The Krystal Klear Aligners System is intended to correct minor discrepancies in the alignment of maloccluded anterior teeth on patients with permanent dentition (second molars) by moving the teeth with a progressive series of clear thin, thermoformed plastic aligners, fabricated in stages to gradually align teeth over a period of several months. The aligners are completely removable by the patient and may be discontinued at any time.
Device Story
Krystal Klear Aligner System uses sequential, thermoformed copolyester plastic trays to reposition teeth. Dental professional assesses patient, takes molds, and submits prescription. Masterpieces (MAP) software designs tray series based on prescription; physician reviews and approves plan. MAP manufactures trays via 3D printed molds. Patient wears trays sequentially to apply corrective force; trays are removable. Professional monitors treatment progress. System intended for clinic use by orthodontists/dentists to treat minor malocclusions. Benefits include gradual, removable tooth alignment.
Clinical Evidence
Bench testing only. Evaluated fit of aligner against treatment plan and process consistency from treatment plan file to 3D printed mold to final aligner. Biocompatibility testing (ISO 10993-1) confirmed non-cytotoxic, non-sensitizing, and non-irritating properties. Physical attributes tested per ASTM D1505, D790, and D882.
Technological Characteristics
Thermoformed copolyester plastic aligners. Principle: sequential application of corrective force via incremental tray changes. Physical properties tested per ASTM D1505 (density), D790 (flexural modulus/strength), D882 (tensile strength/modulus). Software-assisted design and manufacturing. Biocompatible per ISO 10993-1.
Indications for Use
Indicated for patients with permanent dentition (second molars) requiring correction of minor maloccluded anterior teeth discrepancies.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 19, 2019
Shatkin F.I.R.S.T. LLC Kimberly Wiesmore Controller 2495 Kensington Avenue Amherst, New York 14226
Re: K182388
Trade/Device Name: Krystal Klear Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: November 18, 2019 Received: November 19, 2019
Dear Kimberly Wiesmore:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182388
Device Name Krystal Klear Aligners System
#### Indications for Use (Describe)
The Krystal Klear Aligners System is intended to correct minor discrepancies in the alignment of maloccluded anterior teeth on patients with permanent dentition (second molars) by moving the teeth with a progressive series of clear thin, thermoformed plastic aligners, fabricated in stages to gradually align teeth over a period of several months. The aligners are completely removable by the patient and may be discontinued at any time.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21CFR 807.92.
| Submitter: | Shatkin F.I.R.ST. LLC<br>2494 Kensington Avenue<br>Amherst, NY 14226<br>Not Currently FDA Registered |
|-------------------------|------------------------------------------------------------------------------------------------------|
| Primary Contact: | Kimberley Wiesmore, CFO<br>Phone: 716.839.2959 x 149<br>kwiesmore@shatkinfirst.com |
| Secondary Contact: | Todd Shatkin, DDS<br>Phone: 716.839.2959 |
| Date Prepared: | 26 August 2018 (Original)<br>10 November 2019 (Revised) |
| Proprietary Trade Name: | Krystal Klear Aligner System |
| Device Name: | Sequential Aligner |
| Common Name: | Clear Braces / Aligners |
| Classification Name: | Orthodontic Plastic Bracket |
| Regulation Number: | 852.5470 |
| Product Code: | NXC |
| Classification Panel: | Dental Products Panel 76 |
| Classification: | Medical Device, Class II |
### Predicate Devices:
Clear Image Aligners, Specialty Appliance Works, K071970 (Primary Predicate) Clear Correct, Clear Correct Inc., K082556 (Reference Device) Invisalign, Align Technology, K981095 (Reference Device)
# Device Description:
The Krystal Klear device is fabricated of clear, thin, thermoformed copolyester plastic in a sequential series to progressively reposition teeth. Corrective force to straighten the teeth is delivered via minor changes into a position in each subsequent aligner.
{4}------------------------------------------------
#### Indications for Use:
The Krystal Klear Aligners System is intended to correct minor discrepancies in the alignment of maloccluded anterior teeth on patients with permanent dentition (second molars) by moving the teeth with a progressive series of clear thin, thermoformed plastic aligners, fabricated in stages to gradually align teeth over a period of several months. The aligners are completely removeable by the patient and may be discontinued at any time.
### Technological Characteristics:
Treatment of tooth malocclusions via a series of plastic appliances that incrementally moves teeth to a desired end-state is the technological principle for both the subject and predicate devices. A comparison between technological characteristics of the Krystal Klear Aligner System and that of legally marketed predicate devices has been performed. The results of this comparison demonstrate that the design, technology, materials, and composition of the Krystal Klear Aligner System are substantially equivalent to the predicate devices.
A dental health care professional (e.g. orthodontist or dentist), prescribes the Krystal Klear Aligner system based on an assessment of the patient's teeth, determines a course of treatment with the system, takes molds of the patient's teeth, and completes a prescription form. The molds and prescription are created by Krystal Klear. Utilizing standard dental software used for tooth alignment, Masterpieces (MAP) then designs a series of plastic trays intended to gradually realign the patient's teeth in accordance with the physician's prescription. The prescribing physician reviews and approves the treatment plan before the molds are produced. Once approved, MAP produces the trays, which are formed of clear, thin, thermoformed, copolyester plastic. The trays sent back to the dental health care professional, who then provides them to the patient, confirming fit and design. Over a period of months, additional trays are provided sequentially to the patient by the dental health professional to gradually move the target teeth to the desired position. The dental care professional monitors treatment from the moment the first aligner is delivered to when the final aligner is delivered. The trays are held in place by pressure and can be removed by the patient at any time. This technology is essentially identical to that used by a number of sequential alignment systems, including the predicates referenced in Table 1 below.
{5}------------------------------------------------
#### Mechanism of Action:
The mechanism of action is similar to the predicate devices and supports a determination of substantial equivalence. Orthodontic tooth movement occurs through forces applied by the appliance to the dentition as each tooth follows the programmed displacement based on a dental health professional's prescription.
#### Performance Testing:
Bench testing was performed to ensure the accuracy of the final product carried through the entire process from the initial scan through treatment planning and manufacturing of the model to the final thermoformed aligner. Functional characteristics tested and the standard by which they were tested include the following, all pre-established specifications and acceptance criteria:
- Fit of the aligner against the treatment plan as determined by dental health . professional
- Process consistency from the treatment plan file to 3D printed mold to final . aligner was measured against internal acceptance criteria derived by considering all sources of variation
The acceptable results observed during the verification and validation testing that has been conducted on these devices has shown that they are capable of performing to their stated intended use and specification requirements.
Additional physical attributes tested, and the standard by which they were tested are below, which met acceptance criteria:
| Attribute | ASTM Method |
|----------------------------------|-------------|
| Density | D1505 |
| Flexural Modulus | D790 |
| Flexural Strength | D790 |
| Tensile Strength | D882 |
| Tensile Modulus of<br>Elasticity | D882 |
{6}------------------------------------------------
Further testing was not performed given there is sufficient information available from the scientific literature and from the predicate devices to demonstrate that sequential aligners provide reasonable assurance of safety and effectiveness.
### Biocompatibility Testing:
The biological evaluation for the device was conducted in accordance with the US FDA Docket Number FDA-2013-D-0350, "Use of International Standard ISO 10993-1, Biological evaluation and testing within a risk management process – Guidance for Industry and Food and Drug Administration Staff" as recognized by FDA. The aligner is considered mucosal membrane contacting for a duration of greater than 30 days. The battery of testing included the following tests:
- . Cytotoxicity
- Sensitization ●
- Irritation ●
The results of the testing met the requirements of the study protocols and the material is considered non-cytotoxic, non-sensitizing, and is not an intracutaneous irritant. The results of the studies further support the determination of substantial equivalence.
#### Software Consideration
There are two areas in the Krystal Klear Aligner System process where software is used; they are the ordering process and treatment planning. The software usage in these areas is considered of moderate concern. Software testing has been performed to verify the safety and effectiveness of the Krystal Klear Aligner System, which supports a substantial equivalence decision.
### Substantial Equivalence Comparison:
The following table compares the Krystal Klear Aligner System to the predicate devices, Clear Image Aligners (primary predicate), and Clear Correct (reference device) & the Align System (reference device), with respect to intended use, technological characteristics, and principles of operation.
The aligner has the same patient contact, similar materials, and the same manufacturing processes as the legally marketed device on the market. As the material is slightly different, biocompatibility testing was performed to verify safety and effectiveness for the Krystal Klear Aligner system.
{7}------------------------------------------------
## Table 1. Predicate Device Information Comparison
| | KrystalKlear<br>(Submission<br>Device) | Clear Image<br>Aligners<br>(Primary Predicate) | Clear Correct<br>(Reference Device) | Invisalign/Align<br>System<br>(Reference Device) |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | (Submission<br>Device) | (Primary Predicate) | (Reference Device) | (Reference Device) |
| 510(k) Number | K182388 | K071970 | K082556 | K981095 |
| Classification<br>Number & Product<br>Code | 852.5470<br>NXC | 852.5470<br>NXC | 852.5470<br>NXC | 852.5470<br>NXC |
| Indications for Use | The Krystal Klear<br>System is intended to<br>correct minor<br>discrepancies in the<br>alignment of<br>maloccluded teeth on<br>patients with<br>permanent dentition<br>(second molars) by<br>moving the teeth with<br>a progressive series<br>of thermoformed<br>plastic aligners,<br>fabricated in stages<br>to gradually align<br>teeth over a period of<br>several months. The<br>aligners are<br>completely removable<br>by the patient and<br>may be discontinued<br>at any time. | The Clear Image TM<br>Aligners System is<br>intended to correct<br>minor discrepancies<br>in the alignment of<br>maloccluded anterior<br>teeth on patients with<br>permanent dentition<br>(second molars) by<br>moving the teeth with<br>a progressive series<br>of clear thin,<br>thermoformed plastic<br>aligners, fabricated in<br>stages to gradually<br>align teeth over a<br>period of several<br>months. The aligners<br>are completely<br>removable by the<br>patient and may be<br>discontinued at any<br>time. | The ClearCorrect<br>System is indicated<br>for the treatment of<br>tooth malocclusion in<br>patients with<br>permanent dentition<br>(i.e. all second<br>molars). The<br>ClearCorrect System<br>positions teeth by<br>way of continuous<br>gentle force. | The Align System is<br>indicated for the<br>treatment of tooth<br>malocclusion in<br>patients with<br>permanent dentition<br>(i.e. all second<br>molars). The Align<br>System positions<br>teeth by way of<br>continuous gentle<br>force. |
| Mode of Action | Alignment of teeth by<br>sequential use of<br>preformed plastic<br>trays. | Alignment of teeth by<br>sequential use of<br>preformed plastic<br>trays. | Alignment of teeth by<br>sequential use of<br>preformed plastic<br>trays. | Alignment of teeth by<br>sequential use of<br>preformed plastic<br>trays. |
| Material | Thermoformed<br>Copolyester | Thermoformed<br>Polycarbonate | Thermoformed<br>Polycarbonate | Thermoformed<br>Polycarbonate |
| OTC or Rx | Rx | Rx | Rx | Rx |
{8}------------------------------------------------
## Substantial Equivalence Conclusion:
The conclusion drawn from the data included in this submission demonstrates that the Krystal Klear Aligner System is as safe, effective, and substantially equivalent to the predicate devices in indications for use, design, technological characteristics, mechanism of action, performance, materials, and biocompatibility.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.