21 CFR 888.3023 — Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated
Orthopedic (OR) · Part 888 Subpart D—Prosthetic Devices · § 888.3023
Identification
An in vivo cured intramedullary fixation rod is a prescription implanted device consisting of a balloon that is inserted into the medullary canal of long bones for the fixation of fractures. The balloon is infused with, and completely encapsulates, a liquid monomer that is exposed to a curing agent which polymerizes the monomer within the balloon creating a hardened rigid structure.
Classification Rationale
Class II (special controls). The special controls for this device are:
Product Codes
| Product Code | Device Name | Class | Devices | Attributes |
|---|---|---|---|---|
| QAD | Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated | 2 | 7 | Implant |
Special Controls
QAD — Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated
In combination with the general controls of the FD&C Act. the in vivo cured intramedullary fixation rod is subject to the following special controls: - 1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: - a. Mechanical testing must be conducted on the final device to assess burst, abrasion, bending, and torsion in static and dynamic conditions. - b. Mechanical testing must demonstrate the integrity of the balloon including testing for leaks, ruptures, and release of cured/uncured material. - Performance testing must demonstrate that the device can be inserted and C. removed. - d. Performance testing must demonstrate the ability, in the event of a leak, to remove the uncured material from its in vivo location. - Performance testing must demonstrate the reliability and accuracy of the curing e. method used. - Thermal safety testing must be conducted to evaluate the temperature rise during f. curing. - 2. Electrical safety, electromagnetic compatibility (EMC) testing, and electromagnetic interference (EMI) testing must be conducted for all electrical components. - 3. All patient-contacting components must be demonstrated to be biocompatible. - 4. Performance data must demonstrate the sterility and pyrogenicity of patient contacting components of the device that are provided sterile. - 5. Performance data must validate the reprocessing instructions for any reusable components or instruments. - 6. Performance data must support the shelf life of the system by demonstrating continued sterility, package integrity, and system functionality over the established shelf life. - 7. Technological characterization of the device must include materials, curing agents, and a description of the operating principles of the device, including the delivery system and devices which initiate the curing process. - 8. Labeling must include the following: - a. A detailed summary of the device technical parameters. - b. Information describing all materials of the device. - Information describing how to perform the procedure and use the device, C. including the delivery system and devices which initiate the curing process, as well as how to remove the device and any uncured materials. - d. A shelf life. - Validated methods and instructions for reprocessing any reusable components or e. instruments.
De Novo Order DEN160062
QAD — Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Mechanical testing must be conducted on the final device to assess burst, abrasion, bending, and torsion in static and dynamic conditions. (ii) Mechanical testing must demonstrate the integrity of the balloon including testing for leaks, ruptures, and release of cured/uncured material. (iii) Performance testing must demonstrate that the device can be inserted and removed. (iv) Performance testing must demonstrate the ability, in the event of a leak, to remove the uncured material from its in vivo location. (v) Performance testing must demonstrate the reliability and accuracy of the curing method used. (vi) Thermal safety testing must be conducted to evaluate the temperature rise during curing. (2) Electrical safety, electromagnetic compatibility (EMC) testing, and electromagnetic interference (EMI) testing must be conducted for all electrical components. (3) All patient-contacting components must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility and pyrogenicity of patient contacting components of the device that are provided sterile. (5) Performance data must validate the reprocessing instructions for any reusable components or instruments. (6) Performance data must support the shelf life of the system by demonstrating continued sterility, package integrity, and system functionality over the established shelf life. (7) Technological characterization of the device must include materials, curing agents, and a description of the operating principles of the device, including the delivery system and devices which initiate the curing process. (8) Labeling must include the following: (i) A detailed summary of the device technical parameters. (ii) Information describing all materials of the device. (iii) Information describing how to perform the procedure and use the device, including the delivery system and devices which initiate the curing process, as well as how to remove the device and any uncured materials. (iv) A shelf life. (v) Validated methods and instructions for reprocessing any reusable components or instruments.
eCFR
QAD — Rod, Fixation, Intramedullary And Accessories, In-Vivo Cured, Light-Activated
(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Mechanical testing must be conducted on the final device to assess burst, abrasion, bending, and torsion in static and dynamic conditions. (ii) Mechanical testing must demonstrate the integrity of the balloon including testing for leaks, ruptures, and release of cured/uncured material. (iii) Performance testing must demonstrate that the device can be inserted and removed. (iv) Performance testing must demonstrate the ability, in the event of a leak, to remove the uncured material from its in vivo location. (v) Performance testing must demonstrate the reliability and accuracy of the curing method used. (vi) Thermal safety testing must be conducted to evaluate the temperature rise during curing. (2) Electrical safety, electromagnetic compatibility (EMC) testing, and electromagnetic interference (EMI) testing must be conducted for all electrical components. (3) All patient-contacting components must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility and pyrogenicity of patient contacting components of the device that are provided sterile. (5) Performance data must validate the reprocessing instructions for any reusable components or instruments. (6) Performance data must support the shelf life of the system by demonstrating continued sterility, package integrity, and system functionality over the established shelf life. (7) Technological characterization of the device must include materials, curing agents, and a description of the operating principles of the device, including the delivery system and devices which initiate the curing process. (8) Labeling must include the following: (i) A detailed summary of the device technical parameters. (ii) Information describing all materials of the device. (iii) Information describing how to perform the procedure and use the device, including the delivery system and devices which initiate the curing process, as well as how to remove the device and any uncured materials. (iv) A shelf life. (v) Validated methods and instructions for reprocessing any reusable components or instruments.
Ecfr Llm