CORAMATE/SPIROTOME SYSTEM

K060384 · Medinvents · KNW · Jul 13, 2006 · Gastroenterology, Urology

Device Facts

Record IDK060384
Device NameCORAMATE/SPIROTOME SYSTEM
ApplicantMedinvents
Product CodeKNW · Gastroenterology, Urology
Decision DateJul 13, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

The Coramate/Spirotome system is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Coramate/Spirotome system is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surqical procedures.

Device Story

Mechanical biopsy system for soft tissue harvesting, specifically breast tissue. Components: needle set (Spirotome) and powered control unit (Coramate). Operation: trocar needle guides cutting cannula to site; receiving needle with helix penetrates tissue; cutting cannula frees sample. Coramate unit automates needle movement via four motors and applies vacuum for performance. Used in clinical settings for diagnostic sampling. Output: tissue sample for histologic examination. Benefits: enables minimally invasive tissue acquisition for malignancy detection. Healthcare providers use samples to confirm diagnosis; if non-benign, surgical margins must be verified.

Clinical Evidence

Clinical evidence includes published studies in peer-reviewed journals and meetings regarding the Spirotome needle interface, supplemented by recent clinical data specific to the Coramate system. Testing performed on human breast tissues. Results indicate the system performs as intended with maximal safety; no complications reported in clinical tests or subsequent vigilance quality control follow-up. Bench testing and preclinical animal breast tissue testing also conducted to ensure performance.

Technological Characteristics

System comprises a needle set (trocar, receiving needle, cutting cannula) and a powered hand-held device. Features four small motors for automated needle movement, vacuum assistance, and battery power. Interface is identical for manual (Spirotome) and powered (Coramate) operation. Software guides motor movements. Materials and sterilization details not specified. Connectivity: standalone powered device.

Indications for Use

Indicated for diagnostic sampling of breast tissue abnormalities to obtain samples for histologic examination. Applicable to patients requiring biopsy of imaged breast abnormalities. Not for therapeutic use.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Premarket Notification 510(k) Summary (per 21 CFR 807.92) ## Coramate/Spirotome System #### 1. Submitted by: Magda Buttiens CEO Medinvents B-3500 Hasselt - Belgium Phone: +3211271557 Fax: +3211283677 Email: magda.buttiens@medinvents.com Date prepared: April 4, 2006 #### 2. Device Name: Trade Name: Coramate/Spirotome system Classification Name: Biopsy Needle ### 3. Predicate Device: The Mammotome Biopsy System (K033700) is substantially equivalent to other biopsy needle systems on the market, including the Coramate/Spirotome system. ## 4. Device Description: The Coramate/Spirotome system is a mechanical biopsy device to harvest soft tissues from the human body, in particular from the female breast to detect and confirm malignant cells in the earliest phase of progression. The Coramate/Spirotome system consists of 2 major components: the set of needles (Spirotome) and a powered device that operates the needles (Coramate). The set of needles can be operated manually as well (single-use and reusable Spirotome). The needle set contains 3 needles that work in conjunction. The trocar needle brings the cutting cannula up to the diseased site. The receiving needle, with a helix at the top, penetrates the diseased area, and the cutting cannula frees the sample from the surrounding tissues. The following accessories are optionally provided with the system e.g. releasing element, needle spacer and needle protecting tube. ## 5. Intended Use The Coramate/Spirotome system is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. {1}------------------------------------------------ The Coramate/Spirotome system is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surqical procedures. #### 6. Technological Characteristics The Coramate/Spirotome system is equivalent to the currently marketed vacuum assisted powered macrobiopsy systems, including the Mammotome (K033700). It contains a needle set and a hand held powered device. In addition, it has the same intended use. The basic system is a radial cutting of a cutting cannula on a helicoidal shaped receiving element. This system can be operated manually, whereby the system is advanced by the aid of a trocar needle. Combined, this needle set is called the Spirotome and can be delivered single-use or as a reusable device. The Coramate is a powered device that operates the same needles automatically. The interface with the patient is the same both in the Coramate and Spirotome: i.e. the needle set. In addition, the Coramate includes a vacuum to the needles for maximal performance. The Coramate contains a combination of interfaces, motors and software in addition to a battery with loader. The interfaces fix the needle set into the device. The 4 small motors perform the necessary movements of the needles. The software guides the motors into smooth and ordered movements of the needles. The batteries power the device. Accessories may be added to the system: e.g. releasing element, spacer, protecting ring, and battery loader. The technological differences between the Coramate/Spirotome and the predicate device were introduced to improve performance and safety. ### 7. Performance data Thorough preclinical testing was performed to ensure the device performs as intended. In particular most laboratory and preclinical testing was done on animal breast tissues to ensure maximum performance in the human situation. Clinical testing indicates that all performance and safety aims are reached. Since the interface of the system is the same in the Spirotome compared to the Coramate, all the clinical evidence gained by the Spirotome is relevant to the Coramate. Most of the clinical work has been done in human breast tissues as has been published in international PEER reviewed Journals and Meetings. In addition, recent clinical evidence is added relating only to the Coramate. In summary, all this clinical data indicate that the Coramate and Spirotome perform as intended, with aimed performance and maximal safety and similar to the predicate device. In particular, no complication was noted up to now in the clinical tests and subsequent vigilance quality control followup. The Spirotome device complies with the European Medical Device Directive 92/42/EEC. The Coramate device is submitted for approval in the European Community and {2}------------------------------------------------ approval is pending. All requested standard testing according to the claims made have been done. #### 8. Conclusion Based upon the testing and comparison to the predicate device, the Coramate/Spirotome system has the same intended use with similar technological characteristics as the predicate devices. The system performs as intended and raises no new safety or effectiveness issues. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" arranged around the perimeter. Inside the circle is a stylized graphic of a caduceus, a symbol often associated with medicine and healthcare. The caduceus is depicted with a staff entwined by two snakes and topped with wings. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MEDINVENTS % Magda Buttiens CEO Klein Hillststraat 5 Hasselt, Limburg Belgium 3500 JUL 1 3 2006 Re: K060384 Trade/Device Name: Coramate/Spirotome System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: KNW Dated: June 19, 2006 Received: July 6, 2006 Dear Ms. Buttiens: This letter corrects our substantially equivalent letter of May 10, 2006. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the cnclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ #### Page 2 - Ms. Magda Buttiens comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, fartare Briend Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K 060384 Device Name: Coramate/Spirotome system #### Indications For Use: The Coramate/Spirotome system is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Coramate/Spirotome system is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surqical procedures. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The Counter Use (21 CFR 807 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINU ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) LON IN (Division Sign-Off) Division of General. Restorative. and Neurological Devices **510(k) Number** K060384 Page 1 of _1__
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...