K163622 · Bodyvision Medical , Ltd. · LLZ · May 11, 2017 · Radiology
Device Facts
Record ID
K163622
Device Name
Lung Vision
Applicant
Bodyvision Medical , Ltd.
Product Code
LLZ · Radiology
Decision Date
May 11, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Indications for Use
The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.
Device Story
LungVision System assists physicians in guiding endobronchial tools during pulmonary procedures. Input: previously acquired 3D CT datasets (DICOM) and live fluoroscopic X-ray video. Operation: system segments CT data, registers it with live fluoroscopic images, and overlays planned navigation paths onto the live X-ray feed. Used in clinical settings by physicians. Output: visual overlay on external monitor (not provided) showing tool position relative to target anatomy. Benefits: improved navigation accuracy for endobronchial tools toward target areas of interest. System comprises a laptop, video capture box, isolation transformer, and software.
Clinical Evidence
No clinical data. Bench testing only. Verification testing of requirements, hazard mitigations, and system performance conducted. Accuracy in deformable tissue tested using pig lungs. Electrical safety and EMC testing performed per ANSI AAMI ES 60601-1 and IEC 60601-1-2.
Technological Characteristics
PC-based software platform; Windows 10 OS. Hardware includes laptop, video capture box, and IEC 60601-1 compliant isolation transformer. Connectivity: DICOM import, video capture. Image processing: segmentation, registration, and overlay of 3D CT to live fluoroscopy. No patient-contact parts; no biocompatibility required.
Indications for Use
Indicated for use in pulmonary procedures to support catheter/device navigation by segmenting 3D CT datasets and registering/overlaying them with live fluoroscopic X-ray images. No specific age or gender restrictions stated. No contraindications.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, formed by three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 16, 2017
Body Vision Medical Ltd. % Mr. Paul Dryden Consultant PROMEDIC, LLC 24301 Woodsage Drive BONITA SPRINGS FL 34134
Re: K163622
Trade/Device Name: LungVision System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: April 13, 2017 Received: April 14, 2017
Dear Mr. Dryden:
This letter corrects our substantially equivalent letter of May 11, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jeff Balyer
For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K163622
### Device Name
### LungVision System
Indications for Use (Describe)
The Lung Vision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}------------------------------------------------
#### 510(k) Summary April 13, 2017 Page 1 of 5
| Official Contact: | Dorian Averbuch<br>Body Vision Medical Ltd.<br>34 Sokolov St.<br>Ramat Hasharon Israel<br>Tel: 646- 863-7848 (US) |
|----------------------------|-------------------------------------------------------------------------------------------------------------------|
| Proprietary or Trade Name: | LungVision System |
| Common/Usual Name: | System, image processing, radiological |
| Classification Name: | Picture archiving and communications system<br>LLZ, Class II, CFR 892.2050 |
| Predicate Device: | Philips EP-Navigator K062650 |
#### Device Description:
The LungVision System is designed enabling users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.
The Lung Vision System is designed to assist the physician in guiding endobronchial tools towards the target area of interest inside the patient lungs. Prior to the endoscopic procedure the system allows planning the target location and the path to the target area on the CT scan. During the endoscopic procedure the system overlays planned data over fluoroscopic images to support endobronchial tool navigation towards the area of interest. The system does not include the Fluoroscope, Bronchoscope or the external monitor.
Lung Vision image processing algorithms are executed on a PC based hardware platform, which can perform the following functions:
- segment previously acquired DICOM 3D CT image data, .
- register DICOM 3D CT image data with live fluoroscopic X-ray image
- overlay the segmented 3D CT dataset over a live fluoroscopic X-ray image of the same anatomy, ● obtained on a Fluoroscopic system.
### System Components Overview
The following is a list of the LungVision System's main components:
- Laptop with a mouse and DVD reader ●
- . Video Capture Box
- IEC 60601-1compliant Isolation Transformer
- LungVision Board (passive device) ●
- LungVision Software ●
The laptop mouse and DVD reader are off-the-shelf information technology equipment powered from the isolation transformer. The video capture box is manufactured by BodyVision and is also powered from the isolation transformer. The system above has been evaluated to AAIM ANSI ES 60601-12005+ A1:2012 and IEC 60601-1-2: 2014. The system connects to customer supplied imaging and display devices.
{5}------------------------------------------------
#### Indications for Use:
The LungVision System is intended to enable users to segment previously acquired 3D CT datasets and overlay and register these 3D segmented data sets with fluoroscopic live X-ray images of the same anatomy in order to support catheter/device navigation during pulmonary procedures.
#### Contraindications:
None
### Device Comparison
Table 1 compares the subject device to the predicate
| Feature | Predicate<br>K062650 Philips EP-<br>Navigator | Proposed Device<br>LungVision | Comments |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Indications for Use | EP-Navigator is intended to<br>enable users to segment<br>previously acquired 3D<br>CT datasets and overlay and<br>register these 3D segmented<br>data sets with live<br>fluoro X-ray images of the<br>same anatomy in order to<br>support catheter/device<br>navigation during specified<br>procedures. | The LungVision System is<br>intended to enable users to<br>segment previously acquired<br>3D CT datasets and overlay<br>and register these 3D<br>segmented data sets with<br>fluoroscopic live X-ray images<br>of the same anatomy in order to<br>support catheter/device<br>navigation during pulmonary<br>procedures. | Substantially<br>equivalent |
| Classification | Picture archiving and<br>communications system<br>21CFR 892.2050<br>Procut Code: LLZ<br>Class II | Picture archiving and<br>communications system<br>21CFR 892.2050<br>Procut Code: LLZ<br>Class II | Similar |
| Target anatomy | Heart | Lungs | Different anatomy<br>same process |
| Anatomy access | Cardiac vessels | Bronchial airways | Different anatomy<br>same process |
| Windows OS | Yes (Windows 7) | Yes (Windows 10) | LungVision uses<br>latest version of<br>Windows OS |
| Medical imaging<br>software | Yes | Yes | equivalent |
| General Image<br>2D/3D review | Yes | Yes | equivalent |
| 3D rendering view | Yes | Yes | equivalent |
| Multi-modality<br>Support | Yes | Yes | equivalent |
| Image registration | Yes | Yes | equivalent |
| Multi-planar<br>reformatting<br>(MPR) | Yes | Yes | equivalent |
| Feature | Predicate<br>K062650 Philips EP-<br>Navigator | Proposed Device<br>LungVision | Comments |
| DICOM import | Yes | Yes | equivalent |
| Fluoroscopic video | Yes | Yes | equivalent |
| Standard Image<br>viewing tools | Yes | Yes | equivalent |
| Segmentation tool | Yes | Yes | equivalent |
| Video capture | Yes | Yes | equivalent |
| Live Image<br>overlays | Yes | Yes | equivalent |
| Import prior plans | Yes | Yes | equivalent |
| Point<br>marking/ tagging | Yes | Yes | equivalent |
| Navigation type | Visual | Visual | equivalent |
{6}------------------------------------------------
### 510(k) Summary April 13, 2017 Page 3 of 5
### Substantial Equivalence Discussion
We will discuss the table above.
### Indications for Use / Patient Population / Environment of Use:
As in comparison of Indications For Use above, we can conclude that the indications for use for the LungVision and the predicate are substantially equivalent.
Discussion: The differences in proposed indications for use are minor. The minor differences do not raise new risk or safety concerns, and the subject device can be found substantially equivalent.
#### Prescriptive:
Both the LungVision and predicate are prescription devices. Discussion: There are no differences.
### Design and Technology:
The LungVision utilizes the same technological characteristics as the predicate devices. Both:
- are PC based software applications that provide 2D and 3D medical image acquisition including real-. time video image acquisition arid visualization of the anatomy
- use Windows operating systems ●
- allow co-registration of real-time images to previously created 3D image sets based on previously ● collected DICOM CT images
- include image enhancements such as contrast and brightness, zoom and pan capabilities
Discussion: Same design and technology.
### Performance and Specifications:
The performance and specifications demonstrate that the Lung Vision and predicate devices perform the same functions using the same technologies thus can be found substantially equivalent.
Discussion: There are no differences, thus the subject device can be found substantially equivalent.
{7}------------------------------------------------
### 510(k) Summary April 13, 2017 Page 4 of 5
### Compliance with Standards:
LungVision includes hardware and complies with IEC 60601-1:2005 and IEC 60601-1-2:2014. The EP Navigator includes hardware and complies with IEC 60601-1:2005 and IEC 60601-1-2:2007 Discussion: The proposed device comples with the latest standards and thus can be found substantially equivalent.
#### Performance Testing:
#### Nonclinical / Bench:
We have performed bench tests and found that the Body Vision met all requirements specifications and was found to be equivalent in comparison to the predicate. Testing includes verification testing of the requirements, testing of hazards mitigations, and performance testing of the system.
Testing has also been performed with pig lungs to test accuracy in deformable tissue.
The device was also tested to the requirements of ANSI AAMI ES 60601-1 and IEC 60601-1-2.
#### Biocompatibility:
There are no patient contact parts of the LungVision System
#### Animal
No animal testing was performed
### Clinical
No clinical testing was performed
### Substantial Equivalence Conclusion
Based upon the foregoing performance testing and comparison to the legally marketed predicate devices for indications for use, technology, and performance we believe we have demonstrated that the LungVision System is substantially equivalent in safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.