The JADE Mobile X-Ray, is a mobile X-ray device, for the purpose of acquiring X-ray images of the desired parts of a patient's anatomy. This device is not intended for mammography, bone density, or dedicated pediatric applications.
Device Story
Mobile X-ray system; models JADE-32 and JADE-40; non-motorized; used for diagnostic imaging when patient transport to fixed equipment is unfeasible. Inputs: user-selected X-ray technique parameters (kV, mA, exposure time) via console. Operation: high-frequency X-ray generator produces X-rays; collimator directs beam. Output: X-ray exposure for capture by user-supplied film screen cassette or flat panel detector. Used in clinical settings; operated by trained personnel. Benefits: facilitates rapid patient positioning and imaging at point-of-care. Device includes high-frequency generator, collimator with LED timer, user-programmable anatomical programs (APR), and exposure hand switch. Optional accessories include remote controller and mobile/portable stands.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance validation of input/output functions.
Technological Characteristics
Microprocessor-controlled high-frequency inverter; 3.2kW or 4kW output; 40-120 kV range; stationary anode X-ray tube; 0.5/1.5 mm focal spot. Connectivity: USB, Bluetooth (optional), DR interface (optional). Standards: IEC/EN 60601-1, IEC 60601-1-3, IEC 60601-2-54, ISO 14971. Manual collimator with LED lamp timer. Software: DRGEM SDK, HT Frame, and Membrane Console Firmware.
Indications for Use
Indicated for acquiring X-ray images of patient anatomy. Not for mammography, bone density, or dedicated pediatric applications.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
DRGEM Corporation % Mr. Carl Alletto Consultant OTech. Inc. 8317 Belew Drive MCKINNEY TEXAS 75071 February 5, 2019
### Re: K183388
Trade/Device Name: JADE Mobile X-Ray Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: December 1, 2018 Received: December 14, 2018
### Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183388
Device Name JADE Mobile X-Ray
Indications for Use (Describe)
The JADE Mobile X-Ray, is a mobile X-ray device, for the purpose of acquiring X-ray images of the desired parts of a patient's anatomy. This device is not intended for mammography, bone density, or dedicated pediations.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
510(k) number: K183388
Date Prepared:
December 1, 2018
Submitter's Information: 21 CFR 807.92(a)(1)
Mr. Ki-Nam YANG Director | QM representative DRGEM Corporation 7F E-B/D Gwangmyeong Techno-Park 60, Haan-ro, Gwangmyeong-si, Gyeonggi-do, 14322 Korea Email: radcheck@drgem.co.kr
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)
| Product Name: | JADE Mobile X-Ray |
|------------------------|-----------------------|
| Common Name: | Mobile X-Ray System |
| Classification: | System, X-Ray, Mobile |
| Regulation Number: | 892.1720 |
| Product Code: | IZL |
| Device Classification: | Class II |
### Predicate Device: 21 CFR 807. 92(a)(3)
The JADE Mobile X-Ray, device is substantially equivalent to K182317:
| Device Classification Name | System, X-Ray, Mobile |
|------------------------------|------------------------------------|
| 510(K) Number | K182317 |
| Device Name | AMADEO M-DR Mini, AMADEO M-AX Mini |
| Applicant | OEHM UND REHBEIN GMBH |
| Regulation Number | 892.1720 |
| Device Classification | Class II |
| Classification Product Code | IZL |
| Subsequent Product Code | MQB |
| Decision | Substantially Equivalent (SESE) |
| Regulation Medical Specialty | Radiology |
| 510k Review Panel | Radiology |
| Summary | Summary |
| Type | Traditional |
### Device Description: 21 CFR 807 92(a)(4)
The JADE Mobile X-Ray, is a mobile x-ray device that comes in two models: JADE-32 (3.2kw max. output and JADE-40 (4kw max. output). JADE is a non-motorized mobile diagnostic xray device that can facilitate X-ray examinations, in situations where it is not possible or
{4}------------------------------------------------
feasible to transport the patient to a ward with fixed equipment. The unit is stable and precise when using the optional Mobile or Portable Stand. The electric tube unit and wheel locks. column rotation, and a simple user interface to provide for added operator convenience and rapid patient positioning. X-Ray technique presets can be saved. The JADE Mobile X-Ray device can be used with a film screen cassette or a flat panel detector which are supplied by the user and are not part of the JADE Mobile X-Ray, device. The software used with the JADE Mobile system is new and is not based on the predicate device.
"Caution" The JADE Mobile X-ray should not be used as a handheld device.
The JADE device consists of:
- . High-Frequency X-ray Generator
- Collimator with 30 seconds LED lamp timer
- User Programmable APR
- . Exposure Hand Switch
- . Software(SDK, HT Frame and Membrane Console Firmware)
| No. | PESS | Processor | Description |
|-----|------|-----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | JADE | JADE_HTC | JADE_HTC at HT control board in x-ray generator controls whole x-ray generation<br>process by the control of System Control Module in CPC_SDK. This module<br>controls x-ray parameters such as kV, mA and exposure time, and controls the<br>filament and detector interfacing. |
| | | JADE_MCC | JADE_MCC at console board in control console consist of generator control<br>module. JADE_MCC display the state of X-ray parameter and generator. |
| 2 | CPC | CPC_SDK | CPC_SDK is the software provides user interface on generator control. CPC_SDK<br>consists of generator control Module and Display Module, System Diagnosis<br>Module |
- . Optional equipment:
- o Remote Controller (to control the Collimator Lamp, X-ray Exposure)
- Portable Stand o
- Mobile Stand O
Indications for Use: 21 CFR 807 92(a)(5)
The JADE Mobile X-Ray, is a mobile X-ray device, for the purpose of acquiring X-ray images of the desired parts of a patient's anatomy. This device is not intended for mammography, bone density, or dedicated pediatric applications.
Technological Characteristics: 21 CFR 807 92(a)(6)
- . JADE Mobile X-ray specifications are in TABLE 1.
- TABLE 2 is the collimator specifications and
- TABLE 3, compares the predicate device and the new device. Any differences . between the predicate and the new device has no impact on safety or efficacy of the new device and does not raise any new potential or increased safety risks and is equivalent in performance to existing legally marketed devices.
TABLE 1 Mobile X-Ray Unit Specs
{5}------------------------------------------------
| Model | JADE-32 | JADE-40 |
|----------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------------|
| Input Power | 100 - 240V~ (Free voltage input),<br>50/60Hz | 110 - 240V~ (Free voltage input),<br>50/60Hz |
| Output Rating | Max. 3.2 kW (40mA@80kV,<br>32mA@100kV, 25mA@120kV) | Max. 4 kW (50mA@80kV,<br>40mA@100kV, 32mA@120kV) |
| Type | Microprocessor controlled High Frequency inverter | |
| KV Range | 40 ~ 120 KV, 81Step (1KV Step) | |
| mA Range | 10 ~ 80 mA, 10Step (10, 12.5, 16,<br>20, 25, 32, 40, 50, 64, 80mA) | 10 ~ 100 mA, 11Step (10, 12.5, 16,<br>20, 25, 32, 40, 50, 64, 80, 100mA) |
| Exposure Time | 0.01 ~ 10 seconds, 21Step (in 25% Steps) | |
| mAs Range | 0.1 ~ 250 mAs, 35Step (in 25% Steps) | |
| X-ray Tube Type | Stationary Anode | |
| Focal Spot Size (Small / Large) | 0.5 / 1.5 mm | |
| Anode Heat Storage Capacity | 56,000HU (40,000 J) | |
| Power Cord Length | 5 m | |
| Exposure Hand-switch Cord Length | 5 m (Max. Length) | |
| X-ray switching frequency | 100 kHz | |
| Control | 2 Point Control (kV, mAs) | |
| Control Display | LED (7-seg display & indicators) | |
| Anatomical Programs | Preprogrammed 9 APR data - User Programmable | |
| External Interface | USB, Bluetooth (Option), DR Interface (Option) | |
| Max. Input Power<br>Rating | 25 A @ 240 V~, 35 A @ 100 V~ | 24 A @ 240 V~, 34 A @ 110 V~ |
| Weight | 16.8kg (37lb) (Including Collimator) | |
### TABLE 2 - Collimator Specs
| Model | KM1 |
|---------------------|-------------------------------------------|
| Manufacturer | DRGEM |
| Control | Manual with 15, 30, 45, 60sec. Lamp timer |
| Field Shape | Rectangular |
| Max. Field Size | 44x44cm (at 100cm SID) |
| Leakage Radiation | < 40mR/hr. (at SID 1m) |
| Max. kVp shield | 150kV |
| Inherent Filtration | 2.0mmAl eq. |
| Luminosity | Over 160LUX at 100cm SID |
| Light source | 19W LED |
| Standard | Rotating flange |
| Option | Tape measure (Max. 200cm) |
| Electrical Rating | 12 - 45VDC, 20W |
{6}------------------------------------------------
| Dimension / Weight | 172(W) x 172(D) x 97(H) mm / 2.5kg(5.5lb) |
|--------------------|-------------------------------------------|
|--------------------|-------------------------------------------|
TABLE 3 – Predicate and Subject Device Comparison
The following information compares the subject device to the predicate. Any differences are due between the subject device and the predicated device has no impact on safety or efficacy of the modified device and does not raise any new potential safety risks and is equivalent in performance to existing legally marketed device.
| Characteristic | Subject Device | Predicate K182317, AMADEO M-<br>DR Mini, AMADEO M-AX Mini |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The JADE Mobile X-Ray, is a<br>mobile X-ray device, for the<br>purpose of acquiring X-ray images<br>of the desired parts of a patient's<br>anatomy. This device is not<br>intended for mammography, bone<br>density, or dedicated pediatric<br>applications. | They Portable Diagnostic<br>Radiographic System are intended<br>for use by a qualified/trained doctor<br>or technician on both adult and<br>pediatric subjects for taking<br>diagnostic radiographic exposures<br>of the skull, spinal column, chest,<br>abdomen, extremities, and other<br>body parts. Applications can be<br>performed with the patient sitting,<br>standing, or lying in the prone or<br>supine position. (Not for<br>mammography). |
| Configuration | Same as predicate | Line operated portable |
| Generator | High frequency made by DRGEM | High frequency made by POSKOM |
| Generator<br>power | Two power levels: JADE-32<br>(3.2kw max. output and JADE-40<br>(4kw max. output). | One power level: 5 KW |
| Peak Voltage | 120kv | 110kv |
| Collimator | DRGEM Corporation (Model KM1) | POSKOM PCMAX-100CAH |
{7}------------------------------------------------
| Characteristic | Subject Device | Predicate K182317, AMADEO M-DR Mini, AMADEO M-AX Mini |
|----------------------|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image<br>acquisition | None | None or Either the PerkinElmer<br>XRpad2 4336 detector as cleared in<br>K161966 or the XenOR 35CW<br>(CareRay CareView1500CW)<br>cleared in K150929 (our model<br>XenOR 35CW) |
| Software | DRGEM SDK | DICOMPACS DX-R |
| Connections | USB, RS232 or Wi-Fi | Ethernet or Wi-Fi |
| DICOM | None | Yes |
| Electrical<br>Safety | Same | Electrical Safety per IEC-60601.<br>EMC per IEC-60601-1-2; IEC<br>60601-1-3 Radiation protection in<br>diagnostic X-ray equipment IEC<br>60601-2-54 Particular<br>Requirements for The Basic Safety<br>And Essential Performance of X-<br>Ray<br>Equipment for Radiography and<br>Radioscopy |
| Image: X-Ray machine | Image: X-Ray machine | Image: X-Ray machine |
### Clinical Testing:
Clinical testing is not necessary for the JADE Mobile system in order to demonstrate substantial equivalence to the predicate device.
Nonclinical Testing:
The complete system has been assessed and tested at the factory and by Standards testing facilities. The JADE Mobile X-Ray, device has passed all predetermined testing criteria. The Validation Test Plan was designed to evaluate input functions, output functions, and actions performed by JADE, and followed the process documented in the System Validation Test Plan.
Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
The main components of JADE Mobile X-ray, comply with the applicable regulatory requirements and design standards as follows:
#### 1) SAFETY
- IEC/EN 60601-1 3.1 Edition .
{8}------------------------------------------------
- . EN 60601-1:2006/A1:2013 Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance
- EN 60601-1-3:2008/A1:2013. Medical electrical equipment -- Part 1-3: General . requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- . EN 60601-1-6:2010/A1:2015. Medical electrical equipment -- Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- EN 60601-2-54:2009, Medical electrical equipment -- Part 2-54: Particular . requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- EN 60601-2-28:2010, Medical electrical equipment -- Part 2-28: Particular . requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
2) EMC
- EN60601-1-2:2015, Medical electrical equipment Part 1-2: General requirements for . safety - Collateral standard: Electromagnetic compatibility - Requirements and tests
- EN55011:2009/A1:2010, CISPR 11 :2009/A1:2010 ●
- EN 61000-3-2:2014, Electromagnetic compatibility (EMC) -- Part 3-2: Limits Limits Limits -● for harmonic current emissions (equipment input current <= 16 A per phase)
- . EN61000-4-2:2009, Electromagnetic compatibility (EMC) - Part 4-2: Testing and measurement techniques - Electrostatic discharge immunity test
- . EN61000-4-3:2006/A1:2007/A2:2010. Electromagnetic compatibility (EMC) - Part 4-3: Testing and measurement techniques - Radiated, radio-frequency, electromagnetic field immunity test
- . EN61000-4-4:2012, Electromagnetic compatibility (EMC) - Part 4-4: Testing and measurement techniques - Electrical fast transient/burst immunity test
- . EN61000-4-5:201, Electromagnetic compatibility (EMC) - Part 4-5: Testing and measurement techniques - Surge immunity test
- . EN61000-4-6:2014. Electromagnetic compatibility (EMC) - Part 4-6: Testing and measurement techniques - Immunity to conducted disturbances, induced by radiofrequency fields
- EN61000-4-8:2010, Electromagnetic compatibility (EMC) Part 4-8: Testing and . measurement techniques - Power frequency magnetic field immunity test
- . EN61000-4-11:2004, Electromagnetic compatibility (EMC) - Part 4-11: Testing and measurement techniques - Voltage dips, short interruptions and voltage variations immunity tests
4) OTHERS
- EN ISO 15223-1:2016, Medical devices Symbols to be used with medical device ● labels, labelling and information to be supplied - Part 1: General requirements (ISO 15223-1:2016)
- IEC TR60878:2003, Graphical Symbols for electrical equipment in medical practice ●
- IEC60417-1:2000, Graphical Symbols for use on equipment-part1 : overview and application
- . ISO 14971: 2012, Medical devices - Application of risk management to medical devices (ISO 14971:2012)
The 510(k) Pre-Market Notification for the JADE Mobile X-Ray, contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device.
The subject device will be manufactured in accordance with the voluntary standards listed in the voluntary standard survey. The subject device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and
{9}------------------------------------------------
intended use does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device.
Therefore, the JADE Mobile X-Ray, device, is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.