K183236 · Nanofiber Solutions, LLC · MAI · Mar 11, 2019 · Orthopedic
Device Facts
Record ID
K183236
Device Name
Rotium Bioresorbable Wick
Applicant
Nanofiber Solutions, LLC
Product Code
MAI · Orthopedic
Decision Date
Mar 11, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rotium Bioresorbable Wick is intended to be used in conjunction with suture anchors for the reattachment of tendon to bone in rotator cuff repairs.
Device Story
Rotium Bioresorbable Wick is an electrospun, non-woven, microporous, microfiber matrix accessory for rotator cuff repair; used in conjunction with suture anchors to facilitate tendon-to-bone attachment. Device placed between bone and tendon. Used in clinical settings by orthopedic surgeons. Gamma-sterilized; single-use. Provides physical scaffold to support healing. Benefits include improved tendon-bone interface during repair.
Clinical Evidence
No human clinical data. Evidence includes ISO 10993 biocompatibility testing (cytotoxicity, reactivity, sensitization, systemic toxicity, implantation, genotoxicity, pyrogenicity) showing non-toxic/non-irritating results. Bench testing confirmed dimensional, density, tensile, and retention properties. Ovine rotator cuff model study provided biomechanical and histological data confirming safety and effectiveness.
Technological Characteristics
Electrospun, non-woven, microporous, microfiber matrix. Materials: Poly(lactide-co-caprolactone) (PLCL) and Polyglycolic acid (PGA). Gamma-sterilized. Single-use. Mechanical properties include tensile strength, strain at break, and wick retention strength.
Indications for Use
Indicated for patients undergoing rotator cuff repair requiring reattachment of tendon to bone using suture anchors.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
March 11, 2019
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, featuring a stylized eagle. To the right of the HHS logo is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Nanofiber Solutions, LLC Ronald L. Bracken Chief Operating Officer 4389 Weaver Court North Hilliard, Ohio 43026
Re: K183236
Trade/Device Name: Rotium Bioresorbable Wick Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: February 7, 2019 Received: February 8, 2019
Dear Mr. Bracken:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K183236
Device Name Rotium Bioresorbable Wick
Indications for Use (Describe)
The Rotium Bioresorbable Wick is intended to be used in conjunction with suture anchors for the reattachment of tendon to bone in rotator cuff repairs.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
Image /page/3/Picture/1 description: The image shows the logo for Nanofiber Solutions. The logo consists of a purple square with a white design of intersecting curved lines on the left, followed by the word "Nanofiber" in a larger, purple font. Below "Nanofiber" is the word "SOLUTIONS" in a smaller, purple font.
## SUBMITTER'S INFORMATION
| Owner: | Nanofiber Solutions, LLC |
|-------------------------|---------------------------------------------------------------------------|
| Address: | 4389 Weaver Court North<br>Hilliard, OH 43026 |
| Official Correspondent: | Ronald L. Bracken<br>770-597-7656<br>ronny.bracken@nanofibersolutions.com |
| Date Summary Prepared: | February 7, 2019 |
# DEVICE INFORMATION
| Name of Device: | Rotium™ Bioresorbable Wick |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Bioresorbable Wick Accessory |
| Classification Name: | Fastener, fixation, biodegradable, soft tissue fixation<br>appliances and accessories (21 CFR 888.3030) |
| Regulatory Class: | Class II |
| Product Code: | MAI |
| Predicate Device(s): | BioWick™ X Implant with Driver, K172186 |
| Reason for Submission: | New Device |
| Indication for Use: | The Rotium ™ Bioresorbable Wick is intended to be used in<br>conjunction with suture anchors for the reattachment of tendon<br>to bone in rotator cuff repairs. |
| Device Description: | The Rotium™ Bioresorbable Wick is an accessory to be used<br>in conjunction with suture anchors for rotator cuff repair. The<br>wick is placed between the bone and is designed to facilitate<br>tendon-bone attachment. The wick is an electrospun, non-<br>woven, microporous, microfiber matrix. The wick is made from<br>two types of polymer fibers: Poly(lactide-co-caprolactone)<br>(PLCL) and Polyglycolic acid (PGA). The wick is packaged in a<br>primary foil pouch with a desiccant pouch, sealed within a<br>secondary Tyvek pouch. The device is supplied gamma-<br>sterilized. The device is single use only. |
| Technological<br>Characteristics: | The Rotium™ Bioresorbable Wick accessory when utilized in<br>conjunction with suture anchors has very similar technological<br>characteristics compared to the predicate, BioWick™ X Implant<br>with Driver (K172186). Physical and mechanical properties<br>such as wick materials, wick construction, wick thickness, wick<br>density and wick retention are similar between the products. |
{4}------------------------------------------------
#### Performance Data: Biocompatability
The Rotium™ Bioresorbable Wick has been tested according to ISO 10993, Biological Evaluation of Medical Devices. The following testing was conducted: Cytotoxicity. Intracutaneous Reactivity, Sensitization, Acute Systemic Toxicity, 12 Week Systemic Toxicity, 4-Week Muscle Implantation, 8-Week Muscle Implantation, 12-Week Muscle Implantation, Genotoxicity and Pyrogenicity. The Rotium™ Bioresorbable Wick was found to be non-cytotoxic, nonirritating and nonsensitizing. The Rotium™ Bioresorbable Wick exhibited no systemic toxicity. It was non-pyrogenic and non-mutagenic. The Rotium™ Bioresorbable Wick was also tested for the presence of endotoxin via the LAL kinetic chromogenic method. Test results indicate the endotoxin levels were below the recommended limit of 20 EU.
## Bench Testing
The following test were conducted to demonstrate substantial equivalence to the predicate device's wick: dimensional analysis, density, tensile strength, strain at break, flexibility, suturability, cannula compatibility and wick retention strength.
## Animal Testing
Bioresorbable Wick accessory used in The Rotium™ conjunction with suture anchors was evaluated in an ovine rotator cuff model. Biomechanical testing and histological analysis confirmed that the use of the Rotium™ bioresorbable wick did not introduce any new safety or effectiveness issues.
- Substantial Equivalence: The Rotium™ Bioresorbable Wick accessory when utilized in coniunction with suture anchors is substantially equivalent to the BioWick™ X Implant with Driver (K172186). The Rotium™ Bioresorbable Wick accessory raises no new questions of safety or effectiveness as compared to the predicate devices. The Rotium™ Bioresorbable Wick has the same intended use and indications for use, technological characteristics, and principles of operation as the predicate devices. There are no new novel features as compared to the BioWick X predicate device. The minor differences regarding placement of the bioresorbable wick and wick retention between the Rotium™ device and its predicate device raise no new issues of safety or effectiveness.
- Conclusion: The information in this submission demonstrates that the subject device is substantially equivalent to the cited predicate device. The non-clinical data support the safety and effectiveness of the device and demonstrate that Rotium™ Bioresorbable Wick should perform as intended in the specified use conditions.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.