K182660 · Galt Medical Corp. · DRB · May 30, 2019 · Cardiovascular
Device Facts
Record ID
K182660
Device Name
GaltTWS
Applicant
Galt Medical Corp.
Product Code
DRB · Cardiovascular
Decision Date
May 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1380
Device Class
Class 2
Indications for Use
The GaltTWS Stylet is a percutaneous catheter stylet placed inside a catheter, lead, or cannula to stiffen the device for placement or extraction.
Device Story
GaltTWS Stylet consists of stainless steel twisted/braided wire; diameters 0.012-0.015 inch; lengths 30-72cm. Device functions as internal stiffener for catheters, leads, or cannulas during percutaneous placement or extraction. Configured as wire bonded to hub, flushable hub, or connected to flushable adjustable hub via Galt Gateway Hemostasis Valve. Used in clinical settings by healthcare providers. Stiffening action aids device navigation through vasculature; facilitates procedural access. No AI/ML or software components.
Clinical Evidence
Bench testing only. No clinical data. Testing included particulate, mechanical properties, leak, and tortuous path (in vitro) tests. Accelerated aging (4-year equivalent) performed. Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, and material-mediated pyrogen testing. EO residuals testing performed.
Technological Characteristics
Stainless steel twisted/braided wire. Dimensions: 0.012-0.015 inch diameter, 30-72cm length. Hub configurations include standard or flushable options. Sterilization via ethylene oxide (EO) cycle. No software or electronic components.
Indications for Use
Indicated for use as a percutaneous catheter stylet to stiffen catheters, leads, or cannulas to facilitate their placement or extraction in patients.
Regulatory Classification
Identification
A catheter stylet is a wire that is run through a catheter or cannula to render it stiff.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 30, 2019 Galt Medical Corp. David Derrick Director of Quality and Regulatory Affairs 2220 Merritt Drive Garland, Texas 75087
Re: K182660
Trade/Device Name: GaltTWS Regulation Number: 21 CFR 870.1380 Regulation Name: Catheter Stylet Regulatory Class: Class II Product Code: DRB Dated: April 29, 2019 Received: April 30, 2019
## Dear David Derrick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Matthew Hillebrenner Acting Division Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182660
Device Name GaltTWS Stylet
Indications for Use (Describe)
The GaltTWS Stylet is a percutaneous catheter stylet placed inside a catheter, lead, or cannula to stiffen the device for placement or extraction.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Application Date: | November 13, 2018 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Application Type: | Traditional 510(k) |
| Applicant Information | Galt Medical Corporation<br>2220 Merritt Dr.<br>Garland, TX 75041<br>Phone: 214-778-1306<br>Fax: 972-271-4706 |
| Official Contact: | David Derrick<br>Director of Quality and Regulatory Affairs<br>Galt Medical Corporation<br>2220 Merritt Dr.<br>Garland, TX 75041<br>Phone: 214-778-1306<br>Fax: 972-271-4706<br>dderrick@galtmedical.com |
| Device Name: | GaltTWS Stylet |
| Device Model Number: | TBD |
| Classification Name: | Catheter Stylet (DRB),<br>21 CFR 870.1380 |
| Device Classification: | Class II (Cardiovascular) |
| Predicate Device: | TFX Medical Coated Stylet (K992664)<br>Guidewire (K982559)<br>Gateway Hemostasis Valve (152528) |
| Manufacturer: | Galt Medical<br>2220 Merritt Drive<br>Garland, TX 75041<br>Phone: 214-778-5177 |
Establishment Registration Number: 1649395
Intended Use: The GaltTWS Stylet is a percutaneous catheter stylet placed inside a catheter, lead, or cannula to stiffen the device for placement or extraction.
Device Description: The GaltTWS stylet is comprised of a stainless steel twisted/braided wire of varying diameters (0.012-0.015 inch), and varying length (30-72cm). The GaltTWS stylet can be configured as a Twisted/Braided wire bonded to a hub, flushable
{5}------------------------------------------------
hub, or connected to a flushable adjustable hub using the Galt Gateway Hemostasis Valve (K152528).
Comparison of Technological Characteristics: The subject device GaltTWS Stylet and the predicate device TFX Medical Coated Catheter Stylet have same indication and are of similar design. The subject and predicate device are available in similar configurations.
| Mfr. / Product | Subject Device | Predicate Device | Predicate Device | Predicate Device |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Galt Medical /<br>GaltTWS Stylet | TFX Medical / TFX<br>Medical Coated<br>Catheter Stylet | Galt Medical /<br>Guidewire | Galt Medical /<br>Gateway Hemostasis<br>Valve |
| 510(k)<br>Number | N/A | K992664 | K982559 | K152528 |
| Device<br>Classification | 870.1380 | 870.1380 | 870.1340 | 870.4290 |
| Product Code | DRB | DRB | DQX | DTL |
| Intended use | The Galt Medical<br>Corp. GaltTWS Stylet<br>is a percutaneous<br>catheter stylet placed<br>inside a catheter,<br>lead, or cannula to<br>stiffen the device for<br>placement or<br>extraction. | The TFX Medical stylet<br>is intended for use in<br>conjunction with<br>catheters or other<br>devices, which require<br>the aid of a stylet to<br>render it stiff for<br>placement. | The guidewire is<br>intended for use in<br>percutaneous<br>procedures to<br>introduce and<br>position catheter and<br>other interventional<br>devices within the<br>coronary and<br>peripheral<br>vasculature. | The attachable Cath<br>Lab Hemostasis Valve<br>is indicated to<br>minimize blood loss<br>during introduction<br>of catheters,<br>guidewires and other<br>intravascular devices<br>into the vasculature. |
| Design | A stainless steel<br>twisted/braided wire<br>bonded to a hub,<br>flushable hub in a range<br>of sizes and lengths. | A twisted/braided wire<br>bonded to a hub in a<br>range of sizes and<br>lengths. | Guidewire with a tightly<br>wound coil over a<br>tapered core and a<br>ribbon. The ribbon is<br>welded to the coil at<br>both ends. The core is<br>welded to the coil at<br>the proximal end. | Introducer with valved<br>hub and luer lock<br>connector |
| Sizes | Diameter - (0.012-0.015<br>inch)<br>Length (30-72cm) | N/A | Diameter - (0.014-0.065<br>inch)<br>Length (15-400cm) | Maximum 8FR |
{6}------------------------------------------------
Substantial Equivalence and Summary of Bench Testing: The technological differences between the subject devices and predicate devices have been evaluated through bench and biocompatibility tests to provide evidence of substantial equivalence. The GaltTWS Stylet is substantially equivalent to the predicate devices based on comparisons of the devices functionality, compatibility, technological characteristics, and indications for use.
Testing was conducted according to protocols based on international standards and Galt Medical requirements. Functional Testing included the following:
- 0 Particulate Test
- o Mechanical Properties Test
- 0 Leak Test
- 0 Tortuous Path Test (In Vitro)
Additionally, accelerated aging was performed on the subject devices to an equivalent age of 4 years, and the physical and functional testing was repeated.
Biocompatibility testing was performed on the subject device in accordance with ISO 10993-1. This was done in order to update the biocompatibility testing to current standards, and was not done as a result of any changes to the product. Biocompatibility testing included the following:
- Cytotoxicity (MEM Elution) 0
- Sensitization (Magnusson Kligman) 0
- Irritation (Intracutaneous injection) O
- 0 Systemic Toxicity (Systemic injection)
- 0 Hemolysis (In Direct Contact)
- 0 Systemic Toxicity (Materials Mediated Pyrogen)
- o EO Residuals (EO, ECH, & EG)
Additionally the subject device was adopted into the existing ethylene oxide sterilization cycle for the Galt product cleared under K982559 & K152528.
Packaging of the subject device will remain unchanged from the Galt Medical Guidewire and Gateway Hemostasis Valve. Current packaging shelf life testing was provided in the predicate submissions.
Conclusion: The GaltTWS Stylet as designed and manufactured is determined to be substantially equivalent to the current marketed predicate devices.
End of Section
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.