Transcutaneous Electrical Nerve Stimulation (Program 2, 3, 4, 6, 8, 9): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. Powered Muscle Stimulation (Program 1, 5, 7, 10, 11, 12): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
MHD TENS is a portable, battery-powered (DC 3.7V) device providing TENS and Powered Muscle Stimulation (PMS). Input: user-selected programs (12 total) via touch-screen interface. Transformation: electronic stimulatory module generates electrical pulses delivered through adhesive electrode pads to target body areas. Output: biphasic, symmetric, rectangular electrical pulses. Context: OTC use in home healthcare environment. Operation: user selects program, intensity, and timing via LCD interface; screen stylus and lock key provided for navigation. Clinical impact: provides temporary pain relief and muscle conditioning. Benefits: non-invasive, portable pain management and muscle performance enhancement.
Clinical Evidence
Bench testing only. Device performance verified against AAMI/ANSI ES 60601-1 (general safety), IEC 60601-2-10 (nerve/muscle stimulators), IEC 60601-1-2 (EMC), IEC 62133 (battery safety), and IEC 60601-1-11 (home healthcare environment). No clinical data presented.
Technological Characteristics
Materials: ABS and aluminum alloy housing. Energy: DC 3.7V rechargeable lithium battery. Output: 2 channels, alternating, regulated current. Waveform: Pulsed, symmetric, biphasic, rectangular with interphase interval. Connectivity: USB port for charging. Standards: AAMI/ANSI ES 60601-1, IEC 60601-2-10, IEC 60601-1-2, IEC 62133, IEC 60601-1-11. Form factor: 132.8 x 65.8 x 13.8 mm, 110g.
Indications for Use
Indicated for adults experiencing temporary pain in shoulder, waist, back, upper/lower extremities due to exercise or household strain; also indicated for stimulation of healthy muscles to improve/facilitate muscle performance.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 19, 2019
MingHuangDa Electronic Co., Ltd % Doris Dong Manager Shanghai CV Technology Co., Ltd. Room 903, No.19 Dongbao Road Songjiang Area Shanghai, 201613 CN
Re: K190115
Trade/Device Name: MHD TENS Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: March 13, 2019 Received: March 21, 2019
Dear Doris Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Vivek Pinto Assistant Director, Acute Injury Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190115
Device Name MHD TENS
Indications for Use (Describe)
Transcutaneous Electrical Nerve Stimulation (Program 2, 3, 4, 6, 8, 9):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
Powered Muscle Stimulation (Program 1, 5, 7, 10, 11, 12):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Floor6, Building A, Taixinglong Industrial Park, Hezhou Village, Xixiang Town, Baoan District, Shenzhen City, China.
# 510(k) Summary
[As required by 21 CFR 807.92]
### 1. Submission Information:
| 510(k) Number: | K190115 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | June 18, 2019 |
| Type of 510(k) Submission: | Traditional |
| Basis for 510(k) Submission: | New device |
| Submitter/Manufacturer: | MingHuangDa Electronic Co.,Ltd. |
| | Floor6, Building A, Taixinglong Industrial Park, Hezhou Village, Xixiang<br>Town, Baoan District, Shenzhen City, China. |
| Contact: | Doris Dong<br>[Consultant, from Shanghai CV Technology Co., Ltd.]<br>Add: Room 903, No. 19 Dongbao Road, Songjiang Area, Shanghai, 201613 China<br>E-mail: doris_d@126.com<br>Tel: 86 21-31261348 / Fax: 86 21-57712250 |
| 2. Device Description: | |
| Proprietary Name: | MHD TENS |
| Common Name: | TENS & PMS |
| Classification Name: | Stimulator, nerve, transcutaneous, over-the-counter,<br>Stimulator, muscle, powered, for muscle conditioning |
| Regulation Number: | 882.5890, 890.5850 |
| Product Code: | NUH, NGX |
| Device Class: | II |
| Review Panel: | Neurology & Physical Medicine |
| Device Description: | MHD TENS is portable and DC 3.7V battery powered device, offering both<br>transcutaneous electrical nerve stimulator (TENS) and powered muscle<br>stimulator (PMS) qualities in one device.<br>MHD TENS has 12 operation programs, which can give certain electrical<br>pulses through electrode adhesive pads to the suggested area of the body<br>where the electrodes are placed.<br>The electronic stimulatory module has the operating elements of Switch,<br>LCD Display screen, Screen lock key, Intensity Modification keys, Timing<br>key, Output sockets, and USB port for battery charging.<br>The LCD screen can display treatment remaining time, battery power,<br>selected program, output port, current intensity, selected intensity and lock<br>state.<br>The device is equipped with accessories of electrode pads, electrode cables,<br>a screen stylus, a battery charger and a USB cable. The electrode cables are<br>used to connect the pads to the device; the USB cable is used to connect the<br>AC charger and the built-in lithium battery. The screen stylus is used to<br>touch and operate the display screen. All accessories, including the USB<br>cable, electrode pads, electrode cables, and the charger can only be replaced |
{5}------------------------------------------------
Floor6, Building A, Taixinglong Industrial Park, Hezhou Village, Xixiang Town, Baoan District, Shenzhen City, China. The electrodes are interchangeable. The application area of electrode pads must be larger than 12cm². The electrode pads are provided by Shenzhen Mailuokang Technology Co., Ltd. with 510(k) cleared Number K152815. Indications for use: Transcutaneous Electrical Nerve Stimulation (Program 2, 3, 4, 6, 8, 9): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. Powered Muscle Stimulation (Program 1, 5, 7, 10, 11, 12): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
{6}------------------------------------------------
MingHuangDa Electronic Co.,Ltd
Floor6, Building A, Taixinglong Industrial Park, Hezhou Village, Xixiang Town, Baoan District, Shenzhen City, China.
## 3. Substantial Equivalence to Predicate device:
Detailed comparison data is included in "Section 9 - Substantial Equivalence Discussion" of this 510(k) submission.
| Parameters | New Device | Predicate Device | Same/Different |
|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| 1. 510(k) Number: | K190115 | K143268 | |
| 2. Marketing clearance date: | | 07/21/2015 | |
| 3. Device Name | MHD TENS | TENS AND POWERED MUSCLE STIMULATOR | |
| 4. Manufacturer | MingHuangDa Electronic Co.,Ltd | Shenzhen As-Tec Technology Co., Ltd. | |
| 5. Intended use | Transcutaneous Electrical Nerve Stimulation<br>(Program 2, 3, 4, 6, 8, 9):<br>To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.<br>Powered Muscle Stimulation (Program 1, 5, 7, 10, 11, 12):<br>It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. | TENS:<br>To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.<br>PMS:<br>It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. | Same |
| 6. Type of use | OTC | OTC | Same |
| 7. Power Source(s)<br>- Method of Line Current Isolation<br>- Patient Leakage Current<br>- Normal Condition (μΑ)<br>- Single Fault Condition (μΑ) | DC 3.7V lithium battery<br>Type BF<br>--<br>< 10μΑ<br>< 50μΑ | DC 3.7V lithium battery<br>Type BF<br>--<br>0.1μΑ<br>0.1μΑ | Same<br>Same<br>Similar Note 1 |
| 8. Average DC current | < 0.01μΑ | < 0.01μΑ | Same |
{7}------------------------------------------------
MingHuangDa Electronic Co.,Ltd
| Floor6, Building A, Taixinglong Industrial Park, Hezhou Village, Xixiang Town, Baoan District, Shenzhen City, China. | | | | | |
|----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|---------------------|
| when device is on but<br>no pulses are being<br>applied (µA) | | | | | |
| 9. | Number of treatment<br>programs | | 12 | 6 | Similar<br>Note 1 |
| 10. | Number of Output<br>channels: | | 2 | 2 | Same |
| | - Synchronous or<br>Alternating? | | Alternating | Synchronous | Different<br>Note 1 |
| | - Method of<br>Channel Isolation | | Voltage<br>Isolation | Voltage<br>Isolation | Same |
| 11. | Regulated Current or<br>Regulated Voltage? | | Regulated current | Voltage control | Different<br>Note 1 |
| 12. | Software/Firmware/M<br>icroprocessor<br>Control? | | Software | Software | Same |
| 13. | Automatic Overload<br>Trip? | | No | No | Same |
| 14. | Automatic No-Load<br>Trip? | | No | No | Same |
| 15. | Automatic Shut Off? | | Yes | Yes | Same |
| 16. | User Override<br>Control? | | Yes | Yes | Same |
| 17. | Indicator<br>Display: | On/Off<br>Status? | Yes | Yes | Same |
| | | Low<br>Battery? | Yes | Yes | Same |
| | | Voltage/<br>Current<br>Level? | Yes | Yes | Same |
| 18. | Timer Range<br>(minutes) | | 10 ~ 60 minutes, 10<br>min/step | 10 ~ 60 minutes, 10<br>min./step | Same |
| 19. | Compliance with<br>Voluntary Standards? | | Yes.<br>AAMI / ANSI ES<br>60601-1, IEC 60601-1-2,<br>IEC 60601-2-10,<br>IEC 62133, IEC<br>60601-1-11 | Yes.<br>AAMI / ANSI<br>ES60601-1,<br>IEC 60601-1-2, IEC<br>60601-2-10, IEC 62133,<br>IEC 60601-1-11 | Same |
| 20. | Compliance with 21<br>CFR 8988? | | Yes | Yes | Same |
| 21. | Weight (grams) | | 110g | 170g | Similar |
| 22. | Dimensions (mm) [W<br>x H x D] | | 132.8*65.8*13.8mm | 93*50*9mm | Similar |
| 23. | Housing Materials &<br>Construction | | ABS+aluminium alloy | ABS | Similar |
| 24. | Waveform | Pulsed, symmetric, biphasic | Pulsed, symmetric, biphasic | Same | |
| 25. | Shape | Rectangular, with interphase interval | Rectangular, with interphase interval | Same | |
| 26. | Maximum Output<br>Voltage (±15%) | Program 1:<br>51.2V@500Ω<br>94V@2kΩ<br>156V@10kΩ | Program 1:<br>35.5V@500Ω<br>52.5V@2kΩ<br>68.5V@10kΩ | Similar<br>Note 2 | |
| | | Program 2:<br>48V@500Ω<br>39.2V@2kΩ<br>49.6V@10kΩ | Program 3:<br>53.5V@500Ω<br>67V@2kΩ<br>68V@10kΩ | | |
| | | Program 3:<br>69.6V@500Ω<br>98V@2kΩ<br>98V@10kΩ | Program 3:<br>53.5V@500Ω<br>67V@2kΩ<br>68V@10kΩ | | |
| | | Program 4:<br>57.6V@500Ω<br>76V@2kΩ<br>124V@10kΩ | Program 1:<br>35.5V@500Ω<br>52.5V@2kΩ<br>68.5V@10kΩ | | |
| | | Program 5:<br>49.6V@500Ω<br>74V@2kΩ<br>134V@10kΩ | Program 2:<br>30V@500Ω<br>52.5V@2kΩ<br>64.5V@10kΩ | | |
| | | Program 6:<br>57V@500Ω<br>82V@2kΩ<br>138V@10kΩ | Program 1:<br>35.5V@500Ω<br>52.5V@2kΩ<br>68.5V@10kΩ | | |
| | | Program 7:<br>51.2V@500Ω<br>100V@2kΩ<br>154V@10kΩ | Program 1:<br>35.5V@500Ω<br>52.5V@2kΩ<br>68.5V@10kΩ | | |
| | | Program 8:<br>64V@500Ω<br>104V@2kΩ<br>134V@10kΩ | Program 6:<br>33V@500Ω<br>55.5V@2kΩ<br>67.5V@10kΩ | | |
| | | Program 9: | Program 6: | | |
| Floor6, Building A, Taixinglong Industrial Park, Hezhou Village, Xixiang Town, Baoan District, Shenzhen City, China. | | | | | |
| 27. | Maximum Output<br>Current (±15%) | 63.2V@500Ω<br>94V@2kΩ<br>128V@10kΩ<br><br>Program 10:<br>78V@500Ω<br>150V@2kΩ<br>168V@10kΩ<br><br>Program 11:<br>51.2V@500Ω<br>98V@2kΩ<br>154V@10kΩ<br><br>Program 12:<br>76V@500Ω<br>152V@2kΩ<br>154V@10kΩ | 33V@500Ω<br>55.5V@2kΩ<br>67.5V@10kΩ<br><br>Program 1:<br>35.5V@500Ω<br>52.5V@2kΩ<br>68.5V@10kΩ<br><br>Program 1:<br>35.5V@500Ω<br>52.5V@2kΩ<br>68.5V@10kΩ<br><br>Program 1:<br>35.5V@500Ω<br>52.5V@2kΩ<br>68.5V@10kΩ | Similar<br>Note 2 | |
| | | Program 1:<br>102.4mA@500Ω<br>47mA@2kΩ<br>15.6mA@10kΩ<br><br>Program 2:<br>96mA@500Ω<br>19.6mA@2kΩ<br>4.96mA@10kΩ<br><br>Program 3:<br>139.2mA@500Ω<br>49mA@2kΩ<br>9.8mA@10kΩ<br><br>Program 4:<br>115.2mA@500Ω<br>38mA@2kΩ<br>12.4mA@10kΩ<br><br>Program 5:<br>99.2mA@500Ω<br>37mA@2kΩ<br>13.4mA@10kΩ<br><br>Program 6:<br>114mA@500Ω | Program 1:<br>71mA@500Ω<br>26.3mA@2kΩ<br>6.9mA@10kΩ<br><br>Program 3:<br>107mA@500Ω<br>33.5mA@2kΩ<br>6.8mA@10kΩ<br><br>Program 3:<br>107mA@500Ω<br>33.5mA@2kΩ<br>6.8mA@10kΩ<br><br>Program 1:<br>71mA@500Ω<br>26.3mA@2kΩ<br>6.9mA@10kΩ<br><br>Program 2:<br>60mA@500Ω<br>26.3mA@2kΩ<br>6.5mA@10kΩ<br><br>Program 1:<br>71mA@500Ω | | |
| Floor6, Building A, Taixinglong Industrial Park, Hezhou Village, Xixiang Town, Baoan District, Shenzhen City, China. | | | | | |
| | | 41mA@2kΩ | 26.3mA@2kΩ | | |
| | | 13.8mA@10kΩ | 6.9mA@10kΩ | | |
| | | Program 7: | Program 1: | | |
| | | 102.4mA@500Ω | 71mA@500Ω…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.