K123158 · Ion Genius, Inc. · NGX · Jun 21, 2013 · Physical Medicine
Device Facts
Record ID
K123158
Device Name
ION MAGNUM GENIUS
Applicant
Ion Genius, Inc.
Product Code
NGX · Physical Medicine
Decision Date
Jun 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Ion Magnum Genius is intended for muscle conditioning to stimulate healthy muscles. The Ion Magnum Genius is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the Ion Magnum Genius programs is designed for injured or ailing muscles and its use on such muscles is contraindicated. The Ion Magnum Genius is only offered under prescription given by a physician licensed in the state in which he or she practices.
Device Story
Ion Magnum Genius is a powered muscle stimulator used in clinical settings by physicians to provide muscle conditioning for healthy adults. The device delivers biphasic, rectangular electrical pulses to muscles via six independent channels using self-adhesive electrodes. It operates via voltage and current regulated technology, with manual adjustment of frequencies (90, 100, 120 Hz). Each channel is isolated by separate transformers. The device features an automatic overload trip and patient override control for safety. The physician operates the device, applying clinical judgment to treatment administration. By stimulating healthy skeletal muscles, the device aims to facilitate muscle conditioning. It is strictly a business-to-business device, not for home use.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, including electrical safety, output specifications, and comparison of technological characteristics to the predicate device.
Technological Characteristics
Powered muscle stimulator; 6 independent channels; biphasic rectangular waveform; current/voltage regulated. Max output: 100mA, 200V. Frequency: 90-120Hz. Pulse duration: 416-500µs. Isolation: separate transformers per channel. Power: 110-240V AC. Applied part: Type BF (IEC 60417-5333). Electrodes: Axelgaard self-adhesive (K970426), compliant with ISO 10993-5/10. No software or firmware.
Indications for Use
Indicated for muscle conditioning in healthy adults. Contraindicated for use on injured or ailing muscles or for the treatment of medical diseases/conditions. Prescription use only.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for ION GENIUS. The logo consists of the words "ION GENIUS" in a stylized font, with a three-dimensional cube made of interconnected circles above the words. The cube is made up of smaller circles connected by lines, creating a lattice-like structure.
### JUN 2 1 2013
### TRADITIONAL 510 K SUMMARY
1833 Kalakaua Ave, Suite 905 Honolulu, HI 96821 Phone: +1-866-25-YOUNG/808-395-0656 FAX: 877-26-YOUNG/808-395-0787 Email:info@arasysperfector.com Website www.arasysperfector.com
### 510(k) Summary K123158
lon Genius, Inc. / Ion Magnum LTD / UV Innovations Ltd
Corporate Headquarters and Shipping Address 1833 Kalakaua Ave
Suite #905
Honolulu, HI 96815
Manufacturing Facility UV Innovations Ltd BIC 1 Innova Park Mollison Ave, Enfield, Middlesex, EN3 7XU, UK
UK: +442032861886 USA +1 808 395-0656 * +1-866-25-YOUNG +1-808 395 0787 quality@iellios.com research@iellios.com 1-808 6461948
### Submitter's name:
Address:
Telephone:
FAX:
Company E-mail: Contact Person: Contact Person Telephone:
Date of Summary
{1}------------------------------------------------
ls re-submitted:
FDA Establishment Registration No:
Proprietary Name:
Regulation Description:
Panel Code
Product Code:
Submission Type
Regulation Number
Device Class
Trade Name:
Common Name:
Classification Name:
Distributed by:
June, 2013
3003588228 (Ion Genius Inc)
lon Magnum Genius
Powered Muscle Stimulator for Muscle Conditioning
ва
NGX
510(k)
890.5850
ll
ION MAGNUM GENIUS
Powered Muscle Stimulator for Muscle Conditioning
Stimulator, muscle, powered, for muscle conditioning
Ion Genius, INC / Ion Magnum LTD / UV Innovations LTD Corporate Headquarters and Shipping Address: 1833 Kalakaua #905 Honolulu, HI 96815 Tel: 866-25-YOUNG tel:808-395-0656 Fax: 877-26-YOUNG
PREDICATE: Legally Marketed Devices
> 1. Compex Sport Compex, S.A. Regulation No 890,5850 K 011880 Stimulator, muscle, powered, for muscle conditioning Product Code: NGX
| Action Taken to Comply with the | Yes. A new improvement in the device design to |
|---------------------------------------|---------------------------------------------------|
| requirements of section 514 | enhance the device safety was the additional |
| of the act, regarding compliance | insulation of wires and other components near the |
| with applicable performance standards | chassis which is connected to the ground and the |
| in accordance with 121 CFR 807.87 (d) | increase of the gap between wires and all other |
| | components and chassis to 9 milimeters. This new |
| | design significantly reduces the risk that the |
{2}------------------------------------------------
| | connections could be shorted to the ground.<br>Manufacturing facility has been awarded ISO 13485:<br>2003 for the manufacture and sales of muscle<br>stimulators 2011-2014. There are extensive<br>Bench tests prior to each Ion Magnum being<br>released by the manufacturing facility with the pass<br>sticker and the date that it passed next to the serial<br>number of the device |
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
#### Device Description:
lon Magnum Genius is a muscle stimulator with six ports and one output that is the same for all ports. The unit is designed for healthy men & women to provide muscle conditioning. Ion Magnum Genius stimulator provides selections of different programs through manual adjustment of the three different frequencies of the device, 90 Hz, 100 Hz which are within the same range as the predicate that offer a range from 1-120 Hz. The setting that the frequency button is turned shows which frequency the device is set on. The device delivers a voltage of 50 volts at 500 Ohms load and 100ma just like the predicate. It delivers a voltage of 200 volts at 10K Ohms just like the predicate. The lon Magnum maximum output current is 100 ma which is equivalent to the maximum output current of the predicate. The lon Macnum has a maximum output current of 200 volts which is equivalent to the maximum output current of the predicate. The device primarily treats abdominal muscles (ABS). LCD shows the balance treatment time. The output signal is biphasic, rectangular and based on a voltage and current requlated technology. The unit is easy and simple to use. Ion Magnum is suitable for use by all healthy adults. However as with other muscle stimulators some care is needed when using the lon Magnum Genius [see contraindications and warnings in manual]
NOTE 1: All channels outputs are isolated from each other. They have their own transformers and amplifiers which are independent from neighboring channels or outputs. The only common thing between the outputs is their connectors to the power supply, the start button, the programs button and the pulse button.
There is a transformer that translates energy received from the mains into micro currents. Hence there is an insulation of mains from circuitry. From Circuitry to output there is also insulation through the transformer. Therefore there is a double separation between mains and the human body.
3
This unit contains 24 self-adhesive electrodes that already have 510K clearance (K970426)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a device labeled "ION MAGNUM GENIUS". The device has several knobs and buttons on its side. The top of the device has a pattern of dots.
The lon Magnum power source is 110V - 240 V AC Mains, 50-60Hz The patient leakage current is 0.07 microamps under both the normal and fault conditions.. A new improvement in the device design was made in 2012 in order enhance the device safety which entailed the additional insulation of wres and other components near the chassis. The increase of the gap between wires and all other components and chassis is now increased at least 9 millimeters. This new design safety significantly reduces the risk that the connections could be shorted to the ground.
Channels are synchronous and are isolated by separate transformers the way the channels or the predicate are synchronous and isolated by separate transformers. Ion Magnum Genius has an automatic overload trip, just like the predicate but not an automatic no load trip just like the predicate that do not have a no load trip, therefore in terms of the overload trip features the lon Magnum Genius and the predicate are substantially equivalent. The lon Magnum Genius is also equivalent to the predicate in that it has an automatic shut off and patient override control for additional safety. Indicator displays include on and off status just like it does in the predicate. Ion Magnum Genius is an IEC 1 device, IEC 60417-5333, type BF applied part device.
| Question | Yes | No |
|---------------------------------------------------------------------------------------|-----|----|
| Is the device intended for prescription use (21 CFR 801 Subpart D)? | X | |
| Is the device intended for over-the-counter use (21 CFR 807 Subpart | | X |
| Does the device contain components derived from a tissue or<br>other biologic source? | | X |
| Is the device provided sterile? | | X |
| Is the device intended for single use? | | X |
| Is the device a reprocessed single use device? | | X |
| If yes, does this device type require reprocessed validation data? | | |
#### Design and Use of the Device
{4}------------------------------------------------
| Does the device contain a drug? | | X |
|---------------------------------------------------|---|---|
| Does the device contain a biologic? | | X |
| Does the device use software? | | X |
| Does the submission include clinical information? | X | |
| Is the device implanted? | | X |
### Intended Use
lon Magnum Genius is intended for muscle conditioning to stimulate healthy muscles. The lon Magnum Genius is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. None of the lon Magnum Genius programs is designed for injured or ailing muscles and its use on such muscles is contraindicated. The lon Magnum Genius is only offered under prescription given by a physician licensed in the state in which he or she practices.
{5}------------------------------------------------
### COMPARISON OF INDICATIONS FOR USE OF ION MAGNUM GENIUS AND PREDICATE
| Ion Magnum Genius is intended for muscle | Compex Sport" (K01180) is intended for |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| conditioning to stimulate healthy muscles. The Ion | muscle conditioning to stimulate healthy |
| Magnum Genius is not intended to be used in | muscles in order to improve or facilitate muscle |
| conjunction with therapy or treatment of medical | performance. |
| diseases or medical conditions of any kind. None of the Ion Magnum Genius programs is designed for injured or ailing muscles and its use on such muscles is contraindicated. The Ion Magnum Genius is only offered under prescription given by a physician licensed in the state in which he or she practices. | "Compex® Sport" is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the "Compex® Sport" training programs is designed for injured or ailing muscles and its use on such muscles is contraindicated. |
Substantial Equivalence based on the Indicatins of Use Statement: When compared to the predicated device, Compex® Sport K011880 the lon Magnum Genius has the same intended use. Therefore, the lon Magnum Genius is substantially equivalent to the predicate marketed device Compex® Sport KO11880.
б
{6}------------------------------------------------
### COMPARISON OF DEVICE DESCRIPTION, SAFETY AND EFFECTIVENESS OF ION MAGNUM GENIUS TO PREDICATE DEVICES:
| Ion Magnum Genius is a Muscle stimulator with<br>six independent channels which electrical<br>current can be regulated individually<br>Biphasic rectangular impulses with electrical mean<br>equal to zero. Maximum output current is 100 mA<br>and maximum voltage output is 200 volts. The<br>device gives 50V output at 500 Ohms overload and<br>200 volts at 10K Ohms overload. Maximum<br>quantity of electricity per channel:40 µC Impulse<br>width: 400 to 500 µsecs. The three Ion Magnum<br>Genius frequencies, 90 Hz, 100 and 120Hz are<br>within the same range as the predicate. The<br>electrical pulses generated by the Ion Magnum Genius<br>are transmitted to the muscles via self-adhesive<br>electrodes, Axelgaard type electrodes electrodes (<br>510K K970426) The electrodes are equivalent to<br>those used by the predicate. | Compex Sport" (K01180) has:<br>Four independent channels which electrical current can<br>be regulated individually. Biphasic rectangular impulses<br>with electrical mean equal to zero<br>Maximum output current is 100 mA and maximum<br>voltage output is 200 volts. The device gives 50V output<br>at 500 Ohms overload and 200 volts at 10K Ohms<br>overload. Maximum quantity of electricity per channel: 40<br>µC impulse width: 200 or 400 µS. Range of pulse<br>frequency: 1-120 Hz The electrical pulses generated by<br>'Compex® Sport" are transmitted to the muscles via self-<br>adhesive electrodes. |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### Other equivalence aspects between lon Magnum Genius and the Predicate
| Other<br>equivalence<br>aspects<br>between Ion<br>Magnum<br>Genius and<br>the Predicate | 1. Ion Magnum Genius and the predicate devices are identical in terms of the<br>thermal safety (there is no thermal energy for Ion Magnum Genius or the<br>predicate) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. The chemical safety is equivalent (since this is applicable to neither the Ion<br>Magnum Genius nor the predicate) | |
| 3. Sterility is equivalent (since neither the Ion Magnum Genius nor the predicate<br>require sterilization or are supplied sterile.) | |
| 4. The Ion Magnum Genius uses disposable pads that do not need sterilization that<br>are equivalent to the pads of the predicate which also do not need sterilization<br>because they are disposable. | |
| 5. Biocompatibility of the Ion Magnum disposable pads is equivalent to those of the<br>predicate. Biocompatibility is performed by the company from which these<br>disposable pads are purchased (Axelgaard). Ion Magnum uses disposable pads<br>item no: SN2020 which are square 2" x 2" (5cm x 5cm) which have a 510K<br>number K970426 and comply to ISO10993-5 Biological Evaluation of Medical<br>Devices part 5 Test for in vitro cytotoxicity & ISO10993-10. The are made by<br>Axelgaard Manufacturing Co. LtD, 520 Industrial Way Fallbrook California 92028,<br>USA which has a Certificate of Registration Quality Management System - ISO<br>13485:2003 and holds certificate no FM 40363 and operates a Quality<br>Management System which complies with the requirements of ISO 13485: 2003<br>for the following scope: The design and development, manufacture and | |
{7}------------------------------------------------
| distribution of hydrogel and non-invasive electrodes: Effective date 03/11/2012<br>- Expiry Date: 03/12/2015.] |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 6. The cables carrying the impulses to the pads and the connectors are both<br>designed to insulate the client and operator from inadvertent contact with the<br>relatively weak electrical impulses. Should the connectors be dislodged in use, for<br>instance, by the client changing position during treatment, the insulation will<br>ensure no direct contact with the skin will take place. The lon Magnum cables<br>remain screwed on and attached to the device as are the cables of the predicate<br>devices. |
·
#### output specifications comparing ion Magnum Genius With THE PREDICATE DEVICES .
・
| | ION MAGNUM<br>GENIUS | Compex Sport<br>K011880 |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| Waveform | Biphasic<br>synchronous | Biphasic<br>synchronous |
| Synchronous or<br>alternating channels | synchronous | synchronous |
| Dimension of Device | 15.5 inches length<br>8.5 height<br>7.5 width | Length: 142 mm<br>Width 99mm<br>Height 36mm |
| Weight of the<br>Device | 8 pounds | 350 g |
| Weight of<br>Accessories | 4.7 pounds | 95 g |
| Indicator Display | Yes | Yes |
| Low Battery | N/A | Yes |
| Voltage Current<br>Level | Yes | Yes |
| Shape | Rectangular | Rectangular |
| Max Output Voltage | 50V @ 500 Ω | 50V |
| Net Charge | 200 V @10k Ω<br>0.01 amperes at<br>500 Ohms<br><br>0.004 Amperes at<br>2K ohms<br><br>0.0025 Amperes at<br>10K Ohms | 200 V +/-10 |
| Max Output Current | 100 ma at 500<br>Ohms | 100 ma at 500 Ohms |
| Frequency (Hz)<br>for Muscle<br>Conditioning | 90 Hz<br>100 Hz<br>120Hz | 1-120 Hz |
| Pulse Duration | 416 - 500 µsec | 200 to 400 µsec |
| Pulse Width | 416 - 500 µsec | 200 to 400 µsec |
| Maximum Phase<br>Charge | 50 µC @ 500 Ω at<br>90 and 100 Hz<br><br>40 µC @ 500 Ω at<br>120 Hz | 40 µC |
| Maximum Current<br>Density (mA/cm2) | 4 mA / cm2 | 4 mA/cm2 |
| Maximum Power<br>Density (W/Cm2) | 0.0012 W/cm^2 | 0.038 W/cm^2. |
| Current or Voltage<br>Driven | Current Driven<br>Voltage | Current Driven<br>Voltage |
| Symmetrical<br>Phases of Waveform | Yes | Yes |
| Regulated<br>Current/ voltage | Regulated Voltage<br>and Regulated<br>Current. | Constant Current Regulator /<br>regulated current |
| Method of Channel<br>Isolation | Channels are<br>isolated by<br>separated<br>transformers | Channels are isolated by separated<br>transformers |
| Self Adhesive<br>Electrodes Axelgaard<br>Tyoe | identical | Identical |
| Current Leakage | $0.07 \mu A$ | $>100$ microamps |
{8}------------------------------------------------
and the comments of the comments of the comments of the comments of
:
{9}------------------------------------------------
### DIFFERENCES BETWEEN ION MAGNUM AND PREDICATE..
- 1. The lon Magnum Genius has 6 channels while the Compex Sport has 4 channels. However. the exact same output comes out of all lon Genius channels. Risk assessment testing has indicated that output coming out of six channels simultaneously is equally safe as output coming out of four channels simultaneously.
- 2. The lon Magnum Genius does not have software, a microcomputer or firmware, while the predicate device is controlled by a microcomputer. However this difference does neither affect the indications of use or the safety and effectiveness of the lon Magnum Genius.
- 3. THE ION MAGNUM GENIUS IS ONLY OFFERED UNDER PERSCRIPTION AND IT IS ONLY OFFERED AS A BUSINESS TO BUSINESS DEVICE SOLD ONLY TO PHYSICIANS WHICH MEANS THAT THE ION MAGNUM GENIUS WILL BE NEVER BE USED BY AN INDIVIDUAL AT HOME. ADMINISTRATION OF THE ION MAGNUM TREATMENT WILL ALWAYS BE BY A TRAINED PHYSICIAN AND THEREFORE A PATIENT RECEVING A TREATMENT WITH THE ION MAGNUM TREATMENT WILL ALWAYS HAVE THE ADVANTAGE OF THE PHYSICIAN'S CLINICAL JUDGEMENT. A PHYCISIAN WILL ALSO BE AVAILABLE FOR CONSULTATION WITH ANYTHING REGARDING THE ION MAGNUM AND OVERALL ADDITIONAL SAFETY.
- 4. The lon Magnum Genius has a different weight than the predicate. However this difference does neither affect the indications of use or the safety and effectiveness of the lon Magnum Genius.
### SUBSTANTIAL EQUIVALENCE DISCUSSION
#### 1. Substantial Equivalence on the basis of Indication for Use
The lon Maqnum and the predicate devices have the same intended use and equivalent indications. and principles of operation that includes the delivery of small amount of electrical stimulation to healthy skeletal muscles for conditioning of the muscles.
### 2. Substantial Equivalence on the basis of Technological characteristics:
The lon Magnum Genius and the predicate have a number of similarities in terms of their technological characteristics cited below:
- 1 . Biphasic Rectangular Waveform
- 2. Synchronous channels
- 3. Maximul output voltage
- 4. Maximum output current
- 5. range of frequencies
- 6. Pulse Duration
- 7. Pulse width
- 8. Maximum Phase Charge
- 9. Maximum Current Density
{10}------------------------------------------------
10. Channels' method of isolation by separate transformers
. .
.
.
:
.
. . :
:
1 1 . Identical Electrodes
:
12. Similar current leakage
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like bird with three curved lines representing its wings and tail. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird.
Public Health Service
June 21, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
lon Genius c/o Xanya Sofra-Weiss, Ph.D. Director of Research and Training 7192 Kalanianole Hwy #D-204A Honolulu. HI 96821
Re: K123158
Trade/Device Name: Ion Magnum Genius Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: June, 2013 Received: June 7, 2013
Dear Dr. Sofra-Weiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{12}------------------------------------------------
Page 2 - Xanya Sofra-Weiss, Ph.D.
CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/or You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetw/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Joyce M. Whang -S
for Victor Krauthamer. Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
## Indications for Use
510(k) Number (if known): K123158
Device Name: Ion Magnum Genius
Indications For Use:
lon Magnum Genius is intended for muscle conditioning to stimulate healthy muscles. The lon Magnum Genius is not intended to be used in conjunction with therapy or treatment of medical diseases or medical conditions of any kind. None of the lon Magnum Genius programs is designed for injured or ailing muscles and its use on such muscles is contraindicated. The lon Magnum Genius is only offered under prescription given by a physician licensed in the state in which he or she practices.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| .<br> | . | 1<br>1<br> |
|------------------|---|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (Divicion Sinn I | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number _ K123158 __
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.