DR 800 with Tomosynthesis
K183275 · Agfa N.V. · IZF · Feb 1, 2019 · Radiology
Device Facts
| Record ID | K183275 |
| Device Name | DR 800 with Tomosynthesis |
| Applicant | Agfa N.V. |
| Product Code | IZF · Radiology |
| Decision Date | Feb 1, 2019 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1740 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: - Positioning fluoroscopy procedures - Gastro-intestinal examinations - Urogenital tract examinations - Angiography. It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. In addition, the system provides the Agfa Tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-Ray systems. Tomosynthesize tomographic slices from a single tomographic sweep. The DR 800 is not intended for mammography applications.
Device Story
Floor-mounted radiographic, fluoroscopic, and tomographic X-ray system; utilizes X-ray tube, motorized tilting patient table, bucky, and flat-panel detectors. Operates via Agfa NX workstation with MUSICA image processing. Captures digital X-ray images and performs tomographic sweeps to synthesize slices via MUSICA Digital TomoSynthesis (DTS) software. Used in clinical settings by physicians and radiographers. Output consists of diagnostic images for clinical review. Assists in anatomical visualization and diagnostic decision-making. Benefits include replacement of image intensifier technology and enhanced diagnostic capability through tomographic slice acquisition.
Clinical Evidence
Bench testing only. No clinical trials or animal studies performed. Validation included image quality evaluations using anthropomorphic phantoms, compared against primary predicate (K132261). Testing confirmed DTS reconstruction software meets acceptance criteria for adult and pediatric patients. Usability and functionality verified by independent radiographers.
Technological Characteristics
Floor-mounted X-ray system; flat-panel detectors (GOS or CsI scintillator); motorized tilting table. Connectivity via DICOM. Software: MUSICA image processing and MUSICA DTS (Digital TomoSynthesis). Power: 100-240V, 50-60Hz. Compliant with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, and 21 CFR 1020.30-1020.32.
Indications for Use
Indicated for dynamic imaging (fluoroscopy/rapid sequence) and projection radiography of all body parts in adult and pediatric patients. Procedures include positioning, gastro-intestinal, urogenital, and angiography. Includes tomosynthesis for acquiring tomographic slices. Not indicated for mammography.
Regulatory Classification
Identification
A tomographic x-ray system is an x-ray device intended to be used to produce radiologic images of a specific cross-sectional plane of the body by blurring or eliminating detail from other planes. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
- GE Medical System's Discover XR656 with VolumeRAD (K132261)
- Agfa's DR 800 with MUSICA Dynamic (K180589)
Related Devices
- K193262 — DR 600 with Tomosynthesis · Agfa N.V. · Mar 9, 2020
- K212145 — DR 800 with DSA, DR 800 · Agfa N.V. · Aug 31, 2021
- K142341 — X-RAY TV SYSTEM SONIALVISION G4 · Shimadzu Corp. · Mar 6, 2015
- K152244 — RADspeed Pro · Shimadzu Corporation · Jun 20, 2016
- K242395 — Nanox.ARC · Nano-X Imaging , Ltd. · Dec 4, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 1, 2019
Agfa N.V. % ShaeAnn Cavanagh, RAC Regulatory Affairs Manager, North America Agfa US Corp. 10 South Academy Street GREENVILLE SC 29601
Re: K183275
Trade/Device Name: DR 800 with Tomosynthesis Regulation Number: 21 CFR 892.1740 Regulation Name: Tomographic x-ray system Regulatory Class: Class II Product Code: IZF, JAA Dated: November 20, 2018 Received: November 23, 2018
Dear Ms. Cavanagh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hole 2. Nils
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183275
Device Name
DR 800 with Tomosynthesis
### Indications for Use (Describe)
The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures:
- · Positioning fluoroscopy procedures
- · Gastro-intestinal examinations
- · Urogenital tract examinations
- · Angiography
It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts.
In addition, the system provides the Agfa Tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-Ray systems. Tomosynthesize tomographic slices from a single tomographic sweep.
The DR 800 is not intended for mammography applications.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 5px;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span style="padding-right: 5px;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(K) Summary
{4}------------------------------------------------
{5}------------------------------------------------
# 510(K) SUMMARY
### Agfa N.V. DR 800 with Tomosynthesis
#### I. SUBMITTER
Agfa N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Wim Govaerts, Prepared: November 20, 2018 Telephone: + 32 3444 6246
#### DEVICE II.
Name of Device: DR 800 with Tomosynthesis Common Name: System, X-Ray, Tomographic Classification Name: Tomographic X-ray System Regulatory Classification: Class II, 21 CFR 892.1740 Product Code: IZF
Common Name: System, X-Ray, Fluoroscopic Image-Intensified Classification Name: Fluoroscopic X-ray System Regulatory Classification: Class II, 21 CFR 892.1650 Product Code: JAA
#### PREDICATE DEVICES III.
This is a 510(k) for Agfa's DR 800, which is tomographic and fluoroscopic x-ray system. It is substantially equivalent to GE Medical System's Discover XR656 with VolumeRAD (K132261) and Agfa's previous version of the DR 800 with MUSICA Dynamic (K180589).
### Primary Predicate:
Name of Device: Discover XR656 with VolumeRAD Common Name: System, X-Ray, Tomographic Classification Name: Tomographic X-ray System Regulatory Classification: Class II, 21 CFR 892.1740 Product Code: IZF
### Reference Predicate:
Name of Device: DR 800 with MUSICA Dynamic Common Name: System, X-Ray, Fluoroscopic, Image-Intensified Classification Name: Fluoroscopic X-ray System Regulatory Classification: Class II, 21 CFR 892.1650 Product Code: JAA
{6}------------------------------------------------
The DR 800 with MUSICA Dynamic (K180589) predicate has not been subject to a design-related recall.
The Discovery XR656 with VolumeRAD (K132261) primary predicate device was part of a Class II recall initiated on 8/20/2014 and terminated by CDRH on 2/20/2015. The recall was for nylon hooks used to support the FlashPad detector on the wall stand. No injuries were reported.
#### IV. DEVICE DESCRIPTION
Agfa's DR 800 with Tomosynthesis a tomographic and fluoroscopic x-ray system (product codes IZF and JAA) intended to capture tomographic slices of the human body. The DR 800 is a floormounted radiographic, fluoroscopic and tomographic system that consists of a tube and operator console with a motorized tilting patient table and bucky with optional wall stand, FLFS overlay and ceiling suspension. The new device uses Agfa's NX workstation with MUSICA image processing and flat-panel detectors for digital and wide dynamic range capture. It is capable of replacing other direct radiography, tomography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes.
This submission is to add the newest version of the DR 800 with Tomosynthesis to Agfa's radiography portfolio.
The image processing algorithms in the new device are similar to those previously cleared in the DR 800 with MUSICA Dynamic (K180589) and other devices in Agfa's radiography portfolio today which includes DR 600 (K152639) and DR 400 (K141192). The addition of the tomographic image processing is similar to the predicate device (K132261).
Principles of operation and technological characteristics of the new and predicate devices are the same. The new device is virtually identical to Agfa's DR 800 with MUSICA Dynamic (K180589) with the exception that it has additional MUSICA Digital TomoSynthesis (DTS) software for processing tomographic slices. The Discovery XR656 with VolumeRAD (K132261) contains similar software for processing tomographic slices of human anatomy. It uses the same flat panel detectors to capture and digitize the image. Differences in devices do not alter the intended diagnostic effect. Laboratory data and image quality evaluations conducted with independent radiologists confirm that performance is equivalent to the predicates.
Configuration information for the flat-panel detectors can be found in the DR 14s (K161368) and DR 800 (K180589) User Manuals. The DR 14s and RF FL4343 detectors can be integrated in an X-ray system that communicates to a workstation. The Service Manual details the possible configurations and integrations with the NX workstation and X-ray generator. All of Agfa's DR X-ray systems (i.e. DX-D 100-K103597, DX-D 300-K103050, DX-D 600-K112670, DR 400-K141192, DR 600-K152639, DR 800 with MUSICA Dynamic -K180589) will integrate with the detectors. The NX4.x.21 Service Manual, Chapter 4 and associated appendices addresses the installation and configuration with other system components.
#### V. INDICATIONS FOR USE
The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures:
- Positioning fluoroscopy procedures
- Gastro-intestinal examinations
{7}------------------------------------------------
- Urogenital tract examinations
- . Angiography
It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts.
In addition, the system provides the Agfa Tomosynthesis option, which is intended to acquire tomographic slices of human anatomy and to be used with Agfa DR X-Ray systems. Tomosynthesis is used to synthesize tomographic slices from a single tomographic sweep.
The DR 800 is not intended for mammography applications.
NOTE: The mammography applications embedded in the MUSICA software are for previously cleared CR imaging applications (K081963) and not intended for direct radiography (DR), fluoroscopic or tomosynthesis imaging. Furthermore, the additional mammography software is only available through additional license keys that must be purchased. These license keys are only available outside of the USA.
# PEDIATRIC USE SUMMARY
The DR 800 is intended for general populations, including adult and pediatric patients of all ages. Specific design features of the DR 800 for pediatric use include but not limited to using the specific AEC values, NX protocol settings per pediatric age range and automatic brightness control with pediatric settings to lower the dose for pediatric fluoroscopic applications.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH VI. PREDICATE DEVICES
Agfa's DR 800 with Tomosynthesis and GE Medical System's Discovery XR656 with VolumeRAD primary predicate device (K132261) are tomographic x-ray imaging devices, Product Code IZF. The DR 800 with Tomosynthesis also has product code JAA which is identical the Agfa DR 800 with MUSICA Dymanic predicate device (K180589). Agfa's DR 800 is substantially equivalent to the primary predicate device (K132261) in that it uses precisely the same technology to capture and transmit images. The DR 800 is a floor-mounted radiographic, fluoroscopic and tomographic system that consists of a tube and operator console with a motorized tilting patient table and bucky with optional wall stand, FLFS overlay and ceiling suspension. The new device uses Agfa's NX workstation with MUSICA image processing and flat-panel detectors for digital and wide dynamic range image capture. It is capable of replacing other direct radiography, tomography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes.
Principles of operation and technological characteristics of the new and predicate devices are the same. The new device is virtually identical to Agfa's DR 800 with MUSICA Dynamic (K180589) with the exception that it has additional MUSICA DTS software for processing tomographic slices. The Discovery XR656 with VolumeRAD (K132261) contains similar software for processing tomographic slices of human anatomy. It uses the same flat panel detectors to capture and digitize the image. Differences in devices do not alter the intended diagnostic effect.
{8}------------------------------------------------
The optional image processing allows users to conveniently select image processing settings for different patient sizes and examinations. The image processing algorithms in the new device are similar to those previously cleared in the DR 800 with MUSICA Dynamic (K180589) and other devices in Agfa's radiography portfolio today which includes DR 600 (K152639) and DR 400 (K141192). The addition of the tomographic image processing is similar to the predicate device (K132261).
Agfa's DR 800 with Tomosynthesis has an Indications For Use statement virtually identical to the predicate device (K180589) except it includes the addition of tomosynthesis. Intended uses are the same. The devices have the same technological characteristics.
The DR 800 indications for use is equivalent to the primary predicate (K132261) because both include the delineation of anatomical areas and imaging applications and application of tomosynthesis. The DR 800 and both predicate devices (K180589 & K132261) include the statement that the device is not indicated for mammography.
Descriptive characteristics and performance data including image quality evaluations by qualified radiologists are adequate to ensure equivalence. Differences in devices do not alter the intended therapeutic/diagnostic effect.
Table 1 on the next page summarizes the similarities and differences between the new device and predicate.
{9}------------------------------------------------
### Agfa N.V. Premarket Notification: DR 800 with Tomosynthesis
| | DR 800 Tomosynthesis<br>(NEW DEVICE)<br>Agfa | Discovery XR656 Volume RAD<br>(PRIMARY PREDICATE -<br>K132261)<br>GE Medical Systems | DR 800 MUSICA Dynamic<br>(REFERENCE<br>PREDICATE-K180589)<br>Agfa |
|-------------------------|--------------------------------------------------------|--------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Communications | Same as both predicates | DICOM | DICOM |
| Flat Panel<br>Detectors | Same as both predicates | Flat Panel Detectors | Flat Panel Detectors |
| Detector<br>Material | Same as both predicates | Gadolinium Oxysulfide (GOS) or<br>Cesium Iodide (CsI) scintillator | Gadolinium Oxysulfide<br>(GOS) or Cesium Iodide<br>(CsI) scintillator |
| Detector Sizes | Same as predicate K180589 | 40.4 cm x 40.4 cm | 17x17 in.<br>14 x 17 in<br>10 x 10 in |
| Field Sizes | Same as predicate K180589 | 41 x 41 cm<br>35 x 41 cm<br>24 x 30 cm<br>18 x 24 cm | 43 x 43 cm<br>30 x 30 cm<br>20 x 20 cm<br>15 x 15 cm |
| Pixel Size | Same as predicate K180589 | 200 um | 148 um |
| Dynamic Range | Same as predicate K180589 | 14 bit | 16 bit |
| Matrix Size | Same as predicate K180589 | 2022 x 2022 | Static - 2880 x 2880<br>Dynamic - 1024 x 1024 |
| Power Supply | Same as both predicates | 50-60 Hz<br>100-240V auto ranging | 50-60 Hz<br>100-240V auto ranging |
| Operator<br>Workstation | Same as predicate K180589 | Diagnostic Workstation | Agfa NX |
| Image<br>processing | MUSICA Dynamic,<br>MUSICA2<br>MUSICA3/3+<br>MUSICA DTS | VolumeRAD, Dual Energy,<br>Tissue Equalization, Auto Image<br>Paste | MUSICA Dynamic,<br>MUSICA2<br>MUSICA3/3+ |
| Tabletop<br>Features | Same as predicate K180589 | Tilt +/- 90°<br>225 x 89 cm<br>220kg - 320 kg maximum weight | Tilt +/- 90°<br>240 x 80 cm<br>265 kg maximum weight |
| Generators | Same as both predicates | Choice of three models:<br>50, 65KW. 80 KW | Choice of three models:<br>50, 65KW, 80 KW |
| Operating System | Same as both predicates | Windows 7, 8, 8.1, 10 | Windows 7, 8, 8.1, 10 |
| Display System | Same as predicate K180589 | Separately cleared medical<br>display | Separately cleared medical<br>display (K051901) |
{10}------------------------------------------------
### Agfa N.V. Premarket Notification: DR 800 with Tomosynthesis
| Indications for<br>Use Statements | The DR 800 system is<br>indicated for performing<br>dynamic imaging<br>examinations (fluoroscopy<br>and/or rapid sequence) of the<br>following<br>anatomies/procedures:<br>Positioning fluoroscopy<br>procedures, Gastro-intestinal<br>examinations, Urogenital tract<br>examinations, and<br>Angiography. It is intended<br>to replace fluoroscopic<br>images obtained through<br>intensifier technology. In<br>addition, the system is<br>intended for project<br>radiography of all body parts.<br>In addition, the system<br>provides the Agfa<br>Tomosynthesis option, which<br>is intended to acquire<br>tomographic slices of human<br>anatomy and to be used with<br>Agfa DR X-ray systems.<br>Tomosynthesis is used to<br>synthesize tomographic slices<br>from a single tomographic<br>sweep. The DR 800 is not<br>intended for mammography<br>applications. | The Discovery XR656 is<br>intended to generate digital<br>radiographic images of the skull,<br>spinal column, chest, abdomen,<br>extremities, and other body parts<br>in patients of all ages.<br>Applications can be performed<br>with the patient sitting, standing,<br>or lying in the prone or supine<br>position and is intended for use in<br>all routine radiography exams.<br>When the VolumeRAD option is<br>included on the system, the<br>system can generate tomographic<br>images of human anatomy<br>including the skull, spinal<br>column, chest, abdomen,<br>extremities, and other body parts<br>in patients of all ages.<br>VolumeRAD can also be used to<br>detect lung nodules for patients<br>undergoing thoracic imaging.<br>VolumeRAD generates<br>diagnostic images of the chest<br>that aid the radiologist in<br>achieving superior detectability<br>of lung nodule versus posterior-<br>anterior and left lateral views of<br>the chest, at a comparable<br>radiation level. This device is not<br>intended for mammographic<br>applications. | The DR 800 system is<br>indicated for performing<br>dynamic imaging<br>examinations (fluoroscopy<br>and/or rapid sequence) of<br>the following<br>anatomies/procedures:<br>Positioning fluoroscopy<br>procedures, Gastro-intestinal<br>examinations, Urogenital<br>tract examinations, and<br>Angiography. It is intended<br>to replace fluoroscopic<br>images obtained through<br>intensifier technology. In<br>addition, the system is<br>intended for project<br>radiography of all body<br>parts. The DR 800 is not<br>intended for mammography<br>applications. |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Table 1: Device Comparison Table
#### VII. PERFORMANCE DATA
Laboratory testing and software testing (for a moderate level of concern device) using equivalent test protocols were evaluated by qualified individuals employed by the sponsor and independent radiologists to demonstrate that adequate design controls (according to 21 CFR 820.30) were in place.
Verification and validation testing confirmed the device meets performance, safety, usability and security requirements. Pediatric indications were also taken into account. Results were verified and validated.
No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device. No patient treatment was provided or withheld.
### Bench Testing
Clinical image quality evaluations for adults and pediatric patients, performance/functionality and usability data has been provided.
{11}------------------------------------------------
- Technical and acceptance testing was completed on the DR 800 with Tomosynthesis in ● order to confirm the medical device functions and performs as intended. All deviations or variances are documented in a defect database and addressed in the CR&T documentation and verified. All mitigations have been tested and passed. All design input requirements have been tests and passed. All planned verification activities have been successfully completed.
- . Usability and functionality evaluations were conducted with qualified independent radiographers and internal experts. The results of these tests fell within the acceptance criteria for the DR 800 X-ray system therefore, the DR 800 supports a radiographic, fluoroscopic, and tomosynthesis workflow including dynamic and static imaging, continuous and rapid sequence exams, tomographic slices calibration, and positioning.
- Image Quality Validation testing was conducted using anthropomorphic phantoms and ● evaluated by qualified independent radiographers and internal experts. The image quality validation included testing a full range of applications for the DR 800 X-ray system with Tomosynthesis compared to reference images from primary predicate GE Discovery XR656 with VolumeRAD (K132261) using anonymized adult and pediatric phantoms. The pediatric phantom image quality validation testing analyzed five tomographic slices at 5x the dose and five tomographic slices at 10x the dose. Both the 5x the dose and 10x the dose images are clinically sufficient and within the intended use. The test results indicated that the reconstruction software of the image processing for Digital TomoSynthesis (DTS) of the DR 800 X-ray system passed the acceptance criteria and that the DR 800 with Tomosynthesis is capable of making DTS studies for both adult and pediatric patients. The test results showed MUSICA Digital TomoSynthesis (DTS) images were suitable for diagnosis for both adult and pediatric patients.
Performance data including clinical image quality evaluations for adults and pediatric patients, performance/functionality and usability data are adequate to ensure equivalence.
# Software Verification and Validation Testing
Verification and validation plans comprise of test protocols. The complete device has been certified and validated. During the final risk analysis meeting, the risk management team concluded that the medical risk is no greater than with conventional x-ray film previously released to the field.
For the NX4.x.21 (NX Mentor) there are a total of 322 risks in the broadly acceptable region and 27 risks in the ALARP region with only eight of these risks identified. Zero risks were identified in the Not Acceptable Region. Therefore, the device is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk. The software risk assessment is assessed on solution level for the DR 800 and also includes separate risk assessments for the NX4.x.21 software and XRDI.
The term "Level of Concern" means the level of risk that the software device is determined to be if the software were to fail. The Level of Concern for the DR 800 and NX4.x.21 has been determined to be moderate.
{12}------------------------------------------------
# Electrical Safety and Electromagnetic Compatibility (EMC) Testing:
- IEC 60601-1: 2012 Medical Electrical Equipment: General Requirements for Safety and ● Essential Performance.
- IEC 60601-1-2: 2007 Medical Electrical Equipment Part 1-2: General Requirements for ● Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests.
- IEC 60601-1-3: 2008 Medical Electrical Equipment - Part 1-3: General Requirements for Safety and Essential Performance - Collateral Standard: Radiation Protection in Diagnostic X-Ray Equipment
- IEC 60601-2-54: 2009 Medical Electrical Equipment - Part 2-54: Particular Requirements for the Basic Safety and Essential Performance of X-Ray Equipment for Radiography and Radioscopy.
The DR 800 with Tomosynthesis is compliant to the FDA Subchapter J mandated performance standard 21 CFR 1020.30 - 1020.32.
Agfa's in-house standard operating procedures were also used for the development of the device and software; these procedures conform to the following standards:
- ISO 13485:2003 Medical Devices Quality Management Systems ●
- ISO 14971:2012 Application of Risk Management to Medical Devices
- IEC 62304:2006 Medical Device Software - Software life cycle processes
- IEC 62366:2007 Medical Devices Application of Usability Engineering ●
- ACR/NEMA PS3.1-3.20: 2011 Digital Imaging and Communications in Medicine (DICOM) ●
# Guidance Documents
Agfa utilized the following guidance documents in the development of the DR 800 with Tomosynthesis:
- Guidance for the Content of Premarket Submissions for Software Contained in Medical ● Devices (May 2005)
- Cybersecurity for Networked Medical Devices Containing Off-The-Shelf (OTS) Software . (January 2005)
- Off-the-Shelf Software Use in Medical Devices (September 1999) ●
- . Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (October 2014)
- Guidance for Pediatric Information for X-ray Imaging Device Premarket Notifications (November 2017)
# Summarv
Based on the performance data as documented in the above testing. the DR 800 is found to have a safety and effectiveness profile that is similar to the predicate devices.
{13}------------------------------------------------
#### VIII. CONCLUSIONS
Agfa's DR 800 has indications for use that is consistent with those of the legally marketed predicate devices (K132261 and K180589). Intended uses are the same. Where technological characteristics differ lab tests concluded that the device is substantially equivalent to the predicates in that it does not alter the intended therapeutic/diagnostic effect.
The new device and the Discovery XR656 with VolumeRAD primary predicate device (K132261) are tomographic x-ray systems Product Code IZF. The DR 800 with Tomosynthesis also has product code JAA which is identical the Agfa DR 800 with MUSICA Dymanic predicate device (K180589). Agfa's DR 800 with Tomosynthesis is substantially equivalent to both predicate devices (K132261 & K180589) in that it uses precisely the same technology to capture and transmit images.
This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
{14}------------------------------------------------