Spirotome Endo 08 Gauge, Spirotome Endo 10 Gauge, Spirotome Endo 14 Gauge, Spirotome Endo 14 Gauge with 10 mm Helix
Applicant
Bioncise NV
Product Code
KNW · Gastroenterology, Urology
Decision Date
May 24, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Spirotome Endo is a family of large core soft tissue biopsy medical devices to be used in humans to take out adequate samples of tissue. The Spirotome Endo is intended to provide soft tissue for histological and biomolecular detection and confirmation of imaged abnormalities of the internal female reproductive organs such as uterus, ovaries, and retroperitoneal lymph nodes.
Device Story
Spirotome Endo is a manual, single-use, large-core soft tissue biopsy instrument. It consists of a stainless steel helical receiving needle, a stainless steel cutting cannula, and a polypropylene releasing element. The clinician inserts the receiving needle through the cutting cannula to the biopsy site; rotates the needle clockwise to engage tissue; rotates the cannula clockwise to sever the sample; and retracts the device. The releasing element and counter-clockwise rotation of the needle are used to extract the tissue sample. Used in clinical settings for gynecological procedures, the device provides tissue samples for histological and biomolecular analysis. This allows physicians to confirm imaged abnormalities, facilitating diagnosis of benign or malignant conditions in the uterus, ovaries, or retroperitoneal lymph nodes.
Clinical Evidence
No new clinical trials were conducted. Substantial equivalence is supported by bench testing (functional, biocompatibility, sterilization, packaging) leveraged from predicate devices (K060384, K080095) and a review of peer-reviewed clinical literature. The literature reports demonstrate the safe and effective use of the Spirotome Endo for biopsy of internal female reproductive organs, confirming an acceptable safety profile without raising new safety or effectiveness concerns.
Technological Characteristics
Manual core biopsy instrument. Materials: stainless steel (needle/cannula), polycarbonate, polypropylene. Variants: 8G and 14G needles; receiving needle length 375 mm, cutting cannula 325 mm; helical needle length 10-18 mm. Sterilization: steam. Single-use, disposable. Shelf life: 3 years.
Indications for Use
Indicated for humans requiring soft tissue biopsy for histological and biomolecular detection/confirmation of imaged abnormalities in internal female reproductive organs, including uterus, ovaries, and retroperitoneal lymph nodes.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
May 24, 2019
Bioncise NV Jaak Janssens CEO Bodemstraat 16 Wellen, Limburg Belgium 3830
Re: K190378
> Trade/Device Name: Spirotome Endo 08 Gauge, Spirotome Endo 10 Gauge, Spirotome Endo 14 Gauge, Spirotome Endo 14 Gauge with 10 mm Helix Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: Class II Product Code: QHC Dated: February 20, 2019 Received: February 25, 2019
Dear Jaak Janssens:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or (301) 796-7100).
Sincerely,
Sharon Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190378
#### Device Name
Spirotome Endo 08 Gauge, Spirotome Endo 14 Gauge, Spirotome Endo 14 Gauge with 10 mm Helix
#### Indications for Use (Describe)
The Spirotome Endo is a family of large core soft tissue biopsy medical devices to be used in humans to take out adequate samples of tissue.
The Spirotome Endo is intended to provide soft tissue for histological and biomolecular detection and confirmation of imaged abnormalities of the internal female reproductive organs such as uterus, ovaries, and retroperitoneal lymph nodes.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K190378
#### Submitter Information:
Bioncise NV Bodemstraat 16 Wellen, Limburg Belgium 3830
#### Contact:
| Name: | Jaak Janssens<br>CEO Bioncise NV |
|----------------|----------------------------------|
| Telephone: | +3211271557 |
| Fax: | +3211283677 |
| E-mail: | jaak.janssens@bioncise.com |
| Date prepared: | May 23, 2019 |
#### Device Identification:
| Trade Name: | Spirotome Endo 08 Gauge, Spirotome Endo 10 Gauge, Spirotome Endo 14 Gauge, Spirotome Endo 14 Gauge with 10 mm Helix |
|--------------------|---------------------------------------------------------------------------------------------------------------------|
| Common Name: | Core Biopsy Instrument |
| Regulation Number: | 21 CFR 876.1075 |
| Regulation Name: | Gastroenterology-Urology Biopsy Instrument |
| Product Code: | QHC |
| Regulatory Class: | Class II |
### Predicate Device Information
Spirotome Cervicore (K080095)
The predicate device has not been subject to a design-related recall.
### Device Description
The Spirotome Endo is a core biopsy device. It consists of a stainless steel receiving needle with a helical shape, stainless steel cutting cannula, and a polypropylene releasing element. The receiving needle is inserted though the cutting cannula prior to advancing to the biopsy site. The receiving needle is rotated clockwise into the tissue using the plastic hub on the proximal end. The cutting cannula is then rotated clockwise to separate the biopsy sample from the surrounding tissue. Following retraction from the biopsy site, the releasing element is used to remove the tissue sample from the receiving needle. The receiving needle is rotated counter-clockwise to release the tissue. The Spirotome Endo is operated manually.
The Spirotome Endo has four variants: one 8 gauge, and two 14 gauge needles. The receiving needle is 375 mm in length and the cutting cannula is 325 mm long. The helical needle is 18 mm long in all the variants except one 14 gauge variant, where it is 10 mm long. The device is provided sterile and is for single-use only.
{4}------------------------------------------------
### Indications for Use
The Spirotome Endo is a family of large core soft tissue biopsy medical devices to be used in humans to take out adequate samples of tissue.
The Spirotome Endo is intended to provide soft tissue for histological and biomolecular detection and confirmation of imaged abnormalities of the internal female reproductive organs such as uterus, ovaries, and retroperitoneal lymph nodes.
#### Substantial Equivalence Discussion
| Attribute | Spirotome Endo - Subject Device | Spirotome Cervicore -<br>Predicate Device |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Bioncise NV | Medinvents |
| 510(k) Number | K190378 | K080095 |
| Indications for Use | The Spirotome Endo is a family of<br>large core soft tissue biopsy medical<br>devices to be used in humans to take<br>out adequate samples of tissue.<br><br>The Spirotome Endo is intended to<br>provide soft tissue for histological<br>and biomolecular detection and<br>confirmation of imaged<br>abnormalities of the internal female<br>reproductive organs such as uterus,<br>ovaries, and retroperitoneal lymph<br>nodes. | The Cervicore system is a family of<br>core soft tissue biopsy medical<br>devices to be used in humans to take<br>out adequate samples of tissue from<br>the human cervix.<br><br>The Cervicore system is intended to<br>provide soft tissue for histological<br>and biomolecular detection and<br>confirmation of imaged<br>abnormalities by inspection or<br>colposcopy.<br><br>The Cervicore system is intended to<br>provide soft tissue for histological<br>and biomolecular detection and<br>confirmation of palpable<br>abnormalities.<br><br>An early detection and correct<br>diagnosis is a prerequisite for a<br>successful multimodality treatment<br>of malignant disease. Therefore, an<br>adequate sampling of human tissue<br>is necessary to make a reliable<br>diagnosis of benignancy or<br>malignancy and to characterize as<br>complete as possible the nature of<br>the lesion without unnecessary<br>discomfort to the patient.<br><br>In case of malignancy, a proper<br>multimodality treatment is necessary<br>i.e. surgery and/or<br>chemotherapy and/or hormone<br>therapy and/or biological therapy. In<br>case of benignancy, adequate follow-<br>up of the patient can be provided. |
| Biopsy Tissue<br>Types | Uterine, ovarian, and retroperitoneal<br>lymph nodes | Cervical |
#### Table 1: Substantial Equivalence Comparison
Bioncise NV – Bodemstraat 16 – 3830 Wellen – Belgium www.bioncise.com – info@bioncise.com tel + 32 11 271557 - fax + 32 11 283677
{5}------------------------------------------------
| Device<br>Components | Cutting Cannula | Cutting Cannula |
|--------------------------|--------------------------------------------------|--------------------------------------------------|
| | Receiving Needle | Receiving Needle |
| | Releasing Element | Releasing Element |
| Helical Needle<br>Length | 10, 18 mm | 10, 20 mm |
| Component<br>Length | Receiving Needle: 375 mm | Receiving Needle: 384.5 mm |
| | Cutting Cannula: 325 mm | Cutting Cannula: 350 mm |
| Device Materials | Stainless steel, polycarbonate,<br>polypropylene | Stainless steel, polycarbonate,<br>polypropylene |
| Sterilization | Steam Sterilized | Steam Sterilized |
| Number of Uses | Single-use; disposable | Single-use; disposable |
| Shelf Life | Three years | Three years |
The Spirotome Endo and the predicate device (K080095) do not have identical Indications for Use statements. The indication for the subject device includes different biopsy tissue types and does not include statements referring to diagnosis or treatment. This change does not represent a new intended use as the intended use of this device is the same as the predicate – to provide soft tissue biopsy samples.
## Non-Clinical Performance Testing
## Bench Testing
Bench testing of the subject device, Spirotome Endo, including functional testing, biocompatibility, sterilization and package testing was leveraged from testing conducted on the Spirotome SU (K060384) and Spirotome Cervicore (K080095). The subject device has identical packaging materials, materials of construction, manufacturing, and sterilization processes as the cleared devices.
### Literature
As part of demonstrating substantial equivalence to the predicate, literature was provided demonstrating safe and effective use of the subject device for internal female reproductive organs. Clinical use of the Spirotome Endo for gynecological tissues was reported in peer reviewed journals. The reports in clinical literature supported an acceptable safety profile for the subject device during biopsy procedures. The reports did not raise additional concerns regarding the safety and effectiveness of the subject device.
### Conclusion
The results of the information and testing described above demonstrate that the Spirotome Endo is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.