The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.
Device Story
LungVision System assists physicians in guiding endobronchial tools to pulmonary targets. Input: pre-operative 3D CT datasets and live fluoroscopic X-ray images. Operation: system segments CT data; registers and overlays 3D segmented data onto live fluoroscopic images to provide 3D visualization of tool position relative to target. Modifications include virtual bronchoscopy, C-arm based tomography (stacking images for alternative visualization), multi-view registration (using multiple C-arm angles), and real-time compensation for tool movement. Used in clinical settings by physicians. Output: visual overlay on monitor/tablet to support navigation. Benefits: improved accuracy in navigating to target areas in deformable lung tissue. Tablet used for planning; not for diagnostic purposes.
Clinical Evidence
No clinical testing performed. Evidence consists of bench testing, including verification of requirements, hazard mitigation testing, and performance testing in pig lungs to assess accuracy in deformable tissue.
Technological Characteristics
System includes main unit, tablet, router, and software. Hardware complies with ANSI/AAMI/ES 60601-1 and IEC 60601-1-2. Features 2D/3D medical image acquisition, DICOM CT import, segmentation, and real-time fluoroscopic registration. Software-based image processing includes contrast/brightness adjustment, zoom, pan, and multi-planar reformatting. Operates on Windows 10. No patient-contacting parts.
Indications for Use
Indicated for patients undergoing pulmonary procedures requiring endobronchial tool navigation to a target area within the lungs.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Covidien - superDimension™ Navigation System V7.2 (K173244)
Submission Summary (Full Text)
{1}------------------------------------------------
| | 510(k) Summary<br>April 18, 2019 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Dorian Averbuch<br>Body Vision Medical Ltd.<br>34 Sokolov St.<br>Ramat Hasharon Israel<br>Tel: 646- 863-7848 (US) |
| Submission Correspondent: | Paul Dryden<br>ProMedic, LLC<br>131 Bay Point Dr. NE<br>St. Petersburg, FL 33704 |
| Proprietary or Trade Name: | LungVision System |
| Common/Usual Name: | System, image processing, radiological |
| Classification Name: | Picture archiving and communications system<br>LLZ, Class II, CFR 892.2050 |
| Predicate Device: | BodyVision – LungVision – K163622 |
| Reference Device: | Covidien - superDimension™ Navigation System V7.2<br>K173244 |
#### Device Modification
- 1. Virtual Bronchoscopy
- a. A standard technology in the field of pulmonology. This is a visualization technique showing the same information already available in the CT scan.
- 2. Software upgrades
- a. C-Arm based Tomography - visual aid. Collecting a number of successive images produced by a standard C-Arm which are digitally "stacked" together to form an additional alternative visualization of imaging information sourcing from standard C-Arm.
- b. Multi view Setup - an algorithm improvement. Same airway based registration algorithm, extended to multiple C-Arm angles. previous was limited to one angle (AP, LAO, RAO or other C-Arm position)
- 3. Real time compensation
- Predicate displays a static target overlay on the fluoroscopic image, whereas the proposed a. modification performs small adjustments related to the tool movement on the fluoroscopic screen to the target overlay.
- 4. 3D Guidance
- Predicate calculates the tool position with respect to pre-operative CT. The proposed a. modification provides 3D visualization for those calculations in addition to the existing screen.
- 5. Use of a Tablet in addition to PC
- The Tablet image displayed is not for diagnostic purposes. a.
None of these modifications changes the indications for use, environment of use or patient population.
#### Description:
The Lung Vision System (K163622) is designed to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.
{2}------------------------------------------------
### 510(k) Summary April 18, 2019
The Lung Vision System is designed to assist the physician in guiding endobronchial tools towards the target area of interest inside the patient lungs. Prior to the endoscopic procedure the system allows planning the target location and the path to the target area on the CT scan. During the endoscopic procedure the system overlays planned data over fluoroscopic images to support endobronchial tool navigation towards the area of interest. The system does not include the Fluoroscope, Bronchoscope or the external monitor. Lung Vision system includes a main unit and a tablet vs. the previous PC based hardware platform. Image processing algorithms are executed on the main unit and the tablet is used as a primary method of interacting with the system. The Tablet is for planning but is not for diagnostic purposes. Both can perform the following
- functions: segment previously acquired DICOM 3D CT image data, .
- . register DICOM 3D CT image data with live fluoroscopic X-ray image
- . overlay the segmented 3D CT dataset over a live fluoroscopic X-ray image of the same anatomy, obtained on a Fluoroscopic system.
# System Components Overview
The following is a list of the LungVision System's main components: LungVision Main Unit LungVision Tablet LungVision Router LungVision Board (passive device) LungVision Software
Tablet is off-the-shelf information technology equipment powered from the medical grade power supply. The system has been evaluated to AAIM ANSI ES 60601-1 2005+ A1:2012 and IEC 60601-1-2: 2014.
#### Indications for Use:
The Lung Vision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.
#### Contraindications:
None
# Device Comparison
Table 1 compares the subject device to the predicate
| Feature | Proposed Device<br>LungVision | Primary Predicate<br>K163622<br>Body Vision-LungVision |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The LungVision System is intended to<br>enable users to segment previously<br>acquired 3D CT datasets and overlay and<br>register these 3D segmented data sets<br>with fluoroscopic live X-ray images of<br>the same anatomy in order to support<br>catheter/device navigation during<br>pulmonary procedures. | The LungVision System is intended to<br>enable users to segment previously<br>acquired 3D CT datasets and overlay and<br>register these 3D segmented data sets<br>with fluoroscopic live X-ray images of<br>the same anatomy in order to support<br>catheter/device navigation during<br>pulmonary procedures. |
| Classification | Picture archiving and communications<br>system<br>21CFR 892.2050 | Picture archiving and communications<br>system<br>21CFR 892.2050 |
| Feature | April 18, 2019<br>Proposed Device<br>LungVision | Primary Predicate<br>K163622<br>Body Vision-LungVision |
| | Product Code: LLZ | Product Code: LLZ |
| Target anatomy | Lungs | Lungs |
| Anatomy access | Bronchial airways | Bronchial airways |
| Windows OS | Windows 10 | Windows 10 |
| Medical imaging<br>software | Yes | Yes |
| General Image 2D/3D<br>review | Yes | Yes |
| 3D rendering view | Yes | Yes |
| Multi-modality<br>Support | Yes | Yes |
| Image registration | Yes | Yes |
| Multi-planar<br>reformatting (MPR) | Yes | Yes |
| DICOM import | Yes | Yes |
| Fluoroscopic video | Yes | Yes |
| Standard Image<br>viewing tools | Yes | Yes |
| Segmentation tool | Yes | Yes |
| Video capture | Yes | Yes |
| Live Image overlays | Yes | Yes |
| Import prior plans | Yes | Yes |
| Point<br>marking/ tagging | Yes | Yes |
| Navigation type | Visual | Visual |
| Modifications | | |
| Virtual bronchoscopy | Yes | No<br>Reference Covidien -<br>superDimension™ Navigation<br>System V7.2 - K173244 |
| C-Arm based | Yes | No |
| Tomography | | |
| Multi-view set-up | Yes | No |
| Real-time | Yes - software upgrade | No |
| compensation | | |
{3}------------------------------------------------
#### 0(k) Summar 000 0010
# Substantial Equivalence Discussion
We will discuss the table above.
Indications for Use / Patient Population / Environment of Use: As in comparison of Indications For Use above, we can conclude that the indications for use for the LungVision and the predicate are substantially equivalent.
Discussion: The differences in proposed indications for use are minor. The minor differences do not raise new risk or safety concerns, and the subject device can be found substantially equivalent.
Prescriptive: Both the LungVision and predicate are prescription devices.
{4}------------------------------------------------
### 510(k) Summary April 18, 2019
### Discussion: There are no differences.
Design and Technology: The LungVision utilizes the same technological characteristics as the predicate and reference devices. Both include:
- . software applications that provide 2D and 3D medical image acquisition including real-time video image acquisition and visualization of the anatomy
- . allow co-registration of real-time images to previously created 3D image sets based on previously collected DICOM CT images
- . include image enhancements such as contrast and brightness, zoom and pan capabilities
- addition of a Tablet to the PC but not to be used for diagnostic purposes ●
For the Virtual Bronchoscopy feature the reference device - Covidien - superDimension™ Navigation System V7.2 - K173244 has the same feature and intended use.
Discussion: The software modifications including virtual bronchoscopy, C-arm based Tomography, 3D Guidance, and Real-time compensation are similar to the reference device.
### Performance and Specifications:
The performance and specifications demonstrate that the Lung Vision and predicate devices perform the same functions using the same technologies thus can be found substantially equivalent. Discussion: There are no differences, thus the subject device can be found substantially equivalent.
### Compliance with Standards:
LungVision includes hardware and complies with ANSI/AAMI/ES 60601-1:2005(2012) and IEC 60601-1-2:2014.
Discussion: The proposed device complies with the latest standards and thus can be found substantially equivalent.
# Performance Testing:
Nonclinical / Bench: We have performed bench tests and found that the Lung Vision met all requirements specifications and was found to be equivalent in comparison to the predicate. Testing includes verification testing of the requirements, testing of hazards mitigations) and performance testing of the system.
Testing has also been performed with pig lungs to test accuracy in deformable tissue.
Biocompatibility: There are no patient contacting parts of the LungVision System
Animal: No animal testing was performed
Clinical: No clinical testing was performed
# Substantial Equivalence Conclusion
Based upon the foregoing performance testing and comparison to the legally marketed predicate devices for indications for use, technology, and performance we believe we have demonstrated that the Lung Vision System is substantially equivalent in safety and effectiveness to the predicate device.
{5}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement on last page. |
|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
**Indications for Use**
| 510(k) Number ( <i>if known</i> ) | K183593 |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | LungVision System |
| Indications for Use ( <i>Describe</i> ) | The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures. |
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
| XX Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete and
review the collection of information. Send comments regarding this burden estimate or any other aspect of this
information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and
Drug Administration
Office of Chief Information Officer Paperwork
Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION" in a smaller font.
April 18, 2019
Body Vision Medical Ltd. % Mr. Paul Dryden Consultant ProMedic, LLC 131 Bay Point Dr., NE ST. PETERSBURG FL 33704
Re: K183593
Trade/Device Name: LungVision System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: April 1, 2019 Received: April 2, 2019
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{7}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.