Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Blue), Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Blue) - Extended Cuff

K190454 · Hartalega NGC Sdn. Bhd. · LZA · Sep 26, 2019 · General Hospital

Device Facts

Record IDK190454
Device NameNitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Blue), Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Blue) - Extended Cuff
ApplicantHartalega NGC Sdn. Bhd.
Product CodeLZA · General Hospital
Decision DateSep 26, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Nittile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate. Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) - Extended Cuff is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

Device Story

Non-sterile, disposable nitrile examination gloves; worn by healthcare examiners to prevent cross-contamination. Tested for permeation resistance against chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05. Available in standard and extended cuff versions. Used in clinical settings; provides barrier protection during handling of hazardous drugs. Breakthrough detection times vary by chemical; specific warnings provided for Carmustine and Thiotepa due to low permeation resistance.

Clinical Evidence

Bench testing only. Permeation testing performed per ASTM D6978-05 (Reapproved 2013) to determine breakthrough detection times for various chemotherapy drugs and Fentanyl Citrate.

Technological Characteristics

Material: Nitrile. Form factor: Powder-free examination glove (standard and extended cuff). Energy source: None. Sterilization: Non-sterile. Testing standard: ASTM D6978-05 (Reapproved 2013) for chemotherapy drug permeation.

Indications for Use

Indicated for use as a non-sterile, disposable patient examination glove to prevent cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05. Contraindicated for use with Carmustine and Thiotepa.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue. September 26, 2019 Hartalega Ngc Sdn. Bhd. % Nurul Kong Quality Assurance Senior Manager Hartalega Sdn. Bhd No.7. Kawasan Perusahaan Suria Bestari Jaya, 45600 My Re: K190454 Trade/Device Name: Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Blue), Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Blue) - Extended Cuff Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I Product Code: LZA, ODO, LZC Dated: July 5, 2019 Received: July 8, 2019 Dear Nurul Kong: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K190454 ### Device Name Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) ### Indications for Use (Describe) Nittile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs, | Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes | |---------------------------------------|------------------------------------------------| | Carmustine (3.3 mg/ml) | 2.73 | | Cisplatin (1.0 mg/ml) | >240 | | Cyclophosphamide (20.0 mg/ml) | >240 | | Dacarbazine (10.0 mg/ml) | >240 | | Doxorubicin Hydrochloride (2.0 mg/ml) | >240 | | Etoposide (20.0 mg/ml) | >240 | | Fluorouracil (50.0 mg/ml) | >240 | | Methotrexate (25.0 mg/ml) | >240 | | Mitomycin C (0.5 mg/ml) | >240 | | Paclitaxel (6.0 mg/ml) | >240 | | Thiotepa (10.0 mg/ml) | 22.82 | | Vincristine Sulfate (1.0 mg/ml) | >240 | Please note that Carmustine and Thiotepa have extremely low permeation times of 2.73 minutes and 22.82 minutes respectively. | Fentanyl Citrate and Concentration | Minimum Breakthrough Detection Time in Minutes | |------------------------------------------|------------------------------------------------| | Fentanyl Citrate Injection (100 mcg/2ml) | >240 | WARNING: Do Not Use With: Carmustine and Thiotepa Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) [X] Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # Indications for Use 510(k) Number (if known) K190454 ### Device Name Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) - Extended Cuff ### Indications for Use (Describe) Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) - Extended Cuff is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate. These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs, | Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes | |---------------------------------------|------------------------------------------------| | Carmustine (3.3 mg/ml) | 30.2 | | Cisplatin (1.0 mg/ml) | >240 | | Cyclophosphamide (20.0 mg/ml) | >240 | | Dacarbazine (10.0 mg/ml) | >240 | | Doxorubicin Hydrochloride (2.0 mg/ml) | >240 | | Etoposide (20.0 mg/ml) | >240 | | Fluorouracil (50.0 mg/ml) | >240 | | Methotrexate (25.0 mg/ml) | >240 | | Mitomycin C (0.5 mg/ml) | >240 | | Paclitaxel (6.0 mg/ml) | >240 | | Thiotepa (10.0 mg/ml) | 61.5 | | Vincristine Sulfate (1.0 mg/ml) | >240 | Please note that Carmustine and Thiotepa have extremely low permeation times of 30.2 minutes and 61.5 minutes respectively. | Fentanyl Citrate and Concentration | Minimum Breakthrough Detection Time in Minutes | |------------------------------------------|------------------------------------------------| | Fentanyl Citrate Injection (100 mcg/2ml) | >240 | | WARNING: | Do Not Use With: Carmustine and Thiotepa | |----------|------------------------------------------| |----------|------------------------------------------| | Corporation of the Comments of Concellente Children Children Children Compress of Children Comments of Children Commens and Children<br>Type of Use (Select one or both, as applicable) | |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| Prescription Use (Part 21 CFR 801 Subpart D) × Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {5}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...