K182220 · Curbell Medical Products, Inc. · DSA · Nov 7, 2018 · Cardiovascular
Device Facts
Record ID
K182220
Device Name
Curbell branded Patient Monitoring Cables
Applicant
Curbell Medical Products, Inc.
Product Code
DSA · Cardiovascular
Decision Date
Nov 7, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2900
Device Class
Class 2
Indications for Use
Curbell patient cables are used to connect electrodes and/or sensors placed at the appropriate sites on the patient to a monitoring device for general monitoring and/or diagnostic evaluation by a healthcare professional.
Device Story
Curbell Patient Monitoring Cables consist of shielded/unshielded leadwires and trunk cables; connect electrodes, sensors, or catheters to patient monitors; facilitate ECG, EKG, and SpO2 signal transmission. Used in clinical settings by healthcare professionals. Cables feature copper wiring with PVC or thermoplastic polyurethane jackets; various connectors (snap, pinch, Sure Lock). Device acts as passive conduit for physiological signals; enables monitoring/diagnostic evaluation. Benefits include reliable signal transmission for patient monitoring.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ANSI/AAMI EC53:2013 (noise, flex life, tensile strength, connector retention, contact resistance, dielectric withstand) and IEC 60601-1:2005 (patient leads). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Shielded/unshielded copper wiring; PVC or thermoplastic polyurethane jackets. Connectors: snap, pinch, or Sure Lock. Lengths: 7-20ft (trunk), 18-120" (leadwires). Standards: ANSI/AAMI EC53:2013, IEC 60601-1:2005 (3rd Ed). Biocompatibility: ISO 10993-5, ISO 10993-10. Non-sterile. Passive electrical connectivity.
Indications for Use
Indicated for patients requiring ECG, EKG, and SpO2 monitoring via connection of electrodes, sensors, or catheters to a monitoring device for diagnostic evaluation by a healthcare professional.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
Shenzhen Med-link Electronics Tech Co., Ltd. patient cables and lead wires (K120010)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
Curbell Medical Products, Inc. Michael Winter Sr. Director, Quality and Regulatory Affairs 20 Centre Drive Orchard Park, New York 14127
Re: K18220
Trade/Device Name: Curbell Patient Monitoring Cables Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer And Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: August 10, 2018 Received: October 16, 2018
## Dear Michael Winter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Todd D. Courtney -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182220
Device Name Curbell Patient Monitoring Cables
Indications for Use (Describe)
Curbell patient cables are used to connect electrodes and/or sensors placed at the appropriate sites on the patient to a monitoring device for general monitoring and/or diagnostic evaluation by a healthcare professional.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for CURBELL MEDICAL. The word "CURBELL" is written in a large, bold, blue font. Below "CURBELL" is the word "MEDICAL" in a smaller, green font. The logo is simple and professional.
### Special 510(k) Summary
| 1. | Company: | Curbell Medical Products, Inc.<br>20 Centre Drive<br>Orchard Park, NY 14127<br>716-667-3377 x419 |
|----|----------|--------------------------------------------------------------------------------------------------|
|----|----------|--------------------------------------------------------------------------------------------------|
- 2. Contact: Michael Winter
- 2. Date Prepared: October 10, 2018
- 3. Trade Name: Curbell Patient Monitoring Cables
- 4. Common Name: Patient Monitoring Cables SAC-JAJA-007-000, Length: 6' 7'' Models: SAC-JAJA-010-000, Length 10' SAC-JAJA-025-000, Length: 25'
- న్. Classification Name: Patient Transducer and Electrode Cable (including connector) (21CFR870.2900, Product Code DSA)
- Primary Predicate: Curbell Patient Monitoring Cables: 510(k) 6. Predicate device(s): K081762 Secondary Predicate for Expanded Use (SPO2 Monitoring): Shenzhen Med-link Electronics Tech Co., Ltd. patient cables and lead wires: K120010
- 7. Curbell Patient Monitoring Cables comprise cables and lead Device description: wires that connect from sensors and/or electrodes to a monitor for the purpose of conducting ECG diagnosis, monitoring, and SPO2 monitoring. As per 21CFR801.109, this device is a prescription device.
- 8. Intended use: Curbell patient cables are used to connect electrodes and/or sensors placed at the appropriate sites on the patient to a monitoring device for general monitoring and/or diagnostic evaluation by a healthcare professional.
- 9. Technological characteristics to predicate device: Curbell Patient Monitoring Cables are substantially equivalent to the primary predicate device, Curbell Medical Products, Inc. (K081762). Additionally, they are substantially equivalent for expanded use (SPO2 Monitoring) to the secondary predicate device. Shenzhen Medlink Electronics Tech Co., Ltd. patient cables and lead wires: (K120010). This logic follows "Multiple Predicates Example 2", page 11, FDA Guidance Document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]."
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for CURBELL MEDICAL. The word "CURBELL" is written in a large, bold, blue font. Below it, the word "MEDICAL" is written in a smaller, green font. The logo is simple and modern.
| | Proposed Curbell<br>Patients Cables | Current Curbell<br>Patient Cables (Primary<br>Predicate) | Shenzhen Med-link Electronics<br>Tech Co., Ltd.<br>(Secondary Predicate – SPO2<br>Monitoring) |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K182220 | K081762 | K120010 |
| Intended use | Curbell patient cables<br>are used to connect<br>electrodes and/or<br>sensors placed at the<br>appropriate sites on the<br>patient to a monitoring<br>device for general<br>monitoring and/or<br>diagnostic evaluation<br>by a healthcare<br>professional. | Curbell patient cables<br>are used to connect<br>electrodes and/or<br>sensors placed at the<br>appropriate sites on the<br>patient to a monitoring<br>device for general<br>monitoring and/or<br>diagnostic evaluation<br>by a healthcare<br>professional. | The Shenzhen Med-link cables<br>and lead wires are intended to<br>be used with ECG, EKG, SPO2<br>and Invasive Blood Pressure<br>monitoring devices. The patient<br>cables and lead wires are used<br>to connect electrodes, catheters,<br>and/or sensors placed<br>at appropriate sites on the<br>patient to a monitoring device<br>for general monitoring and/or<br>diagnostic evaluation by health<br>care professional. |
| Regulation | 21CFR870.2900 | 21CFR870.2900 | 21CFR870.2900 |
| Product code | DSA | DSA | DSA |
| Common Name | Patient cables and<br>leadwires | Patient cables and<br>leadwires | Patient cables and<br>leadwires |
| Proprietary Name | Curbell branded | Curbell branded | NA |
| Design/Materials | Shielded and unshielded<br>patient leadwires and<br>shielded patient trunk<br>cables with electrode<br>connectors including<br>snap, pinch, or Sure<br>Lock electrode clip.<br>Trunk cable lengths<br>include - 7 to 20ft.<br>Leadwire lengths from<br>18" to 120". Wire<br>material is comprised of<br>shielded and unshielded<br>copper with PVC or<br>thermoplastic<br>polyurethane jacket. | Shielded and unshielded<br>patient leadwires and<br>shielded patient trunk<br>cables with electrode<br>connectors including<br>snap, pinch, or Sure<br>Lock electrode clip.<br>Trunk cable lengths<br>include - 7 to 20ft.<br>Leadwire lengths from<br>18" to 120". Wire<br>material is comprised of<br>shielded and unshielded<br>copper with PVC or<br>thermoplastic<br>polyurethane jacket. | ECG and EKG Cables with<br>various connectors (monitor,<br>trunk/lead wire, electrode<br>grabber & snapper). SPO2<br>adapter cables with<br>various connectors for connect<br>for the OEM Instrument<br>connector of SPO2 monitor and<br>SPO2 probe.<br>Cable is set at various specified<br>standard lengths.<br>Wire material is comprised of<br>shielded and unshielded copper<br>with PVC or thermoplastic<br>polyurethane jacket. |
{6}------------------------------------------------
Image /page/5/Picture/2 description: The image contains the logo for CURBELL MEDICAL. The word "CURBELL" is in large, bold, blue letters. Underneath "CURBELL" is the word "MEDICAL" in smaller, green letters. The logo is simple and professional.
| MEDICAL | | | |
|--------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Biocompatibility | Meets ISO 10993-5<br>Cytotoxicity,<br>ISO 10993-10<br>Sensitization, and<br>ISO 10993-10 Irritation. | Meets ISO 10993-5<br>Cytotoxicity,<br>ISO 10993-10<br>Sensitization, and<br>ISO 10993-10 Irritation | Meets ISO 10993-5<br>Cytotoxicity,<br>ISO 10993-10 Sensitization,<br>and<br>ISO 10993-10 Irritation |
| Sterility | Non sterile | Non sterile | Non sterile |
| Electrical<br>performance<br>testing | ANSI/AAMI EC53:<br>2013 | ANSI/AAMI EC53:<br>1995 (revised 2001) | ANSI/AAMI EC53:1995<br>(revised 2001) |
| Electrical Safety<br>testing | ANSI/AAMI EC53:<br>2013, IEC 60601-1:<br>2005, 3rd Edition<br>standard - Part 8.5.2.3<br>Patient Leads | ANSI/AAMI EC53:<br>1995 (revised 2001) | ANSI/AAMI EC53:1995<br>(revised 2001), IEC 60601-<br>1:1998;Am1, A2:1995 |
| Connector<br>Retention Force | ANSI/AAMI EC53:<br>2013 | ANSI/AAMI EC53:<br>1995 (revised 2001) | ANSI/AAMI EC53:1995<br>(revised 2001) |
The proposed Curbell Patient Monitoring Cables have the following similarities to the current Curbell primary predicate device:
- Same indications for use
- Same manufacturing process ●
- Similar monitor connector and similar yoke connector ●
- Same fundamental scientific technology .
The proposed Curbell Patient Monitoring Cables have the following similarities to the Shenzhen Med-link secondary predicate device:
- Same intended use (patient monitoring).
- Included indication for use (SPO2 monitoring) ●
- Similar/equivalent construction ●
- Same fundamental scientific technology ●
The proposed Curbell Patient Monitoring Cables have the following differences to the current Curbell primary predicate device:
- · Expanded Intended Use (SPO2 monitoring)
The proposed Curbell Patient Monitoring Cables have the following differences to the Shenzhen Med-link secondary predicate device:
- Color
- Length variations .
Curbell Reports that provide data and analysis to support the claims provided within:
- Engineering Verification and Validation Protocols .
- I Med-link device (secondary predicate)
{7}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for CURBELL MEDICAL. The word "CURBELL" is written in large, bold, blue letters. Below "CURBELL" is the word "MEDICAL" in smaller, green letters. The logo is simple and professional.
- ANSI/AAMI EC53:2013 I
- 5.3.2 Cable and Lead Wire 트 Noise
- . 5.3.3 - Flex Life of Trunk Cable Flex Relief
- . 5.3.4 - Tensile Strength of Cable Connections
- I 5.3.5 - Number of Connector Mating/Unmating Cycles
- 트 5.3.6 - Connector Retention Force
- . 5.3.7 - Contact Resistance
- 트 5.3.9 - Dielectric Withstand Voltage
- IEC 60601-1:2005 3rd Edition standard 트
- = Part 8.5.2.3 Patient Leads
- Cable Mate
- Aux Device Compatibility
- Cable Flex Life
#### ■ Proposed Curbell device
- 트 ANSI/AAMI EC53:2013
- 5.3.2 Cable and Lead Wire . Noise
- . 5.3.3 - Flex Life of Trunk Cable Flex Relief
- . 5.3.4 - Tensile Strength of Cable Connections
- . 5.3.5 - Number of Connector Mating/Unmating Cycles
- I 5.3.6 - Connector Retention Force
- I 5.3.7 - Contact Resistance
- 5.3.9 Dielectric Withstand 트 Voltage
- IEC 60601-1:2005 3rd Edition standard .
- = Part 8.5.2.3 Patient Leads
- Cable Mate
- Aux Device Compatibility
- Cable Flex Life
- Current Curbell device (primary predicate)
- ANSI/AAMI EC53:2013
- 5.3.2 Cable and Lead Wire . Noise
{8}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for Curbell Medical. The word "CURBELL" is written in a large, bold, blue font. Below the word "CURBELL" is the word "MEDICAL" in a smaller, green font. The logo is simple and professional.
- 5.3.3 - Flex Life of Trunk Cable Flex Relief
- 트 5.3.4 - Tensile Strength of Cable Connections
- 5.3.5 - Number of Connector Mating/Unmating Cycles
- I 5.3.6 - Connector Retention Force
- 5.3.7 Contact Resistance ■
- I 5.3.9 - Dielectric Withstand Voltage
- IEC 60601-1:2005 3rd Edition standard I
- . Part 8.5.2.3 Patient Leads
- Cable Mate I
- 트 Aux Device Compatibility
- I Cable Flex Life
- . Engineering Part Evaluation
- Proposed Curbell device ■
- I Verification of inspection to revision controlled drawings
- Inspection Report
- Proposed Curbell device ■
- 트 Dimensional inspection
- Biocompatibility (ISO 10993) .
- Proposed Curbell device 트
- I Wire Jacket Biocompatibility
- Overmold Biocompatibility
- I Connector Biocompatibility
- Curbell Design Inputs document: SPO2 Adapter Cable .
- Curbell Risk Analysis document: SPO2 Adapter Cable ●
- 10. Nonclinical performance data:
Curbell Patient Monitoring Cables have been assessed to the following performance standards and special controls:
- IEC 60601-1:2005, Section 8.5.2.3 - Patient Leads
- ANSI/AAMI EC53:2013 ECG Trunk Cables and Patient LeadWires
- 11. Clinical data: Not Applicable
- 12. Conclusion from nonclinical testing: The summary above illustrates that there are no different questions of safety or effectiveness of the proposed Curbell Patient Monitoring Cables compared to the current Curbell primary predicate device. Additionally, with respect to expanded use (SPO2 monitoring), the proposed Curbell
{9}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for Curbell Medical. The word "CURBELL" is written in large, bold, blue letters. Below the word "CURBELL" is the word "MEDICAL" in smaller, green letters. The logo is simple and professional.
Patient Monitoring Cables device is substantially equivalent to the Curbell primary predicate device and the Shenzhen Med-link secondary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.