Small Bone Phantom® Intramedullary Nail System; TTC Phantom® Intramedullary Nail System
K182307 · Paragon 28, Inc. · HSB · Jan 14, 2019 · Orthopedic
Device Facts
Record ID
K182307
Device Name
Small Bone Phantom® Intramedullary Nail System; TTC Phantom® Intramedullary Nail System
Applicant
Paragon 28, Inc.
Product Code
HSB · Orthopedic
Decision Date
Jan 14, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Small Bone Phantom® Intramedulary Nail System is indication and fixation and fixation of the small bones of the feet and ankle for the treatment of fractures, osteotomies, pseudarthroses and malunions by revision. joint fusion or reconstruction procedures. The TTC Phantom® Intramedullary Nail system is intended for tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of the hindfoot and ankle including the transverse tarsal joints coupling the mid-foot to the hindfoot. Examples of specific indications include: - Post-traumatic or degenerative arthritis - Previously infected arthrosis - Revision of failed ankle arthrodesis - Revision of failed total ankle arthroplasty - Talar deficiency conditions such as avascular necrosis of the talus (requiring tibiocalcaneal arthrodesis) - Neuromuscular deformity or other neuromuscular disease with severe deformity or instability of the ankle - Rheumatoid arthritis - Osteoarthritis - Nonunions or pseudarthrosis of hindfoot and distal tibia - Trauma (severe or malunited tibial pilon fracture) - Charcot foot (neuroarthropathy) - Severe end-stage degenerative arthritis - Instability and skeletal defects after tumor resection - Pantalar arthrodesis - Severe foot/ankle deformity
Device Story
System comprises intramedullary nails, locking screws, and threaded pegs; used for bone fixation and stabilization in foot/ankle surgery. Implants inserted into bone to secure construct; varying lengths accommodate patient anatomy. Used by orthopedic surgeons in clinical/OR settings. Provides mechanical stabilization to facilitate bone healing/fusion. Benefits include restoration of alignment and stability in complex foot/ankle pathologies.
Clinical Evidence
Bench testing only. Performance characterized via four-point bend testing per ASTM F1264. No clinical data required to support safety and effectiveness.
Technological Characteristics
Implants: Titanium Alloy (ASTM F136). Instrumentation: Titanium alloy, stainless steel, anodized aluminum, polymer. Mechanical fixation via intramedullary nails, locking screws, and threaded pegs. No software or electronic components.
Indications for Use
Indicated for patients requiring stabilization/fixation of small bones of feet/ankle (fractures, osteotomies, pseudarthroses, malunions, joint fusion, reconstruction) or tibiotalocalcaneal arthrodesis (arthritis, failed arthrodesis/arthroplasty, talar deficiency, neuromuscular deformity, nonunions, trauma, Charcot foot, tumor resection).
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
Phantom™ Small Bone Intramedullary Nail System (K170693)
Piccolo Composite® Nailing System – Ankle Arthrodesis (K123810)
SBi Anatomic Ankle Arthrodesis Interlocking Nail System (K130147)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 14, 2019
Paragon 28, Inc. Mr. Eric Lintula Director of Regulatory Affairs 4B Inverness Ct. E., Ste 280 Englewood, Colorado 80112
Re: K182307
Trade/Device Name: Small Bone Phantom® Intramedullary Nail System; TTC Phantom® Intramedullarv Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB. KTW. HWC Dated: December 20, 2018 Received: December 31, 2018
Dear Mr. Lintula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Peter G. Allen Digitally signed by Peter
G. Allen -s Date: 2019.01.14 13:28:37 -5 -05'00'
FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182307
#### Device Name
Small Bone Phantom® Intramedullary Nail System
Indications for Use (Describe)
The Small Bone Phantom® Intramedulary Nail System is indication and fixation and fixation of the small bones of the feet and ankle for the treatment of fractures, osteotomies, pseudarthroses and malunions by revision. joint fusion or reconstruction procedures.
Type of Use (Select one or both, as applicable)
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------|-----------------------------------------------|
|------------------------------------------------|-----------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known)
K182307
#### Device Name
TTC Phantom® Intramedullary Nail System
#### Indications for Use (Describe)
The TTC Phantom® Intramedullary Nail system is intended for tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of the hindfoot and ankle including the transverse tarsal joints coupling the mid-foot to the hindfoot. Examples of specific indications include:
- · Post-traumatic or degenerative arthritis
- · Previously infected arthrosis
- · Revision of failed ankle arthrodesis
- · Revision of failed total ankle arthroplasty
- · Talar deficiency conditions such as avascular necrosis of the talus (requiring tibiocalcaneal arthrodesis)
- · Neuromuscular deformity or other neuromuscular disease with severe deformity or instability of the ankle
- Rheumatoid arthritis
- · Osteoarthritis
- · Nonunions or pseudarthrosis of hindfoot and distal tibia
- · Trauma (severe or malunited tibial pilon fracture)
- · Charcot foot (neuroarthropathy)
- · Severe end-stage degenerative arthritis
- · Instability and skeletal defects after tumor resection
- Pantalar arthrodesis
- · Severe foot/ankle deformity
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
| <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Date: | October 3rd, 2018 |
|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Paragon 28, Inc.<br>4B Inverness Ct. E., STE 280<br>Englewood, Colorado 80112<br>Phone: (888) 728-1888<br>Fax: (888) 728-1220 |
| Sponsor contact: | Eric Lintula, Director of Regulatory Affairs |
| Trade Names: | Small Bone Phantom® Intramedullary Nail System<br>TTC Phantom® Intramedullary Nail System |
| Small Bone Phantom®<br>Regulatory Class: | Class II |
| TTC Phantom®<br>Regulatory Class: | Class II |
| Small Bone Phantom®<br>Regulation, Product<br>Code, Classification,<br>and Common Name: | 888.3030, KTW, Appliance, fixation, nail/blade/plate<br>combination, single component<br>888.3040, HWC, Smooth or threaded metallic bone fixation<br>fastener, bone screw |
| TTC Phantom®<br>Regulation, Product<br>Code, Classification,<br>and Common Name: | 888.3020, HSB, Rod, Fixation, Intramedullary And<br>Accessories<br>888.3040, HWC, Smooth or threaded metallic bone fixation<br>fastener, bone screw |
| Small Bone Phantom®<br>Device Description: | The Paragon 28® Small Bone Phantom® Intramedullary Nail<br>System is comprised of small bone intramedullary nails,<br>locking screws and threaded pegs. The Small Bone<br>Phantom® nails are offered in a variety of lengths to<br>accommodate variations in patient anatomy. The Small Bone<br>Phantom® threaded pegs and locking screws insert through<br>the intramedullary nail to secure the construct. These are<br>offered in varying lengths to accommodate the anatomical<br>fixation required. The system includes instruments for<br>implantation. |
| TTC Phantom®<br>Device Description: | The Paragon 28® TTC Phantom® Intramedullary Nail<br>System is comprised of intramedullary nails, screws and<br>accessory components. The TTC Phantom® nails are offered<br>in a variety of sizes and lengths to accommodate variations<br>in patient anatomy. The TTC Phantom® screws insert<br>through the intramedullary nail to secure the construct.<br>These are offered in varying lengths to accommodate the<br>anatomical fixation required |
| Small Bone Phantom®<br>Materials: | The Small Bone Phantom® Intramedullary Nail System<br>implants are made from Titanium Alloy (ASTM F136). The<br>instrumentation is manufactured from medical grades of<br>titanium alloy, stainless steel, anodized aluminum and<br>polymer. |
| TTC Phantom®<br>Materials: | The TTC Phantom® Intramedullary Nail System implants are<br>made from Titanium Alloy (ASTM F136). The<br>instrumentation is manufactured from medical grades of<br>titanium alloy, stainless steel, anodized aluminum and<br>polymer. |
| Small Bone Phantom®<br>Indications for Use: | The Small Bone Phantom® Intramedullary Nail System is<br>indicated for use in stabilization and fixation of the small<br>bones of the feet and ankle for the treatment of fractures,<br>osteotomies, nonunions, pseudarthroses and malunions by<br>revision, joint fusion or reconstruction procedures. |
| TTC Phantom®<br>Indications for Use: | The TTC Phantom® Intramedullary Nail system is intended<br>for tibiotalocalcaneal arthrodesis (fusion) and to provide<br>stabilization of the hindfoot and ankle including the<br>transverse tarsal joints coupling the mid-foot to the hindfoot.<br>Examples of specific indications include:<br>Post-traumatic or degenerative arthritis Previously infected arthrosis Revision of failed ankle arthrodesis Revision of failed total ankle arthroplasty Talar deficiency conditions such as avascular<br>necrosis of the talus (requiring tibiocalcaneal<br>arthrodesis) Neuromuscular deformity or other neuromuscular<br>disease with severe deformity or instability of the<br>ankle Rheumatoid arthritis Osteoarthritis Nonunions or pseudarthrosis of hindfoot and distal<br>tibia Trauma (severe or malunited tibial pilon fracture) Charcot foot (neuroarthropathy) Severe end-stage degenerative arthritis Instability and skeletal defects after tumor resection Pantalar arthrodesis Severe foot/ankle deformity |
| Small Bone Phantom®<br>Primary Predicate: | K170693, Phantom™ Small Bone Intramedullary Nail<br>System |
| TTC Phantom®<br>Primary Predicate: | K171376, DynaNail™ Ankle Arthrodesis Nail |
| TTC PhantomⓇ<br>Additional Predicates: | K120419, Dyna Locking Ankle Nail™<br>K123810, Piccolo CompositeⓇ Nailing System – Ankle Arthrodesis<br>K130147, SBi Anatomic Ankle Arthrodesis Interlocking Nail System |
| Small Bone PhantomⓇ<br>Comparison to<br>Predicate Indications: | The subject Small Bone PhantomⓇ Intramedullary Nail System is a modification of the unmodified Phantom™ Small Bone Intramedullary Nail System, and therefore has the same indications. |
| TTC PhantomⓇ<br>Comparison to<br>Predicate Indications: | The subject TTC PhantomⓇ Intramedullary Nail System and predicate DynaNail™ Ankle Arthrodesis Nail, Dyna Locking Ankle Nail™, Piccolo CompositeⓇ Nailing System for Ankle Arthrodesis, and SBi Anatomic Ankle Arthrodesis Interlocking Nail System devices are intended to be used for tibiotalocalcaneal arthrodesis. All devices are indicated for use in the hindfoot and ankle. All indications for the subject device are within the indications of the predicate devices. |
| Small Bone PhantomⓇ<br>Comparison to<br>Predicate<br>Technological<br>Characteristics: | The subject Small Bone PhantomⓇ intramedullary nail and components possess the same technological characteristics as the predicate device. These include:<br>• performance,<br>• basic design,<br>• material, manufacturing and<br>• sizes (dimensions are comparable to those offered by the predicate systems).<br>Therefore, the fundamental scientific technology of the subject Small Bone PhantomⓇ intramedullary nail and components is similar to the previously cleared device. |
| TTC PhantomⓇ<br>Comparison to<br>Predicate<br>Technological<br>Characteristics: | The subject TTC PhantomⓇ intramedullary nail and components possess the same technological characteristics as the predicate devices. These include:<br>• performance,<br>• basic design,<br>• material, manufacturing and<br>• sizes (dimensions are comparable to those offered by the predicate systems).<br>Therefore, the fundamental scientific technology of the subject TTC PhantomⓇ intramedullary nail and components is similar to previously cleared devices. |
| Small Bone PhantomⓇ<br>Performance Data: | Engineering analysis is presented to provide evidence that the subsequent performance of the Small Bone PhantomⓇ Intramedullary Nail System is not adversely affected by the design modifications.<br>The results of the analysis demonstrated the subject devices are substantially equivalent to the Phantom™ Small Bone Intramedullary Nail predicate |
| TTC Phantom®<br>Performance Data: | All necessary testing has been performed on representative<br>TTC Phantom® Intramedullary Nail System components.<br>The device performance was characterized via four-point<br>bend testing per ASTM F1264. Clinical data are not needed<br>to support the safety and effectiveness of the subject devices.<br>Engineering analysis is presented to assure substantial<br>equivalence of the TTC Phantom® Intramedullary Nail<br>System to its predicates.<br>The results of the analysis demonstrated the subject devices<br>are substantially equivalent to the Dyna Locking Ankle<br>NailTM and Piccolo Composite® Nailing System predicates. |
| Conclusion: | Engineering analysis demonstrates the substantial<br>equivalence of the Small Bone Phantom® Intramedullary<br>Nail System to the PhantomTM Small Bone Intramedullary<br>System and the TTC Phantom® Intramedullary Nail System<br>to the Dyna Locking Ankle NailTM and Piccolo Composite®<br>Nailing System. Therefore, the Phantom® Intramedullary<br>Nail Systems are substantially equivalent to the predicate<br>devices with respect to their indications for use, technical<br>characteristics, and function. |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.