K183199 · Omniguide, Inc. · GEX · May 2, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183199
Device Name
TRIO Handpiece
Applicant
Omniguide, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 2, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OmniGuide® TRIO™ handpiece is indicated for the electrosurgical coagulation of soft body tissues and suction of fluids in general and ENT surgical procedures for the specialties of General Surgery, Head and Neck Surgery, Otolaryngology, and Oral surgery (including intra-oral tissues). When used with the OmniGuide® Laser waveguide, it is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues for the added specialties of Dermatology, Neurosurgery, Orthopedic Surgery, Plastic & Reconstructive Surgery, Podiatry and Urology.
Device Story
Handheld, single-use, sterile surgical instrument; integrates fluid suction, bipolar cautery, and CO2 laser energy delivery. Input: CO2 laser energy via OmniGuide fiber; RF energy from external electrosurgical unit (ESU). Operation: Surgeon holds device like a pencil; directs distal tip to target tissue for suction, coagulation, or laser delivery. Suction controlled via handle aperture. Used in OR/clinic by licensed surgeons. Output: Physical tissue effects (coagulation, ablation, etc.). Benefits: Enables precision control, stabilizes hand motion, reduces surgical time by eliminating need to switch instruments. Connects to standard hospital suction and ESU (up to 750 Vp-p).
Clinical Evidence
Bench testing only. No clinical data. Testing included user validation, beam clipping, suction performance, fiber retention, reliability, mechanical robustness, thermal profile, sterilization validation (ISO 11135:2014), and biocompatibility (ISO 10993-5, 10993-10, 10993-11).
Technological Characteristics
Handheld, single-use, sterile device. Materials biocompatible per ISO 10993. Energy: RF (bipolar) from external ESU; CO2 laser energy via waveguide. Connectivity: 12-foot bipolar cable, standard suction tubing. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62366, ISO 14971, ISO 11135. No internal software/algorithm.
Indications for Use
Indicated for electrosurgical coagulation of soft tissue and fluid suction in general, ENT, head/neck, otolaryngology, and oral surgery. With OmniGuide laser waveguide, indicated for incision, excision, ablation, vaporization, and coagulation in dermatology, gynecology, neurosurgery, orthopedic, plastic/reconstructive, podiatry, and urology. Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
OmniGuide® Laser System with FlexGuide Ultra (K140378)
OmniGuide® CO2 Laser fibers (Velocity™, Elite™, and Select™) (K070157)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font, also in blue. The logo is simple and professional, and it is easily recognizable.
OmniGuide, Inc. Samuel Wade VP, Quality Assurance/Regulatory Affairs 4 Maguire Road Lexington, Massachusetts 02412
May 2, 2019
#### Re: K183199
Trade/Device Name: TRIO Handpiece Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX, GEI Dated: April 18, 2019 Received: April 19, 2019
Dear Samuel Wade:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Jennifer Stevenson, Acting Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K183199
Device Name TRIO™ handpiece
#### Indications for Use (Describe)
The OmniGuide® TRIO™ handpiece is indicated for the electrosurgical coagulation of soft body tissues and suction of fluids in general and ENT surgical procedures for the specialties of General Surgery, Head and Neck Surgery, Otolaryngology, and Oral surgery (including intra-oral tissues). When used with the OmniGuide® Laser waveguide, it is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues for the added specialties of Dermatology, Neurosurgery, Orthopedic Surgery, Plastic & Reconstructive Surgery, Podiatry and Urology.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Attachment 1 Premarket Submission: OmniGuide TRIO" handpiece
Confidential and Proprietary Information
#### SECTION 7: 510(K) SUMMARY
#### General Provisions:
| 510(k) Owner's Name: | OmniGuide®, Inc. |
|----------------------|-----------------------------------------------------------------------|
| Address: | 4 Maguire Road |
| | Lexington, |
| | MA 02421, USA |
| Contact Person | Samuel D Wade, Vice President of Quality Assurance and Regulatory |
| | Affairs |
| Phone Number: | (617) 551-8410 |
| | Cell: (978) 844-7515 |
| Fax Number: | (617) 551-8445 |
| Classification Name: | Powered Laser Surgical Instrument (21 CFR §878.4810 Laser surgical |
| | instrument for use in general and plastic surgery and in dermatology, |
| | Product Code GEX/ 21 CFR §878.4400 Electrosurgical cutting and |
| | coagulation device and accessories, Product code GEI) |
| Proprietary Name: | OmniGuide® TRIO™ handpiece (Proposed name) |
| Common Name: | Powered Laser Surgical Instrument |
#### Name of Predicate Devices:
- . OmniGuide® BeamPath® Fiber Optic handpiece (Product Code: GEX, 510(k) Number: K081939)
- . Karl Storz Endoscopy - America. Inc., Briner Bipolar Coagulation Suction Cannula (Product code: EOB, 510(k) Number: K962396/K960009)1
- Kirwan Surgical Products, Inc., Kirwan Bipolar Coagulation Forceps (Product code: GEI, 510(k) ● Number: K965241/K900533)2
## Indications for Use:
The OmniGuide® TRIOTM handpiece is indicated for the electrosurgical coagulation of soft body tissues and suction of fluids in general and ENT surgical procedures for the specialties of General Surgery, Head and Neck Surgery, Otolaryngology, and Oral surgery (including intraoral tissues). When used with the OmniGuide® BeamPath® Laser waveguide, it is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues for the added specialties of Dermatology, Gynecology, Neurosurgery, Orthopedic Surgery, Plastic & Reconstructive Surgery, Podiatry and Urology.
<sup>1</sup> Both 510(k) Numbers are included as we could not find the complete Summaries on the FDA releasable 510(k) database or FOI records
<sup>2</sup> Both 510(k) Numbers are included as we could not find the 510(k) Summary for K900533 on the FDA release or FOI records
{4}------------------------------------------------
#### Attachment 1 Premarket Submission: OmniGuide TRIO" handpiece
Confidential and Proprietary Information
## Device Description:
The TRIO™ handpiece is a handheld single use device designed to act as a laser waveguide guide/holder for previously cleared waveguides. The TRIO™ handpiece combines three fundamental surgical instrument functions into one lightweight and compact handheld disposable surgical instrument. If enables a surgeon to provide fluid suction, bipolar cautery, and CO2 laser energy without the necessity of changing instruments, saving surgical time for both physician and patient and a corresponding reduction in cost. Laser energy is delivered via a flexible OmniGuide® fiber for minimally invasive surgery.
The instrument is provided sterile and is intended for single use only. The package has an accelerated shelf life of two years.
The sterile package includes a cable to connect the handpiece to an electrosurgical generator. The cables are generic in nature and several varieties available at a hospital/ office will function. The package also includes a small brush for cleaning coagulation tips with adhered material. The TRIO™ handpiece is intended to be used only with the family of OmniGuide® CO2 Laser fibers including Velocity™, Elite™ and Select™ fibers previously cleared in 510(k) K070157.
The OmniGuide® handpiece is intended for use by licensed surgeons only for open surgical procedures listed in the indications for use. The objective of the handpiece is to enable precision control and to stabilize surgeon hand motion. The waveguide fiber is inserted through the proximal end of the handpiece and fixated so that it is observed at the handpiece's distal tip. The surgeon wields the handpiece as a pencil and advances the handpiece so that its distal tip is near the target tissue to exert the desired effect of the waveguide fiber: suction and coagulation. The surgeon may use the handpiece distal tip in either non-contact mode or contact mode and may use the distal tip to enable tissue manipulation, electrosurgical portion is used for coagulation due to bleeding during laser surgery or tissue manipulation.
The TRIO™ handpiece can be used with any Electro-Surgical Unit (ESU) with output power of up to 750 peak to peak voltage (Vp-p) and footswitch actuation. The instrument is provided with a 12-foot bipolar cable with fixed leads. The TRIO™ handpiece instrument can be used with any hospital or office suction apparatus. Fluid aspiration is controlled via a suction control port on the handle of the instrument.
Image /page/4/Figure/9 description: The image shows four boxes with text and arrows pointing to different parts of a device. The first box reads "CO2 laser energy delivered at the distal tip", and an arrow points to the lower left. The second box reads "Hole for suction control", and an arrow points down. The third box reads "Channel for OmniGuide® laser fiber", and an arrow points down. The fourth box reads "Proximal end of the Handpiece for fiber", and an arrow points down and to the right. The text "Section 7:510(k) Summary" and "7-2" are also visible.
{5}------------------------------------------------
#### Attachment 1 Premarket Submission: OmniGuide TRIO" handpiece
Confidential and Proprietary Information
Image /page/5/Picture/3 description: The image shows a medical device with a white handle and a blue tip. The device has a long, slender body that curves slightly at the end. The handle is wider and has several ports for connecting to other equipment. The device is labeled with the words "OmniGuide Surgical".
#### Figure 1: OmniGuide® TRIO™ Handpiece
#### Technological Characteristics:
The OmniGuide® TRIO™ handpiece has similar technological characteristics of the Briner Bipolar Coagulation Suction Cannula (K962396/K960009) and the Kirwan Bipolar Coagulation Forceps (K965241/K900533).
All use an electrosurgical generator to supply RF to the electrodes are similar in terms of dimensions and location at the distal end of the device. All three devices are designed to be handheld and use the same operating principles for electrocautery by bringing the electrodes into contact with the tissue.
BRINER and TRIO™ devices use wall suction through a 3/8" tube attached to the unit. Suction is performed by bringing the tip into contact with the fluid in the surgical field. Suction is controlled by closing or opening an aperture in the handle.
The TRIO™ is similar in function to the OmniGuide® BeamPath® fiber optic handpiece (K081939) in that both are used to facilitate delivery of laser energy through a previously cleared fiber optic waveguide (K140378 OmniGuide® Laser System with FlexGuide Ultra) allowing the surgeon to hold the waveguide and direct laser energy from the tip of the waveguide to the tissue. Both handpieces do not generate laser energy, they merely act as a support and guide the laser fiber to the surgical field.
The OmniGuide® TRIO™ handpiece system differs from the OmniGuide® BeamPath® fiber optic handpiece (K081939) in the use of construction materials. The TRIO™ is manufactured with materials suitable for disposable devices while the OmniGuide® BeamPath® fiber optic handpiece is constructed of stainless steel to allow for resterilization and multiple uses.
{6}------------------------------------------------
Confidential and Proprietary Information
The TRIO™ handpiece differs from the Karl Storz Briner device and the Kirwan Bipolar Coagulation Forceps in the use of different construction materials and slightly different design characteristics, such as length, outer diameter of the cannula and geometry of the handle.
## Electrical Safety and Electromagnetic Compatibility and other standard testing:
Evaluation of the device was conducted by an independent testing laboratory (Intertek) and was to be found in compliance with:
- . IEC 60601-1, Medical electrical equipment, Part 1 General requirements for basic safety and essential performance
- . IEC 60601-1-2:2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances --Requirements and tests
- IEC 62366 Medical devices- Application of usability engineering to medical devices
- IEC 14971:2007/(R) 2010 Medical Devices- Application of Risk Management to Medical . Devices
- IEC 60601-1 Medical Electrical Equipment, Part 1-6 General requirement for Safety: Collateral ● Standard Usability
# Performance Testing Summary:
Safety and effectiveness of TRIO™ handpiece was established to be the same as the predicate devices and other similar devices on the market by performance testing. The ability of the handpiece was tested by performing user validation and bench-top testing such as beam clipping, suction, fiber retention, reliability, mechanical robustness and thermal profile.
## Sterilization:
Sterilization Validation was performed which was successful and acceptable. It met the protocol and requirements specified in ISO 11135:2014.
## Biocompatibility:
The following list of tests were performed to establish biocompatibility:
- Cytotoxicity: MEM Elution Test Serum MEM Extract in conformance with ISO 10993-5:2009 . (Recognition Number 2-153), Biological evaluation of medical devices – Part 5: Tests for invitro cytotoxicity.
- ISO 10993-10:2010 Sensitization: Guinea Pig Maximization Test, Saline and Vegetable Oil Extracts ●
{7}------------------------------------------------
#### Attachment 1 Premarket Submission: OmniGuide TRIO" handpiece
Confidential and Proprietary Information
- ISO 10993-10:2002 (Recognition Number 2-87) Irritation: Intracutaneous Reactivity Test, Saline . and Vegetable Oil Extracts
- . ISO 10993-11:2006 (Recognition Number 2-118) Acute Systemic Toxicity: Systemic Injection Test, Saline and Vegetable Oil Extracts
The patient-contacting materials in the OmniGuide® TRIO™ handpiece passed these tests.
## Packaging and Labeling:
The OmniGuide® TRIO™ handpiece is a sterile, single use only, prescription device, in a thermoformed tray along with a suitable cable to allow attachment to the handpiece and electrosurgical generator and the tray is heat sealed to a Tyvek lid. The tray with the Instructions for Use are inserted into a cardboard box imprinted with product information and labeling. The Package has a proposed shelf life of two vears.
## Conclusion:
The OmniGuide® TRIO™ Handpiece is as safe and effective as its predicates. The new device has similar intended use and indications for use, technological characteristics and principle of operation as the predicate devices. The differences in the indications for use do not alter the intended surgical use of the device and do not affect the safety or efficacy when used as labeled. The minor design differences raise no new issues of safety or efficacy. Thus, the OmniGuide® TRIO™ Handpiece and its predicates are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.