Novidia Universal Bond is indicated for: 1) Direct restorations using light-cured composite resin or compomer 2) Cavity sealing as a pretreatment for indirect restorations 3) Treatment of hypersensitive teeth and/or exposed root surfaces 4) Intraoral repairs of fractured crowns/bridges made of porcelain, hybrid ceramics, or composite resin using light-cured composite resin 5) Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics, or cured composite resin 6) Core build-ups using light- or dual-cured composite resin 7) Cavity sealing under amalgam restorations * For indications 1, 2, 3, 6, and 7 the device exhibits an antibacterial cavity cleansing effect.
Device Story
Resin-based dental bonding system; consists of self-etch primer and light-cured bond. Primer contains adhesion promoter/solvent; bond contains resin, solvent, initiators, and antibacterial insoluble particles (covalently bound quaternary ammonium). Used by dentists in clinical settings to enhance adhesion between mineralized dental surfaces (enamel/dentin) and restorative materials (composites, cements, metals, ceramics). Applied in two steps; light-cured at 430-490nm for 10 seconds. Provides antibacterial cavity cleansing effect for specific indications. Improves restoration retention and reduces hypersensitivity.
Clinical Evidence
Bench testing only. No clinical or animal data provided. Performance validated via ISO 10993 and ISO 7405 biocompatibility testing and bench tests including polymerization conversion degree, viscosity, shear bond strength (zirconia, glass ceramics, enamel, dentin, titanium, composite, amalgam), microleakage, and antibacterial activity. All tests met acceptance criteria.
Technological Characteristics
Resin-based, two-bottle system (self-etch primer and light-cured bond). Contains quaternary ammonium silica (QASi) for antibacterial properties. Curing wavelength 430-490nm; curing time 10 seconds. Biocompatibility per ISO 10993 and ISO 7405.
Indications for Use
Indicated for patients requiring direct composite/compomer restorations, cavity sealing for indirect restorations, treatment of hypersensitive teeth/exposed root surfaces, intraoral repair of fractured dental prosthetics (porcelain, hybrid ceramic, composite), core build-ups, and cavity sealing under amalgam. No specific age or gender restrictions.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
September 24, 2019
Nobio Ltd. % Shoshana (Shosh) Friedman Senior Regulatory Consultant Nobio [c/o ProMedoss, Inc.] 3521 Hatwynn Rd. Charlotte, North Carolina 28269
Re: K183391
Trade/Device Name: Novidia Universal Bond Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: June 24, 2019 Received: June 26, 2019
Dear Shoshana (Shosh) Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael Adjodha Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K183391
Device Name Novidia™ Universal Bond
Indications for Use (Describe)
# Novidia Universal Bond is indicated for:
- 1) Direct restorations using light-cured composite resin or compomer
- 2) Cavity sealing as a pretreatment for indirect restorations
- 3) Treatment of hypersensitive teeth and/or exposed root surfaces
- 4) Intraoral repairs of fractured crowns/bridges made of porcelain, hybrid ceramics, or composite resin using light-cured composite resin
- 5) Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics, or cured composite resin
- 6) Core build-ups using light- or dual-cured composite resin
- 7) Cavity sealing under amalgam restorations
- * For indications 1, 2, 3, 6, and 7 the device exhibits an antibacterial cavity cleansing effect.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K183391
## 510(K) SUMMARY [as required by section 807.92(c)] Novidia™ Universal Bond 510(k) Number K183391
#### 5.1 SUBMITTER
#### Applicant's Name:
Nobio Ltd. 8 Hamatechet St. POB 50502 Kadima, Israel 6092000 Phone: +972-3-9059966
### Contact Person:
Shoshana (Shosh) Friedman Regulatory Affairs Consultant 3521 Hatwynn Rd. Charlotte, NC 28269 Phone: 704-430-8695 s.friedman@promedoss.com
### Date Prepared:
January 3, 2019
### 5.2 DEVICE
Trade Name: Novidia™ Universal Bond
Name: Agent, Tooth Bonding, Resin Classification Code: Product Code: KLE Regulation No: 872.3200 Class: 2 Review Panel: Dental
### 5.3 PREDICATE DEVICE
- . Primary predicate device: CLEARFIL™ Protect Bond, manufactured by Kuraray Medical Inc., cleared under K023842 and K033938
- Reference device: 3M™ ESPE™ Adper™ Scotchbond™ multi-purpose, . manufactured by 3M Company, cleared under K920424 and K942493
### 5.4 DEVICE DESCRIPTION
Novidia™ Universal Bond is a resin-based bond system consisting of a self-etch primer and a light-cured bond. It is designed to enhance penetration into prepared
{4}------------------------------------------------
enamel and dentin surfaces and to establish strong adhesion between mineralized dental surfaces and the restorative materials such as resin composites, resin cements, metals, and ceramics (e.g., porcelain, lithium disilicate, zirconia, or hybrid ceramics) restorations.
The chemical composition of the Universal Bond's primer consists of adhesion promoter and solvent, and the Universal Bond consists of resin, solvent, initiators and antibacterial insoluble particles, which contain covalently bound quaternary ammonium (QA).
## 5.5 INDICATIONS FOR USE
Novidia™ Universal Bond is indicated for use in:
- 1) Direct restorations using light-cured composite resin or compomer
- 2) Cavity sealing as a pretreatment for indirect restorations
- 3) Treatment of hypersensitive teeth and/or exposed root surfaces
- 4) Intraoral repairs of fractured crowns/bridges made of porcelain, hybrid ceramics, or composite resin using light-cured composite resin
- 5) Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics, or cured composite resin
- 6) Core build-ups using light- or dual-cured composite resin
- 7) Cavity sealing under amalgam restorations
** For indications 1, 2, 3, 6, and 7 the device exhibits an antibacterial cavity cleansing effect.
## 5.6 SUBSTANTIAL EQUIVALENCE
The Novidia Universal Bond has the same intended use and substantially similar technological characteristics as the Clearfil Protect Bond primary predicate device and its performance characteristics is substantially similar to both the Clearfil Protect Bond and the Adper Scotchbond reference device.
The general comparison between the Novidia Universal Bond, Clearfil Protect Bond and Adper Scotchbond are provided in Table 5-1 and the performance comparison between the devices are provided in Table 5-2 below.
{5}------------------------------------------------
| Element | Subject Device:<br>Novidia Universal Bond | Primary Predicate Device:<br>Clearfil SE Protect Bond | Reference Device:<br>3M Adper Scotchbond |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Nobio, Ltd. | Kuraray Medical, Inc. | 3M ESPE Dental Products |
| 510(k) Number | K183391 | K033938 | K942493 |
| Product Code | KLE | KLE | KLE |
| Indications for Use | 1) Direct restorations using light-<br>cured composite resin or<br>compomer<br>2) Cavity sealing as a pretreatment<br>for indirect restorations<br>3) Treatment of hypersensitive teeth<br>and/or exposed root surfaces<br>4) Intraoral repairs of fractured<br>crowns/bridges made of porcelain,<br>hybrid ceramics, or composite<br>resin using light-cured composite<br>resin<br>5) Surface treatment of prosthetic<br>appliances made of porcelain,<br>hybrid ceramics, or cured<br>composite resin<br>6) Core build-ups using light- or<br>dual-cured composite resin<br>7) Cavity sealing under amalgam<br>restorations<br>** For indications 1, 2, 3, 6, and 7 the<br>device exhibits an antibacterial cavity<br>cleansing effect. | 1) Direct restorations using light-<br>cured composite resin or<br>compomer<br>2) Cavity sealing as a pretreatment<br>for indirect restorations<br>3) Treatment of hypersensitive<br>and/or exposed root surfaces<br>4) Intraoral repairs of fractured<br>crowns/bridges made of porcelain,<br>hybrid ceramics, or composite<br>resin using light-cured composite<br>resin<br>5) Surface treatment of prosthetic<br>appliances made of porcelain,<br>hybrid ceramics, or cured<br>composite resin<br>6) Core build-ups using light- or<br>dual-cured composite resin<br>7) Cavity sealing under amalgam<br>restorations<br>** For indications 1, 2, 3, 6, and 7 the<br>device exhibits an antibacterial cavity<br>cleansing effect. | Bonding all classes of direct<br>composite restorations as well as<br>bonding porcelain veneers, porcelain<br>and composite repairs with light cure<br>composite and bonding light cure<br>composite to amalgam. |
| Components | Primer and Bond | Primer and Bond | Etchant, Primer and Adhesive |
| Element | Subject Device:<br>Novidia Universal Bond | Primary Predicate Device:<br>Clearfil SE Protect Bond | Reference Device:<br>3M Adper Scotchbond |
| Composition of Primer | Adhesive monomer carried in a water-ethanol solution | Adhesive monomer carried in a water-ethanol solution | Aqueous solution of HEMA and a polyalkenoic acid copolymer |
| Composition of Bond (Adhesive) | Matrices of hydrophilic, hydrophobic and amphiphilic nature | Matrices of hydrophilic, hydrophobic and amphiphilic nature | Matrices of hydrophilic, hydrophobic and amphiphilic nature |
| Antibacterial Component | Quaternary Ammonium Silica (QASi) | 12-methacryloyloxy-dodecyl-pyridinium-bromide (MDPB) | None |
| Mode of Use | Two-steps using two bottles of components, the first containing a self-etching primer and the second the bonding agent. | Two-steps using two bottles of components, the first contains a self-etching primer and the second contains the bonding agent. | Three-steps using three bottles, the first contains an etchant, the second contains a primer, and the third contains the bonding (adhesive) agent. |
| Self-Etching | Yes | Yes | No |
| Light-Cured | Yes | Yes | Yes |
| Curing Wavelength | 430-490nm | 400-515nm | Unknown |
| Curing Time | 10 seconds | 10 seconds | 10 seconds |
# Table 5-1: General Comparison between Novidia Universal Bond, Clearfil Protect Bond and Adper Scotchbond
{6}------------------------------------------------
{7}------------------------------------------------
| | Test | Acceptance<br>Criteria | Novidia Universal Bond | Clearfil Protect Bond | 3M Adper Scotchbond |
|-------------------------------------|----------------|----------------------------------------------------------------------------------------------------|------------------------|-----------------------|---------------------|
| Polymerization<br>Conversion Degree | | $\u226570%$ | Passed | Passed | Passed |
| Viscosity | | $550\u00b1450 cP$ | Passed | Passed | Passed |
| Shear Bond Strength | Zirconia | $\u2265 10 MPa$ | Passed | Passed | Passed |
| | Glass Ceramics | | Passed | Passed | Passed |
| | Enamel | | Passed | Passed | Passed |
| | Dentin | | Passed | Passed | Passed |
| | Titanium | | Passed | Passed | Passed |
| | Composite | | Passed | Passed | Passed |
| | Amalgam | $\u2265 7 MPa$ | Passed | Passed | Passed |
| Micro Leakage | | Penetration levels:<br>0 (no penetration)<br>to 1 (only shallow<br>enamel surface<br>penetration). | Passed | Passed | Passed |
| Antibacterial Activity | | Reduction of the<br>viable bacteria | Passed | Passed | NA |
Table 5-2: Performance Comparison between Novidia Universal Bond, Clearfil Protect Bond and Adper Scotchbond
{8}------------------------------------------------
# 5.7 PERFORMANCE DATA
## Non-Clinical Performance Testing:
Non-clinical and biological testing was completed to assess the performance and biocompatibility of the Novidia™ Universal Bond and to support substantial equivalence. The data provided in this 510(k) submission shows that the bonding agent is biocompatible based on the biocompatibility assessment conducted as per ISO 10993 and ISO 7405 and performs as intended based on the bench testing. The list of these tests is provided in Table 5-3.
| Biocompatibility |
|-----------------------------------|
| Cytotoxicity |
| Acute Systemic Toxicity |
| Direct Buehler Sensitization Test |
| Oral Irritation Test |
| Chemical Characterization |
| Material Mediated Pyrogenicity |
| Biological Risk Assessment |
| Bench Testing |
| Polymerization Conversion Degree |
| Viscosity |
| Shear Bond Strength |
| • Zirconia |
| • Glass Ceramics |
| • Enamel |
| • Dentin |
| • Titanium |
| • Composite |
| • Amalgam |
| Visual Inspection |
| Micro Leakage |
| Antibacterial Activity |
Table 5-3: List of Tests Completed on Novidia Universal Bond
# Animal and Clinical Performance Testing:
Animal and clinical performance data was not included.
## 5.8 CONCLUSION
Nobio Ltd. believes that Novidia™ Universal Bond is substantially equivalent to the Clearfil Protect Bond primary predicate device and the 3M Adper Scotchbond reference device. It does not introduce new indications for use, has substantially equivalent technological and performance characteristics, and therefore does not introduce any new safety or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.