Depuy Synthes Femoral Neck System

K182154 · Synthes (USA) Products, LLC · KTT · Oct 16, 2018 · Orthopedic

Device Facts

Record IDK182154
Device NameDepuy Synthes Femoral Neck System
ApplicantSynthes (USA) Products, LLC
Product CodeKTT · Orthopedic
Decision DateOct 16, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Femoral Neck System (FNS) is indicated for femoral neck fractures, including basilar, transcervical, and subcapital fractures, in adults and adolescents (12-21) in which the growth plates have fused or will not be crossed.

Device Story

Modular orthopedic implant system for femoral neck fracture fixation; consists of four connected components forming a fixed-angle gliding device; allows controlled collapse of femoral neck. Used by surgeons in clinical settings; implanted via system-specific insertion instruments. Benefits patients by providing stable fixation for various femoral neck fracture types. Device manufactured from titanium alloy; same construct used for left and right femur.

Clinical Evidence

No clinical data. Substantial equivalence determined via non-clinical performance data and evaluation of simulated device volumes in the femoral neck and head.

Technological Characteristics

Modular implant system; four connected components; fixed-angle gliding fixation mechanism. Material: Titanium alloy. Dimensions: Range of sizes. Application: Left and right femur. No software or energy source.

Indications for Use

Indicated for adults and adolescents (12-21) with fused or non-crossed growth plates suffering from femoral neck fractures, including basilar, transcervical, and subcapital fractures.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ October 16, 2018 Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Synthes (USA) Products, LLC Stephan Jepards Sr. Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380 ## Re: K182154 Trade/Device Name: Depuy Synthes Femoral Neck System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: KTT Dated: August 8, 2018 Received: August 9, 2018 ## Dear Stephan Jepards: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Peter G. Allen -S 2018.10.16 09:51:40 -04'00' - FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182154 Device Name DePuy Synthes Femoral Neck System Indications for Use (Describe) The Femoral Neck System (FNS) is indicated for femoral neck fractures, including basilar, transcervical, and subcapital fractures, in adults and adolescents (12-21) in which the growth plates have fused or will not be crossed. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Sponsor | DePuy Synthes<br>Contact: Stephan Jepards<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>Phone: +41 32 720 41 59 | |--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Prepared | October 15, 2018 | | Proprietary Name | DePuy Synthes Femoral Neck System | | Common Name | Multiple component metallic bone fixation appliance | | Classification Name | Single/multiple component metallic bone fixation appliances and accessories | | Classification | Class II<br>Regulation Number: 21 CFR 888.3030<br>Product Code: KTT | | Primary predicate<br>device | Smith & Nephew PERI-LOC Proximal Femur Locking Bone Plates (K112406) | | Predicate device | Stryker Omega 3 System (K062066)<br>DePuy Synthes Femoral Neck System (K172872) | | Reference device | Asnis III 6.5mm Cannulated Screw (K024060) | | Device Description | The subject Femoral Neck System is comprised of implants designed to treat femoral<br>neck fractures as well as system-specific insertion instruments. The Femoral Neck<br>System is a modular system consisting of four connected implant components forming<br>a fixed-angle gliding fixation device which allows for controlled collapse of the<br>femoral neck.<br>The implants are manufactured from Titanium Alloy and are provided in a range of<br>dimensions. The same construct can be used for the left and right femur. | | Indications for use | The Femoral Neck System (FNS) is indicated for femoral neck fractures, including<br>basilar, transcervical, and subcapital fractures, in adults and adolescents (12-21) in<br>which the growth plates have fused or will not be crossed. | | Technological<br>characteristics of the<br>new device compared to<br>the predicate devices | The Smith & Nephew PERI-LOC Proximal Femur Locking Bone Plates (K112406)<br>is a multiple component device including a variety of plate, cannulated locking and<br>non-locking screw and cable saddle designs and sizes. | | | The Stryker Omega 3 System (K062066) is a compression screw system designed to<br>treat various types of fractures of the proximal and distal femur.<br>The Femoral Neck System is a modular system consisting of four connected implant<br>components forming a fixed-angle gliding fixation device which allows for controlled<br>collapse of the femoral neck.<br>Any differences in technological characteristics from the predicate device do not<br>raise any new questions of safety and effectiveness. | | Non-clinical<br>Performance Data | The evaluation of simulated device volumes in the femoral neck and head<br>demonstrates that the subject device does not present a new worst case for use in<br>femoral neck fracture fixation compared to the predicate and reference devices. Hence<br>it supports the proposed expanded 'Indications for Use' for the subject device. | | Clinical Performance<br>Data | Clinical testing was not necessary for the determination of substantial equivalence. | | Substantial Equivalence | The subject system has similar intended use and indications for use compared to the<br>predicate device. | | | The non-clinical performance data included in this submission supports that any<br>differences in technological characteristics from the predicate device do not raise<br>any new questions of safety and effectiveness. | | | It is concluded that the information provided in this submission supports substantial<br>equivalence. | {4}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...