AtriCure Cryo Module Indication for Use Statement: The AtriCure Cryo Module is a nonsterile, reusable device which delivers cryogenic energy, namely nitrous oxide, to AtriCure's cryo-ablation probes. cryolCE™ Indication for Use Statement: AtriCure's cryoICE™ cryo-ablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway. The probe is also intended for use in blocking pain by temporarily ablating peripheral nerves.
Device Story
System comprises AtriCure Cryo Module (ACM) console and sterile, single-use cryoICE (CRYO2) probes; delivers nitrous oxide to probe tip; utilizes Joule-Thomson effect for cryo-ablation; freezes target tissue to create cryonecrosis; blocks electrical conduction pathways in cardiac tissue or temporarily ablates peripheral nerves for pain management; console controls gas supply and freeze parameters; operated by clinicians; provides therapeutic benefit through targeted tissue destruction; ACM is non-sterile, reusable electro-mechanical/pneumatic unit; CRYO2 probe is malleable, closed-tip aluminum; system includes manual/automatic freeze timer and automatic gas purge.
Clinical Evidence
Bench testing only. No clinical data presented. Testing included CRYO2 mechanical reliability, performance testing, acute animal verification, ASTM testing, and ACM software/controller board validation. All tests passed.
Technological Characteristics
Electro-mechanical/pneumatic system; Nitrous Oxide energy source; Joule-Thomson effect; Malleable aluminum cryoprobe; Gamma irradiation sterilization; Mains powered; Automatic gas purge; Manual/automatic freeze timer; Biocompatible materials per ISO 10993-1.
Indications for Use
Indicated for adults requiring cryosurgical treatment of cardiac arrhythmias via cryonecrosis of electrical conduction pathways or temporary ablation of peripheral nerves for pain management.
Regulatory Classification
Identification
A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.
Cryomedical Instruments LTD. CryoStar™ System (K031482)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which consists of a stylized human figure with three faces in profile, representing the department's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the logo in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
AtriCure Inc., Jonathan McElwee Regulatory Engineer 6217 Centre Park Dr. West Chester, OH 45069
Re: K142203
> Trade/Device Name: AtriCure Cryosurgical System - AtriCure Cryo Module (ACM) and AtriCure cryoICE cryo-ablation probes (CRY02) Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Regulatory Class: Class II Product Code: GXH, GEH Dated: October 30, 2014 Received: October 31, 2014
Dear Mr. Jonathan McElwee:
This letter corrects our substantially equivalent letter of November 25, 2014.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
Page 2 - Mr. Jonathan McElwee:
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/1/Picture/7 description: The image shows the FDA logo. The logo is in light gray and consists of the letters "FDA" in a stylized font. The letters are connected and have a geometric design.
Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142203
Device Name: AtriCure Cryosurqical System, comprising:
- a) AtriCure Crvo Module (ACM)
- b) AtriCure cryolCE™ cryo-ablation probes (CRYO2)
Indications for Use:
AtriCure Cryo Module Indication for Use Statement: The AtriCure Cryo Module is a nonsterile, reusable device which delivers cryogenic energy, namely nitrous oxide, to AtriCure's cryo-ablation probes.
cryolCE™ Indication for Use Statement: AtriCure's cryoICE™ cryo-ablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway. The probe is also intended for use in blocking pain by temporarily ablating peripheral nerves.
× Prescription Use (Part 21 CRF 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CRF 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AtriCure. The logo is in two colors, blue and orange. The word "Atri" is in blue, and the word "Cure" is in orange. There is a blue dot in the middle of the "C" in "Cure."
## 510(k) Summary
| I. Submitter | |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | AtriCure, Inc.<br>6217 Centre Park Dr.<br>West Chester, OH 45069<br>P: 513-755-4100<br>F: 513-755-4108 |
| Contact Person: | Jonathan McElwee, RAC<br>Regulatory Engineer |
| Alternate Contact: | Dennis Hong, JD, RAC<br>Senior Director, Regulatory Affairs |
| Date Prepared: | 08/8/2014 |
| II. Device | |
| Name of Device: | AtriCure Cryosurgical System, comprising:<br>a) AtriCure Cryo Module (ACM)<br>b) AtriCure cryoICE™ cryo-ablation probes (CRYO2) |
| Common Name: | Cryosurgical probe<br>Cryo Surgical Unit and Accessories<br>Cryoanalgesia System |
| Classification Name: | Surgical, General and Plastic Surgery, 21 CFR 878.4350<br>Cryogenic Surgical Device, 21 CFR 882.4250 |
| Regulatory Class: | Class II |
| Product Code: | GEH and GXH |
### lll. Predicate Devices
AtriCure cryo-ablation probe K082074 AtriCure Cryo Module System K111042, K112072, K121507, and K140058 Cryomedical Instruments LTD. CryoStar™ System K031482
The predicate devices have not been subject to a design-related recall.
No reference devices were used in this submission.
{4}------------------------------------------------
### IV. Device Description
The AtriCure Cryo Module (ACM) unit is a non-sterile reusable electro-mechanical and pneumatic cryogenic surgical system that delivers a cryogenic energy source, namely Nitrous Oxide, to the AtriCure cryo-ablation probes. The AtriCure's cryo-ablation probe (CRYO2) is a sterile, single use, cryosurgical device to be used in conjunction with the AtriCure Cryo Module or ACC2 Cardiac Cryosurgical System [K811390] to freeze target tissue, blocking the electrical conduction pathways by creating an inflammatory response or cryonecrosis. The cryolCE™ probe can also be used to temporarily ablate peripheral nerves.
Model numbers:
AtriCure Cryo Module - ACM1 AtriCure cryoICE™ cryo-ablation probe - CRYO2
## Materials:
All materials used in the manufacture of the AtriCure Cryosurgical System are safe and suitable for their intended use and suitable for their use with pressurized nitrous oxide. The ACM is not intended for patient contact. Testing has been previously conducted in accordance with ISO 10993-1 to ensure biocompatibility of all appropriate materials in the AtriCure cryolCE™ cryoablation probe (CRYO2).
#### V. Indications For Use
AtriCure Cryo Module Indication for Use Statement: The AtriCure Cryo Module is a non-sterile, reusable device which delivers cryogenic energy, namely nitrous oxide, to the AtriCure cryoablation probes.
cryolCE™ Indication for Use Statement: AtriCure's cryolCE™ cryo-ablation probes are sterile, single use devices intended for use in the cryosurgical treatment of cardiac arrhythmias by freezing target tissues, creating an inflammatory response (cryonecrosis) that blocks the electrical conduction pathway. The probe is also intended for use in blocking pain by temporarily ablating peripheral nerves.
#### VI. Comparison Of Technological Characteristics With The Predicate Devices
When used for the intended use of "cryosurgical treatment of cardiac arrhythmias", the AtriCure Cryosurgical System has the following similarities to the technological characteristics of the previously cleared predicate AtriCure Cryo Module System (K082074, K111042, K112072, K121507, and K140058):
- Same intended use ●
- . Same operating principle
- Same fundamental scientific technology ●
- Same crvo-ablation probes ●
- Same materials used ●
- . Same module design
- Same electronic solenoid valves ●
- Same 3 year age shelf life of the cryo-ablation probes ●
{5}------------------------------------------------
# AtriCure®
When used for the intended use of "blocking pain by temporarily ablating the peripheral nerve", the AtriCure Cryosurgical System has the following similarities to the technological characteristics of the previously cleared predicate CryoStar™ System K031482:
| | CryoStar™ Predicate | AtriCure Device |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Cryomedical Instruments LTD.<br>(now Kryo Science LTD) | AtriCure |
| Market Product<br>Name | CryoStar™ System<br>(Cryostar™ console, 1 & 2mm<br>cryoprobe, convenience procedure kit<br>for the probe placement) | AtriCure Cryosurgical System<br>(ACM and cryolCE™ (CRYO2) cryo-<br>ablation probe) |
| 510(k) Number | K031482 | Subject of this Submission |
| Intended Use | Cryoanalgesia device intended for use<br>in blocking pain by temporarily ablating<br>the peripheral nerves. | The AtriCure Cryosurgical System is<br>intended for use in blocking pain by<br>temporarily ablating the peripheral<br>nerves. |
| Target Population | Adults | Adults |
| Operating Principle | Joule-Thompson Effect | Joule-Thompson Effect |
| Technology | The system consists of a range of<br>cryoprobes that are used for freezing<br>nerves to block pain by temporary<br>ablation. A console is used to house<br>and control the supply of gas to the<br>cryoprobe and to provide an electrical<br>nerve location device. | The system consists of cryoprobes<br>that are used for freezing target tissue<br>to treat cardiac arrhythmias and for<br>freezing nerves to block pain by<br>temporary ablation. A console is used<br>to control the supply of gas to the<br>cryoprobe. |
| Energy Used | Nitrous Oxide | Nitrous Oxide |
| Operating<br>Temperature | -40C to -70C | -50°C to 70°C |
| Peripheral Nerve<br>Stimulation | Included as part of the system | Not included in the system. |
| Human Factors | Hand-held device containing cryogen<br>with footswitch. | Hand-held device containing cryogen<br>with activation button on console or<br>footswitch. |
| Gas Purge System | Automatic | Automatic |
| Freeze Timer | Manual or automatic | Manual or automatic |
| Patient Contacting<br>Materials | Hemispherical and trocar-tip stainless<br>steel 1mm & 2mm diameter cryoprobe | Closed-tip aluminum 4mm cryoprobe |
| Cryotip | Rigid | Malleable |
| Biocompatibility | Biocompatible patient contacting<br>materials. | Biocompatible patient contacting<br>materials. |
| Probe Packaging | Non-Sterile - Reusable device | Sterile - Single Use disposable device |
| Sterilization | Steam Autoclave; EtO; or Sterrad | Gamma Irradiation |
| Power Source | Mains Powered | Mains Powered |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the AtriCure logo. The word "Atri" is in blue, while "Cure" is in orange. There is a blue dot in the middle of the "C".
#### VII. Performance Data
Appropriate product testing was conducted to evaluate conformance to product specification and substantial equivalence to predicate devices. Design verification and validation, software validation, and electrical safety testing were completed to show that the AtriCure Cryosurgical System is substantially equivalent and in conformance with international standards and device specifications.
## Non-clinical Bench Testing
| Name | Description | Results |
|----------------------------------------------|------------------|---------|
| CRYO2 Comparison testing to Predicate Device | Bench testing | Passed |
| CRYO2 Mechanical Reliability | Bench testing | Passed |
| CRYO2 Performance testing | Bench testing | Passed |
| CRYO2 Acute Animal Verification Lab | Bench testing | Passed |
| CRYO2 ASTM Testing | Bench testing | Passed |
| Evaluation Against Similar Product | Bench testing | Passed |
| ACM Software Validation | Software testing | Passed |
| ACM Display Board Verification | Bench testing | Passed |
| ACM Controller Board Verification | Bench testing | Passed |
| Handpiece Connector Board Verification | Bench testing | Passed |
| ACM Power Board Verification | Bench testing | Passed |
| ACM Reliability Testing | Bench testing | Passed |
| ACM Simulated Use | Bench testing | Passed |
#### VIII. Conclusions
When used for the intended use of "cryosurgical treatment of cardiac arrhythmias", the proposed AtriCure Cryosurgical System is equivalent to the previously cleared AtriCure Cryo Module System, as the intended use, overall function, and materials used are the same.
When used for the intended use of "blocking pain by temporarily ablating the peripheral nerve", the proposed AtriCure Cryosurgical System is equivalent to the previously cleared CryoStar™ System, as the intended use, basic overall function, and materials used are substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.