K183232 · Smith and Nephew, Inc. · HTN · Mar 12, 2019 · Orthopedic
Device Facts
Record ID
K183232
Device Name
Double ENDOBUTTON Fixation Device
Applicant
Smith and Nephew, Inc.
Product Code
HTN · Orthopedic
Decision Date
Mar 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Double ENDOBUTTON Fixation Device is intended for the treatment of anterior glenoid bone loss using the Latarjet or bone block procedure (allograft or autograft).
Device Story
The Double ENDOBUTTON Fixation Device is a sterile, single-use orthopedic implant used during the healing period of osteotomies and fracture fixation. It consists of two stainless steel fixation buttons connected by a UHMW polyethylene suture. The device is used by orthopedic surgeons in a clinical setting to secure bone grafts during Latarjet or bone block procedures for anterior glenoid bone loss. By providing a fixation construct, the device stabilizes the graft, facilitating bone healing and reconstruction. Clinical outcomes, including complication rates and functional scores (ROWE, Walch-Duplay), are comparable to traditional screw fixation methods, with the device offering a potential reduction in hardware-related complications such as migration or breakage.
Clinical Evidence
Clinical literature review comparing the subject device to screw fixation in 299 joints (Double ENDOBUTTON) vs. 1,895 joints (screws). Results showed favorable complication rates for the Double ENDOBUTTON: revision (0% vs 0.85-11.49%), device migration (0% vs 5.26-21.43%), and device breakage (0% vs 4.68%). Clinical outcome scores (ROWE, Walch-Duplay) for the subject device were similar to those reported for screws, with postoperative ROWE scores of 81-95.3 and Walch-Duplay scores of 80-96.
Technological Characteristics
Components: two stainless steel fixation buttons and UHMW polyethylene suture. Sterilization: Gamma or EtO. Single-use, prescription-only. Mechanical testing performed per ANSI/AAMI ST72 and FDA sterility guidance.
Indications for Use
Indicated for patients requiring treatment of anterior glenoid bone loss via Latarjet or bone block procedure (allograft or autograft).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Arthrex Low Profile Screws (3.5mm and larger, cannulated) (K103705)
Smith & Nephew Ultraslide Acromioclavicular and Syndesmotic Repair Device (K082095)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 12, 2019
Smith and Nephew, Inc. Dharaben Desai Regulatory Affairs Specialist II 150 Minuteman Road Andover, Massachusetts 01810
#### Re: K183232
Trade/Device Name: Double ENDOBUTTON Fixation Device Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN, HWC Dated: February 7, 2019 Received: February 8, 2019
Dear Ms. Desai:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K183232
Device Name Double ENDOBUTTON Fixation Device
Indications for Use (Describe)
The Double ENDOBUTTON Fixation Device is intended for the treatment of anterior glenoid bone loss using the Latarjet or bone block procedure (allograft or autograft).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 201 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 201 Subpart C) | ☐ |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitted by: | Smith & Nephew, Inc.<br>Advanced Surgical Devices Division<br>150 Minuteman Road<br>Andover, MA 01810 |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Submission: | February 6, 2019 |
| Contact Person: | Dharaben Desai, Regulatory Affairs Specialist II<br>T (901) 281-8073<br>F (901) 566-7598 |
| Name of Device: | Smith & Nephew, Inc. Double ENDOBUTTON Fixation Device |
| Common Name: | Washer, Bolt Nut<br>Screw, Fixation, Bone |
| Device Classification<br>Name and Reference: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories<br>21 CFR 888.3040 Smooth or threaded metallic bone fixation<br>fastener |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | HTN, HWC |
| Predicate Device: | Arthrex Low Profile Screws (3.5mm and larger, cannulated) -<br>K103705<br>Smith & Nephew Ultraslide Acromioclavicular and<br>Syndesmotic Repair Device - K082095 |
{4}------------------------------------------------
#### Device Description:
The subject of this Traditional 510(k) is the Double ENDOBUTTON Fixation Device. The Smith & Nephew Double ENDOBUTTON Fixation Device is used during the healing period of osteotomies for orthopedic reconstruction and to assist in the management of fracture fixation. The device consists of three components: two stainless steel fixation devices and a UHMW polyethylene suture. This device is provided sterile, for single use only.
#### Indications for Use
The Double ENDOBUTTON Fixation Device is intended for the treatment of anterior glenoid bone loss using the Latarjet or bone block procedure (allograft or autograft).
# Technological Characteristics
The Smith & Nephew Double ENDOBUTTON Fixation Device is substantially equivalent to the predicate screws despite the differences in technological characteristics and design. Performance data and clinical results demonstrate this equivalence and show the subject device does not raise any new issues of safety and efficacy. The technological characteristics comparison between the subject device and predicate devices is provided below.
| Technological<br>Characteristics | Double<br>ENDOBUTTON<br>Fixation Device-<br>Subject device | Arthrex Low Profile<br>Screws (3.5mm and<br>larger, cannulated)-<br>K103705 | Smith & Nephew<br>Ultraslide<br>Acromioclavicular<br>and Syndesmotic<br>Repair Device-<br>K082095 |
|----------------------------------|------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Components | Two<br>ENDOBUTTONS<br>and suture | Screw | Two<br>ENDOBUTTONS and<br>suture |
# Table 1: Summary of Technological Characteristics
{5}------------------------------------------------
| Material | Stainless Steel<br>ENDOBUTTONS,<br>UHMWPE Suture | Titanium or Stainless<br>Steel screw | Titanium<br>ENDOBUTTONS,<br>UHMWPE Suture |
|-------------------------|--------------------------------------------------|-------------------------------------------|-------------------------------------------|
| Labeling | Sterile, single-use<br>only, Rx only | Non-sterile, single-<br>use only, Rx only | Sterile, single-use<br>only, Rx only |
| Sterilization<br>Method | Gamma or EtO | Non-sterile | EtO |
# Performance Data
Mechanical testing demonstrates the device has met the performance specifications for the mechanical integrity of the Smith & Nephew Double ENDOBUTTON Fixation Device. The following performance testing was used as a basis for the determination of substantial equivalence.
- . Interface pressure test
- . Suture construct test with open ended suture lines
- Closed loop suture test
- Suture construct tests with closed looped suture lines .
Bacterial endotoxin testing was completed and met the acceptable endotoxin limits as stated in the FDA Guidance, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72.
# Substantial Equivalence Information
The Smith & Nephew Double ENDOBUTTON Fixation Device is substantially equivalent in intended use and fundamental scientific technology to the following legally marketed devices in commercial distribution:
# Table 2: Predicate Devices
{6}------------------------------------------------
| Manufacturer | Description | Submission<br>Number | Clearance<br>Date |
|----------------|---------------------------------------------------------------|----------------------|-------------------|
| Arthrex, Inc. | Low Profile Screws (3.5mm and<br>larger, cannulated) | K103705 | 03/18/2011 |
| Smith & Nephew | Ultraslide Acromioclavicular and<br>Syndesmotic Repair Device | K082095 | 10/21/2008 |
# Clinical Data
A clinical literature review was completed to compare the subject devices and screws. The literature available for the Double ENDOBUTTON or screws for glenoid bone loss in the shoulder was reviewed. In the literature review, total of 299 joints in 299 patients were treated with a Double ENDOBUTTON, and 1,895 joints in 1,886 patients with various screws.
The complication rate reported after use of the Double ENDOBUTTON compared favorably to screws in most complications such as revision (0% compared to 0.85-11.49% respectively), device migration(0% compared to 5.26-21.43% respectively), device breakage (0% compared to 4.68% respectively) and graft migration (4.5% compared to 3.0-14.89% respectively). No device migrations, breakages, loosening or removals were noted in the publications that reported results of patients treated with the Double ENDOBUTTON.
Clinical outcome scores such as ROWE and Walch-Duplay reported for the subject Double ENDOBUTTON device were similar to those reported by screws. In patients treated with screws, postoperative ASES scores ranged from 86.3 to 93.3 and postoperative ROWE scores ranged from 78.1 to 97.1. In patients treated with the Double ENDOBUTTON, postoperative ROWE score ranging from 81 to 95.3 and Walch-Duplay scores ranged from 80 to 96. Postoperative Walch-Duplay scores for patients treated with screws ranged from 76.7 to 91.7.
# Conclusion
As previously noted, this Traditional 510(k) Premarket Notification is being submitted to request clearance for the Smith & Nephew Double ENDOBUTTON Fixation Device.
{7}------------------------------------------------
Based on the similarities to the predicate components and a thorough review of the clinical and mechanical testing, the devices are substantially equivalent to above predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.