The Encompass™ SRS Headframe provides noninvasive stereotactic head and neck immobilization by using a patient specific thermoplastic mask that conforms to the patient's features to provide accurate, reproducible positioning, repositioning, and immobilization. The Encompass™ SRS Headframe allows the patient to undergo diagnostic imaging in the same position as that of the treatment position enabling more accurate radiation therapy.
Device Story
Encompass™ SRS Headframe is a noninvasive stereotactic head and neck immobilization device. It functions as a removable attachment for patient positioning devices that accept S-type thermoplastic masks. The device uses a patient-specific thermoplastic mask (Encompass™ SRS Fibreplast® System) to conform to patient features, ensuring sub-millimetric accuracy for positioning and repositioning. Used in radiation oncology clinics, it allows patients to maintain the same position during diagnostic imaging and subsequent radiation therapy. Healthcare professionals, including radiation therapists, operate the device. By limiting patient motion to less than 1 mm, it facilitates precise radiation delivery, benefiting patients undergoing SRS for cancer or non-cancerous conditions like trigeminal neuralgia. The device is manufactured from composite materials and includes MR-compatible variants.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing, including hardware verification, motion studies for accuracy validation, ergonomics assessments, aluminum equivalence verification, and compatibility testing with optical tracking systems. Testing confirmed the device meets sub-millimetric immobilization requirements.
Technological Characteristics
Non-sterile, reusable device made of non-conducting composite materials. MR-compatible variants are MR Safe per ASTM F2503-13. Designed as a removable attachment for S-type thermoplastic mask positioning systems. Provides sub-millimetric immobilization.
Indications for Use
Indicated for noninvasive stereotactic head and neck immobilization in patients undergoing diagnostic imaging and radiation therapy, including stereotactic radiosurgery (SRS) for radiation oncology and treatment of noncancerous conditions such as trigeminal neuralgia.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 4, 2018
Qfix Ms. Alexandra Low Smythe Regulatory Affairs Specialist 440 Church Road AVONDALE, PA 19311
### Re: K182189
Trade/Device Name: Encompass™ SRS Headframe : Encompass MR SRS Headframe Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: September 7, 2018 Received: September 10, 2018
Dear Ms. Smythe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182189
Device Name Encompass™ SRS Headframe
#### Indications for Use (Describe)
The Encompass™ SRS Headframe provides noninvasive stereotactic head and neck immobilization by using a patient specific thermoplastic mask that conforms to the patient's features to provide accurate, reproducible positioning. repositioning, and immobilization. The Encompass™ SRS Headframe allows the patient to undergo diagnostic imaging in the same position as that of the treatment position enabling more accurate radiation therapy.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Q-Fix. The logo is a red letter Q with the word FIX in black letters inside the Q. The Q is stylized with a thick line and a curved tail. The word FIX is in a bold, sans-serif font.
# SECTION 6. 510k Summary
| I.GENERAL INFORMATION<br>510k Number: | K182189 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment: | WFR/Aquaplast Corporation/Anholt Technologies Inc., Dba Qfix<br>440 Church Road<br>Avondale, PA 19311 USA |
| Date Prepared: | October 3, 2018 |
| Manufacturer: | Qfix<br>440 Church Road<br>Avondale, PA 19311 USA<br>Registration Number: 2247992 |
| Contact Person: | Alexandra Low Smythe<br>Regulatory Affairs Specialist<br><br>Qfix<br>440 Church Road<br>Avondale, PA 19311 USA<br>Phone: 610 268-0585 Ext 736<br>Fax: 610-268-0588 |
| Device Name:<br>Trade Name: | Encompass™ SRS Headframe<br>Encompass™ SRS Headframe, Encompass™ SRS Headframe<br>for Portrait, Encompass™ Headframe, SRS Headframe,<br>Encompass™-S |
| Common Name: | Stereotactic Radiosurgery (SRS) Immobilization Device |
| Classification Name: | Medical Charged-Particle Radiation Therapy System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR § 892.5050 |
| Device Class: | II |
| Product Code: | Primary: IYE<br>Secondary: LNH, LHN, JAI, JAK, KPS, OUO |
#### II.SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING SUBSTANTIAL EQUIVALENCE
#### Indications for Use
The Encompass™ SRS Headframe provides noninvasive stereotactic head and neck immobilization by using a patient specific thermoplastic mask that conforms to the patient's features to provide accurate, reproducible positioning, repositioning, and immobilization. The Encompass™ SRS Headframe allows the patient to undergo
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Q-Fix. The logo is a large, red letter Q with the word "FIX" in black letters inside the bottom left portion of the Q. There is a small black circle above the "IX" in the word "FIX". There is also a registered trademark symbol to the right of the Q.
diagnostic imaging in the same position as that of the treatment position enabling more accurate radiation therapy.
## Device Description
Qfix, has made a modification to its previously cleared Encompass™ SRS Immobilization System (cleared in K152321). In this modified variant of the Encompass™ SRS Immobilization System, the essential geometry of the head portion of the Encompass™ SRS Standalone Device is isolated to create a separate attachment which can be affixed to patient positioning devices that accept S-type thermoplastic masks, such as Qfix's Portrait™ Intracranial, Head and Neck Device (a device for use with S-type thermoplastics). Ofix intends to initially market the subject device for use with the Encompass™ SRS Fibreplast® System with optional IntegraBite™ (also cleared in K152321) and its devices which share a hole pattern and geometry with its Portrait™ Intracranial, Head and Neck Device. Examples of such devices include the kVue™ Portrait™ Intracranial. Head and Neck Insert and the Symphony® Portrait™ Transfer Device (cleared in K160627). This creates a flexible solution for a variety of imaging and radiotherapy treatment applications using stereotactic radiosurgery, such as radiation oncology and the treatment of noncancerous conditions and malformations such as trigeminal neuralgia.
## Predicate information
The subject device, the Encompass™ SRS Headframe, includes all of the device properties belonging to the predicate device, the Encompass™ SRS Immobilization System, also manufactured by Ofix. The predicate device information follows:
| Predicate Device Name | FDA Clearance<br>Number and Date | Product<br>code | Manufacturer |
|-----------------------------------------|--------------------------------------|-----------------|--------------|
| Encompass™ SRS<br>Immobilization System | K152321, cleared<br>December 4, 2015 | IYE | Qfix |
To date, this predicate device has not been subject to a design-related recall per information that is publicly accessible in the FDA recall database.
## Comparison to Predicate Device
The Encompass™ SRS Headframe bears many of the same features as the predicate device, the Encompass™ SRS Immobilization System. The Encompass™ SRS Headframe has the same intended use and purpose as the predicate device. Both the subiect device and the predicate device are compatible with radiation therapy environments, including stereotactic radiosurgery (SRS). Both devices are non-sterile, reusable devices manufactured from composite materials. Both devices accept the same Encompass™ SRS Fibreplast® System patient immobilization masks. Both devices boast sub-millimetric accuracy.
Notably, the subject device is a removable attachment intended for use with other patient positioning devices that accept S-type thermoplastic masks, such as the
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Q-Fix. The logo consists of a large, red letter "Q" with the word "Fix" in black letters superimposed over the lower portion of the "Q". There is a registered trademark symbol to the right of the word "Fix".
Portrait™ Intracranial. Head and Neck Device. allowing for improved setup flexibility for a variety of clinical applications as compared to the predicate device.
## Performance Standards and Testing
The FDA has not established performance standards for this product under Section 514 of the Food, Drug and Cosmetic Act. Testing and analysis has been conducted to show that the verification, validation, and safety requirements have been met.
- Verification of hardware specifications
- Accuracy validation via motion studies
- Ease of use/ergonomics assessments
- . Verification of aluminum equivalence
- . Verification of compatibility with optical tracking systems
Non-clinical bench testing and testing using healthy volunteers was conducted to support the continued applicability of the intended use and marketing claims of the predicate device with respect to this modification.
The Encompass™ SRS Headframe met all acceptance criteria for testing conducted and was appropriately validated per its intended use. The intended use and indications for use of the predicate device have not changed as a result of this modification.
## Safety and Effectiveness
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in development, testing, and product labeling. To minimize risks. Ofix adheres to recognized and established industry practices and standards to minimize safety and performance risks. Furthermore, the operators and end users of the device are healthcare professionals familiar with and responsible for radiation therapy treatments and other hospital procedures.
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
#### Substantial Equivalence
The subject device, the Encompass™ SRS Headframe, offers the following improvements over the predicate device, the Encompass™ SRS Immobilization System (cleared under K152321 on December 4, 2015).
- The subject device is intended to be used in conjunction with other devices, allowing for improved setup flexibility for a variety of clinical applications.
- The subject device has a smaller footprint.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Q-Fix. The logo is a large, red letter Q with the word "Fix" in black letters inside the Q. The dot of the "i" is a circle. The "Q" is stylized and has a trademark symbol in the upper right corner.
#### The subject device enhances available treatment options using existing clinical . infrastructure.
The fundamental attributes of the subject device and the predicate device are the same.
- . Both devices have the same intended use. The Encompass™ SRS Headframe has the same intended use as the Encompass™ SRS Immobilization System.
- . Both devices provide sub-millimetric immobilization for imaging and radiotherapy. including SRS.
The Encompass™ SRS Headframe, when used in conjunction with the Encompass™ SRS Fibreplast® System provides the same sub-millimetric immobilization as the Encompass™ SRS Immobilization System limiting motion to less than 1 mm in most clinical applications using stereotactic radiosurgery. such as radiation oncology and the treatment of non-cancerous conditions and malformations such as trigeminal neuralgia.
- Both the predicate and the subject device are intended to be used together . with additional accessories, inserts, overlays, and standalone devices which are intended to immobilize, position, and reposition patients undergoing radiation therapy.
The original Encompass™ SRS Immobilization System was intended to be used in conjunction with other devices in the Ofix product portfolio, such as the Encompass™ SRS Fibreplast® System with optional IntegraBite™, the kVue™ Couch Top, and various other patient positioning accessories. The Encompass™ SRS Headframe is intended to be used in conjunction with other devices in the Qfix product portfolio, such as the Portrait™ Intracranial, Head and Neck Device, the Encompass™ SRS Fibreplast® System with optional IntegraBite™, the kVue™ Couch Top, and various other patient positioning accessories.
#### Both devices have MR compatible variants. .
Both the Encompass™ SRS MR Immobilization Standalone Device and the Encompass™ SRS MR Headframe are made of non-conducting composite materials and are considered MR Safe per ASTM F2503-13.
The conclusions from the non-clinical data suggest that the subject device has the same fundamental technological characteristics with respect to the predicate device and exhibits an equivalent safety and performance profile as that of the predicate device.
Therefore, Qfix is of the opinion that the Encompass™ SRS Headframe does not raise new questions of safety or effectiveness and, therefore, is substantially equivalent to the marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.