K183145 · Illuminoss Medical, Inc. · QAD · Dec 13, 2018 · Orthopedic
Device Facts
Record ID
K183145
Device Name
IlluminOss Photodynamic Bone Stabilization System
Applicant
Illuminoss Medical, Inc.
Product Code
QAD · Orthopedic
Decision Date
Dec 13, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3023
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
IlluminOss Photodynamic Bone Stabilization System (PBSS) is indicated for use in skeletally mature patients in the treatment of traumatic, fragility, pathological, and impending pathological fractures of the humerus, radius, and ulna.
Device Story
Minimally invasive bone stabilization system; uses catheter to deploy inflatable, noncompliant PET balloon into medullary canal across fracture site. Physician infuses balloon with photodynamic (light-cured) monomer via syringe; balloon expands to fill intramedullary canal. Visible light delivered via radially emitting light pipe temporarily positioned in central lumen cures monomer in situ. System provides internal fixation/stabilization. Used in clinical settings by physicians. Output is a cured, rigid intramedullary implant providing structural support to fractured bone. Benefits include minimally invasive fracture fixation.
Clinical Evidence
Bench testing only. Testing demonstrated new sizes withstand infusion pressures, monomer achieves sufficient cure and mechanical strength, balloon catheter bonds maintain adequate tensile characteristics, and light pipes function without breakage.
Technological Characteristics
Intramedullary fixation rod; PET balloon; photodynamic monomer; visible light curing system; radially emitting light pipe. System includes catheter, sheath/dilator, and delivery instruments. Minimally invasive; non-implantable components are reusable. No specific ASTM standards cited.
Indications for Use
Indicated for skeletally mature patients with traumatic, fragility, pathological, or impending pathological fractures of the humerus, radius, or ulna.
Regulatory Classification
Identification
An in vivo cured intramedullary fixation rod is a prescription implanted device consisting of a balloon that is inserted into the medullary canal of long bones for the fixation of fractures. The balloon is infused with, and completely encapsulates, a liquid monomer that is exposed to a curing agent which polymerizes the monomer within the balloon creating a hardened rigid structure.
Special Controls
In combination with the general controls of the FD&C Act. the in vivo cured intramedullary fixation rod is subject to the following special controls:
- 1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- a. Mechanical testing must be conducted on the final device to assess burst, abrasion, bending, and torsion in static and dynamic conditions.
- b. Mechanical testing must demonstrate the integrity of the balloon including testing for leaks, ruptures, and release of cured/uncured material.
- Performance testing must demonstrate that the device can be inserted and C. removed.
- d. Performance testing must demonstrate the ability, in the event of a leak, to remove the uncured material from its in vivo location.
- Performance testing must demonstrate the reliability and accuracy of the curing e. method used.
- Thermal safety testing must be conducted to evaluate the temperature rise during f. curing.
- 2. Electrical safety, electromagnetic compatibility (EMC) testing, and electromagnetic interference (EMI) testing must be conducted for all electrical components.
- 3. All patient-contacting components must be demonstrated to be biocompatible.
- 4. Performance data must demonstrate the sterility and pyrogenicity of patient contacting components of the device that are provided sterile.
- 5. Performance data must validate the reprocessing instructions for any reusable components or instruments.
- 6. Performance data must support the shelf life of the system by demonstrating continued sterility, package integrity, and system functionality over the established shelf life.
- 7. Technological characterization of the device must include materials, curing agents, and a description of the operating principles of the device, including the delivery system and devices which initiate the curing process.
- 8. Labeling must include the following:
- a. A detailed summary of the device technical parameters.
- b. Information describing all materials of the device.
- Information describing how to perform the procedure and use the device, C. including the delivery system and devices which initiate the curing process, as well as how to remove the device and any uncured materials.
- d. A shelf life.
- Validated methods and instructions for reprocessing any reusable components or e. instruments.
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Mechanical testing must be conducted on the final device to assess burst, abrasion, bending, and torsion in static and dynamic conditions.
(ii) Mechanical testing must demonstrate the integrity of the balloon including testing for leaks, ruptures, and release of cured/uncured material.
(iii) Performance testing must demonstrate that the device can be inserted and removed.
(iv) Performance testing must demonstrate the ability, in the event of a leak, to remove the uncured material from its in vivo location.
(v) Performance testing must demonstrate the reliability and accuracy of the curing method used.
(vi) Thermal safety testing must be conducted to evaluate the temperature rise during curing.
(2) Electrical safety, electromagnetic compatibility (EMC) testing, and electromagnetic interference (EMI) testing must be conducted for all electrical components.
(3) All patient-contacting components must be demonstrated to be biocompatible.
(4) Performance data must demonstrate the sterility and pyrogenicity of patient contacting components of the device that are provided sterile.
(5) Performance data must validate the reprocessing instructions for any reusable components or instruments.
(6) Performance data must support the shelf life of the system by demonstrating continued sterility, package integrity, and system functionality over the established shelf life.
(7) Technological characterization of the device must include materials, curing agents, and a description of the operating principles of the device, including the delivery system and devices which initiate the curing process.
(8) Labeling must include the following:
(i) A detailed summary of the device technical parameters.
(ii) Information describing all materials of the device.
(iii) Information describing how to perform the procedure and use the device, including the delivery system and devices which initiate the curing process, as well as how to remove the device and any uncured materials.
(iv) A shelf life.
(v) Validated methods and instructions for reprocessing any reusable components or instruments.
Predicate Devices
IlluminOss Photodynamic Bone Stabilization System (K181228)
Submission Summary (Full Text)
{0}------------------------------------------------
December 13, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
IlluminOss Medical, Inc. % Hollace Rhodes Senior Director, Orthopedic Regulatory Affairs Musculoskeletal Clinical Regulatory Advisors, LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001
Re: K183145
Trade/Device Name: IlluminOss Photodynamic Bone Stabilization System Regulation Number: 21 CFR 888.3023 Regulation Name: In Vivo Cured Intramedullary Fixation Rod Regulatory Class: Class II Product Code: QAD Dated: November 12, 2018 Received: November 13, 2018
Dear Hollace Rhodes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183145
#### Device Name
IlluminOss Photodynamic Bone Stabilization System ("IlluminOss PBSS")
#### Indications for Use (Describe)
IlluminOss Photodynamic Bone Stabilization System (PBSS) is indicated for use in skeletally mature patients in the treatment of traumatic, fragility, pathological, and impending pathological fractures of the humerus, radius, and ulna.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
# 510(k) Summary
| Manufacturer: | IlluminOss Medical, Inc.<br>993 Waterman Avenue<br>East Providence, RI 02914<br>Phone: 401.714.0008 |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mr. Robert Rabiner<br>Chief Technical Officer<br>IlluminOss Medical, Inc.<br>993 Waterman Avenue<br>East Providence, RI 02914<br>Phone: 401.714.0008 x207<br>Fax: 401.714.0009<br>rrabiner@illuminoss.com |
| Prepared By: | Musculoskeletal Clinical Regulatory Advisers, LLC<br>1050 K Street, NW, Suite 1000<br>Washington, DC 20001<br>Phone: 202.552.5800<br>Fax: 202.552.5798 |
| Date Prepared: | November 12, 2018 |
| Device Trade Name: | IlluminOss Photodynamic Bone Stabilization System<br>("IlluminOss PBSS") |
| Classifications: | 21 CFR 888.3023, In vivo cured intramedullary fixation<br>rod |
| Class: | Class II |
| Product Code: | QAD |
#### Predicate Device:
The modified IlluminOss Photodynamic Bone Stabilization System is substantially equivalent to the predicate IlluminOss Photodynamic Bone Stabilization System (K181228) with respect to indications, design, materials, function, and performance. The information summarized in the Design Control Activities Summary demonstrates that the modified IlluminOss Photodynamic Bone Stabilization System met the pre-determined acceptance criteria for the verification activities.
{4}------------------------------------------------
## Indications for Use:
IlluminOss Photodynamic Bone Stabilization System (PBSS) is indicated for use in skeletally mature patients in the treatment of traumatic, fragility, pathological, and impending pathological fractures of the humerus, radius, and ulna.
The Indications for Use statement for the IlluminOss Photodynamic Bone Stabilization System device is identical to the predicate device.
### Device Description:
The IlluminOss Photodynamic Bone Stabilization System provides an important treatment option in the fixation and stabilization of fractures through a minimally invasive procedure. The system uses a catheter to deploy an inflatable, noncompliant, thin wall PET balloon into the medullary canal of the bone across the fracture site. The balloon is infused using a standard 20cc syringe with a photodynamic (light cured) monomer that causes the balloon to slowly expand and fill the intramedullary canal of the fractured bone. Activation of the light system allows for visible spectrum light to be delivered through a radially emitting light pipe that is temporarily positioned into a central lumen of the catheter that runs the length of the balloon. With this design, the liquid monomer within the balloon is exposed to light along the entire length of the balloon during the curing process. The purpose of this Special 510(k) is to add smaller, intermediate, and larger sizes of the implant, an intermediate size of the Sheath/Dilator Set, and a smaller size of the Delivery Set.
### Comparison of Technological Characteristics with Predicate Device:
The subject device and the IlluminOss PBSS predicate device are identical, with the exception of the addition of new small, intermediate, and large sizes of the implant and the auxiliary components to support these new sizes, specifically a new intermediate Sheath/Dilator Set and smaller reusable instruments and light pipe. These additional parts are not anticipated to introduce any new issues regarding device safety or effectiveness.
#### Performance Testing Summary:
Testing of the IlluminOss PBSS device includes:
- 1. Testing to demonstrate that the new device sizes could withstand the pressures necessary to complete the infusion process with the liquid monomer.
- 2. Testing to demonstrate that monomer sufficiently cures during the given cure time for the new implant sizes, providing sufficient mechanical strength
- 3. Mechanical testing to demonstrate the strength of the balloon catheter bond possess adequate tensile characteristics.
- 4. Mechanical testing to demonstrate the light pipe does not break upon removal and can sufficiently cure the monomer.
#### Substantial Equivalence:
The subject device was demonstrated to be substantially equivalent to the predicate cited above with respect to indications, design, function, and performance.
{5}------------------------------------------------
## Conclusion:
The IlluminOss PBSS device is substantially equivalent to the previously cleared device with respect to its indications for use, design, function, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.