K182440 · A-1 Engineering · NGX · Feb 26, 2019 · Physical Medicine
Device Facts
Record ID
K182440
Device Name
Body System
Applicant
A-1 Engineering
Product Code
NGX · Physical Medicine
Decision Date
Feb 26, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Body System is intended for muscle conditioning to stimulate healthy muscles. The Body System is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. The Body System is intended to be operated by a trained professional who is present to monitor treatment.
Device Story
Body System (models SX101, OS2911) is a powered muscle stimulator for muscle conditioning; utilizes 16 channel ports with self-adhesive pad applicators. Device delivers electrical currents to contract muscles; features manual adjustment of five frequency programs (200 Hz to 1250 Hz). Output signal is biphasic, rectangular, voltage and current regulated; channels are isolated via separate transformers. Operated by trained professional in clinical setting; professional monitors treatment. Device powered by 110-240V AC mains. Output parameters (voltage, current, frequency) adjusted manually; display shows treatment time. Benefits include muscle conditioning through electrical stimulation of healthy muscle tissue.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included biocompatibility (cytotoxicity, irritation, sensitization per ISO 10993), output waveform analysis, electromagnetic compatibility (EN/IEC 60601-1-2), and electrical safety (IEC 60601-1).
Indicated for muscle conditioning to stimulate healthy muscles in adults. Contraindicated for use in conjunction with therapy or treatment of medical diseases or conditions of any kind; not for use on injured or ailing muscles.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 26, 2019
A-1 Engineering % Peter Knauer Principal Consultant Sage BioPartners 1741 Little Kate Rd Park City, Utah 84060
Re: K182440
Trade/Device Name: Body System Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: November 26, 2018 Received: November 27, 2018
Dear Peter Knauer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182440
Device Name Body System
Indications for Use (Describe)
The Body System is intended for muscle conditioning to stimulate healthy muscles.
The Body System is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. The Body System is intended to be operated by a trained professional who is present to monitor treatment.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><span style="margin-right:5px;"> Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div style="display:flex; align-items:center;"><span style="margin-right:5px;"><input type="checkbox"/></span>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5. 510(k) Summary / Statement - 21 CFR 807.92(a)
This 510(k) Summary for the Body System is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and follows the Office of Device Evaluation (ODE) guidance concerning the organization and content of a 510(k) Summary.
| Applicant Name: | A-1 Engineering |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Address: | 30 Mauchly, Suite A<br>Irvine, CA 92618 |
| Official Contact: | Tony Picciano, Owner |
| Telephone: | (949) 748-7488 |
| Email: | tony6213@yahoo.com |
| Submission Date: | November 26, 2018 |
| Subject Device: | Trade Name: Body System<br>Common Name: Muscle stimulator for conditioning. |
| Regulation Description: | Powdered Muscle Stimulator |
| Classification: | Regulatory Class II, 21 CFR 890.5850 |
| Classification Name: | Stimulator, Muscle, Powered, For Muscle Conditioning |
| Product Code: | NGX |
| Device Description | Body System (available in models: SX101 and OS2911). Both models<br>utilize the same internal electrical components. The models differ only in<br>the control/display features and exterior casing/dimensions. The Body |
{4}------------------------------------------------
System is a Powered Muscle Stimulator with 16 channel ports using self adhesive pad applicators attached to the body for muscle conditioning. The pad applicator is a 510K cleared device (K970426), itself.
The Body System uses the same technological characteristics as the predicate devices (listed below); it is an electrically powered device that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area. The Body System muscle stimulators provide selections of different programs through manual adjustment of five different frequency programs of the device from 200 Hz, to 1250 Hz. The Body System has a maximum output voltage of 88V @ 2k Ω, rated at 200 volts. A display shows the balance treatment time. The output signal is biphasic, rectangular and based on a voltage and current regulated technology. All channel outputs are isolated from each other; they have their own transformers and amplifiers which are Independent from neighboring channels or outputs. The only commonality between the outputs are their connectors to the power supply, the start button, the programs button, and the pulse button. The Body System power source is 110 V - 240 V AC Mains, 50-60 Hz. The patient leakage current tested at 240VAC 60Hz is 5 microamps for both the normal and fault conditions. The Body System Channels are synchronous and are isolated by separate transformers.
The Intended Use of the Body System has the same key elements as the predicates: for muscle conditioning to stimulate healthy muscle. There are no differences in Intended Use amongst the predicates.
- The Body System is intended for muscle conditioning to stimulate healthy Indications for use: muscles.
The Body System is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. The Body Sy is intended to be operated by a trained professional who is present to monit treatment.
- Predicate Device: The Body System is substantially equivalent to the predicate devices listed below:
- . K123158, Ion Magnum Genius, under 21 CFR 890.5850 and Product Code NGX
Predicate Trade Name: Ion Magnum Genius
{5}------------------------------------------------
Conditioning Premarket Notification: K123158 Manufacturer: Ion Genius, Inc. . K143551, Compex Wireless USA, under 21 CFR 890.5850 and Product Code NGX Predicate Trade Name: Compex Wireless USA Classification Name: Stimulator, Muscle, Powered for Muscle Conditioning Premarket Notification: K143551 Manufacturer: DJO, LLC Based on the intended use/Indications for use, design, materials, principle of operation, energy used and materials, the Body System is substantially equivalent to the legally marketed predicate devices. The non-clinical performance data demonstrates substantial equivalence and includes an analysis of output waveforms. Non-Clinical Tests The Body System has been tested to establish is safety and performance. Performed on the The following tests have been performed on the patient contacting parts: Device: ● Biocompatibility - Cytotoxicity (ISO-10993-5) - result: Pass . Biocompatibility – irritation (ISO-10993-10) – result: Pass ● Biocompatibility - Sensitization (ISO-10993-10) - result: Pass . Performance – including Output Waveforms of the Body System as compared to a predicate . Electromagnetic Compatibility - according to EN 60601-1-2:2015, IEC 60601-1-2:2014, IEC 60601-2-10 2012 + A1 2016 . Electrical Safety – according to IEC 60601-1-6:2010, AMD1:2013, IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) Summary: There are no differences in the technological characteristics between the subject device and the predicate devices, which raise issues concerning safety, performance or efficacy. The Body System is substantially equivalent to the predicate device(s) in terms of safety and effectiveness.
Classification Name: Stimulator, Muscle, Powered for Muscle
{6}------------------------------------------------
| Comparison of the Subject and Predicate Devices | | | | | |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Elements of<br>Comparison | Subject device | Primary<br>Predicate Device | Secondary<br>Predicate Device | Similar/Different | |
| Device Name and<br>Model | Body System | Ion Magnum<br>Genius | Compex Wireless<br>USA | N/A | |
| 510 (K) Number | Subject of this<br>submission | K123158 | K143551 | N/A | |
| Product Code | NGX | NGX | NGX | Similar | |
| Regulation<br>Number | 890.5850 | 890.5850 | 890.5850 | Similar | |
| Indications for<br>Use | The Body System<br>is intended for<br>muscle<br>conditioning to<br>stimulate healthy<br>muscles.<br><br>The Body System<br>is not intended to<br>be used in<br>conjunction with<br>therapy or<br>treatment of<br>medical diseases<br>or medical<br>conditions of any<br>kind. The Body<br>System is<br>intended to be<br>operated by a<br>trained<br>professional who<br>is present to<br>monitor<br>treatment. | Ion Magnum<br>Genius is<br>intended for<br>muscle<br>conditioning to<br>stimulate healthy<br>muscles. The Ion<br>Magnum Genius<br>is not intended to<br>be used in<br>conjunction with<br>therapy or<br>treatment of<br>medical diseases<br>or medical<br>conditions of any<br>kind. None of the<br>Ion Magnum<br>Genius programs<br>is designed for<br>injured or ailing<br>muscles and its<br>use on such<br>muscles is<br>contraindicated.<br>The Ion Magnum | The Compex<br>Wireless USA is an<br>Over-The-Counter<br>device intended<br>to stimulate<br>healthy muscles in<br>order to improve<br>or facilitate<br>muscle<br>performance. It is<br>to be used by<br>adults only.<br><br>The Compex<br>Wireless USA is<br>not intended for<br>adjunctive<br>therapy in the<br>treatment of<br>medical diseases<br>and conditions of<br>any kind. None of<br>the Compex<br>Wireless USA<br>stimulation<br>programs are<br>designed for<br>injured or disease<br>afflicted muscles.<br>Its use on such | Similar - although<br>there are some<br>differences in the<br>exact language of<br>each Intended<br>Use, each is<br>indicated for the<br>Stimulation of<br>Healthy Muscles | |
| Comparison of the Subject and Predicate Devices | | | | | |
| Elements of<br>Comparison | Subject device | Primary<br>Predicate Device | Secondary<br>Predicate Device | Similar/Different | |
| | | offered under<br>prescription given<br>by a physician<br>licensed in the<br>state in which he<br>or she practices. | contraindicated.<br>The work imposed<br>on the muscles by<br>the Compex<br>Wireless USA<br>programs is<br>definitely not<br>suitable for<br>rehabilitation and<br>physiotherapy.<br>The Compex<br>Wireless USA<br>electrical impulses<br>allow the<br>triggering of<br>action potentials<br>on motoneurones<br>of motor nerves<br>(excitations).<br>These excitations<br>of motoneurones<br>are transmitted to<br>the muscle fibers<br>via the motor<br>endplate where<br>they generate<br>mechanical<br>muscle fiber<br>responses that<br>correspond to<br>muscle work.<br>Depending on the<br>parameters of the<br>electrical impulses<br>(pulse frequency,<br>duration of<br>contraction,<br>duration of rest,<br>total session<br>duration),<br>different types of | | |
| Comparison of the Subject and Predicate Devices | | | | | |
| Elements of<br>Comparison | Subject device | Primary<br>Predicate Device | Secondary<br>Predicate Device | Similar/Different | |
| | | | be imposed on<br>the stimulated<br>muscles.<br>The Compex<br>Wireless USA may<br>therefore be<br>considered a<br>technique of<br>muscle training. | | |
| Body Application<br>Areas | Body | Body | Body | Similar | |
| Power Sources | 110vac — 240 vac<br>(AC wall plug-In) | 110vac — 240 vac<br>(AC wall plug-In) | Lithium Polymer<br>(LiPo)<br>rechargeable<br>batteries | Different -<br>differences do not<br>present different<br>questions of/or<br>effect safety and<br>effectiveness; the<br>similar currents<br>generated by<br>each channel and<br>applied to the<br>patient is isolated<br>from the power<br>source. | |
| Patient Leakage<br>Current | 0.05 µA and | 0.07 μΑ | N/A | Similar | |
| Number of<br>Output Modes for<br>Micro current<br>Stimulation | 5 | 3 | 1 | Different -<br>differences do not<br>present different<br>questions of/or<br>effect safety and | |
| Comparison of the Subject and Predicate Devices | | | | | |
| Elements of<br>Comparison | Subject device | Primary<br>Predicate Device | Secondary<br>Predicate Device | Similar/Different | |
| | | | | effectiveness; all<br>outputs meet the<br>requirements of<br>the FDA Guidance<br>on Powered<br>Muscle<br>Stimulators. | |
| Number of<br>Channels for<br>Micro current<br>Stimulation | 16 | 6 | 4 | Different -<br>differences do not<br>present different<br>questions of/or<br>effect safety and<br>effectiveness;; all<br>channels are<br>isolated and<br>generate the<br>identical signal,<br>which are similar<br>to the predicates. | |
| Channel Output:<br>Synchronous or<br>Alternating | Synchronous | Synchronous | Synchronous | Similar | |
| Channel Output:<br>Method of Line<br>Current Isolation | Power Supply<br>isolation | Power Supply<br>isolation | Power Supply<br>isolation | Similar | |
| Regulated<br>Current or<br>Regulated<br>Voltage | Both | Voltage | Both | Similar (voltage<br>regulated devices) | |
| Software/Firm<br>ware/Microproce<br>ssor control | No | No | Yes | Different -<br>differences do not<br>present different<br>questions of/or | |
| Comparison of the Subject and Predicate Devices | | | | | |
| Elements of<br>Comparison | Subject device | Primary<br>Predicate Device | Secondary<br>Predicate Device | Similar/Different | |
| | | | | effect safety and<br>effectiveness; the<br>similar currents<br>generated by<br>each unit is not<br>related to<br>software/<br>firmware. | |
| Automatic<br>Overload Trip | Yes | Yes | Yes | Similar | |
| Automatic No<br>load Trip | Yes | Yes | Yes | Similar | |
| Indicator Display:<br>On / Off Status | Yes | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.