K152528 · Galt Medical Corp. · DTL · Dec 2, 2015 · Cardiovascular
Device Facts
Record ID
K152528
Device Name
Attachable Cath Lab Hemostasis Valve
Applicant
Galt Medical Corp.
Product Code
DTL · Cardiovascular
Decision Date
Dec 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4290
Device Class
Class 2
Indications for Use
The Attachable Cath Lab Hemostasis Valve is indicated to minimize blood loss during the introduction of catheters, guidewires and other intravascular devices into the vasculature.
Device Story
Device is a single-patient-use, disposable hemostasis valve assembly; consists of a molded hub with integrated valve, sideport tubing, and 3-way stopcock. Used in clinical settings (cath lab) by physicians or trained clinical staff during intravascular procedures. Device attaches to existing introducer systems to provide a seal around inserted catheters or guidewires. Function is to minimize blood loss during device exchange or insertion. Output is a physical seal; healthcare providers observe blood loss reduction directly. Benefits include maintenance of hemostasis during complex vascular interventions.
Clinical Evidence
Bench testing only. No clinical data provided. Functional testing included pressure leak, vacuum leak, guidewire insertion, dilator insertion, sidearm tubing to hub pull force, and spinlock collar adapter pull tests. Biocompatibility and shelf-life testing were leveraged from predicate device submissions (K043525 and K140028) based on identical raw materials and packaging.
Technological Characteristics
Molded hub with integrated hemostasis valve, sideport tubing, and 3-way stopcock. Accepts up to 8F devices. Pressure withstand: 7 PSI; Vacuum withstand: 0.47 in-Hg. Sterilization: Ethylene oxide (EO) cycle adopted from K043525. Materials identical to predicate devices.
Indications for Use
Indicated for patients undergoing percutaneous procedures requiring the introduction of catheters, guidewires, and other intravascular devices into the vasculature, where minimization of blood loss is required.
Regulatory Classification
Identification
A cardiopulmonary bypass adaptor, stopcock, manifold, or fitting is a device used in cardiovascular diagnostic, surgical, and therapeutic applications to interconnect tubing, catheters, or other devices.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Galt Medical VTI Valved Tearaway Introducer (K140028)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, resembling a bird-like shape.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 02, 2015
Galt Medical Corp. David Derrick Director Of Quality And Regulatory Affairs 2220 Merritt Dr. Garland, Texas 75041
Re: K152528
Trade/Device Name: Attachable Cath Lab Hemostasis Valve Regulation Number: 21 CFR 870.4290 Regulation Name: Cardiopulmonary Bypass Adaptor, Stopcock, Manifold, Or Fitting Regulatory Class: Class II Product Code: DTL Dated: September 1, 2015 Received: September 3, 2015
Dear David Derrick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{2}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrenner
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE
510(K) number (if known): K152528
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The Attachable Cath Lab Hemostasis Valve is indicated to minimize blood loss during the introduction of catheters, guidewires and other intravascular devices into the vasculature.
Prescription Use_ X (Per 21 CFR 801 Subpart D) OR
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of _ 1
{4}------------------------------------------------
| Application Date: | September 01, 2015 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Application Type: | Traditional 510(k) |
| Applicant Information | Galt Medical Corporation<br>2220 Merritt Dr.<br>Garland, TX 75041<br>Phone: 214-778-1306<br>Fax: 972-271-4706 |
| Official Contact: | David Derrick<br>Director of Quality and Regulatory Affairs<br>Galt Medical Corporation<br>2220 Merritt Dr.<br>Garland, TX 75041<br>Phone: 214-778-1306<br>Fax: 972-271-4706<br>dderrick@galtmedical.com |
| Device Name: | Attachable Cath Lab Hemostasis Valve |
| Device Model Number: | TBD |
| Classification Name: | Cardiopulmonary bypass adaptor, Stopcock, Manifold, or<br>Fitting (DTL),<br>21 CFR 870.4290 |
| Device Classification: | Class II (Cardiovascular) |
| Predicate Device: | Merit Medical IN-LINE Hemostasis Valve (K990975)<br>Galt Medical Cath Lab Introducer Sheath (K043525)<br>Galt VTI® Valved Tearaway Introducer Sheath (K112398) |
| Manufacturer: | Galt Medical<br>2220 Merritt Drive<br>Garland, TX 75041<br>Phone: 214-778-5177<br>Fax: 972-271-4706 |
{5}------------------------------------------------
Intended Use: The Attachable Cath Lab Hemostasis Valve is indicated to minimize blood loss during the introduction of catheters, guidewires and other intravascular devices into the vasculature.
Device Description: The Galt Medical Attachable Cath Lab Hemostasis Valve assembly consists of a molded hub with an integrated hemostasis valve, sideport tubing with 3 way stopcock. The design of the Attachable Cath Lab Hemostasis Valve is derived directly from the current line of Galt Medical Cath Lab Introducer Sheath (K043525). The subject and predicate device are both have an integrated hemostasis valve, and side port tubing with stopcock.
The Galt Cath Lab Introducer (K043525) has been previously determined to be substantially equivalent. In this submission, Galt Medical Corp will establish that the subject device Attachable Cath Lab Hemostasis is substantially equivalent to the current marketed devices; Merit Medical IN-LINE Hemostasis Valve, Galt Cath Lab Introducer Sheath and Galt VTI Valved Tearaway Introducer.
Comparison of Technological Characteristics: The Attachable Cath Lab Hemostasis Valve is substantially equivalent to the unmodified predicate in construction, materials, and device performance.
| | Subject device | Predicate Device | Predicate Device | Predicate Device |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mfr. /<br>Product | Galt Medical<br>Attachable Cath<br>Lab Hemostasis<br>Valve | Merit Medical IN-<br>LINE Hemostasis<br>(K990975) | Galt Medical Cath<br>Lab Introducer<br>(K043525) | Galt Medical VTI<br>Valved Tearaway<br>(K140028 ) |
| 510(k)<br>Number | N/A | K112398 | K043525 | K140028 |
| Device<br>Classification | 870.4290 | 870.4290 | 870.1340 | 870.1340 |
| Product<br>Code | DTL | DTL | DYB | DYB |
| Intended use | Intended to<br>minimize blood<br>loss during the<br>introduction of<br>catheters,<br>guidewires and<br>other intravascular<br>devices into the<br>vasculature | Intended to<br>minimize blood<br>loss during the<br>introduction of<br>catheters,<br>guidewires and<br>other intravascular<br>devices into the<br>vasculature | The seath<br>introducer system<br>is indicated for use<br>in percutaneous<br>procedures to<br>introduce<br>catheters and<br>other intravascular<br>devices into the<br>vasculature. | These introducers<br>are used for the<br>percutaneous<br>introduction of<br>diagnostic or<br>therapeutic<br>devices, such as<br>catheters and<br>pacing leads into<br>the vasculature. |
{6}------------------------------------------------
| | Subject device | Predicate Device | Predicate Device | Predicate Device |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | The Galt Attachable<br>Cath Lab Hemostasis<br>Valve assembly<br>consists of a molded<br>hub with an<br>integrated<br>hemostasis valve,<br>sideport tubing with<br>3 way stopcock. | The Merit Medical<br>IN-LINE Hemostasis<br>Valve assembly<br>consists of a molded<br>hub with an<br>integrated<br>hemostasis valve,<br>sideport tubing with<br>3 way stopcock. | Cath Lab introducer<br>has a molded<br>hub/sheath with<br>integrated<br>hemostasis valve,<br>and side port tubing<br>with stopcock. The<br>device includes a<br>locking dilator to<br>assist in the insertion<br>of the device into<br>the vascular system. | The VTI valved<br>tearaway introducer<br>has a molded<br>hub/sheath with<br>integrated<br>hemostasis valve,<br>and side port tubing<br>with stopcock. The<br>device includes a<br>locking dilator to<br>assist in the insertion<br>of the device into<br>the vascular system |
| Color | White or Green<br>Colored hub and<br>cap with clear<br>sideport tubing<br>and stopcock | White Colored hub<br>with clear sideport<br>tubing and<br>stopcock | White Colored<br>hub, color coded<br>cap with clear<br>sideport tubing<br>and stopcock | Color coded hub,<br>clear cap with<br>clear sideport<br>tubing and<br>stopcock |
| Shape | Round hub with<br>spinlock collar | Round hub with<br>spinlock collar | Cylindrical cannula<br>with round hub | Cylindrical cannula<br>with winged hub |
| Sizes | Accepts up to 8F<br>device | Accepts up to 8F<br>device | 6cm, to120cm<br>lengths, sizes 4F -<br>9F | 13cm, and 23cm<br>lengths, sizes 6F -<br>16F |
| Pressure<br>withstand | 7 PSI | 7 PSI | 7 PSI | 6.1 PSI |
| Vacuum<br>Withstand | 0.47in-Hg | 0.47in-Hg | 0.47in-Hg | 0.47in-Hg |
| Dilator Lock | N/A | N/A | Dilator is retained | Dilator is retained |
Use Type: The Galt Attachable Cath Lab Hemostasis Valve is a single patient use, disposable device.
Summary of Non-Clinical Data Submitted: Functional testing on 4 year aged product was conducted to verify that the Attachable Cath Lab Hemostasis Valve met product specifications. Testing was conducted according to protocols based on international standards and Galt Medical requirements. Functional Testing included the following:
- . Pressure Leak Test
- Vacuum Leak Test .
- . Guidewire Insertion Test
- . Dilator Insertion Test
- Sidearm Tubing to Hub Pull Force Test
- . Spinlock Collar Adapter Pull Test
Additionally the Attachable Cath Lab Hemostasis Valve was adopted into the existing ethylene oxide sterilization cycle for the Galt Cath Lab Introducer Sheath cleared under K043525.
{7}------------------------------------------------
Biocompatibility testing was provided in Galt Cath Lab Introducer Sheath submission K043525 and for the optional valve - Galt VTI Valved Tearaway Introducer submission K140028. The devices tested included identical raw materials and package configuration and materials utilized in the Galt Attachable Cath Lab Hemostasis Valve and Galt VTI Valved Tearaway Introducer.
Packaging shelf life testing was provided in the predicate submissions.
Conclusion: It will be shown in this 510(k) submission that the differences between the Galt Attachable Cath Lab Hemostasis Valve and the predicate device do not raise any questions regarding safety and effectiveness. The Galt Attachable Cath Lab Hemostasis Valve as designed and manufactured is determined to be substantially equivalent to the referenced predicate device.
End of Section
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.