K121719 · Hong Qiangxing (Shenzhen) Electronics Limited · NGX · May 10, 2013 · Physical Medicine
Device Facts
Record ID
K121719
Device Name
SM TENS & PMS
Applicant
Hong Qiangxing (Shenzhen) Electronics Limited
Product Code
NGX · Physical Medicine
Decision Date
May 10, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TENS: To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS: It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Device Story
SM TENS & PMS is a portable, battery-powered device providing transcutaneous electrical nerve stimulation (TENS) and powered muscle stimulation (PMS). Device uses electrode adhesive pads placed on body areas to deliver electrical pulses. User interface includes ON/OFF switch, display screen, mode selection, intensity adjustment, and timer. Device features 6 operation modes and 2 output channels. Powered by DC 3.7V lithium battery, rechargeable via USB. Intended for over-the-counter use to relieve muscle pain or improve muscle performance. Output parameters (voltage, current, frequency) are controlled by internal software. Healthcare providers or patients use the device to manage muscle soreness or facilitate conditioning. Benefits include non-invasive pain relief and muscle stimulation.
Clinical Evidence
Bench testing only. Device performance verified through compliance with IEC 60601-1 (General Safety), IEC 60601-2-10 (Nerve/Muscle Stimulator Safety), IEC 60601-1-2 (EMC), IEC 62133 (Battery Safety), and FCC Part 18 (Conducted Emissions). No clinical data provided.
Technological Characteristics
Portable, DC 3.7V lithium battery-powered stimulator. Delivers pulsed, symmetric, biphasic rectangular waveforms. 2 output channels with voltage transformer isolation. Type BF applied part. Connectivity via USB for charging. Software-controlled operation with 6 modes and adjustable timer (10-60 min). Complies with IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, IEC 62133, and FCC Part 18.
Indications for Use
Indicated for temporary relief of pain in muscles of shoulder, waist, back, neck, arms, and legs due to exercise or household strain (TENS); and stimulation of healthy muscles to improve/facilitate muscle performance (PMS).
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
l long Qiangxing (Shen Zhen) Electronics Limited 4F. Jingcheng Building. Xicheng Industrial Zone. Xixiang Road. Bao an District. Shenzhen. Guangdong. China
Tcl: 86-755 26423605 / Url: www.sunmas.hk
### Section 5 510(k) Summary [As required by 21 CFR 807.92]
1. Submission Information:
Type of 510(k) Submission:
Basis for 510(k) Submission:
Submitter/Manufacturer:
510(k) Number:
Date:
Contactor:
·K121719 October 23td, 2012 Traditional New device Hong Qiangxing (Shen Zhen) Electronics Limited 4F, Jingcheng Building, Xicheng Industrial Zone, Xixiang Road, Bao'an District, Shenzhen City, Guangdong, China 518126 Doris Dong [Consultant, from Shanghai CV Technology Co., Ltd.] Add .: Room 1706 Yuesha, No. 128 Songle Rd., Songjiang, Shanghai, China 201600 E-mail: doris_d@l26.com Tel: 86 21-31261348 Fax: 86 21-37824346
2. Device Description: Proprietary Name: Common Name: . Classification Name:
Regulation Number: Product Code: Device Class: Review Panel: Device Description:
TENS & PMS Stimulator, nerve, transcutaneous, over-the-counter, Stimulator, muscle, powered, for muscle conditioning 882.5890. 890.5850 NUH, NGX II
Neurology & Physical Medicine
SM TENS & PMS
SM TENS & PMS is a portable and DC 3.7V battery powered multi function device, offering both Transcutancous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) qualities in one device. '
SM TENS & PMS has 6 operation modes, which can give certain electrical pulses through electrode adhesive pads to the suggested area of the body where the electrodes are placed.
The electronic stimulatory module has the operating elements of ON/OFF Switch, Display screen, Mode Selection key, Intensity Modification keys, Timing key, Pause key, Output socket, and USB port for battery charging
The display screen can show battery power, selected mode, current intensity, time remaining of an application mode, and indication of a pause.
The device is equipped with accessories of electrode pads, electrode cables, battery chargers, and USB cables. The electrode cables are used to connect the pads to the device; the USB cable is used to connect the charger and the built-in lithium battery. All accessories, including USB cables, electrode pads, electrode cables, chargers can only be changed by special
# MAY 1 0 2013
{1}------------------------------------------------
Hong Qiangxing (Shen Zhen) Electronics Limited
Hong Qiangxing (Shen Zhen) Electronics Limited
4ti, Jingsteng Building, Xicheng Industrial Zone, Xixiang Road, Bao' an District, Shenzhen, Ghangdong. China
Tel: 86-755 264
| Indications for use: | person.<br>The electrodes are interchangeable. The application area of electrode pads<br>must be larger than 4cm². The electrode pads are provided by GMDASZ<br>Manufacturing Co., Ltd. with 510(k) cleared Number K092546.<br>TENS:<br>To be used for temporary relief of pain associated with sore and aching<br>muscles in the shoulder, waist, back, neck, upper extremities (arm), and<br>lower extremities (leg) due to strain from exercise or normal household<br>work activities.<br>PMS:<br>It is intended to be used to stimulate healthy muscles in order to improve |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### 3. Substantial Equivalence to Predicate device:
Detailed comparison data is included in "Section 9 - Substantial Equivalence Discussion" of this 510(k) submission. .
and facilitate muscle performance.
| Parameters | | New Device | Predicate Device |
|-------------|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| 1. | 510(k) Number | K121719 | K102598 |
| 2. | Marketing clearance date: | - | May 13, 2011 |
| 3. | Device Name | SM TENS & PMS | Powered Muscle Stimulator, JQ-SC |
| 4. | Manufacturer | Hong Qiangxing (Shenzhen) Electronics<br>Limited | Hi-Dow International, Inc. |
| 5. | Accessories | Self-adhesive electrodes, electrode wires. | Self-adhesive electrodes, electrode wires, |
| | | Battery charger, USB cable. | Battery charger, USB cable |
| 6. | Intended use | TENS (Mode 1, 3, 4, 5, 6): | TENS (Mode 1. 3. 4. 5. 6): |
| | | To be used for temporary relief of pain | To be used for temporary relief of pain |
| | | associated with sore and aching muscles in | associated with sore and aching muscles |
| | | the shoulder, waist, back, neck, upper | in the shoulder, waist, back, neck, upper |
| | | extremities (arm), and lower extremities | extremities (arm), and lower extremities |
| | | (leg) due to strain from exercise or normal | (leg) due to strain from exercise or |
| | | household work activities. | normal household work activities. |
| | | PMS (Mode 1, 2, 3, 6): | PMS (Mode 1, 2, 3, 6): |
| | | It is intended to be used to stimulate healthy | It is intended to be used to stimulate |
| | | muscles in order to improve and facilitate | healthy muscles in order to improve and |
| | | muscle performance. | facilitate muscle performance. |
| 7. | Power Source(s) | DC 3.7V lithium battery | DC 3.7V lithium battery |
| | - Method of Line Current | Type BF | Type BF |
| | Isolation | | |
| | - Patient Leakage Current | - | - |
| | - Normal Condition | 2μA | 2μA |
| | - Single Fault | < 10μA | < 10μA |
| | Condition | | |
| 8. | Average DC current | < 0.01 μA | < 0.01 μA |
| | through electrodes when | | |
| | Tel: 86-755 26423605 / Url: www.sunmas.hk | | |
| | device is on but no pulses<br>are being applied (µA) | | |
| 9. | Number of Output Modes | 6 | 6 |
| 10. | Number of Output<br>channels: | 2 | 2 |
| | - Synchronous or<br>Alternating? | Synchronous | Synchronous |
| | - Method of Channel<br>Isolation | Voltage transformer Isolation | Voltage transformer Isolation |
| 11. | Timer Range (minutes) | 10 - 60 minutes, 10 min./step | 10 - 60 minutes, 10 min./step |
| 12. | Compliance with Voluntary<br>Standards? | IEC 60601-1, IEC 60601-1-2,<br>IEC 60601-2-10, IEC 62133,<br>FCC 47 CFR Part 18 | IEC 60601-1, IEC 60601-1-2,<br>IEC 60601-2-10 |
| 13. | Waveform | Pulsed, symmetric, biphasic | Pulsed, symmetric, biphasic |
| 14. | Shape | Rectangular, with interphase interval, | Rectangular, with interphase interval |
| 15. | Maximum Output Voltage<br>(volts) | $42V±10% @500Ω$<br>$84V±10% @2kΩ$<br>$130V±10% @10kΩ$ | $62.4V @500Ω$<br>$76V @2KΩ$<br>$84V @10KΩ$ |
| 16. | Maximum Output Current<br>(specify units) | $84mA±10% @500Ω$<br>$42mA±10% @2KΩ$<br>$13mA±10% @10KΩ$ | $124.8mA @500Ω$<br>$38mA @2KΩ$<br>$8.4mA @10KΩ$ |
| 17. | Pulse width (µsec) | 100µs | 100µs |
| 18. | Max. pulse frequency (Hz) | 110Hz | 61.3Hz |
| 19. | Net Charge (µC per pulse) | 0µC @500Ω; Method: Balanced waveform | 0µC @500Ω; Method: Balanced<br>waveform |
| 20. | Maximum Phase Charge,<br>(µC) | 16.8µC @500Ω | 17.92µC @500Ω |
| 21. | Maximum Average Current,<br>(mA) | 0.924mA @500Ω | 1.248mA @500Ω |
| 22. | Maximum Current Density,<br>(mA/cm², r.m.s ) | 0.462mA/cm² @500Ω | 9.92mA/cm² @500Ω |
| 23. | Maximum Average Power<br>Density. (W/cm³) | 9.702mW/cm² @500Ω | 2.72mW/cm²@500Ω |
| | Similarities between New device<br>and Predicate Device: | Same intended use, power supply, components, 6 modes, 2 channels, software controlled,<br>standards, same waveform and wave shape, same pulse width, net charge, similar phase<br>charge and maximum average current | |
| | Differences between New device<br>and Predicate Device: | Different weight and dimensions, different values of Maximum Average Power Density<br>because of different smallest surface area of electrodes;<br>The new device provides safety test report on battery | |
| Conclusion: | | The SM TENS & PMS is substantially equivalent to the Powered Muscle Stimulator,<br>JQ-5C (K102598). This conclusion is based upon comparison on design, technical<br>characteristics, output mode, intended use, and safety standards complied with. Any<br>differences in the technological characteristics do not raise any new safety and<br>effectiveness issues. | |
2-4
{2}------------------------------------------------
**Hong Qiangxing (Shen Zhen) Electronics Limited**
l l one Qiangxing (Shen Zhen) Electronics Limited
4F. Jingcheng Building. Xicheng Industrial Zone. Xixiang Roud. Bao'an District. Shenzhen, Guangdong. China
Tel: 86-755 264
3-4
{3}------------------------------------------------
Hong Qiangxing (Shen Zhen) Electronics Limited
4F. Jingcheng Building, Xicheng Industrial Zone. Xixiang Road, Bao'an District, Shenzhen, Guangdong, China Tel: 86-755 26423605 / Url: www.sunmas.hk
#### 4. Safety and Effectiveness of the device:
SM TENS & PMS is safe and effective as the predicate devices cited above.
The new device has passed testing according to the safety standards:
1) IEC 60601-1: 2005, Medical Electrical Equipment - Part 1: General Requirements for Safety 2) IEC 60601-2-10: 2001, Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators;
3) IEC 60601-1-2: 2001, Medical Electrical Equipment - Part 1-2: General Requirements for Safety -
Collateral Standard: Electromagnetic Compatibility -- Requirements and Tests;
4) IEC 62133: 2002, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
5) FCC 47 CFR Part 18 Telecommunication: Industrial, Scientific, and Medical Equipment; Conducted Emissions
The conclusion drawn from the safety testing is that the new device is substantially equivalent to the predicate device. Furthermore, the new device complies with the recognized standards and performs its intended tasks as well as the legally marketed predicate devices.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
May 14,2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Hong Qiangxing (Shen Zhen) Electronics Limited % Shanghai CV Technology Co., Ltd. Attn: Ms. Doris Dong Room 1706 Yuesha, No. 128 Songle Rd., Songjiang Area Shanghai, 201600 China
Re: K121719
Trade/Device Name: SM TENS & PMS Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX, NUH Dated: April 24, 2013 Received: May 1, 2013
Dear Ms. Dong:
This letter corrects our substantially equivalent letter of May 10, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
Page 2 - Ms. Doris Dong
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Section 4 Indications for Use Statement
510(k) Number (if known):
Device Name: SM TENS & PMS
### Indications for Use:
TENS:
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
PMS:
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
Prescription Use_ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use __ J (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
| Joyce M. Whang -S |
|---------------------|
| (Division Sign Off) |
Division of Neurological and Physical Medicine Devices (DNPMD)
K121719 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.