K182565 · AtriCure, Inc. · GXH · Nov 9, 2018 · Neurology
Device Facts
Record ID
K182565
Device Name
AtriCure cryoICE cryoSPHERE cryoablation probe
Applicant
AtriCure, Inc.
Product Code
GXH · Neurology
Decision Date
Nov 9, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use blocking pain by temporarily ablating peripheral nerves.
Device Story
Single-use, sterile, hand-held cryoablation probe; used by surgeons to block pain via temporary peripheral nerve ablation. Device connects to AtriCure Cryo Module (ACM) to circulate nitrous oxide; utilizes Joule-Thomson effect to freeze target tissue via 8mm spherical ball tip. Malleable shaft allows manual bending for anatomical access. Surgeon inserts probe through incision, places tip on target nerve, and activates cooling via console or footswitch. Temporary freezing creates cryonecrosis to interrupt pain signaling. Benefits include targeted, reversible nerve block for pain management. Used in clinical settings.
Clinical Evidence
Bench testing only. Testing included reliability, transit, shelf-life, cryogen performance/thermal insulation, mechanical testing, and biocompatibility. All tests met predetermined acceptance criteria, demonstrating equivalence to the predicate device.
Indicated for blocking pain by temporarily ablating peripheral nerves in patients requiring cryosurgical intervention.
Regulatory Classification
Identification
A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
November 9, 2018
AtriCure, Inc. Jonathan McElwee Regulatory Affairs Manager 7555 Innovation Way Mason, Ohio 45040
Re: K182565
Trade/Device Name: AtriCure cryoICE cryoSPHERE cryoablation probe Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: Class II Product Code: GXH Dated: September 17, 2018 Received: September 18, 2018
Dear Jonathan McElwee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Matthew C. Krueger -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182565
Device Name cryoICE® cryoSPHERE™ cryoablation probe
Indications for Use (Describe)
AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use in blocking pain by temporarily ablating peripheral nerves.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| <input type="checkbox"/> | |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "AtriCure" in a stylized font. The first three letters, "Atri", are in blue, while the rest of the word, "Cure", is in orange. The "C" in "Cure" has a blue dot in the middle. There is a registered trademark symbol to the right of the word.
## 510(k) Summary
#### l. Applicant Information
| Manufacturer: | AtriCure®, Inc.<br>7555 Innovation Way<br>Mason, Ohio 45040<br>P: 513-644-4736<br>F: 513-895-9013 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jonathan McElwee, RAC<br>Manager, Regulatory Affairs |
| Date Prepared: | 9/17/2018 |
| II.<br>Device Information | |
| Proprietary Name: | cryoICE™ cryoSPHERE® cryo-ablation probe (CRYOS) |
| Common Name: | Cryosurgical Probe |
| Classification: | Cryogenic surgical device<br>Regulatory Class: Class II; per 21 CFR 882.4250<br>Product Code: GXH<br>Classification Panel: Neurodiagnostic and Neurosurgical Devices |
| Predicate Device: | cryoICE cryoablation probe (CRYO2)<br>(K180138, February 15, 2018) |
| Reference Device: | cryoICE cryoablation probe (CRYO2)<br>(K142203, November 25, 2014) |
#### III. Device Description
The cryolCE cryoSPHERE cryoablation (CRYOS) probe is a single use device that achieves cryoablation of peripheral nerves by allowing a surgeon to insert the probe through an incision to reach the target tissue, place the probe tip on the target site, and in conjunction with an AtriCure Cryo Module (ACM), temporarily freeze the tissue in contact with the probe tip by circulating a cryogenic agent, nitrous oxide, through the device. The CRYOS probe is offered in two probe length configurations: approximately 11" and 18" long. The probe is malleable, and is capable of being bent by the end user. At the distal end, the CRYOS device features an 8 mm diameter ball tip shaped probe.
#### IV. Intended Use/ Indications for Use
AtriCure's cryolCE cryoSPHERE cryoablation probes are sterile, single use devices intended for use blocking pain by temporarily ablating peripheral nerves.
{4}------------------------------------------------
# AtriCure®
#### Comparison of Technological Characteristics (CRYO2 K180138) V.
- The devices include the same intended use, and; ●
- No changes were made in operating principle, or specifications of performance, and; .
- The results of the verification and validation testing: .
- o Demonstrated equivalency in performance
- O Did not raise any new issues of safety
Modifications included in the cryoSPHERE probe end-effector were for differing body habitus and surgeon preference for the use of the device specifically intended to blocking pain.
| | CRYO2 (Predicate) | CRYOS (Subject) |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Market Product<br>Name | cryoICE cryoablation probe | cryoICE cryoSPHERE cryoablation probe |
| 510(k) Number | K180138 | Subject of this Submission |
| Intended Use | AtriCure's cryoICE cryo-ablation probes are<br>sterile, single use devices intended for use in the<br>cryosurgical treatment of cardiac arrhythmias by<br>freezing target tissues, creating an inflammatory<br>response (cryonecrosis) that blocks the electrical<br>conduction pathway. The probe is also intended<br>for use in blocking pain by temporarily ablating<br>peripheral nerves. | AtriCure's cryolCE cryoSPHERE cryoablation<br>probes are sterile, single use devices intended<br>for use blocking pain by temporarily ablating<br>peripheral nerves. |
| Operating<br>Principle | Joule-Thompson Effect | Joule-Thompson Effect |
| Technology | The system consists of cryoprobes that are used<br>for freezing target tissue. A console is used to<br>control the supply of gas to the cryoprobe. | The system consists of cryoprobes that are used<br>for freezing target tissue. A console is used to<br>control the supply of gas to the cryoprobe. |
| Energy Used | Nitrous Oxide | Nitrous Oxide |
| Operating<br>Temperature | -50°C to 70°C | -50°C to 70°C |
| Human Factors | Hand-held device containing cryogen with<br>activation button on console or footswitch. | Hand-held device containing cryogen with<br>activation button on console or footswitch. |
| Cryotip | Material: aluminum alloy | Material: aluminum alloy |
| | Construction: smooth malleable cylindrical linear<br>probe | Construction: smooth spherical ball welded to a<br>malleable cylindrical aluminum shaft. |
| | Diameter: 4 mm | Diameter: ball - 8mm; aluminum shaft - 5mm |
| | Length: 10cm | Length: reference the "end-effector" length |
| End-effector<br>Length – Shaft<br>and Cryotip | 28 cm | 28cm (CRYOS)<br>46cm (CRYOS-L) |
| Biocompatibility | Biocompatible patient contacting materials. | Biocompatible patient contacting materials. |
| Probe Packaging | Sterile - Single Use disposable device | Sterile - Single Use disposable device |
| Sterilization | Gamma Irradiation | Gamma Irradiation |
| Power Source | Mains Powered | Mains Powered |
{5}------------------------------------------------
#### VI. Performance Data
The following bench testing was conducted for design and performance elements deemed appropriate to demonstrate equivalence to the previously cleared CRYO2 device met the predetermined acceptance criteria ensuring substantial equivalence to the previously cleared CRYO2 device. No new safety or performance issues were raising during testing.
## Non-clinical Bench Testing:
- Reliability Testing ●
- Transit ●
- Shelf-life
- Cryogen Performance/Thermal Insulation ●
- Mechanical testing .
- Biocompatibility .
#### VII. Conclusions
AtriCure has demonstrated that the cryolCE cryoSPHERE cryoablation probe is substantially equivalent in fundamental design, technology, function, device materials, packaging, sterilization, operating principal, and intended use/ indication for use as the previously cleared device: cryolCE cryoablation probe.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.