<br>Revascularization<br>Device (Modified) |
| 510(k) No. | K162901 | K190464 |
| Classification | Class II, NRY | SAME |
| Indication for Use | Penumbra Reperfusion Catheters and Separators<br>As part of the Penumbra System, the Reperfusion Catheters and Separators are<br>indicated for use in the revascularization of patients with acute ischemic stroke<br>secondary to intracranial large vessel occlusive disease (within the internal<br>carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries)<br>within 8 hours of symptom onset. Patients who are ineligible for intravenous<br>tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are<br>candidates for treatment.<br>Penumbra 3D Revascularization Device<br>As part of the Penumbra System, the Penumbra 3D Revascularization Device<br>is indicated for use in the revascularization of patients with acute ischemic<br>stroke secondary to intracranial large vessel occlusive disease (within the<br>internal carotid, middle cerebral – M1 and M2 segments) within 8 hours of<br>symptom onset. Patients who are ineligible for intravenous tissue plasminogen<br>activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.<br>Penumbra Aspiration Tubing<br>As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is<br>indicated to connect the Penumbra Reperfusion Catheters to the Penumbra<br>Aspiration Pump.<br>Penumbra Aspiration Pump<br>The Penumbra Aspiration Pump is indicated as a vacuum source for<br>Penumbra Aspiration Systems. | SAME |
| | Reperfusion Catheter and Separators | SAME |
| System Components | 3D Revascularization Device | SAME |
| | Aspiration Tubing | SAME |
| | Aspiration Pump & Pump Canister | SAME |
| Materials | Penumbra 3D Revascularization Device | Penumbra 3D<br>Revascularization<br>Device (Modified) |
| Delivery-Wire | | |
| Core wire | Stainless Steel | SAME |
| Coating | PTFE Green PC 8-403G | SAME |
| Outer coil | Stainless Steel | SAME |
| Proximal Solder | Silver Solder | SAME |
| System Name | Penumbra System® | |
| Distal Inner Coil | N/A | Platinum Alloy |
| Distal Solder | Silver Solder | SAME |
| Distal Tip | | |
| Outer coil | Stainless Steel | SAME |
| Radiopaque Markers | Platinum Alloy | SAME |
| Attachment Tip | Stainless Steel | SAME |
| Solder joint | Gold Solder / Silver Solder | Silver Solder |
| 3D Nitinol Device | | |
| 3D lattice | Nitinol | SAME |
| Introducer sheath | FEP | SAME |
| Dimensions | | |
| - Device Nominal<br>Total Length | 26 mm | SAME |
| - Device Nominal OD | 4.5 mm | SAME |
| - Device Nominal<br>Usable Length | 20 mm | SAME |
| - Delivery Wire Distal<br>OD | 0.014" | SAME |
| - Delivery Wire<br>Length | 200 cm | SAME |
| Sterilization | EO | SAME |
| Shelf-Life | 36 Months | SAME |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the left side of the "P" to the edge of the circle.
#### Device Description 1.8
The Penumbra System 3D Revascularization Device is designed to revascularize patients experiencing acute ischemic stroke as a component of the Penumbra System. The Penumbra System's fundamental mechanism of action is aspiration removes clot by drawing it into the Penumbra Reperfusion Catheter. The 3D Revascularization Device is designed to assist the Reperfusion Catheter with thrombus removal if needed. Intended users for this device are physicians who have received appropriate training in interventional neuro-endovascular techniques and treatment of acute ischemic stroke. The device is provided sterile, non-pyrogenic, and intended for single use only.
#### 1.9 Indications for Use
# Penumbra Reperfusion Catheters and Separators
As part of the Penumbra System, the Reperfusion Catheters and Separators are indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments, basilar, and
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Penumbra company logo. The logo consists of the word "Penumbra" in a red sans-serif font, followed by a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending to the edge of the circle.
vertebral arteries) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
### Penumbra 3D Revascularization Device
As part of the Penumbra System, the Penumbra 3D Revascularization Device is indicated for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral - M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
### Penumbra Aspiration Tubing
As part of the Penumbra System, the Penumbra Sterile Aspiration Tubing is indicated to connect the Penumbra Reperfusion Catheters to the Penumbra Aspiration Pump.
### Penumbra Aspiration Pump
The Penumbra Aspiration Pump is indicated as a vacuum source for Penumbra Aspiration Systems.
# 1.10 Summary of Non-Clinical Data
As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the device follows.
Included in this section are descriptions of the testing, which substantiates the safe and effective performance of the subject Penumbra System 3D Revascularization Device as well as its substantial equivalence to the predicate device:
- Biocompatibility
- Design Verification (Bench-Top Testing) ●
- Design Validation (Animal Study) ●
- . Performance Data
The subject, Penumbra System 3D Revascularization Device met all established requirements.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The word "Penumbra" is written in a bold, red font. To the right of the name is a circular logo, also in red, with a stylized white "P" inside.
## 1.10.1 Previously Completed Biocompatibility Testing
The materials used in the subject Penumbra System 3D Revascularization Device [modified] are identical to those in the predicate device, Penumbra System 3D Revascularization Device [K162901]. The subject Penumbra System 3D Revascularization Device [modified] uses identical materials, equivalent processing, and identical sterilization methods as products that Penumbra has already successfully conducted biocompatibility testing per BS EN ISO 10993. The studies were selected in accordance with ISO 10993-1 guidelines (Biological Evaluation of Medical Devices) for a limited exposure (< 24 hours), external communicating device with circulating blood contact. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices (GLP).
The classification for the predicate, Penumbra System 3D Revascularization Device [K162901] and subject device, Penumbra System 3D Revascularization Device [modified] are identical. All biocompatibility risks have been mitigated through successful completion of prior biocompatibility testing per ISO 10993-1:2009/AC: 2010 which will be used to support the biocompatibility for the subject Penumbra System 3D Revascularization Device [modified]. Therefore, no further biocompatibility testing was conducted on the subject device.
### 1.10.2 Design Verification - Bench-top Testing
The physical and mechanical properties of the 3D Revascularization Device was assessed using standard test methods and pre-determined acceptance criteria. The following tests were performed and all tests passed successfully:
| Attribute | Specification | Result |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Pouch Seal Strength | Minimum value per specification | Pass |
| Dimensional / Visual<br>Inspection | These evaluations confirm that the units used in this Design Verification<br>testing meet all product specifications. | Pass |
| Simulated Use<br>[Intracranial Access,<br>Vessel Access Entry<br>Performance & Clot<br>Removal] | Simulated use testing of the Reperfusion Catheter and 3D Revascularization<br>Device was performed with accessory devices in an anatomical model<br>which simulated the tortuosity of the neurovasculature. Devices were<br>delivered through the tortuous anatomical model to evaluate the<br>effectiveness of the devices to remove clots. | Pass |
| Physician Evaluation<br>[Deliverability & Clot<br>Removal] | Multiple Physician performance evaluation of the subject Penumbra System<br>3D Revascularization Device in a simulated neurovascular tortuosity model<br>with the Penumbra System 3D Revascularization Device used as a baseline. | Pass |
| Particulate Testing | The maximum number of particles: | Pass |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "Penumbra" in a red font, followed by a red circle with a white "P" inside. The "P" is stylized with a line extending from the top of the "P" to the edge of the circle. The logo is simple and modern.
| Attribute | Specification | Result |
|--------------------------------------|-------------------------------------------------------------------------|--------|
| | $\ge$ 10 μm will be $\le$ 6000 particles | |
| | $\ge$ 25 μm will be $\le$ 600 particles | Pass |
| | $\ge$ 75 μm will be measured for informational purposes only<br>(FIFO) | FIFO |
| | $\ge$ 125 μm will be measured for informational purposes only<br>(FIFO) | FIFO |
| Radial Pressure | Value within the specification range | Pass |
| Torsion | Minimum value per specification | Pass |
| Kink Resistance | No kinking when formed in a defined radius | |
| Device Af | Maximum value per specification | Pass |
| Corrosion | Specimens will be free from signs of corrosion | |
| Tensile Strength -<br>Markerband | Minimum value per specification | Pass |
| Tensile Strength -Wire<br>Attachment | Minimum value per specification | Pass |
### 1.10.3 Design Validation - Previously Completed Animal Study
The subject, 3D Revascularization Device [modified] uses the same materials, has the same indication for use, and has the same intended use as the predicate, 3D Revascularization Device. In addition, the subject, 3D Revascularization Device [modified] will be held to the same maximum radial pressure specification as the predicate, 3D Revascularization Device [K162901]. Therefore, the subject, 3D Revascularization Device [modified] presents an equivalent challenge for vessel response compared to the predicate device. Rabbit vessel response studies have been performed for the predicate, 3D Revascularization Device [K162901] with successful results. Based on the existing GLP test results, the interaction of the predicate, 3D Revascularization Device [K162901] with the rabbit vasculature is well understood. Therefore, the existing rabbit vessel response animal study results mitigate all vessel response risks for the subject, 3D Revascularization Device [modified].
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Penumbra, a medical device company. The word "Penumbra" is written in a bold, red font. To the right of the word is a red circle with a white "P" inside of it.
### 1.10.4 Performance Data - Clinical
No clinical study was conducted as bench and previously performed animal testing was determined sufficient for verification and validation purposes. A review was conducted considering published clinical study articles that featured the predicate device and other devices with similar dimensions used for clot removal. The literature review was used to support the determination of substantial equivalence by using clinical outcomes from devices that are considered technologically equivalent.
### 1.11 Summary of Substantial Equivalence
The subject, Penumbra System 3D Revascularization Device is substantially equivalent to the predicate device with regard to intended use, operating principle, design concept, materials, shelf-life, packaging and sterilization processes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.