The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizers and cycles: Lumen, Non Lumen, Flexible, Fast and Fast Non Lumen Cycles of the V-PRO® 1, 1 Plus, maX, maX 2, 60 and s 2 Sterilizers STERRAD® 100S Sterilizer (default cycle) STERRAD® 200 Sterilizer (default cycle) Standard and Advanced Cycles of the STERRAD® NX Sterilizers with All Clear Technology Express, Flex Scope and Standard Cycles of the STERRAD® 100NX Sterilizers with All Clear technology. The VERIFY V24 BI Challenge Pack is intended for qualification testing of the V PRO 1 Plus, VPRO maX, V-PRO maX 2, V-PRO 60 and V PRO s 2 Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included. The challenge pack is not intended for routine of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
Device Story
Device consists of a self-contained biological indicator (SCBI) containing Geobacillus stearothermophilus spores and growth media, and a challenge pack containing the SCBI and a chemical indicator. Used in healthcare facilities to monitor low-temperature hydrogen peroxide sterilization cycles. User places SCBI or challenge pack in sterilizer; after cycle, SCBI is activated using a mechanical activator to rupture the media ampoule. Incubation at 55-60°C for ≥ 24 hours. Visual output: media color change (orange to yellow) and/or turbidity indicates sterilization failure; orange/non-turbid indicates pass. Provides objective verification of sterilization efficacy, ensuring medical devices are safe for patient use.
Clinical Evidence
Bench testing only. Performance validated via simulated use testing, viable spore population verification (2.0-3.4 x 10^6 spores/BI), D-value (4-8 sec), survival/kill time testing, and reduced incubation time (RIT) testing. All tests met acceptance criteria.
Technological Characteristics
SCBI contains Geobacillus stearothermophilus spores (2.0-3.4 x 10^6/BI) in a plastic vial with a glass media ampoule. Resistance: D-value 4.0-8.0 sec at 2.7 mg/L H2O2. Incubation: 55-60°C. Visual colorimetric detection (magenta to orange/yellow). Challenge pack includes a Class 1 chemical indicator. No software or electronic components.
Indications for Use
Indicated for routine monitoring of specific V-PRO and STERRAD low-temperature sterilization cycles, and for qualification testing of V-PRO systems following installation, relocation, or repair. No specific patient population age or gender restrictions.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 3, 2019
STERIS Corporations Gregory Land Senior Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060
Re: K183300
Trade/Device Name: VERIFY V24 Self-Contained Biological Indicator, VERIFY V24 Biological Indicator Challenge Pack Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: November 26, 2018 Received: November 27, 2018
Dear Gregory Land:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Elizabeth F. Claverie -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183300
Device Name
VERIFY V24 Biological Indicator Challenge Pack
### Indications for Use (Describe)
The VERIFY V24 BI Challenge Pack is intended for qualification testing of the V PRO 1 Plus, VPRO maX, V-PRO maX 2, V-PRO 60 and V PRO s 2 Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs.
The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The challenge pack is not intended for routine of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183300
#### Device Name
VERIFY V24 Self-Contained Biological Indicator
Indications for Use (Describe)
The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizer cycles:
· Lumen, Non Lumen, Flexible, Fast and Fast Non Lumen Cycles of the V-PRO® 1, 1 Plus, maX, maX 2, 60 and s 2 Sterilizers
· STERRAD® 100S Sterilizer (default cycle)
• STERRAD® 200 Sterilizer (default cycle)
· Standard and Advanced Cycles of the STERRAD® NX Sterilizers with All Clear Technology
· Express, Flex Scope and Standard Cycles of the STERRAD® 100NX Sterilizers with All Clear technology
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary For VERIFY® V24 Self-Contained Biological Indicator And VERIFY® V24 Self-Contained Biological Indicator Challenge Pack
# Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
# Manufacturing Facility
STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896
Contact: Gregory Land Senior Regulatory Affairs Specialist
Telephone: (440) 392-7424 Fax No: (440) 357-9198 e-mail: greg_land@steris.com
Janary 2, 2019 Submission Date:
Premarket Notification Number: K183300
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{5}------------------------------------------------
#### Device Name - Self-Contained Biological Indicator(SCBI) 1.
| Trade Name: | VERIFY® V24 Self-Contained Biological Indicator |
|------------------------|-----------------------------------------------------------------------|
| Common/usual Name: | Biological Indicator (BI, SCBI) |
| Device Classification: | Class II |
| Classification Name: | Indicator, Biological Sterilization Process<br>(21 CFR 880.2800, FRC) |
#### 2. Predicate Device - Self-Contained Biological Indicator(SCBI)
VERIFY® V24 Self-Contained Biological Indicator, K172748
#### 3. Description of Device - Self-Contained Biological Indicator(SCBI)
The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare facilites to monitor the V-PRO® Low Temperature Sterilization Systems. It is designed to accompany medical devices placed in the sterilizer.
The user places the VERIFY V24 SCBI into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.
The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.
The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.
{6}------------------------------------------------
#### Intended Use/ Indications for Use – Self-Contained Biological Indicator(SCBI) 4.
The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizers and cycles:
- Non Lumen, Flexible, Lumen, Fast and Fast Non Lumen Cycles of the V-PRO® 1, 1 Plus, maX, maX 2, 60 and s 2 Sterilizers
- Standard Cycle of the STERRAD® 100S Sterilizer ●
- . Standard and Advanced Cycles of the STERRAD® NX Sterilizer with or without All Clear Technology.
- Express, Flex Scope and Standard Cycles of the STERRAD® 100 NX . Sterilizer to include sterilizers with or without All Clear Technology
#### 5. Summary of Technical Characteristics - Self-Contained Biological Indicator(SCBI)
A comparison of technical characteristics versus the predicate is summarized in Table 5-1.
| Feature | VERIFY V24 SCBI (proposed) | VERIFY V24 SCBI Predicate (K172748) | Comparison |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The VERIFY® V24 Self-Contained Biological<br>Indicator is intended for routine monitoring of the<br>following sterilizers and cycles:<br>• Non Lumen, Flexible,<br>Lumen, Fast and Fast Non Lumen Cycles of<br>the V-PRO® 1, 1 Plus, maX, maX 2, 60 and s<br>2 Sterilizers<br>• Standard Cycle of the<br>STERRAD® 100S<br>Sterilizer<br>• Standard and Advanced<br>Cycles of the<br>STERRAD® NX<br>Sterilizer with or without ALLClear<br>Technology<br>• Express, Flex Scope and Standard Cycles of<br>the STERRAD® 100<br>NX Sterilizer to | The VERIFY V24 Self-Contained Biological<br>Indicator is intended for routine monitoring of the<br>following sterilizers and cycles:<br>• Non Lumen, Flexible,<br>Lumen and Fast Non Lumen Cycles of the V-<br>PRO 1, 1 Plus, maX, 60 and maX 2 Sterilizers<br>• Standard Cycle of the<br>STERRAD 100S<br>Sterilizer<br>• Standard and Advanced<br>Cycles of the<br>STERRAD NX<br>Sterilizer<br>• Express, Flex Scope and Standard Cycles of the<br>STERRAD 100 NX<br>Sterilizer | The proposed and predicate<br>devices are identical. The<br>Fast Cycle is a new cycle in<br>the V-PRO s2 Low<br>Temperature Sterilizer,<br>which has been submitted in<br>this premarket notification.<br>The addition of use in<br>STERRAD cycles with All<br>Clear has been added |
| Feature | VERIFY V24 SCBI<br>(proposed) | VERIFY V24 SCBI<br>Predicate (K172748) | Comparison |
| | include sterilizers with<br>or without All Clear<br>Technology | | |
| Indicator<br>organism | Geobacillus<br>stearothermophilus | Geobacillus<br>stearothermophilus | Same |
| Mechanism of<br>action | Visual detection of growth<br>based on media color<br>change in the presence of<br>surviving indicator<br>organisms. | Visual detection of growth<br>based on media color<br>change in the presence of<br>surviving indicator<br>organisms. | Same. simulated use testing<br>demonstrate appropriate<br>monitoring of indicated<br>sterilization cycles. |
| Accessories | VERIFY Incubator and<br>VERIFY SCBI HP<br>Activator (optional) | VERIFY Incubator and<br>VERIFY SCBI HP<br>Activator (optional) | Same |
| Viable spore<br>population | 2.0 - 3.4 x 106 spore/BI | 2.0 - 3.4 x 106 spore/BI | Same. Both contain greater<br>than 106 spores/BI. |
| Resistance<br>characteristics | Resistance @ 2.7 mg/L<br>H2O2:<br>• D-value 4.0 – 8.0 sec<br>• Survival Time 4 - 30 sec<br>• Kill Time ≤ 16 min | Resistance @ 2.7 mg/L<br>H2O2:<br>• D-value 4.0 – 8.0 sec<br>• Survival Time 4 - 30 sec<br>• Kill Time ≤ 16 min | Same. Simulated use testing<br>verifies suitability for use in<br>claimed cycles. |
| Culture<br>Conditions | 55-60°C, media included<br>in SCBI, 24 hour<br>incubation time. | 55-60°C, media included<br>in SCBI, 24 hour<br>incubation time. | Same |
| Primary<br>Packaging | Direct inoculum on plastic<br>vial, glass ampoule with<br>recovery media. | Direct inoculum on plastic<br>vial, glass ampoule with<br>recovery media. | Same |
| Process<br>indicator | VERIFY V24 Self-<br>Contained Biological<br>Indicator Vail Label<br>(subject of a separate<br>submission); magenta to<br>orange/yellow color<br>change. | VERIFY V-PRO<br>Chemical Indicator<br>magenta to yellow color<br>change. | Same |
| Shelf-life | 6 months: An 18-month<br>shelf life is established for<br>the SCBI however the<br>throughput process<br>indicator (subject to a<br>separate submission) has a<br>shelf life of 6 months so<br>the maximum labeled<br>shelf life for the SCBI is 6<br>months | 9 months: An 18-month<br>shelf life is established for<br>the SCBI however the<br>throughput process<br>indicator (K140515) has a<br>shelf life of 9 months so<br>the maximum labeled<br>shelf life for the SCBI is 9<br>months | Same |
# Table 5-1. SCBI Physical Description and Technological Properties vs the Predicate Device
{7}------------------------------------------------
#### Summary of Nonclinical Tests – Self-Contained Biological Indicator(SCBI) 6.
Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-3 below.
Table 5-3. Summary of Non-clinical Testing
{8}------------------------------------------------
| Test | Acceptance Criteria | Conclusion |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| $\frac{1}{2}$ & $\frac{3}{4}$ Cycle<br>Performance | All test SCBIs exposed in either the $\frac{1}{2}$ or $\frac{3}{4}$ cycle display passing results | PASS |
| Simulated Use | All processed SCBIs exhibit negative growth results All processed SCBI Label PIs exhibit a “pass” result All processed CIs exhibit a “pass” result | PASS |
| Verification of<br>Viable Spore<br>Population after<br>exposure to<br>ALLClear | The mean population of V24 SCBI exposed to the ALLClear pre-conditioning was -50% to +300% of the unexposed V24 SCBI population | PASS |
| | Table 5-4. Summary of Testing Previously Submitted for VERIFY V24 SCBI |
|-----------|------------------------------------------------------------------------|
| (K140499) | |
| Test | Acceptance Criteria | Conclusion |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------|
| Reduced Incubation<br>Time (RIT) Testing | Meets FDA's requirement of > 97% alignment of the<br>20-minute results with the conventional incubation<br>time of 7 days | PASS |
| Viable spore<br>population | 2.0 - 3.4 x 106 spore/SCBI | PASS |
| D-value | 4-8 sec | PASS |
| Survival Time | 4-30 sec | PASS |
| Kill Time | ≤ 16 min | PASS |
| Hold Time | Performance not affected if incubated within 72 hours<br>of exposure to VHP sterilization | PASS |
#### 7. Device Name – Challenge Pack
| Trade Name: | VERIFY® V24 Self-Contained Biological Indicator<br>Challenge Pack |
|------------------------|-----------------------------------------------------------------------|
| Common/usual Name: | Biological Indicator (BI) Process Challenge Device |
| Device Classification: | Class II |
| Classification Name: | Indicator, Biological Sterilization Process<br>(21 CFR 880.2800, FRC) |
#### 8. Predicate Device – Challenge Pack
VERIFY® V24 Self-Contained Biological Indicator, K172748
{9}------------------------------------------------
#### 9. Description of Device - Challenge Pack
The VERIFY V24 Biological Indicator Challenge Pack is used by healthcare facilities for qualification testing of the V-PRO Low Temperature Sterilization Systems following installation, relocations, malfunctions or major repairs. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital defined challenge load is not included.
The user places the VERIFY V24 Biological Indicator Challenge Pack into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the VERIFY HPU Chemical Indicator (CI) and the VERIFY V24 Self-Contained Biological Indicator (SCBI) contained in the challenge pack are retrieved. The CI is assessed for a passing color change immediately and the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation.
The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activator automatically seals the SCBI vial and releases the growth media.
The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.
#### 10. Intended Use/ Indications for Use - Challenge Pack
The VERIFY V24 BI Challenge Pack is intended for qualification testing of the V-PRO 1, V-PRO 1 Plus, VPRO maX, V-PRO maX 2, V-PRO 60 and V-PRO s 2 Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs.
The challenge pack is placed in an otherwise empty sterilizer chamber; a hospitaldefined challenge load is not included.
The challenge pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
#### Summary of Technical Characteristics - Challenge Pack 11.
A comparison of technical characteristics versus the predicate is summarized in Table 5-1.
{10}------------------------------------------------
| Feature | VERIFY V24 Biological<br>Indicator Challenge<br>Pack<br>(proposed) | VERIFY V24 Biological<br>Indicator Challenge Pack<br>Predicate (K172748) | Comparison |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The VERIFY V24<br>Challenge Pack is<br>intended for qualification<br>testing of V-PRO® 1,<br>V-PRO 1 Plus, V-PRO<br>maX, V-PRO maX 2, V-<br>PRO 60 and V-PRO s 2<br>Low Temperature<br>Sterilization Systems<br>following installation,<br>relocation, malfunctions or<br>major repairs.<br><br>The challenge pack is<br>placed in an otherwise<br>empty sterilizer chamber; a<br>hospital-defined challenge<br>load is not included.<br>The challenge pack is <b>not</b><br>intended for routine<br>monitoring of V-PRO<br>Sterilizers. It has been<br>tested and validated solely<br>for use in Sterilizer<br>qualification testing. | The VERIFY V24<br>Challenge Pack is intended<br>for qualification testing of<br>V-PRO® 1, V PRO 1 Plus,<br>V-PRO maX, V-PRO maX<br>2 and V-PRO 60 Low<br>Temperature Sterilization<br>Systems following<br>installation, relocation,<br>malfunctions or major<br>repairs.<br><br>The challenge pack is<br>placed in an otherwise<br>empty sterilizer chamber; a<br>hospital-defined challenge<br>load is not included.<br>The challenge pack is not<br>intended for routine<br>monitoring of V-PRO<br>Sterilizers. It has been<br>tested and validated solely<br>for use in Sterilizer<br>qualification testing. | The proposed and predicate<br>devices are identical. The<br>V-PRO s 2 Low<br>Temperature Sterilizer, is a<br>new model Sterilizer which<br>has been submitted in a<br>separate premarket<br>notification (K182568). |
| Biological<br>Indicator | VERIFY V24 Self-<br>Contained Biological<br>Indicator (subject of this<br>submission) | VERIFY V24 Self-<br>Contained Biological<br>Indicator | Same |
| Class 1<br>Chemical<br>Indicator | The CELERITY HP<br>Chemical Indicator<br>(subject of a separate<br>Premarket Notification) is<br>placed inside the pouch.<br>A throughput process<br>indicator is also located on<br>the VERIFY V24 SCBI<br>label. | The VERIFY HPU<br>Chemical Indicator is<br>placed inside the pouch.<br>A throughput process<br>indicator is also located on<br>the VERIFY V24 SCBI<br>label. | Same |
| Means to<br>distinguish<br>processed<br>from<br>unprocessed | Chemical indicator of<br>proposed device visible<br>through the pack. | Chemical indicator of<br>proposed device visible<br>through the pack. | Same |
Table 5-1. Challenge Pack Physical Description and Technological Properties vs the Predicate Device
{11}------------------------------------------------
#### Summary of Nonclinical Tests – Challenge Pack 12.
Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-3 below.
| Test | Acceptance Criteria | Conclusion |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Comparitive Dose<br>Response to<br>Biological Model | The challenge pack shall demonstrate equal or greater<br>resistance as compared to the worst case biological<br>model | PASS |
| Simulated Use | • All processed SCBIs exhibit negative growth results<br>• All processed SCBI Label PIs exhibit a “pass” result<br>• All processed CIs exhibit a “pass” result | PASS |
Table 5-3. Summary of Non-clinical Testing
#### 13. Conclusion
Based on the intended use, technological characteristics and nonclinical performance data, the subject device is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate device, VERIFY V24 Self-Contained Biological Indicator (cleared in K172748), Class II (21 CFR 880.2800) product code FRC.
Based on the intended use, technological characteristics and nonclinical performance data, the subject device is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate device, VERIFY V24 Biological Indicator Challenge Pack (cleared in K172748), Class II (21 CFR 880.2800), product code FRC.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.